SECURVIEW DX DIAGNOSTIC WORKSTATION, MODELS SV-0001, SV-0002 AND SV-0003
K062107 · Hologic, Inc. · LLZ · Aug 18, 2006 · Radiology
Device Facts
Record ID
K062107
Device Name
SECURVIEW DX DIAGNOSTIC WORKSTATION, MODELS SV-0001, SV-0002 AND SV-0003
Applicant
Hologic, Inc.
Product Code
LLZ · Radiology
Decision Date
Aug 18, 2006
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The Hologic SecurView DX device is intended for selection, display, manipulation, filming and media interchange of multi-modality images from a variety of different modality systems. It also interfaces to various image storage and printing devices using DICOM or similar interface standards. The device used with FDA cleared monitors may be used by a trained physician for display, manipulation and interpretation of lossless compressed or non-compressed mammographic images for screening and diagnostic mammography, as well as any other DICOM multi-modality image. The SecurView DX is typically used by trained professionals, including, but not limited to physicians, radiologists, nurses, medical technicians and assistants.
Device Story
SecurView DX is a multi-modality review workstation software for digital radiology images, focused on mammography. Software accepts standard or multi-frame DICOM images (US, MR, DR, CR, SC, CT, PET) via network interface. Operates on Windows XP PC with high-resolution gray scale displays (5 MP) or color displays (2 MP). Physician/technician uses mouse, keyboard, and workflow keypad to manipulate and interpret images. Image processing is external to software. Output facilitates clinical decision-making by providing visualization tools for screening and diagnostic interpretation. Benefits include efficient review of multi-modality images and support for mammographic diagnostic workflows.
Clinical Evidence
Bench testing only. No clinical data provided. Device validated under ISO 13485 and 21 CFR Part 820 Quality System. Compliance with ACR/NEMA DICOM standards (PS 3.1-3.18) verified.
Technological Characteristics
Software-based workstation for DICOM image display. Runs on Windows XP; Intel CPU >= 2.0 GHz; 2.0 GB RAM; 140 GB HDD. Supports high-resolution 5 MP gray scale displays and 2 MP color displays. Connectivity via 10/100 Base TX Network Interface. Complies with DICOM standards. ISO 13485 and 21 CFR Part 820 compliant development.
Indications for Use
Indicated for trained physicians to display, manipulate, and interpret lossless compressed or non-compressed mammographic images for screening and diagnostic mammography, and other DICOM multi-modality images. Used by trained professionals including radiologists, nurses, and technicians.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Hologic, Inc. Special 510(k) for SecurView DX
Kob 2107
# 510(k) Summary SecurView DX
AUG 1 8 2006
Product Name: SecurView DX
Product Classification Name: Picture archiving and communication system
Product Classification Code: LLZ CFR Section: 892.2050
Classification Panel: Radiology Class II
Manufacturer: Hologic, Inc. 36 Apple Ridge Road Danbury, CT 06810 USA
- Contact Person: Gail Yaeker-Daunis Telephone Number: (203) 731-8337 Fax Number: (203) 731-8440
July 18, 2006 Date Prepared:
K041555 SecurView DX, Hologic Inc. Predicate Device:
Device Description: The Hologic Secur View DX is mainly a software product. It is used for visualization and manipulation of digital radiology images. The SecurView DX is a multi-modality review workstation software focused on mammographic applications. The software can drive at least two high-resolution gray scale displays (FDA cleared for Mammography), a PC mouse, a keyboard and an optional dedicated workflow keypad.
The software accepts standard or multi-frame mammography images that have been created according to the "FOR PRESENTATION" specification of the DICOM Standard with no compression or using lossless compression. Image processing is external to the SecurView DX software.
The software accepts multimodality images of US, MR, DR, CR, SC, CT, PET, and other DICOM formats for display and manipulation on the high resolution 5 MP gray scale displays or on a color 2 MP display in 2-D or 3-D view.
The SecurView DX software can be used in a single or in a multi-workstation configuration.
The minimum computer requirements to run the Secur View DX software are: Windows XP® Operating System Intel CPU with a clock rate of 2.0 GHz or greater 2.0 GB RAM or greater 140 GB Hard drive or greater, operating at 160 Mbs or greater
{1}------------------------------------------------
## CD-ROM/R 10/100 Base TX Network Interface
## Intended Use:
The Hologic SecurView DX is intended for selection, display, manipulation, filming and media interchange of multi-modality images from a variety of different modality systems. It also interfaces to various image storage and printing devices using DICOM or similar interface standards.
The device may be used by a trained physician for display, manipulation and interpretation of lossless compressed or non-compressed mammographic images using FDA cleared displays for screening and diagnostic mammography, as well as any other DICOM multi-modality image.
The Secur View DX software is typically used by trained professionals, including, but not limited to physicians, radiologists, nurses, medical technicians and assistants.
## Comparison with Predicate Device:
The modified SecurView DX is substantially equivalent to the SecurView DX cleared August 9, 2004 as 510(k) #K041555 and has the same intended use as stated above,
| SecurView DX (K)041555 | SecurView DX Modified |
|-------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intel CPU with clock rate of 2.0 GHz or<br>greater | Same |
| 2.0 GB RAM or greater | Same |
| 140 GB Hard drive or greater operating at 160<br>Mbs or greater | Same |
| CD ROM/read/write / DVD +/- read/write | CD ROM/R |
| Windows 2000 Operating System | Windows XP Operating System |
| US QWERTY Keyboard | Same |
| 10/100 Base TX Network Interface | 10/100 Base TX Network Interface |
| 3-button Mouse | Pointing Device w/wheel |
| Workstation Keypad | Workstation Keypad |
| (2) High-Resolution Grey Scale Display<br>(FDA cleared for Mammography)<br>21" Portrait Orientation,<br>Resolution 2058/2560 pixels (5MP) | At least two High-Resolution Grey<br>Scale Displays (FDA cleared for<br>Mammography)<br>Optional Color at least 2 MP, High<br>Contrast Display for Multimodality<br>viewer |
{2}------------------------------------------------
| SecurView DX (K)041555 | SecurView DX Modified |
|------------------------------------------------------------|-----------------------|
| Bar Code Reader (Code 39 Full ASCII, Code<br>128, Code 39) | Same |
### Software:
The SecurView DX software is specified, validated and tested by Hologic, Inc., under an ISO 13485 and 21 CFR Part 820 Quality System.
## Safety and Effectiveness - Performance Standards:
The SecurView DX software is specified, validated and tested by Hologic, Inc., under a registered ISO 13485 and 21 CFR Part 820 Quality System.
The Secur View DX software complies with ACR/NEMA Digital Imaging Communications in Medicine version PS 3.1-3.18 (DICOM) Set
An Operator's Manual will be provided with each system that is user friendly and comprehensive, thus ensuring safe and effective operation.
### Conclusion:
Similar to the predicate device, the SecurView DX software displays and manipulates DICOM images to be interpreted by a physician or trained medical personnel, providing ample opportunity for competent human intervention. The device and the predicate device share similar performance standards Device failures, which might result in partial or failed transmissions, images, or data, may be recovered from storage or retransmission after correcting the problem(s). Passwords are required for operation and to protect against unauthorized use. Potential hazards have been studied and controlled by a Risk Management Plan. Based on the information supplied in this 510(k), Hologic concludes that the device is substantially equivalent to the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread. The eagle is facing to the right.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
## AUG 1 8 2006
Ms. Gail Yeaker-Daunis Senior Regulatory Specialist Hologic, Inc. 36 Apple Ridge Rd. DANBURY CT 06810
Re: K062107
Trade/Device Name: SecurView DX Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: July 20, 2006 Received: July 24, 2006
#### Dear Ms. Yeaker-Daunis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/3/Picture/10 description: The image is a black and white circular logo. The logo contains the text "1826-1926" at the top, followed by the letters "PA" in a bold, blocky font. Below the letters, the word "Centennial" is written in a smaller font. Three stars are arranged in a horizontal line at the bottom of the logo. The logo has a vintage or commemorative appearance.
Protecting and Promoting Public Health
{4}------------------------------------------------
Page 2 -
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrlv/industry/support/index.l1tml.
Sincerely yours,
Nancy C. Hogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Hologic, Inc. Special 510(k) for SecurView DX
K06 2167
510(k) No.
Indications For Use
Device Name: SecurView DX
The Hologic SecurView DX device is intended for selection, display, manipulation, filming and media interchange of multi-modality images from a variety of different modality systems. It also interfaces to various image storage and printing devices using DICOM or similar interface standards.
The device used with FDA cleared monitors may be used by a trained physician for display, manipulation and interpretation of lossless compressed or non-compressed mammographic images for screening and diagnostic mammography, as well as any other DICOM multi-modality image.
The SecurView DX is typically used by trained professionals, including, but not limited to physicians, radiologists, nurses, medical technicians and assistants.
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use<br>21 CFR 801.109 | <div> <span> <svg height="15" width="15"> <path d="M4.5 9 L10 3 L11 4 L5 10 Z" fill="none" stroke="black"></path> </svg> </span> </div> |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| OR | |
| Over-the-Counter Use | |
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
| 510(k) Number | K062102 |
|---------------|---------|
|---------------|---------|
0131 11 11/2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.