K022872 · General Electric Co. · KPS · Sep 12, 2002 · Radiology
Device Facts
Record ID
K022872
Device Name
GE ROAD WARRIOR PET CT IMAGING SYSTEM
Applicant
General Electric Co.
Product Code
KPS · Radiology
Decision Date
Sep 12, 2002
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1200
Device Class
Class 2
Intended Use
The GE Road Warrior PET CT Imaging System is intended for use in head and whole body attenuation corrected Positron Emission Tomography (PET) imaging, facilitation of emission activity in the patient anatomy by means of integrated CT and PET images, and stand alone head and whole body multislice X-ray computed tomography diagnostic imaging.
Device Story
The GE Road Warrior PET CT Imaging System integrates a Positron Emission Tomography (PET) scanner and a LightSpeed Ultra CT scanner. The device acquires PET emission data and CT transmission data. The CT images are used to correct for non-uniform attenuation in PET images and to provide anatomical localization of PET emission activity. The system also functions as a standalone multislice CT scanner. Operated by trained clinical staff in a diagnostic imaging environment, the system provides fused or standalone images to physicians. These images assist in clinical decision-making by providing both metabolic (PET) and anatomical (CT) information, potentially improving diagnostic accuracy and patient management.
Clinical Evidence
No clinical data. Bench testing only; device evaluated for electrical and radiation safety in conformance with applicable medical device safety standards.
Technological Characteristics
Integrated PET and multislice X-ray CT system. Utilizes CT-based attenuation correction for PET imaging. Design conforms to 21 CFR 820, ISO 9001, and EN 46001 quality systems. Complies with applicable medical device safety standards for electrical and radiation safety.
Indications for Use
Indicated for head and whole-body attenuation-corrected PET imaging, anatomical localization of emission activity via integrated CT/PET, and standalone head and whole-body multislice X-ray CT diagnostic imaging.
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
Predicate Devices
GE Discovery LS Positron Emission Tomography System (K010641)
Related Devices
K030199 — GE MOBILE DISCOVERY ST PET CT IMAGING SYSTEM · General Electric Co. · Feb 11, 2003
K040172 — MODIFICATION TO GE DISCOVERY LS SYSTEM · GE Medical Systems · Apr 7, 2004
K041543 — GE DISCOVERY ST · GE Healthcare · Jun 22, 2004
K042257 — DISCOVERY ST PET/CT IMAGING SYSTEM WITH AN ALTERNATE DETECTOR OPTION · General Electric Co. · Sep 8, 2004
K082244 — PCA-8000A, MARK 11, PET / CT SYSTEM · Toshiba America Medical Systems, In.C · Aug 21, 2008
Submission Summary (Full Text)
{0}------------------------------------------------
Special 510(k) Premarket Notification GE Medical Systems - GE Road Warrior PET CT System August 23, 2002
FP 1 2 2002
# KO22812
## Attachment B:
Summary of Safety and Effectiveness Prepared in accordance with 21 CFR Part 807.92(c).
Image /page/0/Picture/5 description: This document contains information about GE Medical Systems. It lists the submitter as GE Medical Systems, located in Milwaukee, WI. The contact person is D. Duersteler, with contact information including a telephone number of 262-312-7029 and a fax number of 262-312-7144. The document also mentions the device name as GE Road Warrior PET CT Imaging System and the marketed device as GE Medical Systems GE Discovery LS Positron Emission Tomography System, with the number K010641.
Device Description: The GE Road Warrior PET CT Imaging System integrates a GE Positron Emission Tomography system and a GE LightSpeed Ultra CT System. In addition to providing CT and PET standalone capabilities, it uses the CT images to correct for non-uniform attenuation of the PET images and to facilitate localization of the emission activity in the patient anatomy.
Indications for Use: The GE Road Warrior PET CT Imaging System is intended for use in head and whole body attenuation corrected Positron Emission Tomography (PET) imaging, facilitation of emission activity in the patient anatomy by means of integrated CT and PET images, and stand alone head and whole body multislice X-ray computed tomography diagnostic imaging.
Comparison with Predicate Device: The GE Road Warrior PET CT Imaging System is of a comparable type and substantially equivalent to the currently marketed GE Discovery LS CT-PET System. It has the same technological characteristics, is comparable in key safety and effectiveness features, uses the same basic design, construction, and materials, and has the same intended uses as the predicate device.
Summary of Studies: The device has been evaluated for electrical, and radiation safety, and conforms with applicable medical device safety standards.
Clinical Tests: None required.
Conclusion: Intended uses and other key features are consistent with traditional clinical practice, FDA guidelines, and established methods of patient examination. Intended uses and fundamental scientific technology are the same as the legally marketed GE Discovery LS CT-PET System. The design and development process of the manufacturer conforms with 21 CFR 820, ISO 9001 and EN 46001 quality systems. The device conforms to applicable medical device safety standards and compliance is verified through independent evaluation with factory surveillance. Therefore, it is the opinion of GE Medical Systems that the GE Road Warrior PET CT Imaging System is substantially equivalent with respect to safety and effectiveness to devices currently cleared for market.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows a circular logo with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the edge. In the center of the circle is a stylized image of an eagle or other bird with three curved lines representing its wings or feathers. The logo is black and white and appears to be a seal or emblem.
Food and Drug Administration 0 Corporate Boulevard ockville MD 20850
### SEP 1 2 2002
Mr. David Duersteler Safety and Regulatory Engineering GE Medical Systems P.O. Box 414 MILWAUKEE WI 53201
Re: K022872
Trade/Device Name: GE Road Warrior PET CT Imaging System Regulation Number: 21 CFR 892.1200 Regulation Name: Emission conputed tomography system
Regulatory Class: II Product Code: 90 KPS Dated: August 23, 2002 Received: August 29, 2002
Dear Mr. Duersteler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (sec above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050,
{2}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Snigdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Special 510(k) Premarket Notification GE Medical Systems - GE Road Warrior PET CT System August 23, 2002
#### STATEMENT OF INTENDED USE
510(k) Number (if known): K022872
Device Name: GE Road Warrior PET CT Imaging System
### Indications for Use
The GE Road Warrior PET CT Imaging System is intended for use in head and whole body attenuation corrected Positron Emission Tomography (PET) imaging, facilitation of emission activity in the patient anatomy by means of integrated CT and PET images, and stand alone head and whole body multislice X-ray computed tomography diagnostic imaging.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801-109)
OR Over-The-Counter Use
David A. Ingram
and Rad 510(k) Number
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.