Myrian is a multi modality medical diagnostic device for the review and analysis of anatomy and pathology in multi-dimensional digital images acquired from a variety of imaging devices. It also includes DICOM communication capabilities and media interchange features (printing, CD burning, storing). It runs on any standard PC including laptops that might be purchased independently by the end user. Typical end users are trained medical professionals .. Myrian includes tools which enable the reviewing physician to provide any selected relevant information for diagnosis, surgery and treatment planning to the referring physician. These toolsets are categorised as follow: - Enhanced imaging tools such as: - Multi-Planar Reformation (MPR) views in any plane (orthogonal, oblique or curved), 3D views in any rendering mode (MIP, MiniP, Average, Volume Rendering) - · Cross-sectional or Endoscope Exploration Modes along a centerline (e.g. of a vessel, a colon ... etc.) - · Filet Visualization Mode, to visualize as a flat surface any tubular hollow organ (such as a colon) - Manual or interactive Objects Of Interest such as : - · Annotations of Interest, for information or measurement purposes - · Paths (considered as Annotations of Interest); - · Regions of Interest, for anatomical and pathological structure isolation (such as liver, spleen, lungs, colon ... etc.) through which any measurement can be performed - · Points of Interest, for marking areas such as lesions, tumors ... etc. - Reporting tools : Objects of Interest generates reports which may be viewed and sent to standard film or paper printers or sent electronically to an intranet web server or any other DICOM device. - Manual or assisted image alignment tools : - · for multiphasic or time-based image comparisons - · Cutting Surface Tool based on ROI for preoperative evaluation of surgery strategies (such as for the liver). This device is not indicated for mammography use. Lossy compressed mammography-images and digitized film screen images must not be used for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 mega pixel resolution and meets other technical specifications approved by the FDA.
Device Story
Myrian is a multi-modality diagnostic software platform for reviewing/analyzing multi-dimensional medical images. It operates on standard off-the-shelf PC hardware. Input consists of DICOM images from various imaging devices. The software provides tools for Multi-Planar Reformation (MPR), 3D rendering (MIP, MiniP, Volume Rendering), cross-sectional/endoscope exploration, and 'Filet' visualization for tubular organs. It allows manual/interactive creation of Objects of Interest (annotations, paths, regions, points) for isolation and measurement. It includes manual/assisted image alignment for multiphasic comparisons and a virtual cutting surface tool for preoperative surgical planning. Output includes processed images, measurements, and generated medical reports, which can be printed or exported via DICOM. Used by trained medical professionals in clinical settings to facilitate diagnosis and treatment planning.
Clinical Evidence
Bench testing only. User site testing and benchmarking demonstrated that the device meets required specifications. No clinical data or adverse effects reported.
Technological Characteristics
Software-based image processing system running on standard off-the-shelf PC hardware. Features include MPR, 3D rendering, DICOM communication, and manual/assisted measurement tools. Complies with FDA guidance for software contained in medical devices.
Indications for Use
Indicated for trained medical professionals to review and analyze multi-dimensional digital images for anatomy and pathology assessment, diagnosis, and treatment planning. Not indicated for mammography use.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
### PREMARKET NOTIFICATION 510(K) SUBMISSION
MYRIAN 1.4
Image /page/0/Figure/3 description: This document is a 510(K) summary for a medical device. The preparation date is April 2nd, 2009, and the submitter is INTRASENSE with registration number 3006546169. The device name is System, Image Processing, trade name MYRIAN, and model number N/A. The classification name is System, Image Processing, Radiological, with code product LLZ, panel 892, regulation number 892.-2050, and regulatory class II.
Follow-up of patient examination
Generation of medical reports
Export of DICOM images to any format, towards any DICOM entity or recognized media.
Virtual Cutting surface tool for preoperative evaluation of surgery strategies.
{1}------------------------------------------------
PREMARKET NOTIFICATION 510(K) SUBMISSION
MYRIAN 1.4
Explanation of how the device operates
Myrian® with its modules is designed to run on standard PC hardware, through the installed operating system. The hardware is all "off-the-shelf" standard computer components and may be purchased independently by the end user.
Intended use
Mvrian is a multi modality medical diagnostic device for the review and analysis of anatomy and pathology in multi-dimensional digital images acquired from a variety of imaging devices. It also includes DICOM communication capabilities and media interchange features (printing, CD burning, storing). It runs on any standard PC including laptops that might be purchased independently by the end user. Typical end users are trained medical professionals ..
Myrian includes tools which enable the reviewing physician to provide any selected relevant information for diagnosis, surgery and treatment planning to the referring physician.
These toolsets are categorised as follow:
- Enhanced imaging tools such as:
-
.
- Multi-Planar Reformation (MPR) views in any plane (orthogonal, oblique or curved), 3D views in any rendering mode (MIP, MiniP, Average, Volume Rendering)
- · Cross-sectional or Endoscope Exploration Modes along a centerline (e.g. of a vessel, a colon ... etc.)
- · Filet Visualization Mode, to visualize as a flat surface any tubular hollow organ (such as a colon)
- Manual or interactive Objects Of Interest such as :
- · Annotations of Interest, for information or measurement purposes
- · Paths (considered as Annotations of Interest);
- · Regions of Interest, for anatomical and pathological structure isolation (such as liver, spleen, lungs, colon ... etc.) through which any measurement can be performed
- · Points of Interest, for marking areas such as lesions, tumors ... etc.
- Reporting tools :
Objects of Interest generates reports which may be viewed and sent to standard film or paper printers or sent electronically to an intranet web server or any other DICOM device.
- Manual or assisted image alignment tools :
- · for multiphasic or time-based image comparisons
- · Cutting Surface Tool based on ROI for preoperative evaluation of surgery strategies (such as for the liver).
This device is not indicated for mammography use. Lossy compressed mammography-images and digitized film screen images must not be used for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 mega pixel resolution and meets other technical specifications approved by the FDA.
All Myrian functionalities can be packaged, licensed and marketed as individual modules. The Myrian System allows the OEM customization of both the graphical user interface and the available functionalities, while implying no impact on the system performance or system intended use.
#### Performance data
Performance data were verified versus the requirements of the FDA "Guidance of the Content of Pre Market Submissions for Software Contained in Medical Devices'
User Site Testing and Benchmarking demonstrate that MYRIAN meet the required specifications. No adverse affects have been detected.
{2}------------------------------------------------
### PREMARKET NOTIFICATION 510(K) SUBMISSION
Substantial equivalence summary
The technological characteristics, features, specifications, materials, mode of operation, and intended use of MYRIAN device are equivalent to those of the predicate devices quoted above.
MYRIAN is the same as the predicate devices in K061624, K052995, K071000 and K082228
The differences that exist between the devices do not raise new issues of safety or effectiveness regarding MYRIAN Device.
It is substantially equivalent in terms of safety and effectiveness to the predicate devices.
The "Substantial Equivalence Decision Making Process" provided by the FDA has been followed, see dedicated section in this document.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of an eagle.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
JUN 2 9 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Frederic Banegas Chief Technical Officer Intrasense CAP OMEGA. Rond Point Benjamin Franklin CS 39521 Montpellier, 34960 FRANCE
Re: K091001
Trade/Device Name; Myrian Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: June 10, 2009 Received: June 16, 2009
#### Dear Mr. Banegas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical
{4}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Janine M. Morris
Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### PREMARKET NOTIFICATION 510(K) SUBMISSION
# A . . . . . . . . . . . INDICATIONS FOR USE .
510{k) Number (if known) :
K04100
Device Name:
Myrian
Indications for Use (no change):
Myrian is a multi modality medical diagnostic device. It is aimed at reviewing and matorny and pathology. It also includes DICOM communication capabilities and media interchange features (printing, storing). It runs on any standard PC including laptops that might be purchased independently by the end user a set of tools meant to create and modify objects of interest : points of interest, annotations of interest and volumes of interest.
This device is not indicated for mammography use. Lossy compressed mammography images and digitized film screen images must not be used for primary image interpretations. Mamnographic images may only be interpreted using an FDA approved monitor that offers at least 5 mega pixel resolution and meets other technical specifications approved by the FDA.
Prescription Use
(Part 21 CFR 801 Subpart D)
510/k) Number
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
A MULA WASHING BANGURA CASA BASANA A POWER
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Hula Reum
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Dev
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.