E-scan XQ is a magnetic resonance (MR) system that produces transversal, sagittal and coronal and oblique cross-section images of the limbs and joints. It is intended for imaging the arm, including the hand, wrist, forearm, elbow, upper arm and shoulder, and imaging the leg, including the foot, ankle, calf, knee, thigh and hip. E-scan XQ MR images correspond to the spatial distribution of protons (hydrogen nucle) that determine magnetic resonance properties and are dependent on the MR parameters, including spin-latice relaxation time (TI), spin-spin relaxation time (T2), nuclei density, flow velocity and "chemical shift". When interpreted by a medical expert trained in the use of MR equipment, the images can provide diagnostically useful information.
Device Story
E-scan XQ is a magnetic resonance imaging (MRI) system for limb and joint diagnostics. It captures MR signals from hydrogen nuclei in patient tissues; transforms these signals via pulse sequences and digital signal processing into cross-sectional images (transversal, sagittal, coronal, oblique). Operated by trained medical experts in clinical settings; output is viewed on a monitor to assist in clinical decision-making. Modifications from the predicate (K020164) include updated electronics (DSP SHARC 21161), higher performance PC, improved RF amplifier (900W), and new high-resolution pulse sequences. Benefits include enhanced image quality and diagnostic utility through technological updates to the acquisition and reconstruction pipeline.
Clinical Evidence
Bench testing only. No clinical data presented. Performance verified through technical comparison of pulse sequences, acquisition matrices, and system hardware upgrades against the predicate device.
Technological Characteristics
MR imaging system; magnetic unit with patient table; RF power amplifier (900W); digital control system using DSP SHARC 21161 (100 MHz); A/D conversion (14-bit, 20 MHz sampling); pulse sequences include Spin Echo, Turbo Spin Echo, Gradient Echo, and Inversion Recovery; high-resolution acquisition matrix up to 512x512; connectivity via DICOM server/visualizer (ESA-VIEW Rel. 2).
Indications for Use
Indicated for imaging limbs and joints, including arm (hand, wrist, forearm, elbow, upper arm, shoulder) and leg (foot, ankle, calf, knee, thigh, hip) in patients requiring MR diagnostic imaging.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
K032/2/
AUG 1 3 2003
510(k) Summary E-Scan XQ Biosound Esaote
# 510(k) Summary
The following 510(k) summary has been prepared pursuant to requirements specified in 21CFR¶807.92(a).
807.92(a)(1)
# Submitter Information
| Colleen Densmore, Official Correspondent<br>8000 Castleway Drive<br>Indianapolis, IN 46250 | | |
|--------------------------------------------------------------------------------------------|--------------------------------------------|---------|
| Phone: | (317) 849-1916 | |
| Facsimile: | (317) 577-9070 | |
| Contact Person: | Colleen Densmore | |
| Date: | June 30, 2003 | |
| 807.92(a)(2) | | |
| Trade Name: | E-Scan XQ | |
| Common Name: | Magnetic resonance diagnostic device | |
| Classification Name(s): | System, Nuclear Magnetic Resonance Imaging | |
| Classification Number: | 90LNH | |
| 807.92(a)(3) | | |
| | Predicate Device(s) | |
| Esaote | Artoscan M | K963262 |
| Esaote | E-Scan | K990968 |
| Esaote | E-Scan | K001894 |
| Esaote | Hip Coil | K012728 |
| Esaote | E - Scan XQ | K020164 |
{1}------------------------------------------------
01.92(a)(4)
## Device Description
### Summary of E-scan XQ modifications
The changes performed on the E-scan XQ device, with respect to the cleared version - E-scan XQ K020164 -Are due to the improvement of the system performance. These modifications, that do not affect the intended use or alter the fundamental scientific technology of the device are the following.
- 1. Modified system configuration.
- 2. New operating tables.
- Upgrading of the electronics. งา
- 4. A new software release.
## System configuration
## Unmodified E-scan XQ
The unit is composed of these main parts:
- 1 Patient positioning table.
- 2. Magnetic unit.
- Console, made of the Operating console (mouse, keyboard, monitor, ODD, FDD and CD-RW) and the 3. Electronics box.
- Modular shielding box with filter panel. 4.
## Modified E-scan XQ
The system is composed of these main parts:
- 1. Patient positioning table.
- 2. Magnetic unit.
- Operating console that consists of the PC unit (including keyboard and mouse), the monitor and the 3. operating table.
- 4. Electronics box with filter panel.
- ഗ Modular shielding box.
### 807.92(a)(5)
# Intended Use(s)
E-scan XQ is a magnetic resonance (MR) system that produces transversal, sagittal and coronal and oblique cross-section images of the limbs and joints. It is intended for imaging the arm, including the hand, wrist, forearm, elbow, upper arm and shoulder, and imaging the leg, including the foot, ankle, calf, knee, thigh and hip.
E-scan XQ MR images correspond to the spatial distribution of protons (hydrogen nucle) that determine magnetic resonance properties and are dependent on the MR parameters, including spin-latice relaxation time (TI), spin-spin relaxation time (T2), nuclei density, flow velocity and "chemical shift". When interpreted by a medical expert trained in the use of MR equipment, the images can provide diagnostically useful information.
{2}------------------------------------------------
# Technological Characteristics
# Comparison to the cleared device E-scan XQ K020164
# Imaging system
| Characteristics | E-scan XQ<br>K020164 | E-scan XQ<br>modified | Comments |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Pulse sequences | Orthogonal Multi-planar Scout<br>Spin Echo T1 (set1)<br>Spin Echo T2 (set2)<br>Multi-Echo (se_pd_t2)<br>Inversion Recovery (ir)<br>Short TI Inversion Recovery (stir)<br>Spin Echo Half Echo (set1he)<br>Spin Echo Half Scan (set1hf)<br>Turbo SE T2 weighted and Turbo<br>ME (tse, tme)<br>Gradient Echo (ge)<br>Short Time Inversion Recovery<br>Gradient Echo (ge_stir)<br>Gradient Echo 3D (t3d_t1)<br>Gradient Echo 3D contrast<br>enhancement (3d_ce)<br>Real Time | Orthogonal Multi-planar Scout<br>Spin Echo T1 (set1)<br>Spin Echo T2 (set2)<br>Multi-Echo (se_pd_t2)<br>Inversion Recovery (ir)<br>Short TI Inversion Recovery (stir)<br>Spin Echo Half Echo (set1he)<br>Spin Echo Half Scan (set1hf)<br>Turbo SE T2 weighted and Turbo ME (tse, tme)<br>Gradient Echo (ge)<br>Short Time Inversion Recovery Gradient Echo<br>(ge_stir)<br>Gradient Echo 3D (t3d_t1)<br>Gradient Echo 3D contrast enhancement (3d_ce)<br>Real Time | Unchanged |
| Sequence<br>parameters | | High Resolution se_26:<br>TR from 60 ms to 5000 ms, step 1 ms<br>TE fixed at 26 ms<br>minimum FOV 100 mm<br>minimum slice thickness 2.0 mm<br><br>High Resolution se_26_hf:<br>TR from 60 ms to 5000 ms, step 10 ms<br>TE fixed at 26 ms<br>minimum FOV 100 mm<br>minimum slice thickness 2.0 mm | The high resolution sequences are a<br>particular version of the E-scan XQ<br>standard sequences with maximum<br>acquisition matrix 512x512 instead of<br>256x256. All these sequences have a<br>fixed TE for obtaining the best<br>compromise between the S/N and the<br>high resolution.<br><br>High Resolution se_26: it is a Spin Echo<br>T1 sequence with TE=26 msec. |
| Characteristics | E-scan XQ<br>K020164 | E-scan XQ<br>modified | Comments |
| | | High Resolution se_18_he<br>TR from 60 ms to 5000 ms, step 10 ms<br>TE fixed at 18 ms | High Resolution se_26_hf: it is a Spin<br>Echo T1 Half Fourier sequence with<br>TE=26 msec. |
| | | minimum FOV 120 mm<br>minimum slice thickness 2.0 mm<br>High Resolution tse_80:<br>TR from 200 ms to 5000 ms, step 10 ms<br>TE fixed at 80 ms | High Resolution se_18_he: it is a Spin<br>Echo T1 Half Echo sequence with<br>TE=18 msec. |
| | | minimum FOV 120 mm<br>minimum slice thickness 3.0 mm<br>High Resolution tse_50: | High Resolution tse_80: it is a Turbo<br>Spin Echo T2 sequence with TE=80<br>msec. |
| | | TR from 200 ms to 5000 ms, step 10 ms<br>TE fixed at 50 ms<br>minimum FOV 120 mm<br>minimum slice thickness 3.0 mm | High Resolution tse_50: it is the high<br>resolution version of the standard<br>sequence Turbo Spin Echo T2, TE=50<br>ms. |
| | | High Resolution tme:<br>TR from 200 ms to 5000 ms, step 10 ms<br>TE: first echo 28 ms, second echo 90 ms<br>minimum FOV 120 mm | High Resolution tme: it is the high<br>resolution version of the standard<br>sequence Turbo Multi-echo. |
| | | minimum slice thickness 3.0 mm<br>High Resolution ge_16:<br>TR from 35 ms to 5000 ms, step 5 ms<br>TE fixed at 16 ms<br>FA from 10° to 90°, step 5°<br>minimum FOV 130 mm<br>minimum slice thickness 2 mm | High Resolution ge_16: it is a Gradient<br>Echo sequence with TE=16 msec. |
| | | High Resolution ge_stir_25:<br>TR from 150 ms to 5000 ms, step 10 ms<br>TE fixed at 25 ms<br>minimum FOV 130 mm<br>minimum slice thickness 3 mm | High Resolution ge_stir_25: it is the<br>high resolution version of the standard<br>sequence Gradient Echo STIR, TE=25<br>ms. |
| Characteristics | E-scan XQ<br>K020164 | E-scan XQ<br>modified | Comments |
| Acquisition Matrix: | l 2D FT: from 192x 128 to<br>256x256; phase encoding step 8<br>l 3D FT: from 192x128 to 256x256;<br>l slice encoding from 24 to 128,<br>l step 8; phase encoding step 8 | 2D FT for non High Resolution: from 192x 128 To perform High resolution sequences is p<br>l to 256x256; phase encoding step 8<br>12D FT for High Resolution: from 192x 128 to<br> 512x512; frequency encoding step 32, phase<br>encoding step 8<br>FT: from 192x128 to 256x256; slice p<br>3D<br>encoding from 24 to 128, step 8; phase encoding ,<br>step 8 | that the raw data matrix ,<br>necessary<br> dimensions are increased, i.e. it's <br>necessary to increase the number of<br>readout (frequency) sampling points and<br> the number of phase encoding steps. |
{3}------------------------------------------------
{4}------------------------------------------------
{5}------------------------------------------------
| Magnetic System | | | |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|
| Characteristics | E-scan XQ<br>K020164 | E-scan XQ<br>modified | Comments |
| Fringe Field (0,5<br>mT line): | X direction (horizontal) : 1,5 m front; 1,2 rear<br>Y direction (vertical) : 1,3 m<br>Z direction (horizontal right/left):<br>1.5 m | X direction (horizontal) : 1,34 m front; 1,06 rear<br>Y direction (vertical) : 1,25 m<br>Z direction (horizontal right/left): 1,29 m | More precise characterization of the data |
the submit was an a submit and the submit of the subsequence is a
{6}------------------------------------------------
| Gradients System | | | |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristics | E-scan XQ<br>K020164 | E-scan XQ<br>modified | Comments |
| Control System: | Digital, based on DSP SHARC 66<br>Mips, 132 MFlops, 0.5 Mbit<br>Memory on the chip<br>1.1515 nsec Instruction rate<br>4 independent channels (X - Y - Z<br>- Bo)<br>DAC 18 bit - updating every 7.2<br>µs<br>Ramp generation - preemphasis of<br>eddy current compensation -<br>adjustable delay | Digital electronic, based on DSP SHARC<br>21161@100 MHz, 400 MFLOPs, 128 KB On-<br>Chip SRAM<br>4 independent channels (X,Y,Z, Bo)<br>DAC 18 bit - update every 7,2 µs<br>Ramp generation -- pre-emphasis of eddy current<br>suppression - adjustable delay | Technological updating |
| Magnetic<br>compensation<br>system: | "open loop" control of the<br>magnetic field variation with<br>external measurement (AC and DC<br>probes); digital elaboration of the<br>signal in separated channels (DC -<br>50/60 Hz - 16.6. Hz) and<br>correction with Bo coil.<br>Digital electronic based on DSP<br>SHARC 66 Mips,132 MFlops, 0.5<br>Mbit Memory on the chip | "open loop" control of the magnetic field<br>variation with external measurement (AC and<br>DC probes); digital elaboration of the signal in<br>separated channels (DC - 50/60 Hz - 16.6. Hz)<br>and correction with Bo coil.<br>Digital electronic, based on DSP SHARC<br>21161@100 MHz, 400 MFLOPs, 128 KB On-<br>Chip SRAM | Technological updating |
| Characteristics | E-scan XQ<br>K020164 | E-scan XQ<br>modified | Comments |
| A/D Conversion: | baseband conversion and<br>demodulation of the RF received<br>signal in phase and quadrature<br>components<br>2 A/D converter 16 bit 78 kHz<br>sampling<br>2 digital low-pass filter for<br>analogical phase and quadrature<br>components; bandwidth: gain 1 up<br>to 0.464xfc, con fc = 78.125<br>KHz/n, n=1+10; stop-band: from<br>0.5xfc; attenuation 90 dB | For each channel:<br>3 MHz conversion of the RF received signal<br>A/D converter 14 bit 20 MHz sampling<br>digital demodulation in phase and quadrature<br>components<br>digital low-pass filter; bandwidth: gain 1 up to<br>0.43 x fc, with fc from 156.25 to 4.882 kHz;<br>stop-band: from 0.5 x fc; attenuation 90 dB | Technological updating.<br>A more detailed decription is in the the<br>section "Device modification<br>description". |
| Synthesizer: | digital, through DSP SHARC 66<br>Mips, 132 MFlops, 0.5 Mbit<br>Memory on the chip, with<br>frequency, amplitude and phase<br>modulation<br>resolutions: 1,2 Hz frequency, 256<br>levels amplitude, 1' 4º phase<br>stability : < 1 ppm into the<br>operative temperature range<br>transmission variable gain: 256<br>levels | digital, through DSP SHARC 21161@100 MHz,<br>400 MFLOPs, 128 KB On-Chip SRAM, with<br>frequency, amplitude and phase modulation<br>resolutions: 28.4 µHz frequency, 4096 levels<br>amplitude, 5' 16 " phase<br>stability : < 1 ppm into the operative temperature<br>range<br>transmission variable gain: 256 levels | Technological updating |
| Transmission<br>Coil: | linear saddle shape coil<br>IN-impedance = 50 ohm<br>passive detuning during receiving<br>through BYW29 diodes | linear saddle shape coil<br>IN-impedance = 50 ohm<br>active detuning during receiving through PIN<br>diodes -5 V 100 mA | The diodes, detuning the transmitting coil to the<br>NMR frequency during the reception, avoid that<br>the transmitting coil could receive partially the<br>NMR signal. On the other hand the lateral lobes<br>of the transmitting pulse are distorted by the<br>BYW29 diodes.<br>This effect is avoided with the introduction of the<br>PIN diodes. |
| Transmission<br>chain: | RF power amplifier until 600 W<br>pep<br>bandwidth 7÷9 MHz<br>gain stability 0.1 dB | RF power amplifier until 900 W pep<br>bandwidth 7÷9 MHz<br>gain stability 0.1 dB | The modifications of the RF amplifier<br>are described in the section "Device<br>modification description". |
| | Image Processing and Display System | | Comments |
| Characteristics | E-scan XQ | E-scan XQ | |
| | K020164 | modified | |
| Central<br>Processing Unit | ISA and PCI Bus<br>CPU Pentium III 850 MHz or faster<br>main memory: 256 MB<br>secondary cache memory: 256 KB or<br>upper | PCI bus<br>CPU Pentium IV 2.4 GHz or more<br>Main memory: 1 GB<br>Secondary cache memory: 512 KB or more | Technological updating |
| Control processor | 36 MIPS, 72 MFlops, 0.25 MB<br>memory | DSP SHARC 21161@100 MHz, 400 MFLOPs, 128 KB<br>On-Chip SRAM + 384 KB SSRAM | Technological updating |
| Acquisition and<br>reconstruction | 36 MIPS, 72 MFlops, 0.5 MB<br>memory + 128 MB DRAM | DSP SHARC 21161@100 MHz, 400 MFLOPs,<br>128 KB On-Chip SRAM + from 256 to 512 MB<br>SDRAM + 1.28 MB SSRAM | Technological updating |
| processor:<br>Hard Disk Unit | 3"1/2 hard disk; at least 20 GB, 7200 rpm | 3"1/2 hard disk; at least 40 GB, 7200 rpm | Technological updating |
| Image<br>Reconstruction<br>Matrix: | 2D: 256x256<br>3D: 256x256x8 to 256x256x128<br>step 8 | 2D: 128x128, 256x256, 512x512<br>3D: 256x256x24 to 256x256x128 step 8 | The visualization matrix of the images<br>obtained by the high resolution<br>sequences can be 128x128, 256x256 or<br>512x512 pixels. |
| Display monitor | 19" CRT Colour<br>17" TFT Colour<br>1280 x 1024 pixel<br>at least 70 Hz. not-interleaved<br>high contrast | 18.1" TFT Color<br>1280x 1024 pixel<br>not interleaved 60 Hz<br>high contrast | Technological updating |
{7}------------------------------------------------
{8}------------------------------------------------
# Image Processing and Display System
{9}------------------------------------------------
| Patient positioning | E-scan XQ | E-scan XQ | Comments |
|------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| Characteristics | K020164 | modified | |
| Cushion of the<br>rotating portion of the<br>Patient Table | Composition of the material<br>covering the cushion:<br>Wild Heather Ambla (DBL)<br>foamed PVC on a fire retardant<br>cotton support | Composition of the material covering the<br>cushion:<br>Coverlim 27 New (LIMONTA) 56% Cotton<br>44% Polyurethane | The evaluation of the new material is described in<br>the section "Biocomaaptibility Summary". |
・
・
{10}------------------------------------------------
| Characteristics | E-scan XQ<br>K020164 | Installation Area Conditions<br>E-scan XQ<br>modified | Comments |
|-----------------|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Vibrations: | < -70 dBg da 0 a 100 Hz | < -65 dBg from 0 to 100 Hz<br>warning range with standard feets: 9÷13 Hz<br>warning range with vibration-damping springs:<br>4÷7 Hz | The vibration-damping springs can be used,<br>instead of the standard feets, to install the system<br>in sites with important vibrations in the frequency of<br>range 9 ÷13 Hz for lowering the resonance<br>frequency of the magnet from 11 Hz to 5.9 Hz.<br>The final effect is the image quality enhancement<br>from the artifacts point of view, as the<br>mechanical vibrations cause a variation of the<br>magnetic field intensity and consequently ghost<br>artifacts on the images. |
and the comments of the comments of the comments of the comments of
the submit and the state of the states and the subsequence and
and the control control control control control controllers and
{11}------------------------------------------------
| Power supply | | | Comments |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|
| Characteristics | E-scan XQ<br>K020164 | E-scan XQ<br>modified | |
| Power<br>consumption: | 1100 VA during quick magnet<br>heating<br>800 VA during normal working<br>200 VA when unit is powered off<br>(thermal control on) | 1000 VA during quick magnet heating<br>800 VA during normal working<br>200 VA when unit is powered off (thermal<br>control on) | More precise characterization of the data |
:
·
.
21
and the comments of the country of
{12}------------------------------------------------
| Dimensions and weights<br>Characteristics | E-scan XQ<br>K020164 | E-scan XQ<br>modified | Comments |
|-----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Composition of<br>the device | Magnetic Unit with Patient<br>Positioning table<br>Console | Magnetic Unit with Patient Positioning table<br>Electronic box<br>Operating Table with Personal Computer | The new composition is described in the<br>section "Device Modification<br>Description". |
| Console: | Width 0.844 m; Depth 0.940 m;<br>Height 0.735 m;<br>Weight 150 Kg | Not available | |
| Electronics box: | Not available | Width 0.793 m; Depth 0.632 m; Height 1.563<br>m;<br>Weight 150 Kg | The electronics box is described in the<br>section "Device modification<br>description". |
| Personal<br>Computer: | Not available | Width 26 cm; Depth 44 cm; Height 43 cm;<br>Weight 15 Kg | The PC is described in the section<br>"Device modification description". |
| Accessories | | E-scan XQ | Comments |
| Characteristics | E-scan XQ | modified | |
| | K020164 | | |
| Operating table: | Not available | Available in two models<br>High version: width 108 cm; depth 88 cm;<br>height 74 cm;<br>weight: 78 kg<br>Comfort version: width 108 cm; depth 80 cm;<br>height 74 cm;<br>weight: 59 kg | The operating tables are described in the<br>section "Device modification<br>description". |
| Shielding box | Shielding box made of Fe<br>attenuation:<br>RF 70 dB<br>DC magnetic field 1.0<br>AC 16.6 Hz magnetic field 1.5<br>AC 50/60 Hz magn. field 2.6<br>weight: 1300 Kg<br>Shielding box made of Al<br>attenuation:<br>RF 70 dB<br>DC magnetic field 1.0<br>AC 16.6 Hz magnetic field 1.8<br>AC 50/60 Hz magn. field 4.0<br>weight: 500 Kg | Shielding box made of Al<br>attenuation:<br>RF 70 dB<br>DC magnetic field 1.0<br>AC 16.6 Hz magnetic field 1.8<br>AC 50/60 Hz magn. field 4.0<br>weight: 500 Kg | The Shielding Box made of Fe is not<br>available any more. |
| DICOM<br>server/visualizer<br>dedicated to PC | ESA-VIEW Rel. 1 | ESA-VIEW Rel.2 | ESA-VIEW Rel.2 is a DICOM<br>server/visualizer dedicated to Personal<br>Computers and is part of the DICOMed<br>family of products. It is picture archiving<br>and communications system software<br>that provides acquisition, storage,<br>transfer, and display of medical image<br>data.<br>The DICOMed family is a class II device<br>in accordance with 21CFR892.2050 -<br>Federal Register 01/04/01 and was<br>cleared via K012093. |
{13}------------------------------------------------
Note: Any deviations from original E-scan XQ specifications are substantiated within Quality System documentation at Essacte S.p.A.
.
{14}------------------------------------------------
Image /page/14/Picture/1 description: The image shows a circular logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling an abstract symbol. The symbol is a stylized representation of a human figure embracing a bird, which is meant to represent the department's mission of protecting the health of all Americans and providing essential human services.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 1 3 2003
Esaote, S.p.A. % Ms. Colleen J. Densmore The Anson Group LLC 7992 Castleway Drive INDIANAPOLIS IN 46250
#### Re: K032121
Trade/Device Name: E-Scan XQ MRI System Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device
Regulatory Class: II Product Code: 90 LNH Dated: July 8, 2003 Received: July 14, 2003
Dear Ms. Densmore:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{15}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{16}------------------------------------------------
#### Indications for Use
510(k) Number (if known): K1032121
Device Name:
Indications for Use:
E-scan XQ is a magnetic resonance (MR) system that produces transversal, sagittal and coronal and oblique cross-section images of the limbs and joints. It is intended for imaging portions of the arm, including the hand, wrist, forearm, elbow, upper arm and shoulder, and imaging portions of the leg, including the foot, ankle, calf, knee, thigh and hip.
E-scan XQ MR images correspond to the spatial distribution of protons (hydrogen nuclei) that determine magnetic resonance properties and are dependent on the MR parameters, including spin-lattice relaxation time (T1), spin-spin relaxation time (T2), nuclei density, flow velocity and "chemical shift". When interpreted by a medical expert trained in the use of MR equipment, the images can provide diagnostically useful information.
*Prescription Use* ✓
David A. Ingram
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.