SIDEXIS 4

K132773 · Sirona Dental Systems GmbH · LLZ · Dec 11, 2013 · Radiology

Device Facts

Record IDK132773
Device NameSIDEXIS 4
ApplicantSirona Dental Systems GmbH
Product CodeLLZ · Radiology
Decision DateDec 11, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

SIDEXIS 4 is software that offers functions for the acquisition, administration, analysis, diagnosis, presentation and transfer of digital or digitized image data, e.g. X-ray images or video recordings, for medical use, predominantly in dentistry.

Device Story

SIDEXIS 4 is a PACS software platform for dental imaging; integrates 2D and 3D functions from legacy Sirona products (SIDEXIS XG and GALAXIS). Inputs include digital/digitized X-ray images and video recordings. Software provides tools for image administration, analysis, diagnosis, presentation, and transfer. Features include volume comparison, 2D projection comparison, and image display adjustments (brightness, contrast, filters). Used in dental clinics by physicians/dentists to view and interpret diagnostic images. Output assists clinicians in dental diagnosis and treatment planning.

Clinical Evidence

No clinical data. Bench testing performed to verify system functions and compare image display performance between SIDEXIS 4 and legacy systems (SIDEXIS XG/GALAXIS) with input from technicians and physicians.

Technological Characteristics

PACS software for dental imaging. Features 2D and 3D image viewing, manipulation (brightness, contrast, filters), and data management. Operates as a digital image processing system. No specific hardware materials or energy sources defined as it is a software-only device.

Indications for Use

Indicated for dental professionals to acquire, manage, analyze, diagnose, present, and transfer digital or digitized medical image data, including X-rays and video recordings.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ### 510(k) Summary for Sirona Dental Systems SIDEXIS 4 ## 1. SPONSOR | Sirona Dental Systems GmbH | | |----------------------------|------------------| | Fabrikstrasse 31 | | | D-64625 Bensheim | | | Germany | | | Contact Person: | Fritz Kolle | | Telephone: | +49 6251 16 3294 | | Date Prepared: | August 30, 2013 | ## 2. DEVICE NAME | Proprietary Name: | SIDEXIS 4 | |----------------------|---------------------------------------------| | Common/Usual Name: | Picture archiving and communications system | | Classification Name: | System, Image Processing, Radiological | # 3. Predicate Devices - SIDEXIS XG (K013659) . - GALAXIS cleared with (GALILEOS family K0123070) . - . Cybermed OnDemand3D (K113543) ## 4. Intended Use SIDEXIS 4 is software that offers functions for the acquisition, administration, analysis, diagnosis, presentation and transfer of digital or digitized image data, e.g. X-ray images or video recordings, for medical use, predominantly in dentistry. Sirona Dental Systems 510(k) SIDEXIS 4 APPENDIX G Page 1 {1}------------------------------------------------ ## 5. Device Description and Function SIDEXIS 4 combines the two Sirona products GALAXIS and SIDEXIS XG and now integrates 2D and 3D functions. New functions have been implemented (e.g. comparison of two volumes, comparison of two 2D projections). ## 6. SCIENTIFIC CONCEPT Not applicable ## 7. PHYSICAL AND PERFORMANCE CHARACTERISTICS Not applicable ## 8. SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS SIDEXIS 4 offers similar functions and views as the predicate device Cybermed OnDemand3D (K113543). It offers similar functions to the predicate software parts of SIDEXIS XG and GALAXIS together. All systems (SIDEXIS XG and GALAXIS are always taken into account together) are typical PACS systems with 3D and 2D image-viewing capabilities for dental imaging devices. They have lots of common options to change the display of the images. With adjustment of brightness, contrast and special filters for SIDEXIS 4, the physician can adjust the display to the preferred style. ## 9. Non-clinical Testing The SIDEXIS 4 system functions have been tested. Additional tests with technicians and physicians have been performed to compare images from SIDEXIS XG/GALAXIS with SIDEXIS 4. Sirona Dental Systems 510(k) SIDEXIS 4 APPENDIX G Page 2 {2}------------------------------------------------ # 10. Clinical Testing Clinical tests have not been performed. # 11. Conclusion ﺮ ﺳﻌﺪ ﺍﻟﻤﻌﺎﺭﺿﺎ ﺍﻟﻤﻮﺿﻮﻋﺎﺕ ﺍﻟﻤﺘﻮﻗﻌﺔ ﺍﻟﻤﺘﻮﻗﻒ ﺍﻟﻤﻮﺿﻮﻋﺎﺕ ﺍﻟﻤﺘﻮﺍﺻﻞ ﺍﻟﻤﺘﻮﺍﺻﻞ ﺍﻟﻤﺘﻮﺍﺻﻞ ﺍﻟﻤﺘﻮﺍﺻﻞ ﺍﻟﻤﺘﻮﺍﺻﻞ ﺍﻟﻤﺘﻮﺍﺻﻞ ﺍﻟﻤﺘﻮﺍﺻﻞ ﺍﻟﻤﺘﻮﺍﺻﻞ ﺍﻟﻤﺘﻮﺍﺻﻞ ﺍﻟﻤﺘﻮﺍﺻﻞ ﺍﻟﻤﺘﻮﺍﺻﻞ ﺍﻟﻤﺘﻮﺍﺻﻞ ﺍﻟﻤﺘﻮﺍﺻﻞ ﺍﻟﻤﺘ Based on a comparison of intended use, indications, principle of operation, features and technical data, the SIDEXIS 4 is safe and effective to perform its intended use and is substantially equivalent to the predicate devices. Sirona Dental Systems 510(k) SIDEXIS 4 APPENDIX G Page 3 {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized human figure with outstretched arms, representing care and protection. The figure is positioned above the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the figure. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-0 Silver Spring, MD 20993-0002 December 11, 2013 Sirona Dental Systems GmbH % Mr. Fritz Kolle Fabrikstrasse 31 Bensheim. D-64625 GERMANY Re: K132773 Trade/Device Name: Sidexis 4 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: September 19, 2013 Received: September 23, 2013 Dear Mr. Kolle: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {4}------------------------------------------------ Page 2- Mr. Kolle device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.lda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, ![](https://i.imgur.com/99999999999999999999999999999999.png) for Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ### Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page. ### 510(k) Number (if known) K132773 ### Device Name .............................................................................................................................................................................. ### SIDEXIS 4 #### Indications for Use (Describe) SIDEXIS 4 is software that offers function, administration, analysis, diagnosis, presentation and transfer of digital or digitized image data, e.g. X-ray images or video recordings, for medical use, predominantly in dentistry. Type of Use (Select one or both, as applicable) 区 Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED. ### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) Sm.h.7) FORM FDA 3881 (9/13)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...