Kanteron DICOM Station (KDS) is a software device intended for viewing of images acquired from computer tomography (CT), magnetic resonance (MR), computed radiography (CR), digital radiography (DR), ultrasound (US), and other DICOM compliant medical imaging systems when installed on suitable commercial standard hardware. Images and data can be captured, stored, communicated, processed and displayed within the system and or across computer networks at distributed locations. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary diagnosis or image interpretation. For primary diagnosis, post process DICOM "for presentation" images must be used. Mammographic images must only be viewed with a monitor approved by the FDA for viewing mammographic images. It is the User's responsibility to ensure monitor quality, ambient light conditions, and image compression ratios are consistent with clinical application. The KDS is not intended for use as a preoperative surgical planning system and must not be used for this purpose.
Device Story
KDS is a software-based Picture Archiving and Communications System (PACS) running on Apple Mac OSX. It ingests medical images from DICOM-compliant modalities (CT, MR, CR, DR, US). The system enables capture, storage, communication, processing, and display of these images across distributed networks. Operated by clinicians in clinical settings, the software provides tools for image review, navigation, and post-processing. It does not contact patients or control life-sustaining equipment. Healthcare providers use the displayed output to assist in clinical decision-making; the system relies on the practitioner to interpret images and intervene as necessary. It facilitates efficient access to diagnostic data across facilities.
Clinical Evidence
No clinical data. Bench testing only. Verification and validation activities were performed per hazard analysis, confirming the system met all predetermined acceptance criteria.
Technological Characteristics
Software-only device; runs on Apple Mac OSX. Complies with DICOM 3.0 and JPEG standards for image communication and compression. Does not contact patients. No hardware components. Connectivity via standard computer networks.
Indications for Use
Indicated for viewing medical images (CT, MR, CR, DR, US, and other DICOM-compliant systems) for diagnostic review. Not for preoperative surgical planning. Mammographic images require FDA-approved monitors; lossy compressed mammographic and digitized film images are excluded from primary diagnosis.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness *
### KANTERON DICOM STATION (KDS) (Software) Version 3.0
### PICTURE ARCHIVING AND COMMUNICATIONS SYSTEM
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990:
#### Submitter's Information (per 21 CFR 807 92(a)(1)):
| Trade Name: | Kanteron DICOM Station (KDS) |
|----------------------|-----------------------------------------|
| Common Name: | Picture Archiving Communications System |
| Classification Name: | System, image processing, radiological |
| Product Code: | LLZ |
#### Predicate Device (21 CFR 807.92(a)(3):
FDA has classified the predicate device (K063470) as Class II; 21 CFR 892.2050, LLZ. I believe that Kanteron DICOM Station (KDS) device should have the same classification as the predicate device. Predicate device details are as follows:
| Device Classification Name: | System, Image processing, Radiological |
|------------------------------------|----------------------------------------|
| 510(k) Number: | K063470 |
| Regulation Number: | 892.2050 |
| Device Name: | Aycan Workstation OsiriX |
| Applicant: | Aycan Digitalsysteme GmbH |
| Classification Product Code: | LLZ |
| Decision date: | 01/05/2007 |
| Classification Advisory Committee: | Radiology |
#### Applicable mandatory and voluntary standards:
Declaration of Conformity or certification statement to applicable voluntary standards that are cited for the device.
* Any statement made in conjunction with this submission regarding substantial equivalence to any other product only relates to whether the product can be lawfully marketed without pre-market approval or reclassification and is not to be interpreted as an admission or used as evidence in patent infringement litigation. As the Commissioner of the FDA has indicated, " . . a determination of substantial equivalence under the Federal Food, Drug, and Cosmetic Act that the product can be lawfully marketed without premarket approval or reclassification. The determination is not intended to have any bearing whatsoever on the resolution of patent infringement suits." 42 Fed. Reg. 42.520 et seg. (1977).
See Declaration of Conformity (DICOM and JPEG standards) in Appendix 1 (Section 21).
{2}------------------------------------------------
DICOM (Digital Imaging and Communications in Medicine) - Developed by the American College of Radiology and the National Electrical Manufacturer's Association. Specifies the format for the communication of digital images between individual devices and over networks.
JPEG (Joint Photographic Experts Group) Standard - Specifies methods for the compression (reversible and irreversible) of digital medical images. References -ISO/IEC 10918-1 (1994-02) Digital Compression and Coding of Continuous-Tone Still Images (JPEG), G.K. Wallace, "The JPEG Still Picture Compression Standard", Communications of the ACM, Vol. 34, No. 4, April 1991.
## Indications for Use (21 CFR 807 92(a)(5))
Indications for Use:
Kanteron DICOM Station (KDS) is a software device intended for viewing of images acquired from computer tomography (CT), magnetic resonance (MR), computed radiography (CR), digital radiography (DR), ultrasound (US), and other DICOM compliant medical imaging systems when installed on suitable commercial standard hardware.
Images and data can be captured, stored, communicated, processed and displayed within the system and or across computer networks at distributed locations.
Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary diagnosis or image interpretation.
For primary diagnosis, post process DICOM "for presentation" images must be used. Mammographic images must only be viewed with a monitor approved by the FDA for viewing mammographic images.
It is the User's responsibility to ensure monitor quality, ambient light conditions, and image compression ratios are consistent with clinical application.
#### The KDS is not intended for use as a preoperative surgical planning system and must not be used for this purpose.
#### Device Description: 21 CFR 807.92(a)(4)
The KDS provides services for review and post processing of diagnostic medical images and information. It conforms to the DICOM 3.0 standard to allow the sharing of medical information with other digital imaging systems. Kanteron DICOM Station (KDS) runs on Apple Mac OSX systems and provides high performance review, navigation and post processing functionality for multidimensional and multimodality images. Language files are separate from image software, allowing software to be provided in various languages.
{3}------------------------------------------------
Currently, the software is provided in English, Spanish, French, German, Japanese and Chinese.
KDS software facilitates viewing of images acquired from CT, MR, CR, DR, and other DICOM-compliant medical imaging systems when installed on suitable commercial standard hardware. Such images and data can be captured, stored, communicated, processed and displayed within the system and/or across computer networks at distributed locations. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary diagnosis or image interpretation. For primary diagnosis, post process DICOM "for presentation" images must be used. Mammographic images should only be viewed with a monitor approved by the FDA for viewing mammographic images. It is the User's responsibility to ensure monitor quality, ambient light conditions, and image compression ratios are consistent with clinical application.
## Technical Characteristics and Principles of Operation:
Kanteron DICOM Station (KDS) is a software device that handles and manipulates digital medical images. The device does not contact the patient, nor does it control any life sustaining devices.
## Technological Characteristics and Specifications: 21 CFR 807 92(a)(6)
Kanteron DICOM Station (KDS) is a software device that handles and manipulates digital medical images.
The device does not contact the patient, nor does it control any life sustaining devices. Prior to any medical decisions, a licensed medical practitioner reviews the output displays of the KDS system, providing ample opportunity for competent human intervention for the interpretation of images and information being displayed and printed.
# Testing
As required by the hazard (risk) analysis, designated individuals performed all verification and validation activities and results demonstrated that the predetermined acceptance criteria were met. The System passed all testing criteria.
Kanteron uses standard irreversible compression techniques, laboratory data are not required.
Conclusion 21 CFR 807 92(b)(1)
{4}------------------------------------------------
The 510(k) Pre-Market Notification for the Kanteron Dicom Station contains adequate information and data to enable FDA/CDRH to determine substantial equivalence to the predicate device.
The submission contains the results of a hazard analysis and the "Level of Concern" for potential hazards has been classified as "Minor" (See Appendix 2 (Section 22)).
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure with outstretched arms, symbolizing care and protection. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
MAR 1 9 2008
Ms. Monica Veytia 7195 SW 101 Street MIAMI FL 33156-3247
Re: K073176
Trade/Device Name: KANTERON DICOM Station (KDS) Version 3.0 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: March 4, 2008 Received: March 12, 2008
Dear Ms. Veytia:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/4/Picture/10 description: The image is a circular logo with the letters "FDA" in large, bold font in the center. Above the letters, the numbers "1906-2006" are printed. Below the letters, the word "Centennial" is written in cursive. The logo is surrounded by a circular border with text and stars. The text is difficult to read due to the image quality, but it appears to be part of the FDA's centennial celebration.
Protesting and Promoting Public Health
{6}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{7}------------------------------------------------
#### Indications for Use
510(k) Number (if known): K073176
Device Name: KANTERON DICOM Station (KDS) Version 3.0
Indications for Use:
Kanteron DICOM Station (KDS) is a software device intended for viewing of images acquired from computer tomography (CT), magnetic resonance (MR), computed radiography (CR), digital radiography (DR), ultrasound (US), and other DICOM compliant medical imaging systems when installed on suitable commercial standard hardware,
Images and data can be captured, stored, communicated, processed and displayed within the system and or across computer networks at distributed locations.
Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary diagnosis or image interpretation.
For primary diagnosis, post process DICOM "for presentation" images must be used. Mammographic images must only be viewed with a monitor approved by the FDA for viewing mammographic images.
It is the User's responsibility to ensure monitor quality, ambient light conditions, and image compression ratios are consistent with clinical application.
The KDS is not intended for use as a preoperative surgical planning system and must not be used for this purpose.
17 Prescription Use Prescription Use _____________________________________________________________________________________________________________________________________________________________
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Arm hithang
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number _
Page of of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.