SView 1.0 is a medical image management device intended for viewing, manipulating and analyzing DICOM-compliant medical images acquired from MRI scanners and other DICOM -compliant imaging devices. SView™ 1.0 can be used for real-time image viewing, image manipulation, and analysis that aid a cardiologist or radiologist in their diagnosis. In addition, it facilitates the physician's reporting and reviewing of patient studies.
Device Story
SView 1.0 is a software tool for 2D viewing, manipulation, and analysis of DICOM-compliant medical images; inputs include MRI and other DICOM-compliant image data. Operated on a standard Windows PC, the device supports dual-monitor configurations for simultaneous image viewing and reporting. It provides real-time visualization and analysis capabilities to assist cardiologists and radiologists in clinical diagnosis and study reporting. The device does not perform image digitization and utilizes lossless compression. It is intended for use in clinical settings by healthcare professionals to facilitate diagnostic workflows and patient study management.
Clinical Evidence
Bench testing only. No clinical data was required as the device is a medical image management tool using lossless compression and no image digitizers. Extensive testing was performed by programmers, non-programmers, quality assurance staff, and potential customers.
Technological Characteristics
PC-based software; runs on standard Windows operating system; supports DICOM-compliant images; dual-monitor capability; lossless compression; no image digitizers; standalone medical image management device.
Indications for Use
Indicated for cardiologists and radiologists to view, manipulate, and analyze DICOM-compliant medical images from MRI and other imaging devices to aid in diagnosis, reporting, and review of patient studies.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
## Ko60840 MRI Cardiac Services, Inc. 510(k) Summary
#### Submitter:
MRI Cardiac Services, Inc. 8 West Third Street, M9 Winston-Salem, NC 27101 336-831-1908 (v) 336-727-0919 (f)
7 2006 APR
Date Prepared:
June 1, 2004
#### Contact Person(s):
Scott Huber, President 336-831-1908 (v) 336-727-0919 (f)
Device Trade Name: SView™
#### Device Common Name:
PACS / Medical Image Management Device
#### Classification Name:
Class II - System, Image Processing
Product Code/Regulation Number:
LLZ/892.2050
#### Substantially Equivalent To:
AccuView Diagnostic Imaging Workstation Software Package (K990241) AccuImage Diagnostics Corporation 400 Oyster Point Boulevard, Suite 114 S. San Francisco, CA 94080
#### Device Description:
SView " 1.0 is a tool for 2D (two-dimensional) viewing and manipulation of DICOM compliant MR images and other DICOM-compliant images. The proposed device provides real-time image viewing, manipulation, analysis and reporting.
#### Indications for Use:
SView 1.0 is a medical image management device intended for viewing. manipulating and analyzing DICOM-compliant medical images acquired from MRI scanners and other DICOM -compliant imaging devices. SView™ 1.0 can be used for real-time image viewing, image manipulation, and analysis that aid a cardiologist or radiologist in their diagnosis. In addition, it facilitates the physician's reporting and
8 West Third Street, Suite M9 Winston-Salem, NC 27101 (t)336-831-1908 (f) 336-727-0919 (e) info@mricardiovascular.com
{1}------------------------------------------------
# MRI Cardiac Services, Inc.
reviewing of patient studies.
#### Technological Comparison to Predicate Device:
The proposed and predicate medical image management devices are both devices that can be used for manipulation of DICOM-compliant MR images. The proposed and predicate device can be operated from a personal computer. SView™ 1.0 and a standard Windows operating system. SView is a subset of the AccuView Diagnostic Imaging Workstation features with an optional added monitor to allow a Cardiologist or Radiologist the convenience of using two monitors, one for image viewing and manipulation, and the other for reporting. SView™ 1.0 has substantially equivalent features and specifications versus the existing AccuView Diagnostic Imaging Workstation for those features and specifications the two devices have in common.
#### Laboratory and Clinical Testing
SView 1.0 is intended for use with existing MRI images for real-time image viewing, image manipulation, and analysis that aid a cardiologist or radiologist in their diagnosis. In addition it facilitates the physician's reporting and reviewing of patient studies. The SView™ medical image management device contains no image digitizers and uses only lossless compression. On this basis, MRI Cardiac Services, Inc. believes that clinical investigation is not necessary.
Extensive testing of the device will be performed by programmers, by nonprogrammers, quality assurance staff and by potential customers prior to commercial release.
#### Adverse Effects on Health:
The potential hazards are identified in the Hazard Analysis and are controlled by:
- Designing controls directed at the cause and/or
- Introducing protective measures and/or
- o Warning the Users
All identified hazards are mitigated to minor level of concern.
See Summary of Safety and Effectiveness on the following page.
#### Conclusions:
We conclude that the subject device, SView™ 1.0 is as safe and effective as the predicate device and posses no new questions of safety and effectiveness. SView 114 1.0 performs in accordance with its intended use as well as the AccuView Diagnostic Imaging device currently on the market. MRI Cardiac Services, Inc. considers features of the SView™ 1.0 to be substantially equivalent to the subset of features in common with the AccuView. Diagnostic Imaging device (510(k)990241.)
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with three stripes on its back, representing the three branches of government. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - U.S.A." is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MRI Cardiac Services, Inc. % Mr. Daniel W. Lehtonen Responsible Third Party Official Intertek Testing Services NA, Inc. 2307 East Aurora Rd., Unit B7 TWINSBURG OH 44087
Re: K060840 Trade/Device Name: SView™ Version 1.0 PACS Medical Image Management Device Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II
Product Code: LLZ Dated: March 27, 2006 Received: March 28, 2006
Dear Mr. Lehtonen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling. and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (2) CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
APR 7 2006
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address
http://www.fda.gov/cdrh/industrv/support/index.html.
Sincerely vours.
Nancy C. Brigdon
Nanev C. Broudon Director. Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Indications for Use
501(k) Number: _______________________________________________________________________________________________________________________________________________________________
Device Name: SView™ Version 1.0 PACS/Medical Image Management Device
Indications For Use:
SView 1.0 is a medical image management device intended for viewing, manipulating and analyzing DICOM-compliant medical images acquired from MRI scanners and other DICOM -compliant imaging devices. SView™ 1.0 can be used for real-time image viewing, manipulation and analysis that aid a cardiologist or radiologist in their diagnosis. In addition, it facilitates the physician's reporting and reviewing of patient studies.
Prescription Use X (Per 21 CFR 801 Subpart D) OR
Over- The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Ingram
(Division Sign-Off) Division of Reproductive. Ab and Radiological Devices 510(k) Number
Page 1 of 1
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.