Myrian® with its modules is a software application that is used for reviewing medical images. Myrian® with its modules is designed to run on standard PC hardware. The hardware is all "off-the-shelf' standard computer components and may be purchased independently by the end user. Myrian with its modules receives digital images and data from various sources (including but not limited to CT, MR, US, RF modalities, computed and direct radiographic devices, secondary capture devices, scanners, imaging gateways or imaging sources like PACS). Myrian® and its Modules can be used to communicate, to process, print and display medical images. Users have access to various image processing, Volume of Interest editing and measurement tools to assist them in viewing and quantifying images. Myrian® and its Modules can be integrated with an institution's existing HIS or RIS for a fully integrated electronic patient record. Typical users of Myrian® with its Modules are trained medical professionals, including but not limited to radiologists, technologists and clinicians. When interpreted by a trained physician, filmed or displayed images on the Myrian® and its Modules may be used as a basis for diagnosis, except in the case of mammography images. Myrian® with its Modules does not support the display of mammography images for diagnosis.
Device Story
Myrian is a software suite for medical image review and analysis; runs on off-the-shelf PC hardware. Inputs include DICOM images from modalities (CT, MR, US, RF, PACS). Features include thin/thick MPR, 3D volume rendering, VOI creation, and quantification (volume, surface, density). Used by radiologists, clinicians, and technologists in clinical settings. Output is displayed on monitors for diagnostic review and reporting. Facilitates clinical decision-making by providing visualization and measurement tools. Integrates with HIS/RIS for electronic patient records. Benefits include improved workflow for image interpretation and patient examination follow-up.
Clinical Evidence
Bench testing only. Performance verified against FDA "Guidance of the Content of Pre Market Submissions for Software Contained in Medical Devices." User site testing and benchmarking demonstrate compliance with specifications. No clinical data or adverse effects reported.
Technological Characteristics
Software-only system; runs on standard off-the-shelf PC hardware. Supports DICOM standard for image communication and processing. Features include MPR, 3D volume rendering, and measurement tools. No specific hardware materials or sterilization required as it is a software application.
Indications for Use
Indicated for trained medical professionals (radiologists, technologists, clinicians) to review and analyze medical images (CT, MR, US, RF, CR/DR) for anatomy and pathology assessment. Contraindicated for mammography image interpretation.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
## Intrasense PREMARKET NOTIFICATION 510(K) SUBMISSION
MYRIAN
060830
.
| 5. 510(K) SUMMARY | | | |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|--|
| | [As Required by 21 CFR 807.92]<br>Summary of Safety and Effectiveness | MAY 14 2007 | |
| Preparation date | April 6th, 2007 | | |
| Submitter | | | |
| Name | INTRASENSE | | |
| Registration number | TBD | | |
| Address | CAP OMEGA, Rond Point Benjamin Franklin CS 39521 34960 Montpellier FR | | |
| Tel Fax | (+33) 467 130 130 / (+33) 467 130 132 | | |
| Contact Person | Mr Stéphane CHEMOUNY<br>Phone number : (+33) 467 130 130<br>Fax number : (+33) 467 130 132 | | |
| Device name | | | |
| Common Name | System, Image Processing | | |
| Trade Name | MYRIAN | | |
| Model number | N/A | | |
| Device classification | | | |
| Classification name | System, Image Processing, Radiological | | |
| Code product | LLZ | | |
| Panel | 892 | | |
| Regulation number | 892.=2050 | | |
| Regulatory class | II | | |
| Predicate devices | [K052995] Cleared [November 8, 2005] [General Electric Medical Systems]<br>[Advantage Workstation Version 4.3], manufactured by [General Electric Medical<br>Systems]<br>[K061624] Cleared [June 27, 2006] [Vital Images, Inc.] [Vitrea2 Version 3.9],<br>manufactured by [Vital Images, Inc.] | | |
| Description | Myrian® system is a software suite providing the following services :<br>Import of DICOM images from any DICOM modality, workstation or PACS<br>Visualization of DICOM images in thin MPR, thick MPR and full 3D volume rendering<br>Creation of VOI (Volume Of Interest) with dedicated tools<br>Calculation of volumes, surface and of average, minimum and maximum densities of<br>VOI<br>Follow-up of patient examination<br>Generation of medical reports<br>Export of DICOM images to any format, DICOM entity or media | | |
| Explanation of how the device operates | Myrian® with its modules is designed to run on standard PC hardware. The hardware<br>is all "off-the-shelf' standard computer components and may be purchased<br>independently by the end user. | | |
| Intended use | Myrian® with its modules is a software application that is used for reviewing medical<br>images. Myrian® with its modules is designed to run on standard PC hardware. The<br>hardware is all "off-the-shelf' standard computer components and may be purchased<br>independently by the end user. Myrian with its modules receives digital images and<br>data from various sources (including but not limited to CT, MR, US, RF modalities,<br>computed and direct radiographic devices, secondary capture devices, scanners,<br>imaging gateways or imaging sources like PACS).<br>Myrian® and its Modules can be used to communicate, to process, print and display<br>medical images. Users have access to various image processing, Volume of Interest<br>editing and measurement tools to assist them in viewing and quantifying images.<br>Myrian® and its Modules can be integrated with an institution's existing HIS or RIS for<br>a fully integrated electronic patient record. | | |
| | Typical users of Myrian® with its Modules are trained medical professionals, including<br>but not limited to radiologists, technologists and clinicians. | | |
| | When interpreted by a trained physician, filmed or displayed images on the Myrian®<br>and its Modules may be used as a basis for diagnosis, except in the case of<br>mammography images.<br>Myrian® with its Modules does not support the display of mammography images for<br>diagnosis. | | |
| Performance data | Performance data were verified versus the requirements of the FDA "Guidance of the<br>Content of Pre Market Submissions for Software Contained in Medical Devices" | | |
| | User Site Testing, Benchmarking and clinical data analysis demonstrate that MYRIAN<br>meet the required specifications. No adverse affects have been detected. | | |
| Substantial equivalence summary | The technological characteristics, features, specifications, materials, mode of<br>operation, and intended use of MYRIAN device are equivalent to those of the<br>predicate devices quoted above. | | |
| | MYRIAN is the same as the predicate devices in K061624 and K052995 | | |
| | The differences that exist between the devices do not raise new issues of safety or<br>effectiveness regarding MYRIAN Device. | | |
| | It is substantially equivalent in terms of safety and effectiveness to the predicate<br>devices. | | |
{2}------------------------------------------------
Intrasense PREMARKET NOTIFICATION 510(K) SUBMISSION 060830
{3}------------------------------------------------
Image /page/2/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged in a circular pattern around the symbol. The logo is black and white.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
INTRASENSE % Mr. Nicolas Clary Project Leader CEISO "Les Portes d'Espagne" - Bât. B - 99, route d'Espagne 31100 Toulouse 31100 FRANCE
MAY 1 4 2007
Re: K071000
Trade/Device Name: Myrian Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: April 6, 2007 Received: April 13, 2007
Dear Mr. Clary:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/2/Picture/10 description: The image is a circular logo for the FDA Centennial, celebrating 1906-2006. The logo features the letters "FDA" in a stylized font, with the word "Centennial" written below. The logo is surrounded by a circular border with text and stars.
moting Public J
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I toase or acressed a determination that your device complies with other requirements of the Act that I Dederal statutes and regulations administered by other Federal agencies. You must or any I vith all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (2) CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) Premarket notification. The FDA finding of substantial equivalence of your device to a legally premainer nedicated device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you use office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under (21 CFF at 80 ! Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours,
Nancy C. Hogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
PREMARKET NOTIFICATION 510(K) SUBMISSION MYRIAN Intrasense 060830
| 4. INDICATIONS FOR USE | |
|------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number (if known): | K07/000 |
| Device Name: | Myrian |
| Indications for Use: | Myrian is a multi modality medical diagnostic device. It is aimed at reviewing and analysing anatomy and pathology. It also includes DICOM communication capabilities and media interchange features (printing, CD burning, storing). It runs on any standard PC including laptops that might be purchased independently by the end user. It provides user a set of tools meant to create and modify volumes of interest. |
| | This device is not indicated for mammography use. Lossy compressed mammography images and digitized film screen images must not be used for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 mega pixel resolution and meets other technical specifications approved by the FDA. |
| <div style="text-align: left;">Prescription Use</div> <div style="text-align: left;">(Part 21 CFR 801 Subpart D)</div> | <div style="text-align: left;">1.1 AND/OR</div> <div style="text-align: left;">Over-The-Counter Use</div> <div style="text-align: left;">(21 CFR 801 Subpart C)</div> |
| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) | |
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
.
.
Nancy C Hogdon
(Division Sign-Off)
Division of Reproductive, Abdominal, and Radiological Devices 071000 510(k) Number ________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.