K183038 · Corin USA · LLZ · Nov 26, 2018 · Radiology
Device Facts
Record ID
K183038
Device Name
Corin Optimized Positioning System (OPS) Plan
Applicant
Corin USA
Product Code
LLZ · Radiology
Decision Date
Nov 26, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The Corin Optimized Positioning System Plan (OPS™ Plan) is indicated for pre-operative planning for primary total hip arthroplasty. OPS™ Plan is intended to be used as a pre-operative tool to assist the Surgeon in the selection (for example, lateralised or standard stem), sizing and positioning of components of primary total hip arthroplasty. The Corin OPS™ Plan is intended to be used with the Corin TriFit TS (K121563, K153772), MetaFix (K082525, K121439, K131952, K153731), TaperFit (K142761, K153725), MiniHip (K083312, K11046, K131986) and TriFit CF (K173880) Femoral Hip Stems.
Device Story
Pre-operative templating tool for primary total hip arthroplasty; assists surgeons in implant selection, sizing, and positioning. Inputs: patient CT scans and AP X-rays in DICOM format. Process: data transferred via OPS™ Portal to manufacturer; Corin Simulation Engineer performs image processing, identifies anatomical landmarks, and determines implant positioning; Surgeon reviews and accepts generated OPS™ Plan Report and Appendix. Output: static report detailing implant selection and placement. Used in clinical settings; operated by manufacturer engineers and surgeons. Benefits: provides patient-specific pre-operative planning to optimize implant fit and positioning.
Clinical Evidence
No clinical data. Bench testing performed to verify accuracy, repeatability, and reproducibility of implant selection, sizing, and placement, and to assess system usability.
Technological Characteristics
Image processing software for pre-operative planning. Utilizes CT and X-ray imaging data. Generates 2D patient-specific bone models. System is networked via OPS™ Portal for data transfer. Software-based analysis performed by simulation engineers.
Indications for Use
Indicated for pre-operative planning for primary total hip arthroplasty. Intended to assist surgeons in selection, sizing, and positioning of femoral hip stem components.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
Corin Optimized Positioning System (OPS™) Plan (K171847)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA acronym is in blue, followed by the words "U.S. Food & Drug Administration" in blue.
November 26, 2018
Corin USA Martina Cecconi Regulatory and Clinical Affairs Team Leader Corin (Australia) 17 Bridge Street PYMBLE, NEW SOUTH WALES 2073 AUSTRALIA
Re: K183038
Trade/Device Name: Corin Optimized Positioning System (OPS) Plan Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ, LZO, MEH Dated: October 30, 2018 Received: November 1, 2018
Dear Martina Cecconi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Michael D. O'Hara
Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### 2. INDICATION FOR USE
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
K183038
Device Name
Corin Optimized Positioning System (OPS™) Plan
Indications for Use (Describe)
The Corin Optimized Positioning System Plan (OPS™ Plan) is indicated for pre-operative planning for primary total hip arthroplasty. OPS™ Plan is intended to be used as a pre-operative tool to assist the Surgeon in the selection (for example, lateralised or standard stem), sizing and positioning of components of primary total hip arthroplasty.
The Corin OPS™ Plan is intended to be used with the Corin TriFit TS (K121563, K153772), MetaFix (K082525, K121439, K131952, K153381), TaperFit (K142761, K153725), MiniHip (K083312, K11046, K131986) and TriFit CF (K173880) Femoral Hip Stems.
| <b>Type of Use (Select one or both, as applicable)</b> |
|--------------------------------------------------------|
|--------------------------------------------------------|
| <div> <span> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
|----------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
FORM FDA 3881 (7/17)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
| 1. | Applicant/Sponsor<br>Distributor: | Corin USA Limited<br>Distributor<br>12750 Citrus Park Lane<br>Suite 120<br>Tampa, Florida 33625<br>Establishment Registration No.: 1056629 |
|----|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. | Manufacturer: | Optimized Ortho Pty Ltd<br>17 Bridge Street<br>Pymble NSW<br>2073 Australia<br>Establishment Registration No: 3012916784 |
| 3. | Contact Person: | Martina Cecconi<br>Regulatory and Clinical Affairs Team Leader<br>Corin (Australia) Pty Limited<br>Martina.Cecconi@coringroup.com<br><br>Lucinda Gerber<br>Global Regulatory Affairs Manager<br>Corin USA Limited<br>1 (772) 321-2478<br>Lucinda.Gerber@coringroup.com |
- 4. Date: Tuesday, 30 October 2018
- 5. Trade Name: Corin Optimized Positioning System (OPS™) Plan
- 6. Common Name: Image Processing Software
# 7. Classification Product Code(s): LLZ
Additional Product Code(s): LZO, MEH
# 8. Classification Name:
21 CFR 892.2050 – Picture archiving and communications system
Additional Classification Names:
21 CFR 888.3353 – Hip joint metal/ceramic/polymer semi-constrained cemented or
{4}------------------------------------------------
nonporous uncemented prosthesis.
21 CFR 888.3353 – Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis.
# 9. Substantially Equivalent (predicate) device(s):
- . Corin Optimized Positioning System (OPS™) Plan (K171847)
## 10. Device Description:
The Corin Optimized Positioning System™ Plan (OPS™ Plan) is a pre-operative templating tool to assist the Surgeon in the planning of primary total hip arthroplasty using any Corin femoral stem. The Corin OPS™ Plan assists in selection (for example, lateralised or standard stem), sizing and placement of implant components (stem, head, cup) using anatomical landmarks of the femur and pelvis obtained from pre-operative CT scan and AP X-ray. The OPS™ Plan consists of the OPS™ Plan Report and the OPS™ Plan Appendix. The OPS™ Plan report is a static report which displays the selection (for example, lateralised or standard stem), sizing and placement of implant components. OPS™ Plan Appendix presents the steps taken by the Engineer on behalf of the Surgeon during the preoperative planning process, which include the identification of anatomical landmarks as well as sizing and positioning of the implants.
The Corin OPS™ Plan process begins with booking the OPS case to final release of the OPS Plan Report. OPS™ case booking details and required CT and X-ray imaging in DICOM format are electronically transferred to the manufacturer via the online end-user interface, the OPS™ Portal. Image processing is completed by an Corin Simulation Engineer with results of the OPS™ analysis provided in the OPS™ Plan Report. The Surgeon does not actively interact with the image processing, implant positioning and/or report generation. Final acceptance of the OPS™ Plan by the Surgeon is confirmed following any modifications or clarifications of the reported results.
The OPS™ Plan was originally cleared in K171847. The purpose of this submission is to include TriFit CF (K173880) to the Femoral Hip systems intended to be used with the OPS™ Plan.
The design and all the other compatible components for use with the OPS™ Plan subject of this submission are identical to that of the predicate device K171847.
{5}------------------------------------------------
## 11. Indications for Use / Intended Purpose:
The Corin Optimized Positioning System Plan (OPS™ Plan) is indicated for preoperative planning for primary total hip arthroplasty. OPS™ Plan is intended to be used as a pre-operative tool to assist the Surgeon in the selection (for example, lateralised or standard stem), sizing and positioning of components of primary total hip arthroplasty.
The Corin OPS™ Plan is intended to be used with the Corin TriFit TS (K121563, K153772), MetaFix (K082525, K121439, K131952, K153381), TaperFit (K142761, K153725), MiniHip (K083312, K11046, K131986) and TriFit CF (K173880) Femoral Hip Stems.
## 12. Summary of Technologies/Substantial Equivalence:
The Corin OPS™ Plan subject of this submission is identical to the predicate Corin OPS™ Plan (K171847) in design, technological characteristics, principles of operation and indication of use, with both the subject device and predicate device utilizing CT scans to generate 2D patient specific bone models and X-Ray imaging to identify bony landmarks. The end user and site of operation (hip) is identical for both the subject device and predicate device. Based on these similarities, Corin believes that the Corin OPS™ Plan is substantially equivalent to the predicate device.
#### 13. Non-Clinical Testing:
Non-clinical testing was performed to assess the safety and effectiveness of the device, to demonstrate the processing of patient imaging to produce accurate, repeatable and reproducible implant selection, sizing and placement of components, and the usability of the OPS™ Plan that is provided to the Surgeon. Testing verified that the accuracy and performance of the system is adequate to perform as intended.
### 14. Clinical Testing:
Clinical testing was not necessary for this Traditional 510(k).
#### 15. Conclusion:
The subject device has similar intended uses and indications, technological characteristics, and principles of operation as its predicate device. The differences between the device and its predicate devices raise no new issues in terms of safety or effectiveness.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.