K012728 · Biosound Esaote, Inc. · MOS · Aug 29, 2001 · Radiology
Device Facts
Record ID
K012728
Device Name
E-SCAN MRI SYSTEM
Applicant
Biosound Esaote, Inc.
Product Code
MOS · Radiology
Decision Date
Aug 29, 2001
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The E-scan is intended for diagnostic nuclear magnetic resonance imaging of the hip, knee, ankle, foot, shoulder, elbow, wrist, hand, calf, thigh, arm and forearm. The device produces transverse, sagittal, coronal and oblique cross-sectional images, displaying the internal structure of the limbs and joints being imaged. The images that are produced correspond to the spatial distribution of protons (hydrogen nuclei) that check the magnetic resonance properties and depend upon the MR parameters (spin-latice relaxation time (T1), spin-spin relaxation time (T2), nuclei density, flow velocity and chemical shift). If interpreted by a medical expert, these images can provide diagnostically useful information.
Device Story
E-Scan is a magnetic resonance imaging (MRI) system designed for extremity imaging. The device utilizes a multipurpose flexible receiving coil to capture magnetic resonance signals from limbs and joints (hip, knee, ankle, foot, shoulder, elbow, wrist, hand, calf, thigh, arm, forearm). The system processes these signals based on proton density, T1/T2 relaxation times, flow velocity, and chemical shift to generate cross-sectional images (transverse, sagittal, coronal, oblique). Operated by trained professionals in clinical settings, the output images are interpreted by medical experts to assess internal anatomical structures. The flexible coil design improves patient comfort, particularly for those with painful limbs or those in plaster casts, facilitating diagnostic imaging in challenging clinical scenarios.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
MRI system utilizing nuclear magnetic resonance imaging. Features a multipurpose flexible receiving coil (22 x 32.5 x 32.5 cm) to replace the standard hip coil. Operates based on proton density, T1/T2 relaxation, flow velocity, and chemical shift. Standalone diagnostic imaging system.
Indications for Use
Indicated for diagnostic MRI of hip, knee, ankle, foot, shoulder, elbow, wrist, hand, calf, thigh, arm, and forearm. For use by medical experts to visualize internal limb and joint structures.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
510(k) Summar E-Scan Biosound Esaote
# 510(k) Summary
K012728
The following safety and effectiveness summary has been prepared pursuant to requirement for 510(k) summaries specified in 21CFR¶807.92(a).
807.92(a)(1)
### Submitter Information
| Colleen Hittle, Official Correspondent<br>8000 Castleway Drive<br>Indianapolis, IN 46250 | |
|------------------------------------------------------------------------------------------|--------------------------------------------|
| Phone: | (317) 849-1916 |
| Facsimile: | (317) 577-9070 |
| Contact Person: | Colleen Hittle |
| Date: | August 6, 2001 |
| 807.92(a)(2) | |
| Trade Name: | E-Scan |
| Common Name: | Magnetic resonance diagnostic device |
| Classification Name(s): | System, Nuclear Magnetic Resonance Imaging |
| Classification Number: | 90LNH 90 MOS |
| 807.92(a)(3) | |
Predicate Device(s)
| Esaote | E-Scan K001894 |
|--------|----------------|
|--------|----------------|
Additional Substantial Equivalence Information is provided in the following substantial Equivalence Comparison Table.
{1}------------------------------------------------
510(k) Summary E-Scan Biosound Esaote
#### 807.92(a)(5)
#### Intended Use(s)
The E-scan is intended for diagnostic nuclear magnetic resonance imaging of the hip, The E-Stan Is Inchood for dagliourist, hand, calf, thigh, arm and forearm The device Knee, ankle, foot, shoulder, cloow, wils, hand, edit, sectional images, displaying the internal structure of the limbs and joints being imaged. The images that are produced correspond to the spatial distribution of protons (hydrogen nuclei) that check the magnetic resonance properties and depend upon the MR parameters (spin-latice relaxation time (T1), spin-spin relaxation time (T2), nuclei density, flow velocity and chemical shift). If interpreted by a medical expert, these images can provide diagnostically useful information.
807.92(a)(6)
#### Technological Characteristics
The Multipurpose Flexible Receiving Coil improves the system performance and does not alter the fundamental scientific technology of the cleared device.
{2}------------------------------------------------
3 Fork) Summar.
E-Sean
Biosound Esaote
| E-scan<br>With Multipurpose<br>Flexible Receiving<br>Coil | Comments |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Shoulder coil No.1:<br>17.5 x 12.6 x 14.5 cm<br>(w x d x h)<br>Shoulder coil No. 5:<br>16.8 x 7.9 x 15.5 cm<br>(w x d x h)<br>Multipurpose flexible coil:<br>22 x 32.5 x 32.5<br>(short axis x long axis) | The Multipurpose flexible coil substitutes<br>completely the Hip Coil. For its<br>characteristics the flex coil can be used to<br>exam the shoulder and the knee too,<br>especially for patient with painful or in<br>plaster limb.<br>See example for the Flexible body Coil<br>(hip and other regions) – 20x150 cm (wxd)<br>- of the AIRIS HITACHI (K945155) |
## Substantial Equivalence Comparison Table
{3}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
BioSound Esaote, Inc. % Ms. Colleen J. Hittle The Anson Group 7992 Castleway Drive Indianapolis, Indiana 46250 AUG 2 9 2001
-und Esaote, inc.
- Colleen J. Hittle
_nson Group
Castleway Drive
anolis Indiana 46250
Re: K012728 E-Scan MRI System (Receiving Coil) Dated: August 6, 2001 Received: August 15, 2001 Regulatory Class: II 21 CFR 892.1000/Procode: 90 MOS
#### Dear Ms. Hittle:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we device is we nave revel your section 710(s) nothedited with the enclosure) to legally marketed predicate devices marketed in interstate subsancery of the monenent to the notenent of the Medical Device Amendments, or to devices that have been reclassified in commerce with the provisions of the Federal Food, Divg, and Cosmetic Act (Act). You may, therefore, market the device, subject to ation and white provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to succ If your device to classinot (coo no vegulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalion assumes compliance with the Current Good Manufacturing Practice requirements, 800 to 07.1 A substantial.) equality System Regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may Qs illsponolis, all 1 od alle Drag Pramination (Slish futher announcements concerning your device in the Federal Register. Please result it tegulatory action. In addition submission does not affect any obligation you might have under sections \$31 through note: and response to your premaine Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marked predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
> (301) 594-4591 8xx.1xxx 876.2xxx, 3xxx, 4xxx, 5xxx (301) 594-4616 (301) 594-4616 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx (301) 594-4654 892.2xxx, 3xxx, 4,xxx, 5xxx (301) 594-4692 Other
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
### Indications For Use
510(k) Number (if known):
Kc/2728
Device Name:
E-Scan
Indications for Use:
The E-scan is intended for diagnostic nuclear magnetic resonance imaging of the hip, knee, ankle, foot, The E-scan is included for dagatosto nacious naments of the device produces transverse, sagittal, soronal and oblique cross-sectional images, displaying the internal structure of the limbs and joints being imaged. The images that are produced correspond to the spatial distribution of protons (hydrogen nuclei) that check the magnetic resonance properties and depend upon the MR parameters (spin-lattice relaxation that check the naginette resoluntion properties cle depsity, flow velocity and chemical shift). If interpreted by a medical expert, these images can provide diagnostically useful information.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use _
OR
Over-The-Counter Use
Nancy Choodon
(Division Sign-Off)
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K012728
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.