ONDEMAND3D

K113543 · Cybermed, Inc. · LLZ · Dec 16, 2011 · Radiology

Device Facts

Record IDK113543
Device NameONDEMAND3D
ApplicantCybermed, Inc.
Product CodeLLZ · Radiology
Decision DateDec 16, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device, 3rd-Party Reviewed

Indications for Use

The OnDemand3D92 is intended for use as a software package which loads DICOM images from CT, MR, X-Ray, stores those and provides 3D visualization and 2D analysis, various MPR (Multi-Planar Reconstruction) functions for further rapid and precise diagnosis. It also provides the following functions as shown below. 1) Functions for implant surgery planning and fabricating a surgical guide according to the plan 2) Functions to combine and show the image data from different modalities such as CT, MRI and PET in one window at the same time 3) Functions to create orthodontic analysis using 3D volume data 4) Functions to sort and save DICOM files, Project Files and 2D image files from different modality [Panorama, Cephalometric, intraoral] under a single patient name

Device Story

OnDemand3D92 is PC-based dental imaging software; inputs DICOM images from CT, MR, X-ray, and other modalities. Transforms data via MPR, volume rendering, and registration techniques to provide 2D/3D visualization, implant planning, and orthodontic analysis. Outputs include processed images, surgical plans, and reports. Used in clinical settings by dental professionals; facilitates diagnosis, surgical guide fabrication, and treatment planning. Enables multi-modality image fusion (CT, MRI, PET) for comparative analysis. Healthcare providers use output to guide surgical procedures and orthodontic interventions, potentially improving diagnostic precision and surgical outcomes.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

PC-based software; supports DICOM 3.0 standard. Features include database management, multi-planar reconstruction (MPR), volume rendering (VR), MIP/MinIP, and image fusion. Connectivity via local/remote PACS. Image formats supported: DICOM (input/output), BMP, JPG, TIF. Lossless compression used.

Indications for Use

Indicated for use by clinicians to load, store, visualize, and analyze medical images (CT, MR, X-Ray, PET) for diagnostic purposes, implant surgery planning, surgical guide fabrication, and orthodontic analysis.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K113543 ## 510(k) Summary ## DEC 1 6 2011 This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92. And Production Comments of Research ### Date: _May. 30, 2011 1. Company and Correspondent making the submission: · Name - CyberMed, Inc. Address - #504 SJ Technoville, Gasan-dong 60-19, Geumcheon-gu, Seoul, 153-710, Korea Telephone - +82-2-3397-3970 Fax - +82-2-3397-3971 Contact – Mr. Yong Ki Im / RA Manager - 2. Device : | Trade/proprietary name | : OnDemand3DG2 | |------------------------|-----------------------------------------------| | Common Name | : Picture archiving and communications system | | Classification Name | : System, image processing, radiological | ## 3. Predicate Devices : | Manufacturer: | : CyberMed, Inc. | |----------------|----------------------------------------| | Device: | : OnDemand3D™ | | 510(k) Number: | : K070464 (Decision Date - 03/16/2007) | | Manufacturer: | : Materialise, N.V. | | Device: | : SimPlant System | | 510(k) Number: | : K033849 (Decision Date - 05/25/2004) | | Manufacturer: | : Anatomage, Inc. | | Device: | : Invivo Dental | | 510(k) Number: | : K070803 (Decision Date - 04/06/2007) | | Manufacturer: | : GE Medical systems SCS | {1}------------------------------------------------ | Device: | : CardIQ Fusion | |----------------|----------------------------------------| | 510(k) Number: | : K061370 (Decision Date - 05/30/2006) | - 4. Classifications Names & Citations : 21CFR 892.2050, LLZ, Picture archiving and communications system, Class2 - 5. Description : - 1) General Description OnDemand3DS2 is a personal computer based dental imaging software which I oads DICOM images taken from CT, MR and provides 3D visualization and 2D analysis, various MPR (Multi-Planar Reconstruction) functions for further rapid and precise diagnosis. It is also intended to plan an implant surgery, fabricate a surgical guide according to the plan, superimpose and stitch multi DICOM da ta sets, mark anatomical landmark from CT data and provide analyzed data for orthodontic treatment while software provides 2D image viewing module enabli ng 2D image acquisition specialized in 2D scanned image supporting DICOM, BITMAP, JPG, TIF. OnDemand3D22 is designed to provide with users and familiar user-interface easy. Also OnDemand3D22 makes it possible to manage medical images more easily and provides advanced tools for 2D and 3D analysis with various rendering functions. The main functions of OnDemand3D62 are as follows. - 2) Main Function - (1) DBM(Data base Manager) The operator can use DICOM images stored in Master database, Local databases or Remote PACS servers more conveniently with user-familiar interface such as Window explorer. DBM supports CD-R/RW to keep necessary DICOM images in CD directly. DBM supports to make multiple Local databases so it is very convenient for multiple users on one system. - (2) LightBox LightBox supports to load different types of DICOM images such as 8/12/16 bits gray images and color images. LightBox provides 'CINE Player' to display multiframe DICOM images. LightBox supports a preview image on Window print and {2}------------------------------------------------ DICOM print screen. #### (3) Dental Reformat Dental Reformat makes it possible to reconstruct Panoramic and Cross sectional images. Conventional imaging solutions are supported, and also new imaging solutions such as Volume Rendering, 3D Scout are supported. The layout and images are optimized for Reporting. (4) 3D 3D makes it possible to switch rendering mode such as VR(Volume Rendering), MIP/MinIP more easily and conveniently. 3D provides more accurate images by using various rendering functions such as MPR Rotating, Curve, 3D Zoom, Clip. #### (5) X-ray Generation X-ray Generation makes it possible to create X-ray images for cephalometric analysis. Lateral X-ray image and frontal X-ray image can be generated in X-ray Generation module. #### (6) Report Report makes it possible to make a report more easily by using intuitive user interface and also export it to HTML format document. The operator can capture necessary images on all screens of OnDemand3D62 with Pane-Capture/Region-Capture function and insert the captured images to Report very easily. #### (7) In2quide In2Guide is designed for the purpose of easy implant planning and short operating time with high accuracy. In2Guide provides functionality for diagnosis, implant planning and implant, abutment library of many manufacturers. #### (8) Fusion Fusion is a visualization tool using a registration technique to combine and show the image data from different modalities such as CT,MRI and PET in one window at the same time. A Fusion window consists of each MPR image of the Primary, Secondary and Fused pane matched between two images. In the case of images {3}------------------------------------------------ ## CyberMed. Inc. comes from same patients we can compare the changes between pre and postoperation. Fusion provides functions for loading two series of Primary and Secondary images and saving the fused image as a new DICOM series. のお気になるときです。 私は私は気になるところした。 このため、 ということで、 この と コー #### (9) 3D Ceph 3D Ceph provides precise analysis result using formulas based on anatomical landmarks of patient. Landmarks and formulas can be customizable. The key features of 3D Ceph are comparing pre-operation with post-operation by superimposition, photo mapping and 2D X-ray image generation, #### (10) X-Image X-Image module is management solution for both 2D and 3D images. X-Image sorts images by patient name and it handles images from Panorama, Cephal ometric, Intraoral, Conebeam CT. The operator can also view 2D images with built in filter and image tools. #### 3) Information of the image format OnDemand3D92 can load only DCM files and save results as DCM, BMP and JPG files. - DCM : DICOM (Digital Image Communication in Medicine) is a Standard Protocol to exchange and transfer the data acquired by Medical Image devices such as a CT, MR, 3D US, etc. It is designated as a Standard Protocol by ACR-NEMA (American College of Radiology-National Electrical Manufacturers Association) and now adopted by most Medical Imaging Devices. OnDemand3D02 2.0 is adaptable technically for all data of DICOM 3.0. Reference : Digital Imaging and Communications in Medicine (DICOM) ACR-NEMA Standards Publication PS 3.1~PS 3.16 2003. - BMP : The standard bit-mapped graphics format used in the Windows environment. By convention, graphics files in the BMP format end with a BMP extension. BMP files store graphics in a format called device-independent bitmap (DIB). - JPG/JPEG : Short for "Joint Photographic Experts Group", the original name of the committee that wrote the standard. JPG is one of the image file formats supported {4}------------------------------------------------ on the Web. JPG images support 16 million colors and are best suited for photographs and complex graphics. 100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000 - 4) Compression - Compression Method : Lossless Compression - 6. Indication for use : The OnDemand3DG2 is intended for use as a software package which loads DICOM images from CT, MR, X-Ray, stores those and provides 3D visualization and 2D analysis, various MPR (Multi-Planar Reconstruction) functions for further rapid and precise diagnosis. - It also provides the following functions as shown below. - 1) Functions for implant surgery planning and fabricating a surgical guide according to the plan - 2) Functions to combine and show the image data from different modalities such as CT, MRI and PET in one window at the same time - 3) Functions to create orthodontic analysis using 3D volume data - 4) Functions to sort and save DICOM files, Project Files and 2D image files from different modality [Panorama, Cephalometric, intraoral] under a single patient name - 7. Comparison with predicate device : CyberMed, Inc., ensures that the OnDemand3D92, Picture archiving and communications system is substantially equivalent to OnDemand3D™ of CyberMed, Inc., SimPlant System of Materialise N.V., Invivo Dental of Anatomage Inc., CardIQ Fusion of GE Medical systems SCS. The similarities are as follows. SALES THE COLLECT THE COLLECT User, Computer Platform, Communication, PACS Server, DICOM data transferring, Archiving data Format, Image format In, Image Compression, Image Archive, Image Display, Image Reformation, 3D Image Construction, Image Projection, Image Edit, Basic Image Operation, Measurements, Other Function, Report, Printing And the additional similarities are as follows. 1) Similarity between OnDemand3D92 In2Guide module and SimPlant System {5}------------------------------------------------ - 2) Similarity between OnDemand3D02 Fusion module and CardIQ Fusion のお気になるところです。ここではないとなっているというとなると、その - 3) Similarity between OnDemand3D62 3D Ceph module and Invivo Dental And the Differences are as follows. - 1) Differences between OnDemand3D 92 In2Guide module nd SimPlant System - 2) Differences between OnDemand3DS2 3D Ceph module and Invivo Dental (3D Analysis) There are a little bit of differences between the OnDemand3D62 and Predicate Devices (OnDemand3D™, SimPlant System_Invivo Dental, CardIQ Fusion) as above. But eventually, these differences do not cause any problems in the safety and effectiveness. 8. Conclusions : In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification CyberMed, Inc. concludes that OnDemand3D2 is safe and effective and substantially equivalent to predicate devices as described herein. - 9. CyberMed, Inc. will update and include in this summary any other information deemed seasonably necessary by the FDA. END {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, representing the department's mission to protect the health of all Americans. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room – WO66-G609 Silver Spring, MD 20993-0002 CyberMed Inc. % Mr. Marc M. Mouser Engineering Leader & FDA Office Coordinator, Program Reviewer Underwriters Laboratories Inc. 2600 NW Lake Road CAMAS WA 98607 DEC 1 6 2011 Re: K113543 Trade/Device Name: Picture Archiving Communications System/OnDemand3D902 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: November 1, 2011 Received: December 1, 2011 Dear Mr. Mouser: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of {7}------------------------------------------------ medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely Yours, Mary Patel Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure Page 2 {8}------------------------------------------------ # Indications for Use # 510(k) Number(if known): K | 3543 Device Name: Picture Archiving Communications System / OnDemand3D62 Indications for Use: The OnDemand3D92 is intended for use as a software package which loads DICOM images from CT, MR, X-Ray, stores those and provides 3D visualization and 2D analysis, various MPR (Multi-Planar Reconstruction) functions for further rapid and precise diagnosis. It also provides the following functions as shown below. - 1) Functions for implant surgery planning and fabricating a surgical guide according to the plan - 2) Functions to combine and show the image data from different modalities such as CT, MRI and PET in one window at the same time - 3) Functions to create orthodontic analysis using 3D volume data - 4) Functions to sort and save DICOM files, Project Files and 2D image files from different modality [Panorama, Cephalometric, intraoral] under a single patient name Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use AND/OR (Part 21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Mary S Patel Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Page 1 of 1 ..
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...