HyperBand

K162403 · Ge Medical Systems, LLC · LNH · Nov 18, 2016 · Radiology

Device Facts

Record IDK162403
Device NameHyperBand
ApplicantGe Medical Systems, LLC
Product CodeLNH · Radiology
Decision DateNov 18, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1000
Device ClassClass 2

Intended Use

HyperBand is a software option intended for use on GE MR 1.5T and 3T Systems. HyperBand consists of an acquisition and reconstruction technique allowing simultaneous excitation of multiple slices at multiple locations to accelerate imaging acquisition times or increase slice coverage without increasing scan time. HyperBand is indicated for echo-planar imaging (EPI) of the head and breast. HyperBand is indicated for imaging all patients not otherwise contraindicated for MRI examination.

Device Story

HyperBand is a software-only optional feature for GE 1.5T and 3T MR systems; it functions as an acquisition and reconstruction technique. By simultaneously exciting multiple slices at multiple locations, it accelerates imaging acquisition or increases slice coverage without increasing scan time. It is used in clinical settings by MRI technologists/radiologists to improve productivity and patient experience through shorter breath-holds and higher resolution. The output is MR images, which are reviewed by clinicians to inform diagnostic decision-making. The device is intended to provide diagnostic-quality images comparable to standard parallel imaging methods.

Clinical Evidence

No external clinical studies required. Internal validation scans were conducted to assess workflow and image quality. Results demonstrated that HyperBand maintains imaging performance equivalent to the predicate device (K012970). Sample clinical images were provided.

Technological Characteristics

Software-only MR application; compatible with GE 1.5T and 3T systems. Complies with AAMI/ANSI 62304, AAMI/ANSI ES60601-1, IEC 60601-2-33, and NEMA PS3.1-3.18 (DICOM). Employs simultaneous multi-slice excitation for parallel imaging acceleration.

Indications for Use

Indicated for echo-planar imaging (EPI) of the head and breast in all patients not otherwise contraindicated for MRI examination.

Regulatory Classification

Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines connecting them. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 November 18, 2016 GE Medical Systems, LLC % Mr. Glen Sabin Regulatory Affairs Leader 3200 North Grandview Blvd. WAUKESHA WI 53188 Re: K162403 Trade/Device Name: HyperBand Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: August 26, 2016 Received: August 29, 2016 Dear Mr. Sabin: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Robert Oolo Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use ### 510(k) Number (if known) K162403 Device Name HyperBand #### Indications for Use (Describe) HyperBand is a software option intended for use on GE MR 1.5T and 3T Systems. HyperBand consistion and reconstruction technique allowing simultaneous excitation of multiple locations to accelerate imaging acquisition times or increase slice coverage without increasing scan time. HyperBand is indicated for echo-planar imaging (EPI) of the head and breast. HyperBand is indicated for imaging all patients not otherwise contraindicated for MRI examination. Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) | <div> <span style="font-size: 20px;">☑</span> </div> | |----------------------------------------------|------------------------------------------------------| | Over-The-Counter Use (21 CFR 801 Subpart C) | | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are decorative swirls around the perimeter of the circle. The logo is simple and recognizable, and it is associated with a well-known and established company. GE Healthcare 510(k) Premarket Notification Submission Section 5: 510(k) Summary HyperBand {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized script, enclosed within a blue circle. There are also some white water droplet shapes around the circle. ## GE Healthcare 510(k) Premarket Notification Submission | In accordance with 21 CFR 807.92 the following summary of information is provided: | | | |------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--| | Date: | November 16, 2016 | | | Primary Contact Person: | Glen Sabin<br>Requlatory Affairs Director, MR<br>GE Healthcare, (GE Medical Systems, LLC)<br>Phone: 262-521-6848<br>Fax: 414-908-9585 | | | Device Trade Name: | HyperBand | | | Common/Usual Name: | MR Software Application | | | Classification Names: | Magnetic Resonance Diagnostic Device<br>(21 CFR 892.1000) | | | Product Code: | LNH | | | Predicate Devices: | Predicate Device: ASSET (K012970) | | | Device Description: | HyperBand is a clinical software application which is<br>offered as an optional feature for GE Healthcare's 1.5T<br>and 3T MR systems. It provides a reduction in scan<br>time by simultaneously exciting multiple slices at<br>multiple locations. It can lead to higher acceleration<br>reduction factors when combined to other methods of<br>parallel imaging. The benefits of HyperBand<br>acceleration include enhancements on productivity and<br>patient experience by shortened breath holds,<br>increased anatomy coverage and higher resolution<br>image acquisition. | | | Indications for Use: | HyperBand is a software option intended for use on GE<br>MR 1.5T and 3T Systems. HyperBand consists of an<br>acquisition and reconstruction technique allowing<br>simultaneous excitation of multiple slices at multiple<br>locations to accelerate imaging acquisition times or<br>increase slice coverage without increasing scan time.<br>HyperBand is indicated for echo-planar imaging (EPI) of<br>the head and breast.<br>HyperBand is indicated for imaging all patients not<br>otherwise contraindicated for MRI examination. | | | Comparison of Intended<br>Use | Both HyperBand and the predicate device are magnetic<br>resonance imaging devices intended for diagnostic use. | | | | Both indications for use statements are functional in<br>nature, and do not list specific diseases or conditions.<br>HyperBand and the predicate device are indicated for<br>the same patient population, and for the same clinical<br>setting. HyperBand and the predicate device are<br>indicated for similar anatomical regions: HyperBand<br>for the head and breast, and ASSET for the head or<br>body. | | | | Therefore, GE Healthcare believes that HyperBand has<br>the same intended use as the predicate device in<br>accordance with the FDA's guidance document "The<br>510(k) Program: Evaluating Substantial Equivalence in<br>Premarket Notifications [510(k)]", dated 28 July 2014. | | | Comparison of<br>Technological<br>Characteristics: | The most notable technological difference between<br>HyperBand and the predicate device is that HyperBand<br>is capable of simultaneously exciting multiple slices<br>leading to higher acceleration reduction factors when<br>combined with other methods of parallel imaging. The<br>predicate device acquires one slice or one slab at a time. | | | | However, per bench and clinical data collected to<br>validate the indications for use, this technological<br>difference does not raise any different questions of<br>safety and effectiveness. Both devices must address<br>questions of whether they provide an adequate level of<br>image quality appropriate for diagnostic use. The<br>performance data described in this submission include<br>results of both bench testing and clinical testing that<br>show the image quality performance of HyperBand<br>compared to the predicate device. | | | Performance Data: | Summary of Non-Clinical Tests:<br>HyperBand is a software only feature and complies<br>with the following voluntary standards:<br>AAMI/ANSI 62304 AAMI/ANSI ES60601-1 IEC 60601-2-33 | | | | In addition, HyperBand complies with NEMA PS3.1-3.18<br>for DICOM conformance. | | | | The following quality assurance measures were applied<br>to the development of the device:<br>Risk Analysis Requirements Reviews Design Reviews Testing on unit level (Module verification) Integration testing (System verification) Performance testing (Verification) Safety testing (Verification) Simulated use testing (Validation) | | | | Signal to Noise Ratio (SNR) testing was performed with<br>and without the HyperBand feature, and provided to<br>support substantial equivalence. | | | | The non-clinical tests have been summarized in the<br>verification and validation testing for HyperBand. The<br>testing was completed with passing results per the<br>pass/fail criteria defined in the test cases. This supports<br>substantial equivalence to its predicates because it was<br>also developed under quality assurance Design<br>Controls. In addition, the software complies with the<br>same applicable Standards. | | | | Summary of Clinical Tests:<br>The subject of this premarket submission, HyperBand,<br>did not require external clinical studies to support<br>substantial equivalence. Internal scans were conducted<br>as part of validation for workflow and image quality for<br>the addition of the new features. The clinical results<br>demonstrated that HyperBand maintain the same<br>imaging performance results as its predicate devices<br>(K012970). Sample clinical images are included in this<br>submission. | | | Conclusion: | GE Healthcare believes that the HyperBand software<br>feature has substantially the same intended use as the<br>predicate ASSET. This 510(k) submission includes<br>information on the technological characteristics of the<br>HyperBand feature, as well as performance data<br>demonstrating that HyperBand is as safe and effective<br>as the predicate, and does not raise different questions<br>of safety and effectiveness | | | | In conclusion, GE Healthcare believes that the<br>HyperBand feature is substantially equivalent to the<br>predicate device. | | {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a blue circle. There are water droplet-like shapes surrounding the circle. The logo is simple and recognizable, representing the company's brand identity. GE Healthcare 510(k) Premarket Notification Submission {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo, which consists of the letters 'GE' intertwined within a blue circle. The letters are stylized and appear to be hand-drawn or calligraphic. The circle has four water droplet-like shapes extending outward at the cardinal directions, giving the logo a dynamic and fluid appearance. The color scheme is simple, with the letters and circle in a light blue hue against a white background. # GE Healthcare 510(k) Premarket Notification Submission {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized script, enclosed within a blue circle. The circle has decorative, wave-like elements around its perimeter, giving it a distinctive and recognizable appearance. GE Healthcare 510(k) Premarket Notification Submission
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...