SCENIUM 2.0

K121074 · Siemens Medical Solutions USA, Inc. · LLZ · Jun 8, 2012 · Radiology

Device Facts

Record IDK121074
Device NameSCENIUM 2.0
ApplicantSiemens Medical Solutions USA, Inc.
Product CodeLLZ · Radiology
Decision DateJun 8, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

The Scenium display and analysis software has been developed to aid the Clinician in the assessment and quantification of pathologies taken from PET and SPECT scans. The software is deployed via medical imaging workplaces and is organized as a series of workflows which are specific to use with particular drug and disease combinations. The software aids in the assessment of human brain scans enabling automated analysis through quantification of mean pixel values located within standard regions of interest. It facilitates comparison with existing scans derived from FDG-PET, amyloid-PET, and SPECT studies and calculation of uptake ratios between regions of interest.

Device Story

Scenium 2.0 is post-processing software for medical imaging workplaces; used by clinicians to visualize and analyze PET and SPECT brain scan data. Input consists of multimodality image data; software performs automated quantification of mean pixel values within standard regions of interest (ROIs). It calculates uptake ratios between ROIs and facilitates comparison with existing patient scans. Output is displayed to the clinician to assist in the assessment of pathologies and support clinical decision-making regarding drug and disease combinations. The device does not control scanner hardware.

Clinical Evidence

No clinical data provided. The submission relies on bench testing and compliance with established standards (ISO 14971, ISO 62304) and regulations (21 CFR 820, 892.1200, 892.2050) to demonstrate safety and effectiveness.

Technological Characteristics

Post-processing software for medical imaging workstations. Features include automated ROI quantification and uptake ratio calculation. Compliant with ISO 14971 (risk management) and ISO 62304 (software lifecycle). Operates on standard medical imaging hardware platforms.

Indications for Use

Indicated for clinicians to aid in the assessment and quantification of pathologies in human brain PET and SPECT scans. Supports FDG-PET, amyloid-PET, and SPECT studies.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K121074 JUN - 8 2012 Scenium 2.0 510(k) Premarket Notification # 510(k) Summary as required by 21 CFR Part 807.87(h) Submitter: Elaine Chang Regulatory Technical Specialist Siemens Medical Solutions USA, Inc. 810 Innovation Dr Knoxville, TN 37932 USA Telephone Number: Fax Number: Name / Address of Manufacturer (865) 218-3019 (865) 218-2873 Siemens Medical Solutions USA, Inc Molecular Imaging 2501 N. Barrington Road Hoffman Estates, IL 60192 USA Date of Submission: April 6, 2012 Scenium 2.0 Identification of the product Device Proprietary Name: Common Name: Classification Name: Picture Archiving and Communication System Picture Archiving and Communication System per 21 CFR 892.2050 Emission Computed Tomography System per 21 CFR 892.1200 Product Code: Classification Panel: Device Class: Radiology LLZ and KPS Class II {1}------------------------------------------------ Scenium 2.0 510(k) Premarket Notification ## Marketed Devices to which Equivalence is claimed | Device | Manufacturer | 510(k) Number | |---------------|--------------------------------|---------------| | Scenium 1.1 | Siemens Medical Solutions, LTD | K061545 | | Neuro Trans3D | Segami Corporation | K043441 | | Brass | Hermes Medical Solution | K021656 | ### Device Description: Scenium 2.0 display and analysis software enables visualization and appropriate rendering of multimodality data, providing a number of features which enable the user to process the acquired image data. Scenium 2.0 is post processing and does not control the scanning features of the system. ### Indications for Use: The Scenium display and analysis software has been developed to aid the Clinician in the assessment and quantification of pathologies taken from PET and SPECT scans. The software is deployed via medical imaging workplaces and is organized as a series of workflows which are specific to use with particular drug and disease combinations. The software aids in the assessment of human brain scans enabling automated analysis through quantification of mean pixel values located within standard regions of interest. It facilitates comparison with existing scans derived from FDG-PET, amyloid-PET, and SPECT studies and calculation of uptake ratios between regions of interest. #### Technological characteristics: The software is similar in uses and applications to the predicate devices. Both the device and predicates are used to assist the Clinician with the visual evaluation, assessment and quantification of pathologies derived from brain scans. #### Safety and Effectiveness: The device is designed and manufactured under Quality System Regulations as outlined in 21 CFR 820. All requirements of Emission Computed Tomography system standards (21 CFR 892.1200) and Picture Archiving and Communications System (21 CFR 892.2050) are met, and software is in compliance with ISO 14971 and ISO 62304. #### Substantial Equivalence: {2}------------------------------------------------ Scenium 2.0 510(k) Premarket Notification Based on the above considerations, Siemens Medical Solutions USA, Inc believes that the Scenium 2.0 software is substantially equivalent to the predicate devices. The device and the predicate devices are all post-processing and provide similar features of visualization and numerical data. {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three stripes representing the department's mission of promoting health, well-being, and human services. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. ## DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002 Ms. Elaine Chang Regulatory Technical Specialist Siemens Medical Solutions USA, Inc. 810 Innovation Drive KNOXVILLE TN 37932 Re: K121074 Trade/Device Name: Scenium 2.0 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ and KPS Dated: April 6, 2012 Received: April 9, 2012 ## Dear Ms. Chang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device; subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of JUN - 8 2012 {4}------------------------------------------------ medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 11 you desire openers the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely Yours, Janine M. Morris Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K121074 Device Name: Scenium 2.0 Indications for Use: The Scenium display and analysis software has been developed to aid the Clinician in the assessment and quantification of pathologies taken from PET and SPECT scans. The software is deployed via medical imaging workplaces and is organized as a series of workflows which are specific to use with particular drug and disease combinations. The software aids in the assessment of human brain scans enabling automated analysis through quantification of mean pixel values located within standard regions of interest. It facilitates comparison with existing scans derived from FDG-PET, amyloid-PET, and SPECT studies and calculation of uptake ratios between regions of interest. Prescription Use X (Part 21 CFR 801 Subpart D) Over the Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) · OR Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-O, Division of Radiological Device: Office of In Vitro Diagnostic Device Evaluation 47 510K K1a1074 Page 1 of
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...