SOFTVIEW, MODEL 2.01

K092363 · Riverain Medical Group · LLZ · Mar 18, 2010 · Radiology

Device Facts

Record IDK092363
Device NameSOFTVIEW, MODEL 2.01
ApplicantRiverain Medical Group
Product CodeLLZ · Radiology
Decision DateMar 18, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesAI/ML, Software as a Medical Device, Real-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K092363 · Mar 18, 2010SOFTVIEW, MODEL 2.01Riverain Medical GroupIndependent dataset of clinical Dual Energy Subtraction (DES) soft tissue images (GE and Fuji modalities)Retrospective clinical images were used to conduct a comparative analysis of contrast-to-noise ratio (CNR) to establish substantial equivalence and validate the performance of the SoftView mathematical model against existing clinical DES systems.Comparative analysis; Contrast-to-noise ratio (CNR); Dual Energy Subtraction (DES); Retrospective clinical dataset

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Comparative analysis of CNR; Retrospective comparative analysisChest radiographic images (stratified by modality: CR/Fuji DES and DR/GE DES; and lung regions: pleural, mid-lung, and hilum)GE and Fuji dual energy subtraction (DES) soft tissue imagesContrast-to-noise ratio (CNR) of residual bone in soft tissue images; equivalency and non-inferiority testing

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
Lung Nodule Detection (standalone)Mathematical model of a DES system using image features.—Sensitivity: 66.3% (95% CI: 63.1-69.7); Specificity: 91.8% (95% CI: 89.5-93.5)——Reader study to assess benefit of SoftView for detecting actionable lung nodules.>1 (radiologist)
Lung Nodule Detection (reader improvement)Mathematical model of a DES system using image features.—Mean difference in LROC curve: -0.098 (95% CI: -0.116 to -0.080)——Reader study to assess benefit of SoftView for detecting actionable lung nodules.>1 (radiologist)

Indications for Use

SoftView is intended to generate an enhanced, secondary digital radiographic image of the chest. The enhanced AP or PA image of the chest provides improved visibility of the lung parenchyma through bone suppression and tissue equalization, and may facilitate discerning the presence or absence of nodules. The SoftView image provides adjunctive information and is not a substitute for the original PA/AP image. This device is intended to be used by trained professionals, such as physicians, radiologists, and technicians on patients with risk of having lung nodules and is not intended to be used on pediatric patients.

Device Story

SoftView is a post-processing software application for digital chest radiographs; inputs are standard PA/AP chest X-rays. Device uses a mathematical model to suppress bone structures and perform tissue equalization, simulating dual-energy subtraction (DES) results without requiring additional radiation exposure. Operated by physicians, radiologists, or technicians in clinical settings. Output is an enhanced secondary image provided as adjunctive information to the original radiograph; assists clinicians in discerning lung nodules. Clinical benefit includes improved nodule detection sensitivity without increased patient radiation dose.

Clinical Evidence

Reader study (n=not specified) evaluated SoftView's impact on lung nodule detection using LROC curves. Mean difference in LROC area was -0.098 (95% CI: -0.116 to -0.080), indicating statistically significant improvement. Sensitivity increased from 49.5% (95% CI: 45.9-53.0) to 66.3% (95% CI: 63.1-69.7) with SoftView. Specificity was 96.1% (95% CI: 95.0-97.1) without and 91.8% (95% CI: 89.5-93.5) with SoftView.

Technological Characteristics

Post-processing software application; mathematical model of dual-energy subtraction (DES) process. Operates on standard digital chest radiographs (CR/DR). No hardware components; no additional radiation source. Connectivity via standard digital imaging workflows.

Indications for Use

Indicated for adult patients at risk of lung nodules. Used to generate enhanced secondary digital chest radiographs (AP/PA) via bone suppression and tissue equalization to assist in nodule detection. Not for pediatric use.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ KOS2363 Traditional 510(k) Premarket Notification SoftView™ Riverain Medical ## 510(K) SUMMARY | Submission Date: | August 4, 2009 | MAR 1 8 2010 | |------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------| | Submitter Information: | | | | Company Name: | Riverain Medical Group, LLC. | | | Company Address: | 3020 South Tech Blvd., Miamisburg, OH 45342-4860 | | | Contact Person: | Jennifer Vetter<br>Director, Regulatory Affairs and Quality Assurance<br>Riverain Medical<br>800.990.3387<br>937.425.6493<br>jvetter@riverainmedical.com | | | Device Information: | | | | Trade Name: | SoftView™ | | | Regulation Number: | 21 CFR §892.2050 | | | Regulation Name: | System, Image Processing, Radiological | | | Regulatory Class: | Class II | | | Product Code: | LLZ | | | | Predicate Device:<br>Dual Energy and Tissue Equalization Software Options<br>(K013481)<br>GE Medical Systems<br>Class II | | | Device Description: | SoftView is a dedicated post-processing application which<br>suppresses bone structures from digital radiographic image<br>of the chest. | | | Intended Use: | SoftView is intended to generate a bone-suppressed image<br>from a original PA/AP chest radiograph. | | | Indications for Use: | SoftView is intended to generate an enhanced, secondary<br>digital radiographic image of the chest. The enhanced AP<br>or PA image of the chest provides improved visibility of<br>the lung parenchyma through bone suppression and tissue | | {1}------------------------------------------------ equalization, and may facilitate discerning the presence or absence of nodules. The SoftView image provides adjunctive information and is not a substitute for the original PA/AP image. This device is intended to be used by trained professionals, such as physicians, radiologists, and technicians on patients with risk of having lung nodules and is not intended to be used on pediatric patients. #### Comparison to Predicate Device: SoftView is substantially equivalent to the cited predicate device. Differences in the design and performance from the cited predicate device do not affect either the safety or effectiveness of SoftView for its intended use. Conclusion: SoftView is a mathematical model of a DES system that operates on a standard chest X-ray. It is an accurate representation of the soft tissue image produced by the predicate device's hardware/software process. The model is built from DES data by using simple image features extracted from the standard PA, along with target values derived from a DES soft tissue image. Thus, although SoftView does not generate the soft tissue image based on two exposures to the patient in real time, it is an accurate mathematical model of the process. The result of SoftView processing is a soft tissue image of the patient, consistent with that produced by a DES device but without requiring any additional radiation dose to the patient. Effectiveness of this model was demonstrated both by a reader study and by a comparative analysis of the contrast-to-noise ratio (CNR) of the residual bone in the predicate device's images in the soft tissue rib and clavicle regions, relative to the subject device. Reader Study Results: A reader study was conducted to assess the benefit of SoftView to a radiologist for detecting actionable lung nodules. Reader performance was quantified by the area under the Localization Receiver-Operating Characteristic (LROC), curve which measures the conjoint ability to detect and correctly localize a true positive location on the radiograph. The difference in the average area under the LROC curve with and without the aid of SoftView was used to assess performance. The mean difference in the area under the curves was -0.098 (95% CI: -0.116 to -0.080), a statistically significant improvement. Sensitivity was 49.5% (95% CI: {2}------------------------------------------------ Riverain Medical ### Traditional 510(k) Premarket Notification SoftViewTM 45.9-53.0) without SoftView and 66.3% (95% CI: 63.1-69.7) with the addition of the SoftView image. Specificity was 96.1% (95% CI: 95.0-97.1) with the standard image and 91.8% (95% CI: 89.5-93.5) with the SoftView image. Substantial Equivalence: To establish substantial equivalence, a comparative analysis of the contrast-to-noise ratio (CNR) of the residual bone in the predicate device's soft tissue images versus the subject device was performed. A CNR analysis was performed for both the GE and Fuji dual energy subtraction (DES) soft tissue images. > Analysis of scatter plots of the contrast-to-noise ratio (CNR) for the residual rib objects for an independent dataset was used to compare the SoftView correlative relationship relative to DES. Statistical hypothesis tests were also performed on the independent dataset. Two series of tests were performed, equivalency and non-inferiority tests. Data were stratified across modality, CR (Fuji DES), DR (GE DES), and lung regions, i.e., pleural, mid-lung, and hilum. > For the indicated strata of the data, it was demonstrated that SoftView is equivalent to the DES soft tissue images. The exception was the middle area of the lung for the DR modality (GE DES). SoftView was found to be better in this region based on an "ideal" CNR of the residual ribs in the soft tissue image being 0.0. Furthermore, a non-inferiority test demonstrated that SoftView was non-inferior to the DR and CR dual energy devices soft tissue images across all strata. Descriptive statistics of the means and standard deviations of the CNRs across the different modalities and regions of the lungs were in agreement. {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the seal of the Department of Health & Human Services-USA. The seal is circular and contains an abstract image of an eagle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged around the top half of the circle. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring. MD 20993-0002 # MAR 1 8 2010 Ms. Jennifer Vetter Director-Regulatory and Quality Assurance Riverain Medical Group, LLC 3020 South Tech Blvd. MIAMISBURG OH 45342 Re: K092363 Trade/Device Name: Softview™M Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: February 5, 2010 Received: February 12, 2010 Dear Ms. Vetter: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of {4}------------------------------------------------ medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. Donald J. Trump Donald J. St.Pierre Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ #### 2.0 INDICATIONS FOR USE STATEMENT 510(k) Number (if known): K Device Name: SoftView™ Indications for Use: SoftView is intended to generate an enhanced, secondary digital radiographic image of the chest. The enhanced AP or PA image of the chest provides improved visibility of the lung parenchyma through bone suppression and tissue equalization, and may facilitate discerning the presence or absence of nodules. The SoftView image provides adjunctive information and is not a substitute for the original PA/AP image. This device is intended to be used by trained professionals, such as physicians, radiologists, and technicians on patients with risk of having lung nodules and is not intended to be used on pediatric patients. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE OF NEEDED) | Concurrence of CDRH, Office of Device Evaluation (ODE) | | | |--------------------------------------------------------|--|--| | | | | (Division Sign-Off) Page 1 of 1 Division Sign Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety 510K K092363
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...