HyperSense

K162722 · Ge Medical Systems, LLC · LNH · Apr 20, 2017 · Radiology

Device Facts

Record IDK162722
Device NameHyperSense
ApplicantGe Medical Systems, LLC
Product CodeLNH · Radiology
Decision DateApr 20, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1000
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K162722 · Apr 20, 2017HyperSenseGe Medical Systems, LLCPeer-reviewed clinical literatureA peer-reviewed journal article was provided as supporting evidence to demonstrate that the HyperSense technique maintains comparable imaging performance to the predicate device.Peer-reviewed literature; Clinical performance; Imaging quality

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Peer reviewed journal article describing a study of the HyperSense technique—Not applicable for this studyImaging performance/image quality

Indications for Use

HyperSense is a software feature intended for use on GE MR 1.5T and 3.0T Systems. HyperSense is an acceleration technique based on sparse data sampling and iterative reconstruction that allows users to reduce scan times or increase scan resolution. HyperSense can be used for non-contrast enhanced imaging of the head, neck, spine, extremities, pelvis, and abdomen.

Device Story

HyperSense is a software-only acceleration feature for GE 1.5T and 3.0T MR systems; utilizes sparse data sampling and iterative reconstruction to enable faster imaging or higher spatial resolution; operates as an add-on to existing volumetric acquisition workflows; can be combined with other acceleration methods like ARC. Used by radiologists and technologists in clinical settings to optimize scan times and image quality. Output is reconstructed MR images; assists clinicians in diagnostic assessment by maintaining image quality while reducing acquisition time. Benefits patients through shorter exam durations and potentially improved image resolution.

Clinical Evidence

Clinical study evaluated image quality via legibility of morphological features; supported by a peer-reviewed journal article. Results demonstrated comparable imaging performance to the predicate device (K142085). Bench testing included phantom studies evaluating Signal to Noise Ratio (SNR) and spatial resolution, with all tests meeting pass/fail criteria.

Technological Characteristics

Software-only feature for GE 1.5T/3.0T MR systems. Employs compressed sensing (sparse data sampling) and iterative reconstruction. Complies with AAMI/ANSI 62304, AAMI/ANSI ES60601-1, IEC 60601-2-33, and NEMA PS3.1-3.18 (DICOM).

Indications for Use

Indicated for patients undergoing non-contrast enhanced MRI of the head, neck, spine, extremities, pelvis, and abdomen on GE 1.5T and 3.0T systems.

Regulatory Classification

Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles three human profiles facing to the right, with flowing lines connecting them. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 20, 2017 Ge Medical Systems, LLC Glen Sabin Regulatory Affairs Director 3200 Grandview Blvd Waukesha. Wisconsin 53188 Re: k162722 Trade/Device Name: HyperSense Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: LNH Dated: March 15, 2017 Received: March 16, 2017 Dear Glen Sabin: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {2}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, Michael O'Hara For Robert A. Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K162722 Device Name HyperSense #### Indications for Use (Describe) HyperSense is a software feature intended for use on GE MR 1.5T and 3.0T Systems. HyperSense is an acceleration technique based on sparse data sampling and iterative reconstruction that allows users to reduce scan times or increase scan resolution. HyperSense can be used for non-contrast enhanced imaging of the head, neck, spine, extremities, pelvis, and abdomen. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| | Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular frame. The logo is colored in blue, with the letters and the frame having the same shade of blue. The background is white. | In accordance with 21 CFR 807.92 the following summary of information is provided: | | | |------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--| | Date: | 19 April 2017 | | | Primary Contact Person: | Glen Sabin<br>Regulatory Affairs Director, MR<br>GE Healthcare, (GE Medical Systems, LLC)<br>Phone: 262-521-6848<br>Fax: 414-908-9585 | | | Secondary Contact Person: | James McMahon<br>Senior Director, Regulatory Affairs - MR<br>GE Healthcare, (GE Medical Systems, LLC)<br>Phone: 508-382-2858 | | | Device Trade Name: | HyperSense | | | Common/Usual Name: | Magnetic Resonance Diagnostic Device | | | Classification Names: | 892.1000 | | | Product Code: | LNH | | | Predicate Devices: | Predicate Device: ARC (K142085) | | | Device Description: | HyperSense is a software feature used on GE 1.5T and<br>3.0T MR systems.<br>HyperSense is an acceleration technique based on<br>sparse data sampling and iterative reconstruction,<br>enabling faster imaging without the penalties<br>commonly found with conventional parallel imaging.<br>HyperSense is intended to be used with volumetric<br>acquisitions, and can be combined with other methods<br>of acceleration (ARC) for achieving high signal to noise<br>ratio with shorter acquisition times. HyperSense can<br>deliver higher spatial resolution images or reduced<br>scan times. | | {5}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are three white water droplet shapes around the letters. The logo is simple and recognizable. | Indications for Use: | HyperSense is a software feature intended for use on<br>GE MR 1.5T and 3.0T Systems. HyperSense is an<br>acceleration technique based on sparse data sampling<br>and iterative reconstruction that allows users to reduce<br>scan times or increase scan resolution. HyperSense<br>can be used for non-contrast enhanced imaging of the<br>head, neck, spine, extremities, pelvis, and abdomen. | |----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Comparison of Intended<br>Use | Both HyperSense and the predicate device are<br>magnetic resonance imaging devices intended for<br>diagnostic use.<br>Both indications for use statements are functional in<br>nature, and do not list specific diseases or conditions.<br>HyperSense and the predicate device are indicated for<br>the same patient population, and for the same clinical<br>setting. Both indications for use statements include a<br>range of intended anatomies.<br>Therefore, GE Healthcare believes that HyperSense has<br>the same intended use as the predicate device in<br>accordance with the FDA's guidance document "The<br>510(k) Program: Evaluating Substantial Equivalence in<br>Premarket Notifications [510(k)]", dated 28 July 2014. | | Comparison of<br>Technological<br>Characteristics: | The most notable technological difference between<br>HyperSense and the predicate device is that HyperSense<br>adopts compressed sensing acquisition method based<br>on sparse data compressibility allowing scan time<br>reduction while maintaining appropriate image quality.<br>Per bench and clinical data collected to validate the<br>indications for use, this technological difference does<br>not raise any different questions of safety and<br>effectiveness. Both devices must address questions of<br>whether they provide an adequate level of image<br>quality appropriate for diagnostic use. The performance<br>data described in this submission include results of<br>both bench testing and clinical testing that show the<br>image quality performance of HyperSense compared to<br>the predicate device | {6}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a blue circle. There are also some water droplet-like shapes surrounding the circle. The logo is simple and recognizable, representing the company's brand identity. | Performance Data | Summary of Non-Clinical Tests: | |------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | HyperSense is a software only feature and complies with the following voluntary standards:<br>AAMI/ANSI 62304 AAMI/ANSI ES60601-1 IEC 60601-2-33 | | | In addition, HyperSense complies with NEMA PS3.1-3.18 for DICOM conformance. | | | HyperSense complies with the same applicable standards as the software used in the predicate device. | | | As with the predicate device, the following quality assurance measures were applied to the development of HyperSense:<br>Risk Analysis Requirements Reviews Design Reviews Testing on unit level (Module verification) Integration testing (System verification) Performance testing (Verification) Safety testing (Verification) Simulated use testing (Validation) | | | Phantom testing has been conducted to evaluate the impact of HyperSense on Signal to Noise Ratio (SNR), and spatial resolution. | | | The non-clinical testing was completed with passing results per the pass/fail criteria defined in the test cases. | | | Summary of Clinical Tests: | | | A clinical study has been performed to evaluate the impact of HyperSense on image quality, as measured by the legibility of morphological features. Additionally, a peer reviewed journal article describing a study of the HyperSense technique has been provided in this submission as supporting evidence. The clinical results demonstrated that HyperSense maintains comparable imaging performance results as its predicate devices (K142085). Sample clinical images are included in this submission. | | Conclusion: | GE Healthcare believes that the HyperSense software<br>feature has substantially the same intended use as the<br>predicate ARC. This 510(k) submission includes<br>information on the technological characteristics of the<br>HyperSense feature, as well as performance data<br>demonstrating that HyperSense is as safe and effective<br>as the predicate, and does not raise different questions<br>of safety and effectiveness.<br>In conclusion, GE Healthcare believes that the<br>HyperSense feature is substantially equivalent to the<br>predicate device. | {7}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are white swirls around the letters, creating a dynamic and recognizable design. The logo is simple, clean, and easily identifiable.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...