K141480 · Circle Cardiovascular Imaging, Inc. · LLZ · Aug 22, 2014 · Radiology
Device Facts
Record ID
K141480
Device Name
CVI42
Applicant
Circle Cardiovascular Imaging, Inc.
Product Code
LLZ · Radiology
Decision Date
Aug 22, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
Indications for Use
cvi42 vascular analysis add-on is an image analysis software package add-on for evaluating CT and MR images of blood vessels. Combining digital image processing and visualization tools such as multiplaner reconstruction (MPR), thin/think maximum intensity projection (MIP) thin and think, inverted MIP thin and think, volume rendering technique (VRT), curved planner reformation, processing tools such as bone removal (based on both single energy and dual energy) table removal and evaluation tools (vessel centerline calculation, lumen calculation, stenosis calculation) and reporting tools (lesion location, lesion characteristics) and key images), the software package is designed to support the physician in conforming the presence or absence of physician identified lesion in blood vessels and evaluation, documentation and follow up of any such lesions. It shall be used by qualified medical professionals, experienced in examining and evaluating cardiovascular CT or MR images, for the purpose of obtaining diagnostic information as part of a comprehensive diagnostic decision-making process. cvi42 is a software application that can be used as a stand-alone product or in a networked environment. The target population for the cvi42 is not restricted, however the image acquisition by a cardiac CT or MR scanner may limit the use of the device for certain sectors of the general public.
Device Story
Software application for qualitative/quantitative analysis of cardiovascular CT and MR images in DICOM format. Inputs: DICOM CT/MR image datasets. Processing: Digital image processing/visualization including MPR, MIP (thin/thick), inverted MIP, VRT, curved planar reformation, bone/table removal. Evaluation tools: vessel centerline, lumen, and stenosis calculation. Output: Visualizations, lesion location/characteristics, and reports. Used in hospital/private practice by qualified medical professionals. Supports physician in confirming lesion presence/absence and clinical decision-making. Benefits: Assists in comprehensive diagnostic evaluation and documentation of vascular lesions.
Clinical Evidence
Bench testing only. No clinical data presented. Software verification and validation performed according to a Master Software Test Plan to demonstrate safety and effectiveness.
Technological Characteristics
Software-only image analysis package. Operates on standard computing hardware in standalone or networked environments. Processes DICOM standard images. Features include MPR, MIP, VRT, curved planar reformation, and automated vessel/lumen/stenosis calculation tools. Bone removal supports single and dual-energy data. No specific hardware or sterilization requirements.
Indications for Use
Indicated for use by qualified medical professionals to evaluate cardiovascular CT and MR images of blood vessels for lesion identification, documentation, and follow-up. No specific age or gender restrictions; usage limited by patient eligibility for cardiac CT/MR scanning.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The caduceus is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular pattern around the symbol. The logo is simple and clean, with a focus on the caduceus symbol and the department's name.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 22, 2014
Circle Cardiovascular Imaging Inc % Shirantha Samarappuli VP- Regulatory Affairs and Quality Assurance 815 8th Avenue SW Suit 250 CALGARY, CA T2P 3P2 CANADA
Re: K141480
Trade/Device Name: cvi42 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving and Communications System Regulatory Class: II Product Code: LLZ Dated: August 01, 2014 Received: August 04, 2014
Dear Mr. Samarappuli:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,

for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K141480
Device Name cvi42
#### Indications for Use (Describe)
Cvi42 vascular analysis add-on is an image analysis software package add-on for evaluating CT and MR images of blood vessels. Combining digital image processing and visualization tools such as multiplaner reconstruction (MPR), think hink maximum intensity projection (MIP) thin and think, inverted MIP thin and think, volume rendering technique (VRT), curved planner reformation, processing tools as bone removal (based on both single energy and dual energy) table removal and evaluation tools (vessel centerline calculation, stenosis calculation) and reporting tools (lesion location, lesion characteristics) and key images), the software package is designed to support the physician in conforming the presence or absence of physician identified lesion in blood vessels and evaluation, documentation and follow up of any such lesions.
It shall be used by qualified medical professionals, experienced in examining and evaluating cardiovascular CT or MR images, for the purpose of obtaining diagnostic information as part of a comprehensive diagnostic decision-making process. cvi42 is a software application that can be used as a stand-alone product or in a networked environment.
The target population for the cvi42 is not restricted.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) SUMMARY
| Submitter's Name | Circle Cardiovascular Imaging Inc. |
|-----------------------------------|-----------------------------------------------------------|
| Address | Suite 250, 815 8th Avenue SW, Calgary, AB, Canada T2P 3P2 |
| Establishment Registration Number | 3007301305 |
| Date of Summary | May 29, 2014 |
| Telephone Number | 1 403 338 1870 |
| Fax Number | 1 403 338 1895 |
| Email | shirantha@circlecvi.com |
| Contact Person | Shirantha Samarappuli |
| Name of the Device | cvi42 |
| Common or Usual Name | Image Processing System |
| Classification Name | Picture Archiving and Communications System |
| Classification Name: | Picture Archiving and Communications System |
| Device Class: | II |
| Product Code: | LLZ |
| Regulation Number: | 21 CFR 892.2050 |
#### Indications for Use
cvi42 vascular analysis add-on is an image analysis software package add-on for evaluating CT and MR images of blood vessels. Combining digital image processing and visualization tools such as multiplaner reconstruction (MPR), thin/think maximum intensity projection (MIP) thin and think, inverted MIP thin and think, volume rendering technique (VRT), curved planner reformation, processing tools such as bone removal (based on both single energy and dual energy) table removal and evaluation tools (vessel centerline calculation, lumen calculation, stenosis calculation) and reporting tools (lesion location, lesion characteristics) and key images), the software package is designed to support the physician in conforming the presence or absence of physician identified lesion in blood vessels and evaluation, documentation and follow up of any such lesions.
It shall be used by qualified medical professionals, experienced in examining and evaluating cardiovascular CT or MR images, for the purpose of obtaining diagnostic information as part of a comprehensive diagnostic decision-making process. cvi42 is a software application that can be used as a stand-alone product or in a networked environment.
The target population for the cvi42 is not restricted. however the image acquisition by a cardiac CT or MR scanner may limit the use of the device for certain sectors of the general public.
{4}------------------------------------------------
## Device Description
cvi42 vascular add-on is software application for evaluating cardiovascular images in a DICOM Standard format. The software can be used as a stand-alone product that can be integrated into a hospital or private practice environment. cvi42 has a graphical user interface which allows users to qualitatively and quantitatively analyze cardiac CT & MR images.
## Indications for Use
| DEVICE | INDICATIONS FOR USE |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| cvi42 | cvi42 vascular analysis add-on is an image analysis software package add-on for evaluating CT and MR images of blood vessels. Combining digital image processing and visualization tools such as multiplaner reconstruction (MPR), thin/think maximum intensity projection (MIP) thin and think, inverted MIP thin and think, volume rendering technique (VRT), curved planner reformation, processing tools such as bone removal (based on both single energy and dual energy) table removal and evaluation tools (vessel centerline calculation, lumen calculation, stenosis calculation) and reporting tools (lesion location, lesion characteristics) and key images), the software package is designed to support the physician in conforming the presence or absence of physician identified lesion in blood vessels and evaluation, documentation and follow up of any such lesions.<br><br>It shall be used by qualified medical professionals, experienced in examining and evaluating cardiovascular CT or MR images, for the purpose of obtaining diagnostic information as part of a comprehensive diagnostic decision-making process. cvi42 is a software application that can be used as a stand-alone product or in a networked environment.<br><br>The target population for the cvi42 is not restricted, however the image acquisition by a cardiac CT or MR scanner may limit the use of the device for certain sectors of the general public. |
| DEVICE | INDICATIONS FOR USE |
| iNtuition<br>K121916 | To receive, store, transmit, post-process, display and allow<br>manipulation of reports and medical images from acquisition<br>devices, including optical or other non-DICOM format images,<br>DICOM images with modality type XA, US, CR, DR, SPECT, NM<br>and MG, and images from volumetric medical scanning devices such<br>as EST, CT, PET or MRI. To provide access to images derived data<br>and derived images via client-server software, web browser and<br>mobile technology. |
| | Visualization in 2D, 3D and 4D are supported for single or multiple<br>datasets, or combinations thereof. Tools are provided to define and<br>edit paths through structures such as centerlines, which may be used<br>to analyze cross-sections of structures, or to provide flythrough<br>visualizations rendered along such a centerline. Segmentation of<br>regions of interest and quantitative analysis tools are provided, for<br>images of vasculature, pathology and morphology, including<br>distance, angle, volume, histogram, ratios thereof, and tracking of<br>quantities over time. A database is provided to track and compare<br>results using published comparison techniques such as RECIST and<br>WHO. Calcium scoring for quantification of atherosclerotic plaque<br>is supported. |
| | Support is provided for digital image processing to derive metadata<br>or new images from input image sets, for internal use or for<br>forwarding to other devices using the DICOM protocol. Image<br>processing tools are provided to extract metadata to derive<br>parametric images from combinations of multiple input images, such<br>as temporal phases, or images co-located in space but acquired with<br>different imaging parameters, such as different MR pulse sequences,<br>or different CT image parameters (e.g. dual energy). |
| | iNtuition is designed for use by healthcare professionals and is<br>intended to assist the physician in diagnosis, who is responsible for<br>making all final patient management decisions. |
| | Interpretation of mammographic images or digitized film screen<br>images is supported only when the software is used without<br>compression and with an FDA-Approved monitor that offers at least<br>5Mpixel resolution and meets other technical specifications<br>reviewed and accepted by the FDA. |
{5}------------------------------------------------
| Feature | iNtuition | cvi42 | Remarks |
|-----------------------|-----------|-------|---------|
| 510k # | K121916 | TBD | |
| Device Class | II | II | |
| Device Classification | LLZ | LLZ | |
# Device Comparison Table
{6}------------------------------------------------
| Regulation Name | Picture Archiving and<br>communications<br>systems | Picture<br>Archiving and<br>communications<br>systems |
|--------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|-------------------------------------------------------|
| Regulation Number | 21 CFR 892.2050 | 21 CFR<br>892.2050 |
| Receive, store, transmit,<br>post process, display and<br>allow manipulation of<br>medical MR and CT<br>images | YES | YES |
| Client server<br>functionality | YES | YES |
| Visualization in 2D, 3D<br>and 4D of single or<br>multiple datasets | YES | YES |
| Define and edit paths<br>through structures such<br>as centerlines | YES | YES |
| Analysis of cross<br>references of structures | YES | YES |
| Fly-through<br>visualization | YES | YES |
| Segmentation of regions<br>of interest | YES | YES |
| Quantitative analysis<br>including distance,<br>angle, volume,<br>histogram, and tracking<br>quantities over time. | YES | YES |
| Derive metadata or new<br>images from input<br>image sets | YES | YES |
| Creating/forwarding<br>DICOM images | YES | YES |
| DICOM complaint | YES | YES |
### Description and Conclusion of Testing
cvi42 have been tested according to the specifications that are documented in a Master Software Test Plan. Testing is an integral part of Circle Cardiovascular Imaging Inc software development process as described in the company's product development process.
{7}------------------------------------------------
### Conclusion:
The successful non-clinical testing demonstrates the safety and effectiveness of the cvi42 when used for the defined indications for use and demonstrates that the device for which the 510(k) is submitted performs as well as or better than the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.