K041555 · Hologic, Inc. · LLZ · Aug 9, 2004 · Radiology
Device Facts
Record ID
K041555
Device Name
SECURVIEW DX SOFTCOPY WORKSTATION
Applicant
Hologic, Inc.
Product Code
LLZ · Radiology
Decision Date
Aug 9, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The Hologic SecurView DX device is intended for selection, display, manipulation, filming and media interchange of multi-modality images from a variety of different storage and printing devices using DICOM or similar interface standards. The device used with FDA cleared monitors may be used by a trained physician for display, manipulation and interpretation of lossless compressed or non-compressed mammographic images for screening and diagnostic mammography, as well as any other DICOM multi-modality image. The SecurView DX is typically used by trained professionals, including, but not limited to, radiologists, technicians and assistants.
Device Story
SecurView DX is a multi-modality review workstation software; accepts DICOM-compliant medical images from external storage/printing devices; provides visualization, manipulation, and interpretation tools for radiologists and technicians. Operates on Windows 2000-based PC hardware with high-resolution monitors. System facilitates clinical review of mammography and other radiological images; supports lossless compressed or non-compressed image display. Output allows clinicians to interpret diagnostic images to support clinical decision-making. Used in clinical environments by trained medical staff.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Software-based workstation; runs on Windows 2000 OS; requires Intel CPU (>=2.0 GHz), >=2.0 GB RAM, >=140 GB HDD. Supports DICOM interface standards for image input/output. Utilizes high-resolution monitors. No specific algorithm architecture or hardware standards (ASTM) detailed.
Indications for Use
Indicated for trained professionals, including radiologists, technicians, and assistants, for the display, manipulation, and interpretation of DICOM multi-modality images, including screening and diagnostic mammography images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
K041555
## SecurView DX AUG - 9 2004 510(k) Summary
Product Name: SecurView DX
Product Classification Name: Picture archiving and communication system
CFR Section: 892.2050 Product Classification Code: LLZ
Classification Panel: Radiology Class II
| Manufacturer: | Hologic, Inc.<br>36 Apple Ridge Road<br>Danbury, CT 06810 USA | | | | |
|-----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|
| Contact Person: | Gail Yaeker-Daunis<br>Telephone Number: (203) 731-8337<br>(203) 731-8440<br>Fax Number: | | | | |
| Date Prepared: | June 4, 2004 | | | | |
| | R. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1.<br>1 2 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 | | | | |
K033400 Seno Advantage, General Electric Medical Systems Predicate Device:
Device Description: The Hologic SecurView DX is mainly a software product. It is Device Description. The Hologie Docurity hadiology images. The Secur View DX is a used for visualization and processing of argitus two high-resolution gray start on horned multi-modality review workstation: It metades we keyboard and workstation keypad.
(FDA cleared for Mammography), 3-button mouse, keyboard and workstation keypad. (I'DA created for mailinegraphy); . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . The soltware accepts mages which have occarded. Image processing is external to the Presentation Specification in the DIOON Station and and of the system or no compression at all when displaying mammography images.
The minimum computer requirements are:
Windows 2000® Operating System
Intel CPU with a clock rate of 2.0 GHz or greater
2.0 GB RAM or greater
140 GB Hard drive or greater, operating at 160 Mbs or greater
CD-ROM/R/RW, DVD+/-RW
10/100 Base TX Network Interface
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with three stripes extending from its head, representing health, services, and human needs. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG - 9 2004
Mr. Richard Follett Vice President, RA/QA Hologic, Inc. 35 Crosby Drive BEDFORD MA 01730 Re: K041555
Trade/Device Name: SecurView DX Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system
Regulatory Class: II Product Code: 90 LLZ Dated: June 4, 2004 Received: June 16, 2004
Dear Mr. Follett:
We have reviewed your Section 510(k) premarket notification of intent to market the device WE have reviewed your Scotton 910(t) pe device is substantially equivalent (for the indications felerenced above and nave decembined ly marketed predicate devices marketed in interstate for use stated in the encrosure) to regally manating to the Medical Device Americal Device of Dove commerce prior to way 26, 1776, the classified in accordance with the provisions of the Federal Food, DVA devices that have been recuasified in accordains approval of a premarket approval application (PMA). and Cosment Act (Rea) that do not require approvince approvisions of the Act. The Y ou may, dierelore, market the detires, baojet requirements for annual registration, listing of general controls provisions of the tice, labeling, and prohibitions against misbranding, and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classince (see a0010) ins. Existing major regulations affecting your device can
may be subject to such additional controls. Existing major regulations of t may be subject to such additional controlly - Lines 800 to 898. In addition, FDA may be found in the Ood of I outstan ongerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not i Picase oe advised hat I DA s issualloo of a device complies with other requirements of the Act that FDA has made a deceimnation that your istered by other Federal agencies. You must comply or any Pederal Statures and regalations administered of the registration and listing (2 CFR Parth in the with an the Act STCquirements, metading, curfacturing practice requirements as set forth in the 807), laocing (21 CFR Part 601), goods and 1820), and if application of the electronic product quality systems (QD) regulation (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will allow you to begin marketing your antial equivalence of your device to a legally premarket notification. The PDA inding of substantial equive and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please (1), please If you desire specific advice for your dones on one of the following numbers, based on the regulation number at the top of the letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Additionally, for questions on the promotion and and the regulation entitled, "Misbranding Othce of Compliance at (301) 574-1057. Floor, production of obtain. Other general by relerence to prematic nonitation - (21 OF Act may be obtained from the Division of Small Information on your responsionalisms and its toll-free number (800) 638-2041 or
Manufacturers, International and Consumer Assistance at its toll-free in the l Manufacturers, International and Collean.br ----------------------------------------------------------------------------------------------------------------------------------
Sincerely yours,
Nancy C. Boyden
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## 510(k) No. K041555
## Indications For Use
## Device Name: SecurView DX
The Hologic SecurView DX device is intended for selection, display, manipulation,
of different The Hologic Secur View DX device is imended 10 images from a variety of different
filming and media interchange of multi-modality images and printing devices u filming and media merchange of indit-modality mages storage and printing devices using DICOM or similar interface standards.
The device used with FDA cleared monitors may be used by a trained physician for The device used with FDA cleared inomitive into of lossless and on non-compressed or non-compressed or non-compressed display, manipulation and interpretation of Nossiess compressed on the country.
mammographic images for screening and diagnostic mammography, as well as any other DICOM multi-modality image.
The SecurView DX is typically used by trained professionals, including, but not limited The Secur View DX is typically dood of isl technicians and assistants.
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | |
|----------------------|--|
| OR | |
| Over-the-Counter Use | |
Nancy C Brogdon
(Division Sign-Off)
Division of Reproductive, Abdominal, and Radiological Devices
| 510(k) Number | K041555 |
|---------------|---------|
|---------------|---------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.