The Veuron-Brain-pAb is a software for the registration, fusion, display and analysis of medical images from multiple modalities including MRI and PET. The software aids clinician in the assessment and quantification of pathologies from PET Amyloid scans of the human brain. It enables automatic analysis and visualization of amyloid protein concentration through the calculation of standard uptake volume ratios (SUVR) within target regions of interest and comparison to those within the reference regions. The software is deployed via medical imaging workplaces and is organized as a series of workflows which are specific to use with radio-tracer and disease combinations.
Device Story
Veuron-Brain-pAb is standalone software for medical imaging workplaces; used by clinicians to process MRI and PET brain scans. Device imports DICOM/NiFTI data; performs automatic registration, fusion, and visualization of amyloid protein concentration. Core function involves calculating Standardized Uptake Value Ratios (SUVR) within target regions of interest compared to reference regions. Output includes visual displays and quantitative reports (.csv/PNG) to assist clinicians in assessing dementia-related pathologies. Device facilitates clinical decision-making by providing objective quantification of amyloid uptake, aiding in diagnostic accuracy. Used in hospital settings.
Clinical Evidence
Bench testing only. Performance testing verified system functionality and measurement tool accuracy against pre-defined requirements. No clinical data provided.
Technological Characteristics
Standalone software; Windows 10 64-bit OS. Supports DICOM and NiFTI input formats. Performs automated image registration, fusion, and SUVR calculation. Complies with ISO 14971, IEC 62304, and IEC 62366 standards. Data export via USB.
Indications for Use
Indicated for use by trained medical professionals to assist in the assessment and quantification of pathologies from PET Amyloid brain scans in patients undergoing dementia evaluation.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
K221405 — Neurophet SCALE PET · Neurophet., Inc. · Aug 5, 2022
K233620 — MIM - Centiloid Scaling · Mim Software, Inc. · May 20, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
January 15, 2021
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Heuron Co., Ltd. % Mr. Edward Park CEO LightenBridge LLC 4408 Tortuga Lane MCKINNEY TX 75070
Re: K203142
Trade/Device Name: Veuron-Brain-pAb Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: October 20, 2020 Received: October 20, 2020
Dear Mr. Park:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K2O3142
Device Name Veuron-Brain-pAb
Indications for Use (Describe)
The Veuron-Brain-pAb is a software for the registration, fusion, display and analysis of medical images from multiple modalities including MRI and PET. The software aids clinician in the assessment and quantification of pathologies from PET Amyloid scans of the human brain. It enables automatic analysis and visualization of amyloid protein concentration through the calculation of standard uptake volume ratios (SUVR) within target regions of interest and comparison to those within the reference regions. The software is deployed via medical imaging workplaces and is organized as a series of workflows which are specific to use with radio-tracer and disease combinations.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="font-weight: bold;"> <span style="text-decoration: underline;"></span> </span> </span> </span> <span style="font-family: Wingdings;"> <span style="font-size: 10pt;"> <span style="font-weight: bold;"> <span style="text-decoration: underline;"> <span style="color: black;">☑</span> </span> </span> </span> </span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="font-weight: bold;"> <span style="text-decoration: underline;"></span> </span> </span> </span> <span style="font-family: Wingdings;"> <span style="font-size: 10pt;"> <span style="font-weight: bold;"> <span style="text-decoration: underline;"> <span style="color: black;">☐</span> </span> </span> </span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for "Heuron". The logo consists of a symbol on the left that resembles a network of interconnected nodes, possibly representing neurons or connections. To the right of the symbol is the word "heuron" in a stylized, sans-serif font. The color of both the symbol and the text is a muted gray.
# 510(k) Summary K203142
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR §807.92.
Submitter Information
| Submitter Name: | HEURON CO., LTD. |
|-----------------------|--------------------------------------------------------------------------|
| Address: | 10th Floor, 7, Mirae-ro, Namdong-gu, Incheon, 21558 Republic of<br>Korea |
| Phone/Fax | Tel. 82-32-429-8508 / Fax. 82-32-429-8507 |
| Contact Person: | Edward Park, official correspondent of Heuron Co., Ltd. |
| Date of submission: | Sep 01, 2020 |
| Device Information | |
| Proprietary Name(s): | Veuron-Brain-pAb |
| Common Name: | Image Processing Software |
| Classification Name: | Picture Archiving and Communication System per 21 CFR 892.2050 |
| Product Code: | LLZ |
| Classification Panel: | Radiology |
| Device Class: | II |
### Device Description
The Veuron-Brain-pAb is stand-alone software to automatically calculate the "Standardized Uptake Value Ratio (SUVR)" for quantitative analysis of amyloid PET. The calculated result supports the accuracy of the diagnosis of patients' dementia for the medical professional. It also helps with accurate visual interpretation through visualization functions. Various amyloid PET images can be processed by providing a variety of options for users to choose in the image process.
### Predicate Device
< Primary predicate device >
- Scenium 2.0 (Siemens Medical Solutions USA, Inc. K121074, 06/08/2012) ●
< Secondary device >
- . MIM 4.0 (NEURO) (MlMvista Corp. K060816, 05/16/2006)
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Heuron. The logo consists of a stylized graphic to the left of the word "heuron". The graphic is made up of a grid of circles connected by lines, resembling a neural network. The word "heuron" is written in a rounded, sans-serif font.
## Intended Use
The Veuron-Brain-pAb is a display and analysis software to aid the Clinician in the assessment and quantification of pathologies taken from amyloid PET scans. The software is deployed via medical imaging workplaces and is organized as a series of workflows which are specific to use with particular drug and disease combinations. The software aids in the assessment of human brain scans enabling automated analysis through quantitative analysis of amyloid PET. It facilitates comparison with those different regions and calculation of standard uptake value ratios between regions of interest on the medical image.
#### Comparison of Indications for Use Statements
The following table compares Indications for Use Statements between the Veuron-Brain-pAb and the two predicate devices. i.e. Scenium 2.0 and MIM 4.0 (NEURO). Similar to the other predicate devices, the Veuron-Brain-pAb is used for managing patient and case base data, collection, analysis, fusion, and display of the medical images. The device assists the clinician with the visual evaluation, assessment and quantification of pathologies derived from brain sans.
| Veuron-Brain-pAb | Simens Scenium 2.0 | MIMvista Corp. MIM 4.0<br>(NEURO) |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Proposed device | Primary predicate device<br>(K121074) | Secondary predicate device<br>(K060816) |
| The Veuron-Brain-pAb is a<br>software for the registration,<br>fusion, display and analysis<br>of medical images from<br>multiple modalities<br>including MRI and PET.<br>The software aids clinician<br>in the assessment and<br>quantification of pathologies<br>from PET Amyloid scans of<br>the human brain. It enables<br>automatic analysis and<br>visualization of amyloid<br>protein concentration<br>through the calculation of<br>standard uptake volume<br>ratios (SUVR) within target<br>regions of interest and<br>comparison to those within<br>the reference regions. The | The Scenium display and<br>analysis software has been<br>developed to aid the Clinician<br>in the assessment and<br>quantification of pathologies<br>taken from PET and SPECT<br>scans.<br>The software is deployed via<br>medical imaging workplaces<br>and is organized as a series of<br>workflows which are specific<br>to use with particular drug<br>and disease combinations.<br>The software aids in the<br>assessment of human brain<br>scans enabling automated<br>analysis through<br>quantification of mean pixel<br>values located within<br>standard regions of interest. It | MIM 4.0 (NEURO) is a<br>software package that<br>provides the physician with<br>the<br>means to display, register and<br>fuse medical images from<br>multiple modalities.<br>Additionally, it evaluates<br>cardiac left ventricular<br>function and perfusion<br>including left ventricular end-<br>diastolic volume, end-systolic<br>volume, and ejection fraction.<br>The Region of Interest (ROI)<br>feature reduces the time<br>necessary for the physician to<br>define objects in medical<br>image volumes by providing<br>an initial definition of object<br>contours. The objects include |
| software is deployed via | facilitates comparison with | |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Heuron. The logo consists of a stylized graphic to the left of the word "Heuron". The graphic is made up of a series of interconnected circles, resembling a neural network. The word "Heuron" is written in a rounded, sans-serif font.
| Veuron-Brain-pAb | Siemens Scenium 2.0 | MIMvista Corp. MIM 4.0<br>(NEURO) |
|------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Proposed device | Primary predicate device<br>(K121074) | Secondary predicate device<br>(K060816) |
| medical imaging workplaces<br>and is organized as a series<br>of workflows which are<br>specific to use with radio-<br>tracer and disease<br>combinations. | existing scans derived from<br>FDG-PET, amyloid-PET, and<br>SPECT studies and<br>calculation of uptake ratios<br>between regions of interest. | but are not limited to tumors<br>and organs.<br>MIM 4.0 (NEURO) also aids<br>the physician in the<br>assessment of PET/SPECT<br>brain<br>scans. It provides automated<br>quantitative and statistical<br>analysis by automatically<br>registering PET/SPECT brain<br>scans to a standard template<br>and comparing intensity<br>values to a reference database<br>or to other PET/SPECT scans<br>on a voxel by voxel basis,<br>within stereotactic surface<br>projections, or within<br>standardized<br>regions of interest. |
## Summary of the technological characteristics compared to the predicate device new device
The software is similar in uses and applications to the predicate devices. As stated in the comparison table provided below, both the device and predicates are used to assist the Clinician with the visual evaluation, assessment and quantification of pathologies derived from brain scans.
| Comparison<br>Item | Veuron-Brain-pAb | Simens<br>Scenium 2.0<br>(Primary predicate) | MIMvista Corp.<br>MIM4.0 (NEURO)<br>(2nd predicate) |
|-------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Target Users | Trained Medical<br>professionals | Trained Medical<br>professionals | Trained Medical<br>professionals |
| Anatomical<br>Site | Brain | Brain | Brain |
| Where Used | Hospital | Hospital | Hospital |
| Design<br>Features | Import DICOM data<br>Perform automatic post-<br>processing | Import DICOM data<br>Perform automatic post-<br>processing | Import DICOM data<br>Perform automatic post-<br>processing |
| Comparison Item | Veuron-Brain-pAb | Simens Scenium 2.0 (Primary predicate) | MIMvista Corp. MIM4.0 (NEURO) (2nd predicate) |
| | Provide the user confirmation<br>Export the resulting data only through USB | Provide the user confirmation<br>Export the resulting data through data network | Provide the user confirmation<br>Export the resulting data only through data network |
| OS | Windows 10, 64-bit | Windows 7, 64-bit | Windows 2000/XP |
| Non-clinical Performance | Bench testing performed to test the functionality of the system and measurement tools. | Bench testing performed to test the functionality of the system and measurement tools. | Bench testing performed to test the functionality of the system and measurement tools. |
| Standards | • ISO 14971<br>• IEC 62304<br>• IEC 62366 | • ISO 14971<br>• IEC 62304 | • ISO 14971<br>• IEC 62304 |
| SW verification and validation | Tested in accordance with verification and validation processes and planning. The testing results support that all the system requirements have met their acceptance criteria and are adequate for its intended use. | Tested in accordance with verification and validation processes and planning. The testing results support that all the system requirements have met their acceptance criteria and are adequate for its intended use. | Tested in accordance with verification and validation processes and planning. The testing results support that all the system requirements have met their acceptance criteria and are adequate for its intended use. |
| Compatible Input Data Format and Modality | DICOM & NiFTI<br>PET, MRI | DICOM<br>PET, SPECT, MRI, CT | DICOM<br>PET, SPECT, MRI, CT |
| Input Patient Data | Manual through keyboard/mouse | Manual through keyboard/mouse | Manual through keyboard/mouse |
| Output Patient Data Format | Pictures: PNG<br>Report: .csv | Pictures: JPEG<br>Report: DICOM | |
| Study list functionality | Search, Importing, Exporting | Search, Importing, Exporting, Deleting | Search, Importing, Exporting |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Heuron. The logo consists of two parts: a symbol on the left and the word "heuron" on the right. The symbol is made up of interconnected circles, resembling a network or a neural connection. The word "heuron" is written in a rounded, sans-serif font, with the "h" in a lighter shade of gray than the rest of the letters.
## Non-Clinical Study performance
Bench testing is done to show that the system is suitable for its intended use and that the measurement tool performance meets its pre-defined requirements. This did not reveal any
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "Heuron" in a stylized font, with a logo to the left of the word. The logo consists of a network of interconnected circles, resembling a neural network or a cluster of neurons. The word "Heuron" is written in lowercase letters, with a rounded and modern typeface.
Heuron Co., Ltd. 10th Floor, 7, Mirae-ro, Namdong-gu Incheon, 21558 Republic of Korea Tel. +82-32-429-8508 / Fax. +82-32-429-8507
issues with the system, demonstrating that the performance of Veuron-Brain-pAb is as safe and effective as its predicate devices.
#### Conclusion
By virtue of its intended use, design features, and technological characteristics, Veuron-BrainpAb is substantially equivalent to a device that has been approved for marketing in the United States. The non-clinical performance data shows that Veuron-Brain-pAb is as safe and effective as the predicate devices without raising any new safety and/or effectiveness concerns.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.