K221116 · Xact Robotics, Ltd. · JAK · May 13, 2022 · Radiology
Device Facts
Record ID
K221116
Device Name
XACT ACE Robotic System
Applicant
Xact Robotics, Ltd.
Product Code
JAK · Radiology
Decision Date
May 13, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1750
Device Class
Class 2
Indications for Use
The XACT ACE Robotic System is a user-controlled positioning system intended to assist in the planning and advancement of an instrument during Computed Tomography (CT) guided percutaneous procedures. The system is used for trajectory planning and is intended to assist the physician in positioning of an instrument, such as a needle, where CT imaging is used for target trajectory planning and intraoperative tracking.
Device Story
User-controlled robotic positioning system for CT-guided percutaneous procedures; assists physicians in trajectory planning and instrument advancement. System inputs: CT images for registration and target planning. Operation: Physician defines trajectory; system registers coordinate system to CT images; robot positions and steers instrument (needle or ablation probe) along trajectory; physician controls advancement via foot pedal. Includes monitoring of respiration-associated motion. Components: Robot unit (placed on patient) and console (control unit, computer, monitor). Output: Visualized instrument position on workstation for review. Clinical benefit: Enables precise instrument placement to target; reduces manual positioning complexity. Used in CT suites by physicians.
Clinical Evidence
Bench testing and simulated clinical environment validation. Bench testing verified physical characteristics (max curvature) of ablation probes and compatibility with the Insertion Kit. Simulated clinical validation in a CT suite using phantoms evaluated tip-to-target accuracy across various targets and depths. Results demonstrated the device meets performance specifications with accuracy comparable to biopsy introducers, with no changes to clinical workflow.
Technological Characteristics
Robotic positioning system comprising a robot unit and console workstation. Integrates with CT scanners for image-guided percutaneous intervention. Features a registration process between device coordinates and real-time CT images. Employs an instrument-specific Insertion Kit for securing needles or ablation probes. Software-controlled steering along predefined trajectories. Foot-pedal operated advancement. No significant changes to system software architecture from predicate.
Indications for Use
Indicated for physicians performing CT-guided percutaneous procedures requiring instrument positioning and trajectory planning. Applicable for use with needles and ablation probes.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
May 13, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
XACT Robotics Ltd. % Jonathan Kahan, Partner Hogan Lovells US LLP 555 Thirteenth Street NW WASHINGTON DC 20004
Re: K221116
Trade/Device Name: XACT ACE Robotic System Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: JAK Dated: April 15, 2022 Received: April 15, 2022
Dear Mr. Kahan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laurel Burk, Ph.D. Assistant Director Diagnostic X-ray Systems Team DHT 8B: Division of Radiological Imaging and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221116
Device Name XACT ACE Robotic System
#### Indications for Use (Describe)
The XACT ACE Robotic System is a user-controlled positioning system intended to assist in the planning and advancement of an instrument during Computed Tomography (CT) guided percutaneous procedures. The system is used for trajectory planning and is intended to assist the physician in positioning of an instrument, such as a needle, where CT imaging is used for target trajectory planning and intraoperative tracking.
| Type of Use (Select one or both, as applicable) |
|----------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
| <span style="font-family: sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) SUMMARY
#### XACT Robotics' XACT ACE Robotic System K221116
### Submitter
XACT Robotics, Ltd. 8 Hatochen Street, PO Box 3097 Caesarea 3079861 Israel Phone: 972 (0)4 770 0153 Facsimile: 972 (0)4 770 0161
Contact Person: Chen Levin, CEO
Date Prepared: April 15, 2022
#### Name of Device:
XACT ACE Robotic System
#### Common or Usual Name:
CT Stereotactic Accessory
#### Classification Name:
21 CFR 892.1750; Computed tomography X-ray system
#### Regulatory Class:
Class II
#### Product Code:
JAK
#### Predicate Devices:
- XACT Robotic System, ACE Model (K201586) 0
#### Purpose of the Special 510(k) Notice
The XACT ACE Robotic System is a modification to the XACT Robotic System, ACE Model, which involves expansion of the list of instruments supported by the device during CT-guided percutaneous procedures, to include four types of ablation probes. To accommodate these ablation probes, an additional configuration of the Insertion Kit is introduced, to secure the ablation probe's handle, which is different than a biopsy introducer's head.
#### Intended Use
The XACT ACE Robotic System is a user-controlled positioning system intended to assist in the planning and advancement of an instrument during Computed Tomography (CT) guided percutaneous
{4}------------------------------------------------
procedures. The system is used for trajectory planning and is intended to assist the physician in positioning of an instrument, such as a needle, where CT imaging is used for target trajectory planning and intraoperative tracking.
# Device Description
The XACT ACE Robotic System is a user-controlled positioning system intended to assist in the planning and advancement of instruments during Computed Tomography (CT) guided percutaneous procedures. The system is used for trajectory planning based on CT images and is intended to assist the physician in positioning of an instrument, such as a needle, and reviewing instrument position during advancement to the target. The system quides (i.e., positions and steers) the instrument according to a predefined trajectory. The physician controls advancement of the instrument along the trajectory using a foot pedal. The system also allows for monitoring of motion associated with respiration during the procedure.
The XACT ACE Robotic System comprises the following main components:
- XACT ACE Robot which is placed on the patient and includes the robot positioning unit & the insertion module assembly
- XACT ACE Console which includes a Control Unit, central computer (in the Control Unit) and monitor workstation for user trajectory planning, user interface and review of instrument position.
# Technological Characteristics
The XACT ACE Robotic System allows for planning of percutaneous CT-quided procedures and tracking and positioning of the instrument during the procedure.
Both the XACT ACE Robotic System and its previously cleared predicate are designed and intended for planning and positioning of instruments for percutaneous intervention under imaging guidance of CT scanners. The systems position the instrument according to a predefined trajectory following a registration process between the device's coordinate system and real-time CT images. The user advances the instrument through several checkpoints using a foot pedal.
Both systems are comprised of the same components and accessories. Further, the system software for the XACT ACE Robotic System does not introduce any new features or significant changes to existing features. Although there are minor differences between the subject and predicate device, namely supporting use of the device in the positioning of ablation probes, these differences do not raise new or different questions of safety or efficacy.
#### Performance Data
The performance testing included compatibility testing between the instruments (ablation probes), the Instrument Insertion Kit securing them and the robotic system. According to a predefined protocol -"Needle Properties", the physical characteristics of each probe were first measured including their "Max. Curvature", which defines the maximum, non-vertical force that can be applied without deforming the probe. Then, the robotic system was used to steer each of the probes to a variety of targets while closely monitoring the performance of the Instrument Insertion Kits and the overall accuracy in reaching the targets. This verification test demonstrated that Instrument Insertion Kit
{5}------------------------------------------------
efficiently secures the probes and advance them to the target without any physical damage to the probes. Furthermore, all relevant physical characteristics were adequately measured and configured in the robotic system allowing steering to the targets while meeting performance specifications.
Then, the XACT ACE Robotic System was validated in a simulated clinical environment, in a CT suite using a specific phantom. This test was performed to verify it steers the compatible listed ablation probes to a variety of targets ("organs" and target depths), while meeting similar performance characteristics (tip-to-target average accuracy) compared to biopsy introducers. All trajectories included several checkpoints and performance was evaluated based on CT images obtained during the procedures. Test results demonstrated that the device meets all of its performance characteristics with no changes to the clinical workflow.
# Conclusions
The XACT ACE Robotic System has the same intended uses and indications for use, technological characteristics and principles of operation as its predicate device. The limited differences in enabling support of ablation probes in addition to biopsy introducers do not affect the safety or effectiveness of the device. Performance tests demonstrates that the XACT ACE Robotic System is as safe and effective as the predicate device. Thus, the XACT ACE Robotic System is substantially equivalent.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.