K151673 · Toshibamedical Systems Corporation · JAK · Nov 27, 2015 · Radiology
Device Facts
Record ID
K151673
Device Name
Aquilion ONE Vision with FIRST 1.0 (CCRS-001A)
Applicant
Toshibamedical Systems Corporation
Product Code
JAK · Radiology
Decision Date
Nov 27, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Indications for Use
This device is indicated to acquire and display cross sectional volumes of the whole body, to include the head, with the capability to image whole organs in a single rotation. Whole organs include but are not limited to brain, heart, pancreas, etc. The Aquilion ONE has the capability to provide volume sets of the entire organ. These volume sets can be used to perform specialized studies, using indicated software/hardware, of the whole organ by a trained and qualified physician. FIRST 1.0 is an iterative reconstruction algorithm intended to reduce exposure dose and improve high contrast spatial resolution for chest (excluding cardiac), abdomen and pelvis applications. This algorithm is not intended for head or extremity applications.
Device Story
Whole body multi-slice helical CT scanner; consists of gantry, couch, and console. Captures cross-sectional volume data sets; processes and displays images for physician review. Incorporates FIRST 1.0, an iterative reconstruction algorithm, to reduce radiation exposure dose and improve high-contrast spatial resolution. Used in clinical settings by trained physicians to perform specialized organ studies. Output consists of diagnostic-quality cross-sectional images. Benefits include reduced patient radiation dose and improved image quality (spatial resolution, artifact reduction) compared to traditional filtered back projection (FBP).
Clinical Evidence
Bench testing only. Evaluated using phantoms (MITA-FDA phantom with body ellipse) to assess spatial resolution, CT number accuracy, contrast-to-noise ratio, noise properties, uniformity, slice sensitivity profile, and low-contrast detectability. Quantitative dose reduction evaluation showed up to 82.2% dose reduction vs. FBP. Artifact reduction confirmed via visual inspection and image SD values in shoulder regions. Diagnostic quality confirmed by American Board Certified Radiologist review of representative chest, abdomen, and pelvis images.
Technological Characteristics
Multi-slice helical CT scanner. Materials/components based on predicate Toshiba CT systems. Iterative reconstruction (FIRST 1.0) and AIDR 3D algorithms. Standards: IEC 60601-1, 60601-1-1, 60601-1-2, 60601-1-3, 60601-1-4, 60601-1-6, 60601-2-28, 60601-2-32, 60601-2-44, IEC 60825-1, IEC 62304, IEC 62366, NEMA PS 3.1-3.18, NEMA XR-25, NEMA XR-26. Radiation safety per 21 CFR 1010/1020.
Indications for Use
Indicated for patients requiring cross-sectional imaging of the whole body, including head and whole organs (e.g., brain, heart, pancreas). FIRST 1.0 algorithm indicated for chest (excluding cardiac), abdomen, and pelvis applications; not indicated for head or extremity applications.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 27, 2015
Toshiba Medical Systems Corporation % Orlando Tadeo, Jr. Manager, Regulatory Affairs Toshiba America Medical System, Inc. 2441 Michelle Drive TUSTIN CA 92780
Re: K151673
Trade/Device Name: Aquilion ONE Vision with FIRST 1.0 (CCRS-001A) Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: October 20, 2015 Received: October 21, 2015
Dear Mr. Tadeo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert Ools
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151673
Device Name Aquilion ONE Vision with FIRST 1.0 (CCRS-001A)
#### Indications for Use (Describe)
This device is indicated to acquire and display cross sectional volumes of the whole the head, with the capability to image whole organs in a single rotation. Whole organs include but are not limited to brain, heart, pancreas, etc.
The Aquilion ONE has the capability to provide volume sets of the entire organ. These volume sets can be used to perform specialized studies, using indicated software, of the whole organ by a trained and qualified physician.
FIRST 1.0 is an iterative reconstruction algorithm intended to reduce exposure dose and improve high contrast spatial resolution for chest (excluding cardiac), abdomen and pelvis applications. This algorithm is not intended for head or extremity applications.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### TOSHIBA AMERICA MEDICAL SYSTEMS, INC.
2441 Michelle Drive, Tustin, CA 92780 Phone: (714) 730-5000
#### 510(k) SUMMARY
#### 1. SUBMITTER'S NAME:
Toshiba Medical Systems Corporation 1385 Shimoishigami Otawara-Shi, Tochigi-ken, Japan 324-8550
### 2. OFFICIAL CORRESPONDENT:
Akinori Hatanaka Senior Manager, Regulatory Affairs and Vigilance
# 3. ESTABLISHMENT REGISTRATION:
9614698
# 4. CONTACT PERSON:
Orlando Tadeo, Jr. Manager, Regulatory Affairs Toshiba America Medical Systems, Inc 2441 Michelle Drive Tustin, CA 92780 (714) 669-7459
### 5. Date Prepared:
June 19, 2015 (Updated November 24, 2015)
### 6. TRADE NAME(S):
Aquilion ONE Vision with FIRST 1.0 (CCRS-001A)
### 7. COMMON NAME:
System, X-ray, Computed Tomography
### 8. DEVICE CLASSIFICATION:
Class II (per 21 CFR 892.1750, Computed Tomography X-ray System)
### 9. PRODUCT CODE / DESCRIPTION:
90JAK / Computed Tomography X-Ray System
### 10. PERFORMANCE STANDARD:
This device conforms to applicable Performance Standards for Ionizing Radiation Emitting Products [21 CFR, Subchapter J, Part 1020]
{4}------------------------------------------------
# 11. PREDICATE DEVICE:
| Product | Marketed by | Regulation<br>Number | Regulation<br>Name | Product Code | 510(k)<br>Number | Clearance<br>Date |
|---------------------------------------------------|------------------------------------------|----------------------|----------------------------------------|---------------------------------------------------|------------------|-------------------|
| Aquilion<br>ONE Vision,<br>TSX-301C/<br>1-8, V7.0 | Toshiba<br>America<br>Medical<br>Systems | 21 CFR<br>892.1750 | Computed<br>Tomography<br>X-ray System | JAK:<br>System, X-ray,<br>Tomography,<br>Computed | K142465 | 03/10/2015 |
# 12. REASON FOR SUBMISSION:
Modification of an existing device.
# 13. DEVICE DESCRIPTION:
Aquilion ONE Vision with FIRST 1.0 (CCRS-001A) is a whole body multi-slice helical CT scanner, consisting of a gantry, couch and a console used for data processing and display. This device captures cross sectional volume data sets used to perform specialized studies, using indicated software/hardware, by a trained and qualified physician. This system is based upon the technology and materials of previously marketed Toshiba CT systems. In addition, the subject device incorporates the latest iterative reconstruction technology, FIRST 1.0, intended to reduce exposure dose while maintaining and/or improving image quality.
# 14. INDICATIONS FOR USE:
This device is indicated to acquire and display cross sectional volumes of the whole body, to include the head, with the capability to image whole organs in a single rotation. Whole organs include but are not limited to brain, heart, pancreas, etc.
The Aquilion ONE has the capability to provide volume sets of the entire organ. These volume sets can be used to perform specialized studies, using indicated software/hardware, of the whole organ by a trained and qualified physician.
FIRST 1.0 is an iterative reconstruction algorithm intended to reduce exposure dose and improve high contrast spatial resolution for chest (excluding cardiac), abdomen and pelvis applications. This algorithm is not intended for head or extremity applications.
# 15. SUBSTANTIAL EQUIVALENCE:
The Aquilion ONE Vision with FIRST 1.0 (CCRS-001A), is substantially equivalent to the Aquilion ONE Vision, TSX-301C/1-8, V7.0, that received premarket clearance under K142465 and is marketed by Toshiba America Medical Systems. The changes made to the subject device include the addition of FIRST 1.0, CCRS-001A, an iterative reconstruction algorithm that allows the exposure dose to be reduced while maintaining and/or improving image quality as seen when using FBP (filtered back projection) and improves spatial resolution over FBP. A comparison of the technological characteristics between the subject and the predicate devices is included below.
{5}------------------------------------------------
| Item | Aquilion ONE with FIRST 1.0<br>(CCRS-001A) | Aquilion ONE (TSX-301C/1-8) V7.0 |
|--------------------------------------|--------------------------------------------------|----------------------------------|
| 510(k) Number | This submission | K142465 |
| Anatomical Region | Chest (excluding cardiac),<br>abdomen and pelvis | Whole Body |
| Exposure Dose Reduction | AIDR 3D<br>FIRST 1.0 | AIDR 3D |
| Quantitative Dose Reduction<br>Claim | Yes | None |
| Image Quality Claim | Improved Spatial Resolution | No change |
# 16. SAFETY:
The device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards. This device is in conformance with the applicable parts of the following standards IEC60601-1, IEC60601-1-1, IEC60601-1-2, IEC60601-1-3, IEC60601-1-4, IEC60601-1-6, IEC60601-2-28, IEC60601-2-32, IEC60601-2-44, IEC60825-1, IEC62304, IEC62366, NEMA PS 3.1-3.18, NEMA XR-25 and NEMA XR-26. Additionally, this device complies with all applicable requirements of the radiation safety performance standards, as outlined in 21 CFR §1010 and §1020.
# 17. TESTING
Risk analysis and verification/validation testing conducted through bench testing are included in this submission which demonstrates that the established specifications for the device have been met.
# Image Quality Evaluation
CT image quality metrics were performed, utilizing phantoms, which demonstrated that the subject device is substantially equivalent to or demonstrates an improvement to the predicate device with regard to spatial resolution, CT number accuracy, contrast-to-noise ratio, noise properties, uniformity, slice sensitivity profile, low contrast detectability and standard deviation of noise. Studies were also conducted to demonstrate equivalence and/or improvements in CT number accuracy in water phantoms and artifact reduction using body phantoms.
# Quantitative Dose Reduction Evaluation
The subject device demonstrated a dose reduction claim of up to 82.2% compared to filtered back projection as demonstrated by a model observer evaluation which showed that equivalent low contrast detectability to FBP (range from 0.6 - 0.686) can be achieved with 51.7% to 82.2% less dose using FIRST1.0 at Standard setting for thin (0.5 mm) reconstruction slice thickness in simulated body phantom (MITA-FDA phantom with a body ellipse surrounding it).
# Artifact Reduction
For truncation artifacts, a body phantom was scanned at 25/50/150 mA and 80/160/320 rows and FIRST reconstruction was performed. It was confirmed that the image SD values in the shoulders have been improved and streak artifacts have been visually reduced, as compared with AIDR 3D. For streak artifacts, a body phantom was scanned at 25/50/150 mA and 80/160/320 rows and FIRST reconstruction was performed. It was confirmed that the image SD values in the shoulders have been improved and streak artifacts have been visually reduced, as compared with AIDR 3D. Image quality check was performed using the existing clinical data. Images obtained with FIRST were compared with those obtained with FBP and AIDR 3D. It was visually confirmed that no significant artifacts and missing anatomical structures occur in FIRST.
{6}------------------------------------------------
Representative diagnostic images, reviewed by an American Board Certified Radiologist, including chest, abdomen and pelvis exams were also obtained using the subject device which demonstrates that the device produces images of diagnostic quality and; therefore, performs as intended.
Software Documentation for a Moderate Level of Concern, per the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005, is also included as part of this submission.
Additionally, testing of the subject device was conducted in accordance with the applicable standards published by the International Electrotechnical Commission (IEC) for Medical Devices and CT Systems.
# 18. CONCLUSION
The Aquilion ONE Vision with FIRST 1.0 (CCRS-001A), performs in a manner similar to and is intended for the same use as the predicate device, as indicated in product labeling. Based upon this information, conformance to standards, successful completion of software validation, application of risk management and design controls and the performance data presented in this submission it is concluded that the subject device is substantially equivalent in safety and effectiveness to the predicate device.