MediSurf is intended to facilitate access to clinical images and information in distributed locations. Along with image processing capabilities, MediSurf can be used by radiologists for on-call teleradiology over "intranets", Internet, direct or dial-up lines, and for providing second opinion services. It can be also used by physicians for enhanced interpretation of cases through the inclusion of images in imaging related reports and in patient records in general. Similarly it can be used by healthcare organization staff for facilitated access to clinical images throughout the healthcare organization.
Device Story
MediSurf is a teleradiology server acting as an access engine for clinical images. It retrieves DICOM images from servers/networks using adaptive progressive compression for efficient transfer over intranets, Internet, or dial-up lines. The system provides software applets for display, processing, storage, and transmission of clinical data. Operated by radiologists, physicians, and healthcare staff in clinical settings, it enables remote interpretation and second opinions. Users interact with the server via a remote Web browser (e.g., Netscape Navigator, Microsoft Explorer) using Java. The device facilitates clinical decision-making by providing diagnostic-quality images for reports and patient records, improving access to clinical information across distributed healthcare environments.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological and functional comparison to the predicate device.
Technological Characteristics
Standard off-the-shelf UNIX-based workstation. Connectivity via DICOM networks, intranets, Internet, and dial-up lines. Features adaptive progressive compression for image transfer. Software includes archiving, communication, and image processing applets. User interface utilizes Java-based Web browsers.
Indications for Use
Indicated for radiologists, physicians, and healthcare staff to access, display, process, store, and transmit clinical images and information for teleradiology, second opinions, and clinical reporting. No specific age or gender restrictions; intended for professional clinical use.
Regulatory Classification
Identification
A medical image communications device provides electronic transfer of medical image data between medical devices. It may include a physical communications medium, modems, and interfaces. It may provide simple image review software functionality for medical image processing and manipulation, such as grayscale window and level, zoom and pan, user delineated geometric measurements, compression, or user added image annotations. The device does not perform advanced image processing or complex quantitative functions. This does not include electronic transfer of medical image software functions.
{1}
Algotec Systems Ltd.
MediSurf Premarket Notification
K971347
# 2 Summary of Safety and Effectiveness
This summary of 510(k) Safety and Effectiveness information is being submitted in accordance with the requirements of the SMDA 1990 and 21 CFR 807.92.
## Submitter :
JUL - 3 1997
Algotec Systems Ltd.
4 Hamelacha St.
P.O. Box 2408
Industrial Zone, Ra'anana
ISRAEL 43000
Tel: +972-9-748-2442
Fax: +972-9-748-2411
**Name of the Device :** MediSurf.
**Predicate Device :** ProVision Diagnostic Workstation manufactured by Algotec Systems Ltd. (K954678).
**Description of the Device :** The MediSurf server is an access engine to healthcare information. MediSurf provides its end-users with easy access to DICOM devices and networks for the retrieval of images in full diagnostic quality using adaptive progressive compression for efficient transfer over different communication links. The server also provides the user with the software applets needed to display, process, store, and send the retrieved clinical data.
The MediSurf server is based on off-the-shelf UNIX based computer that complies with the accepted international standards for computer systems. The system also comprises software developed and validated by Algotec Systems Ltd..
**Intended Use :** MediSurf is intended to facilitate access to clinical images and information in distributed locations. Along with image processing capabilities, MediSurf can be used by radiologists for on-call teleradiology over "intranets", Internet, direct or dial-up lines, and for providing second opinion services. It can be also used by physicians for enhanced interpretation of cases through the inclusion of images in imaging related reports and in patient records in general. Similarly it can be used by healthcare organization staff for facilitated access to clinical images throughout the healthcare organization.
**Comparison of Technological Characteristics :** MediSurf and its predicate device, the ProVision Diagnostic Workstation, share many implementation and functional characteristics:
- Both systems are implemented on a standard off-the-self UNIX based workstation.
- Both systems share many common software components such as Archiving and Communication subsystems. Most of the MediSurf software was originally developed for the ProVision workstation.
2
{2}
Algotec Systems Ltd.
MediSurf Premarket Notification
- Both systems provide standard image viewing tools such as: window/level, zoom and pan. No tools are implemented in the MediSurf, that do not exist in the ProVision.
- Both systems provide access to any standard DICOM image server.
There are no new features affecting safety or effectiveness in the MediSurf that do not exist in the ProVision. There is, however, a single technical difference between the user interface of the two systems. The user interaction with the ProVision is facilitated by using the system’s mouse, keyboard, monitor and dedicated interface software. The interaction with MediSurf, however, is facilitated by using a remote Web browser running Java, such as Netscape’s Navigator or Microsoft’s Explorer. As a result, the MediSurf has slower interaction with the user but all the other standard viewing capabilities are kept.
The differences between the two systems raise no new issues of safety or effectiveness.
April 7, 1997
Date

Dr. Menashe Benjamin, President
Signature, Title
{3}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
Algotec Systems, Ltd.
c/o Dr. Eli Orbach
International Regulatory Consultants
P.O. Box 6718
Efrat 90435
ISRAEL
JUL - 3 1997
Re: K971347
MediSurf (Teleradiology Device)
Dated: April 7, 1997
Received: April 10, 1997
Unclassified/Procode: 90 LMD
Dear Dr. Orbach:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Enclosure
{4}
P. 2
510(k) Number (if known) K971347
Device Name MediSurf
Indications For Use:
MediSurf is intended to facilitate access to clinical images and information in distributed locations. Along with image processing capabilities, MediSurf can be used by radiologists for on-call teleradiology over "intranets", Internet, direct or dial-up lines, and for providing second opinion services. It can also be used by physicians for enhanced interpretation of cases through the inclusion of images in imaging related reports and in patient records in general. Similarly it can be used by healthcare organization staff for facilitated access to clinical images throughout the healthcare organization.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use ☑ (Per 21 CFR 801.109)
OR
Over The Counter Use ☐ (Optional Format 1-2-90)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.