FUJIFILM DUAL ENERGY SUBTRACTION (DES) SOFTWARE OPTION

K122454 · Fujifilm Medical System U.S.A., Inc. · KPR · Nov 20, 2012 · Radiology

Device Facts

Record IDK122454
Device NameFUJIFILM DUAL ENERGY SUBTRACTION (DES) SOFTWARE OPTION
ApplicantFujifilm Medical System U.S.A., Inc.
Product CodeKPR · Radiology
Decision DateNov 20, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1680
Device ClassClass 2

Intended Use

Fujifilm's FDR Dual Energy Subtraction (DES) Option may be used with Fujifilm's DR X-ray Systems and is intended to be used by a qualified/trained doctor or technologist for acquiring dual energy subtraction images of human anatomy. DES is intended to assist the physician through the visualization of anomalies by reducing the visibility of underlying / overlying anatomical structures. This device is not intended for mammographic applications.

Device Story

FDR Dual Energy Subtraction (DES) is a software option for Fujifilm DR X-ray systems; acquires two images at different X-ray energies; performs weighted subtraction to generate three processed images: standard radiographic, soft-tissue (bone removed), and bone (soft-tissue removed). Includes Multi-Stage Registration (MSR) to compensate for involuntary patient motion (e.g., heartbeat) between exposures, reducing motion artifacts. Operated by physicians or technologists in clinical settings. Output assists physicians in visualizing anomalies by suppressing overlying/underlying anatomical structures, potentially improving detectability of lung nodules. Does not perform mammography.

Clinical Evidence

No prospective clinical trials. Evidence includes a literature review of studies performed on the device outside the U.S., including a reading evaluation where 10 radiologists assessed 20 images, confirming MSR improves lung nodule detectability and reduces artifacts. Three U.S. clinical reference images were provided to demonstrate MSR effectiveness. Bench testing and software verification were also performed.

Technological Characteristics

Software-based image processing option for digital X-ray systems. Uses weighted subtraction of two images acquired at different energy levels. Includes Multi-Stage Registration (MSR) algorithm for motion compensation. Complies with IEC 60601-1, 60601-1-2, 60601-1-3, 60601-1-4, 60601-2-7, 60601-2-28, 60601-2-32, IEC 60825, and DICOM V3.0 standards.

Indications for Use

Indicated for use by qualified doctors or technologists to acquire dual energy subtraction X-ray images of human anatomy to assist in visualizing anomalies by reducing visibility of underlying/overlying structures. Not for mammographic applications.

Regulatory Classification

Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the Fujifilm Medical Systems, USA logo. The logo is in black and white. The word "FUJIFILM" is in large, bold letters, and the words "Fujifilm Medical Systems, USA" are in smaller letters below. ## FUJIFILM FDR Dual Energy Subtraction (DES) Option 510(k) Summary Date: August 10, 2012 Contact Person: | Name: | Debbie Peacock | |------------|----------------------------| | Title: | Regulatory Affairs Manager | | Telephone: | (203) 602-3774 | | Facsimile: | (203) 363-3813 | ### ldentification of Device: - | Proprietary/Trade Name: | FDR Dual Energy Subtraction (DES) Option | |-------------------------|----------------------------------------------| | Classification Name: | Stationary X-ray System w/Solid State Imager | | Classification: | Panel: Radiology | | CFR Section: | 21 CFR 892.1650, 21 CFR 892.1680 | | Product Codes: | KPR/MQB | | Common Name: | Stationary X-Ray System | #### INDICATIONS FOR USE 1. Fujifilm's FDR Dual Energy Subtraction (DES) Option may be used with Fujifilm's DR X-ray Systems and is intended to be used by a qualified/trained doctor or technologist for acquiring dual energy subtraction images of human anatomy. DES is intended to assist the physician through the visualization of anomalies by reducing the visibility of underlying / overlying anatomical structures. This device is not intended for mammographic applications. #### II. DEVICE DESCRIPTION Dual Energy Subtraction (DES) is an advanced radiographic imaging application designed as an option to be used with our cleared FDR AcSelerate X-ray system, and will also be used in future FUJFILM X-ray systems. Two images are acquired at mulually different X-ray energies. These images are then subject to a weighted subtraction that can remove structures (e.g. soft tissue structures or bone structures) of a specific x-ray absorption characteristic. This results in a completed DES exam that typically consists of three processed images, (1) a standard radiographic image, (2) a soft-tissue image with the bone information removed, and (3) a bone image with the soft tissue information removed. To suppress involuntary motion artifacts which are tvpical in dual-exposure energy subtraction, Fujifilm's DES option includes Multi-Stage Registration (MSR) which compensates for involuntary patient motion, such as the patient's heartbeat, that can {1}------------------------------------------------ K122454 P. 2 of 3 # Fuilfilm Medical Systems produce slight differences in anatomic registration (seen as artifacts) between the first exposure (high energy image), and the second exposure (low energy image). MSR is added to suppress the motion artifacts seen in the DES image and maintains the image quality. #### SUMMARY OF STUDIES 011. System testing was performed and certified to the following International Standards: - IEC 60601-1: Medical Electrical Equipment-Part 1: General requirements for A safety, 1988; Amendment 1, 1991-11, Amendment 2, 1995 (General) - > UL 60601-1: Medical Electrical Equipment -Part 1: General requirements for safety. 2003 IEC 60601-1-2: Medical Electrical Equipment-Part 1-2. General requirements A for safety-Collateral standard: Electromagnetic compatibility-requirements and tests. 2007. IEC 60601-1-3: Medical Electrical Equipment-Part 1, General requirements A for safety-3. Collateral standard: General requirements for radiation protection in diagnostic X-rav equipment. 1994. A IEC 60601-1-4: Medical Electrical Equipment-Part 1-4; General requirements for safety-Collateral Standard: Programmable Electrical Medical systems. 1996 +A1, 1999. A IEC 60601-2-7: Medical Electrical Equipment-Part 2-7: Particular requirements for the safety of high-voltage generators of diagnostic X-ray generators. 1998. IEC 60601-2-28: Medical Electrical Equipment-Part 2: Particular A requirements for the safety of X-ray source assemblies and X-ray tube assemblies for medical diagnosis. 1993. - IEC 60601-2-32: Medical Electrical Equipment-Part 2: Particular requirements A for the safety of associated equipment of X-ray equipment. 1994. - A IEC 60825: Safety of Laser Products-Part 1: Equipment classification and requirements. 2007. - > DICOM V3.0: Digital Imaging and Communications in Medicine (DICOM). 2007. A literature review was submitted to discuss clinical and phantom studies performed on the outside US version of the proposed device. Studies were performed on the predicate device in Japan. A reading evaluation was performed. Ten radiologists evaluated 20 subject images and concluded that (1) Multi-Stage Registration feature may improve DES detectability of lung nodules in the lower right lung and (2) it has been confirmed by interviews with doctors that the artifact reduction was effective. Three U.S. clinical DES reference images are included in this submission with and without MSR applied. These images are included to provide a visual demonstration of the effectiveness of Fujifilm's DES with MSR for artifact reduction. Results of the Software Verification, bench testing, clinical image samples and literature review support a substantial equivalence determination. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image contains the Fujifilm logo, followed by the text "Fujifilm Medical Systems, USA". Below that, the text "IV. SUBSTANTIAL EQUIVALENCE" is present. In the upper right corner, handwritten text reads "K122454 P. 3 of 3". Fujifilm's FDR Dual Energy Subtraction (DES) Option is a modification to our currentlycleared FCR 9501ES (K952862) and is also substantially equivalent to General Electric's Dual Energy and Tissue Equalization Software Options for Digital Radiographic Systems (K013481). A detailed Substantial Equivalence Comparison is provided in Section 12. #### V. CONCLUSION The proposed optional FDR Dual Energy Subtraction software tested with the currentlycleared FDR AcSelerate System has very similar Indications for Use, Functional and Technical requirements as the currently-cleared General Electric's Dual Energy and Tissue Equalization Software Options for Digital Radiographic Systems (K013481). The Energy Subtraction algorithm is unchanged from our currently-cleared FCR 9501ES (K952862). After analyzing bench and certified system test data, as well as literature review and the inclusion of confirmatory images with and without MSR applied, we feel that the proposed and predicate devices can be considered to be as safe and effective as the currently-cleared predicate devices. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and body. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular fashion around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002 Ms. Debbie Peacock Regulatory Affairs Manager FUJIFILM Medical Systems, USA, Inc. 419 West Avenue STAMFORD CT 06902 December 5, 2012 Re: K122454 Trade/Device Name: FUJIFILM Dual Energy Subtraction (DES) software option Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: KPR Dated: October 26, 2012 Received: October 31, 2012 Dear Ms. Peacock: This letter corrects our substantially equivalent letter of November 20, 2012. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {4}------------------------------------------------ or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostics and Radiological Health at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Michael D. O'Hara Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use # FUJIFILM FDR Dual Energy Subtraction (DES) Option 510(k) Number (if known): FUJIFILM FDR Dual Energy Subtraction (DES) Option Device Name: Indications For Use: Fujifilm's FDR Dual Energy Subtraction (DES) Option may be used with Fujifilm's DR X-ray Systems and is intended to be used by a qualified/trained doctor or technologist for acquiring dual energy subtraction images of human anatomy. DES is intended to assist the physician through the visualization of anomalies by reducing the visibility of underlying / overlying anatomical structures. This device is not intended for mammographic applications. × Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) Muhammad D. Khan Page 1 of of (Division Sign Division of Radiological Health of in Vitro Diegnostics and Ra 510(k) K122454
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...