K083421 · Ge Medical Systems, LLC · LNH · Jul 24, 2009 · Radiology
Device Facts
Record ID
K083421
Device Name
MR-TOUCH OPTION
Applicant
Ge Medical Systems, LLC
Product Code
LNH · Radiology
Decision Date
Jul 24, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Intended Use
MR-Touch™ is a software and hardware option intended for use with GE Signa® MR systems. It is indicated for magnetic resonance imaging of the human body. MR-Touch™ generates transverse sectional information related to the relative stiffness of soft tissue. It consists of hardware as well as acquisition and reconstruction software components induce vibrations into the scan subject. The MR-Touch™ acquisition software is an evolutionary improvement to the gradient echo sequence. The sequence synchronizes the induced vibrations to acquire a series of phase-contrast images over time. The phase-contrast imaging technique measures motion or displacement. The displacement from the induced vibrations is detected using the time-series of phase-contrast images. The displacement information is reconstructed and presented as strain wave and relative stiffness images. When used with a GE Signa® MR system, MR-Touch™ is capable of producing transverse images of internal body structures such as muscle and liver. When interpreted by a trained physician, these images may provide information that can be useful in determining a diagnosis.
Device Story
MR-Touch™ is a hardware/software accessory for GE Signa® MR systems. Hardware includes an acoustic wave generator and passive transducer connected by flexible tubing to induce low-frequency, low-magnitude vibrations in the patient. Software includes a modified gradient echo sequence that synchronizes these vibrations with phase-contrast image acquisition. The system measures tissue displacement over time; reconstruction algorithms process this displacement data into strain wave and relative stiffness images (Elastograms). Used in clinical settings by trained physicians to visualize internal structures like muscle and liver. Output assists clinicians in differentiating tissues based on stiffness, potentially aiding diagnostic decision-making.
Clinical Evidence
Bench testing only. Verification confirmed functionality with GE Signa® HDx MR System. Confidence studies demonstrated repeatable results and the ability to reliably differentiate between tissues of varying stiffness. Hardware evaluated for safety and electromagnetic compatibility per IEC 60601-1 and IEC 60601-1-2 standards.
Technological Characteristics
Hardware: acoustic wave generator, flexible tubing, passive transducer. Software: gradient echo sequence modification for synchronized vibration acquisition and reconstruction. Connectivity: integrated with GE Signa® MR systems. Safety: compliant with IEC 60601-1 and IEC 60601-1-2.
Indications for Use
Indicated for magnetic resonance imaging of the human body, specifically for generating transverse sectional information related to the relative stiffness of soft tissue (e.g., muscle and liver) to assist trained physicians in diagnosis.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
K083421
## GE Healthcare
MR-Touch™ Option for GE Signa® MR Systems 510(k} Premarket Notification
## 510(k) Summary
This 510|k) summary is submitted in accordance with the requirements of 21 C.F.R. Part 807.92.
JUL 2 4 2009
| Submitter: | GE Medical Systems, LLC |
|-------------------|---------------------------|
| | 3200 N. Grandview Blvd. |
| | Waukesha, WI 53188 |
| Contact Person: | Daniel T. Biank |
| | Regulatory Affairs Leader |
| Telephone: | 262 - 521 - 6467 |
| Fax: | 262 - 521 - 6379 |
| Email: | Daniel.Biank@med.ge.com |
| Date Prepared: | November 12, 2008 |
| Device Name: | |
| Proprietary Name: | MR-TouchTM Option |
Proprietary Name: MK-Touch™ Option
Common Name: Software and Hardware Accessory for Magnetic Resonance Imaging System
Classification Name: Magnetic Resonance Diagnostic Device, (21 C.F.R. 892.1000, LNH)
#### Predicate Devices:
GE 1.5T and 3.0T Signa® HDx MR System (K052293).
Ultrasonix SonixTouch Ultrasound Imaging System for Elastoaraphy (K083095),
#### Device Description:
MR-Touch™ is a combined software and hardware accessory for use with a GE Signa® MR System. It is an evolutionary improvement of the existing phase-contrast imaging feature included with existing GE Signa® MR Systems. The Resoundant hardware component consists of an acoustic wave generator coupled through a length of flexible tubing with a passive transducer to induce small vibrations in the scan subject. The software includes both image acquisition and reconstruction components. The acquisition software is a gradient echo sequence that acquires a series of phase-contrast images over time. It also synchronizes the low frequency and low magnitude induced vibrations generated by Resoundant. The phase-contrast imaging technique measures motion or displacement. During reconstruction, the displacement from the induced vibrations is detected using the timeseries of phase-contrast images. MR-Touch™ then presents the reconstructed displacement information as strain wave and relative stiffness images (referred to as Elastograms).
#### Intended Use:
MR-Touch™ is a software and hardware option intended for use with GE Signa® MR systems. It is indicated for magnetic resonance imaging of the human body.
MR-Touch™ generates transverse sectional information related to the relative stiffness of soft tissue. It consists of hardware as well as acquisition and reconstruction software components induce vibrations into the scan subject. The MR-Touch™ acquisition software is an evolutionary improvement to the gradient echo sequence. The sequence synchronizes the induced vibrations to acquire a series of phase-contrast images over time. The phase-contrast imaging technique measures motion or displacement. The displacement from the induced vibrations is detected using the time-series of phase-contrast images. The displacement information is reconstructed and presented as strain wave and relative stiffness images.
{1}------------------------------------------------
# GE Healthcare
MR-Touch™ Option for GE Signa® MR Systems 510[k} Premarket Notification
When used with a GE Signo® MR system, MR-Touch™ is capable of producing transverse images of internal body structures such as muscle and liver.
When interpreted by a trained physician, these images may provide information that can be useful in determining a diagnosis.
#### Comparison with Predicate Devices:
MR-Touch™ is substantially equivalent to the existing phase-contrast imaging feature of the previously cleared GE 1.5T and 3.0T Signa® HDx MR System (K052293). The primary differences include a more convenient method to visualize a time-series of phase-contrast images and the ability to induce synchronized vibrations in the scan subject during acquisition. The Elastography relative-stiffness images (Elastograms) are substantially equivalent to the output of the previously cleared Ultrasonix SonixTouch Elastography imaging mode, which provides color-coded images to differentiate between tissues based on stiffness.
#### Summary of Studies:
MR-Touch™ has been verified to function with the GE Signa® HDx MR System. Sample images demonstrate the strain wave and Elastogram outputs. Additionally, included confidence studies prove that MR-Touch™ produces repeatable results and can reliably differentiate between tissues of different stiffness.
The Resoundant, new hardware introduced with MR-Touch™, has been evaluated to the appropriate clauses of IEC 60601-1 Medical Electrical Equipment - Part 1: General Requirements for Safety standard and IEC 60601-1-2 Medical Electrical Equipment - Part 1-2: General Requirements for Safety 2001 - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests - Second Edition with Amendment 1.
#### Conclusion:
It is the opinion of GE that MR-Touch™ when used with a GE Signa® MR System is substantially equivalent to the currently cleared and marketed GE 1.5T and 3.0T Signa® HDx MR Systems, providing Elastography image output that is substantially equivalent to the output provided by the Elastography imaging mode of the currently cleared Ultrasonix SonixTouch Ultrasound Imaaing System.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image is a circular seal or logo. The seal contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged around the top half of the circle. In the center of the seal is a stylized image of an eagle or other bird with its wings spread. The bird is depicted in a simple, abstract manner with three curved lines forming its body and wings.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Daniel Biank Regulatory Affairs Leader GE Healthcare, MR 3200 N. Grandview Blvd. WAUKESHA WI 53188
JUL 24 2009
Re: K083421
Trade/Device Name: MR-Touch™ Option Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: June 23, 2009 Received: June 25, 2009
### Dear Mr. Biank:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
> Sincerely yours willy
Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
GE Healthcare
## Indications for Use
510(k) Number (if known):
Device Name:
### MR-Touch™ Option
### Indications for Use:
MR-Touch™, is a software and hardware option intended for use with GE Signa® MR systems. It is indicated for maqnetic resonance imaging of the human body.
MR-Touch™ generates transverse sectional information related to the relative stiffness of soft tissue. It consists of hardware as well as acquisition and reconstruction software. The hardware components induce vibrations into the scan subject. The MR-Touch™ acquisition software is an evolutionary improvement to the gradient echo sequence. The sequence synchronizes the induced vibrations to acquire a series of phase-contrast images over time. The phase-contrast imaging technique measures motion or displacement. The displacement from the induced vibrations is detected using the timeseries of phase-contrast images. The displacement information is reconstructed and presented as strain wave and relative stiffness images.
When used with a GE Signa® MR system, MR-Touch™ is capable of producing transverse images of internal body structures such as muscle and liver.
When interpreted by a trained physician, these images may provide information that can be useful in determining a diagnosis.
Prescription Use × AND/OR Over-the-Counter Use (21 CFR 801 Subpart D) (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Hogart M. Thag
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
Section 4
MR-Touch™ Option
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.