The XACT ACE Robotic System is a user-controlled positioning system intended to assist in the planning and advancement of an instrument during Computed Tomography (CT) guided percutaneous procedures. The system is used for trajectory planning and is intended to assist the physician in positioning of an instrument, such as a needle, where CT imaging is used for target trajectory planning and intraoperative tracking.
Device Story
The XACT ACE Robotic System is a user-controlled robotic positioning system for CT-guided percutaneous procedures. It consists of a robot unit mounted on the patient and a console with a workstation. The system uses CT images for trajectory planning and intraoperative tracking. The physician defines a target trajectory; the robot then steers and positions an instrument (e.g., a needle) along this path. Advancement is controlled by the physician via a foot pedal or a remote-control accessory (ACE Xtend) located in the control room. The system monitors respiratory motion during the procedure. By automating the steering and positioning of the instrument, the device assists the physician in achieving accurate target placement in a single insertion, potentially reducing procedure time and improving clinical outcomes in abdominal and lung interventions.
Clinical Evidence
Clinical evidence includes two prospective trials. Abdominal study (NCT03008603): 60 patients (avg age 66), 100% reached target in one insertion, avg steering time 8.91 min. Lung study (NCT04651517): 20 patients (avg age 73.4), avg steering time 6.7 min. One mild device-related adverse event (skin scratch) reported. Data confirms device performance, safety, and accuracy in navigating instruments to target sites.
Technological Characteristics
System comprises a robot positioning unit, insertion module, and console workstation. Operates via registration of device coordinate system to real-time CT images. Includes ACE Xtend remote-control accessory for control room operation. Software manages trajectory planning, instrument steering, and respiratory motion monitoring. Electrical safety per IEC 60601-1; EMC per IEC 60601-1-2.
Indications for Use
Indicated for patients undergoing CT-guided percutaneous procedures requiring instrument positioning and trajectory planning. No specific age or gender contraindications provided.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
September 23, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
XACT Robotics, Ltd. % Jonathan S. Kahan Partner Hogan Lovells US LLP 555 Thirteenth Street NW WASHINGTON DC 20004
Re: K213759
Trade/Device Name: XACT Robotic System, ACE Model Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: JAK Dated: August 26, 2022 Received: August 26, 2022
# Dear Jonathan Kahan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laurel Burk, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213759
Device Name XACT Robotic System, ACE Model
#### Indications for Use (Describe)
The XACT ACE Robotic System is a user-controlled positioning system intended to assist in the planning and advancement of an instrument during Computed Tomography (CT) guided percutaneous procedures. The system is used for trajectory planning and is intended to assist the physician in positioning of an instrument, such as a needle, where CT imaging is used for target trajectory planning and intraoperative tracking.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------|--|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | |
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### K213759 510(k) SUMMARY
### XACT Robotic Ltd.'s XACT Robotic System, ACE Model
#### Submitter
XACT Robotics, Ltd. 8 Hatochen Street, PO Box 3097 Caesarea 3079861 Israel Phone: 972 (0)4 770 0153 Facsimile: 972 (0)4 770 0161 Contact Person: Tsvia Erlich, Vice President, RA & Clinical Affairs
#### Date Prepared:
August 24, 2022
#### Name of Device:
XACT Robotic System, ACE Model
#### Common or Usual Name:
CT Stereotactic Accessory
#### Classification Name:
21 CFR 892.1750; Computed tomography X-ray system
## Regulatory Class:
Class II
## Product Code:
JAK
#### Predicate Devices
XACT Robotic System, ACE Model (K201586)
{4}------------------------------------------------
## Device Description
The XACT ACE Robotic System is a user-controlled positioning system intended to assist in the planning and advancement of instruments during Computed Tomography (CT) quided percutaneous procedures. The system is used for trajectory planning based on CT images and is intended to assist the physician in positioning of an instrument, such as a needle, and reviewing instrument position during advancement to the target. The system guides (i.e., positions and steers) the instrument according to a predefined trajectory. The physician controls advancement of the instrument along the trajectory using a foot pedal or a pushbutton on the control room console. The system also allows for monitoring of motion associated with respiration during the procedure.
The XACT ACE Robotic System comprises the following main components:
- . XACT ACE Robot which is placed on the patient and includes the robot positioning unit & the insertion module assembly
- . XACT ACE Console which includes a Control Unit, central computer (in the Control Unit) and monitor workstation for user trajectory planning, user interface and review of instrument position
## Intended Use / Indications for Use
The XACT ACE Robotic System is a user-controlled positioning system intended to assist in the planning and advancement of an instrument during Computed Tomography (CT) guided percutaneous procedures. The system is used for trajectory planning and is intended to assist the physician in positioning of an instrument, such as a needle, where CT imaging is used for target traiectory planning and intraoperative tracking.
## Summary of Technological Characteristics
The XACT ACE Robotic System allows for planning of percutaneous CT-guided procedures and tracking and positioning of the instrument during the procedure. The current version of the XACT ACE Robotic System includes a modification to an optional, remote-control accessory (ACE Xtend). The ACE Xtend allows the control of the system and the advancement of an instrument from the control room. The use of ACE Xtend is limited to facilities that have appropriate view from the control room, of both, the patient and the XACT ACE Robot mounted on the patient.
Both the XACT ACE Robotic System and its previously cleared predicate are designed and intended for planning and positioning of instruments for percutaneous intervention under imaging guidance of CT scanners. The systems position the instrument according to a predefined trajectory following a registration process between the device's coordinate system and real-time CT images. The user advances the instrument through several checkpoints using a foot pedal (or advancement button if controlling the XACT ACE Robot with ACE Xtend from the control room). Both systems have the same intended use/indication for use, target patient population, compatible imaging system, clinical procedure and compatible type of instruments.
{5}------------------------------------------------
Both systems are comprised of the same components and accessories. The XACT ACE Robot and Console are the same. The Remote-Control box of the XACT ACE Robotic System (ACE Xtend) and its previously cleared predicate are located in the control room of the CT suit and duplicates the Emergency Stop Button and the monitor while the subject device also duplicates the Console GUI and the needle advancement button (Foot Pedal of the predicate device).
Further, the system software for the XACT ACE does not introduce any new features or significant changes to existing features. The software was updated to facilitate the use of the ACE Xtend and prevent the control from both units at the same time. Although there are minor differences between the subject and predicate device, namely updated labeling to incorporate the results of the abdominal and lung clinical studies, as well as updated remote control of the XACT ACE Robot from the control room, these differences do not raise new or different questions of safety or efficacy. Thus, the current XACT ACE Robotic System, is substantially equivalent to its predicate.
## Performance Data
The following performance/safety tests were conducted with the XACT Robotic System, ACE Model:
- Electrical safety (per IEC 60601-1)
- EMC (per IEC 60601-1-2)
- . Steering Accuracy test evaluated system performance including tip-to-target accuracy when controlled from the ACE Xtend in a simulated clinical setup,
- . Software verification and validation
## Clinical Data
The XACT ACE Robotic System has been assessed in two clinical trials evaluating the device in both abdominal and lung CT-guided percutaneous procedures.
The abdominal study (NCT03008603) was held at 2 centers. Study population consisted of 60 completers, of which 33 (55%) were men and 27 (45%) women, average age was 66 ± 14.3 years (range: 35-89). The average steering time was 8.91±5.05 min. All procedures completed with the ACE system reached the target in one needle insertion.
The Lung study (NCT04651517) was held at 2 clinics in the US. Study population consisted of 20 patients 10 men (50%) and 10 women (50%). The average age was 73.4 ± 6.5 years (63-85). Lesions were distributed throughout each lobe of the lung. Skin to lesion distance was less than 8 cm in 14 patients (70%) and between 8-12 cm in the remaining 6 patients (30%). 6 lesions (30%) were smaller than one (1) cm, 14 (70%) were larger than one cm. The average steering time was 6.7±2.9 min There was one, mild, device related adverse events (skin scratch) which resolved without treatment and no device related serious adverse events.
Clinical data in lung procedures confirm the performance and safety of the device, supporting the same performance profile for the device in other CT guided percutaneous procedures. These data demonstrate that the device can be used for planning and navigating an instrument
{6}------------------------------------------------
(e.g. biopsy introducer) to the target site meeting the system's accuracy specifications with minimal adverse events.
#### Conclusions
The XACT ACE Robotic System has the same intended uses and indications for use, technological characteristics and principles of operation as its predicate device. The limited differences in the labeling and remote operation of the device do not affect the safety or effectiveness of the device. Clinical data demonstrates that the XACT ACE Robotic System is as safe and effective as the predicate device. Thus, the XACT ACE Robotic System is substantially equivalent.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.