Aquilion ONE Vision, v7.0
K142465 · Toshibamedical Systems Corporation · JAK · Mar 10, 2015 · Radiology
Device Facts
| Record ID | K142465 |
| Device Name | Aquilion ONE Vision, v7.0 |
| Applicant | Toshibamedical Systems Corporation |
| Product Code | JAK · Radiology |
| Decision Date | Mar 10, 2015 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.1750 |
| Device Class | Class 2 |
Intended Use
This device is indicated to acquire and display cross sectional volumes of the whole the head, with the capability to image whole organs in a single rotation. Whole organs include but are not limited to brain, heart, pancreas, etc. The Aquilion ONE has the capability to provide volume sets of the entire organ. These volume sets can be used to perform specialized studies, using indicated software, of the whole organ by a trained and qualified physician.
Device Story
Whole-body CT scanner; captures cross-sectional volume data sets via X-ray. Operates in clinical settings; used by trained physicians. Input: X-ray projections; Output: reconstructed cross-sectional volume images. Features include AIDR 3D Enhanced (noise reduction), Single Energy Metal Artifact Reduction (SEMAR) for helical/volume scans, SurekV (automated kV selection), and SURE Exposure 3D (dose modulation). New clinical applications include 4D Orthopedic Analysis and 4D Cerebral Artery Morphological Analysis. Physician reviews images to perform specialized organ studies; aids in clinical diagnosis and treatment planning. Benefits include improved image quality, reduced noise, and optimized radiation dose.
Clinical Evidence
Clinical evaluations performed for 4D Orthopedic Analysis and 4D Cerebral Artery Morphological Analysis applications. Results confirmed performance comparable to manual measurements and segmentations. Bench testing utilized phantoms to validate AIDR 3D Enhanced, SurekV, and SEMAR features against predicates.
Technological Characteristics
Whole-body CT X-ray system. Features: AIDR 3D Enhanced noise reduction, SURE Exposure 3D, SurekV, SEMAR. Conforms to IEC 60601-1, -1-1, -1-2, -1-3, -1-4, -1-6, -1-9, -2-28, -2-32, -2-44, IEC 60825-1, IEC 62304, IEC 62366, NEMA PS 3.1-3.18, NEMA XR-25, NEMA XR-26. Radiation safety per 21 CFR 1010/1020.
Indications for Use
Indicated for patients requiring cross-sectional volume imaging of the head and whole organs (e.g., brain, heart, pancreas). Used by trained physicians for specialized diagnostic studies.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
- Aquilion ONE Vision, TSX-301C/1 and 301C/2, v6.00 (K132222)
- Aquilion ONE Vision, TSX-301C/3, 301C/4 and 301C/5, v6.00 (K133497)
Related Devices
- K222819 — Aquilion Serve (TSX-307A/1) V1.2 with AiCE-i · Canon Medical Systems Corporation · Mar 3, 2023
- K151833 — Aquilion Lightning (TSX-035A/2, V7.0) · Toshibamedical Systems Corporation · Oct 9, 2015
- K231281 — Aquilion Serve (TSX-307A/1) V1.3 · Canon Medical Systems Corporation · Sep 19, 2023
- K132222 — AQUILION ONE VISION, V6.0 · Toshibamedical Systems Corporation · Nov 7, 2013
- K153263 — Aquilion Lightning, TSX-035A/4 and /5, V7.0 · Toshibamedical Systems Corporation · Apr 1, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right. The profiles are stacked on top of each other and are connected by a flowing line.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 10, 2015
Toshiba Medical Systems Corporation % Mr. Orlando Tadeo, Jr. Manager, Regulatory Affairs 2441 Michelle Drive TUSTIN CA 92780
Re: K142465
Trade/Device Name: Aquilion ONE Vision, TSX-301C/1, /2, /3, /4, /5, /6, /7, /8, v7.0 Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: February 9, 2015 Received: February 11, 2015
Dear Mr. Tadeo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K142465
Device Name
Aquilion ONE Vision, TSX-301C/1, /2, /3, /4, /5, /6, /7, /8, v7.0
Indications for Use (Describe)
This device is indicated to acquire and display cross sectional volumes of the whole the head, with the capability to image whole organs in a single rotation. Whole organs include but are not limited to brain, heart, pancreas, etc.
The Aquilion ONE has the capability to provide volume sets of the entire organ. These volume sets can be used to perform specialized studies, using indicated software, of the whole organ by a trained and qualified physician.
Type of Use (Select one or both, as applicable)
| <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|-------------------------------------------------------------|
| <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
TOSHIBA AMERICA MEDICAL SYSTEMS, INC.
2441 Michelle Drive, Tustin, CA 92780 Phone: (714) 730-5000
#### 510(k) SUMMARY
- 1. SUBMITTER'S NAME:
Toshiba Medical Systems Corporation 1385 Shimoishigami Otawara-Shi, Tochigi-ken, Japan 324-8550
# 2. OFFICIAL CORRESPONDENT:
Akinori Hatanaka Senior Manager, Regulatory Affairs and Vigilance
- 3. ESTABLISHMENT REGISTRATION: 9614698
#### 4. CONTACT PERSON:
Orlando Tadeo, Jr. Manager, Regulatory Affairs Toshiba America Medical Systems, Inc. 2441 Michelle Drive Tustin, CA 92780 (714) 669-7459
- 5. Date Prepared:
February 9, 2015
- 6. TRADE NAME(S): Aquilion ONE Vision, TSX-301C/1, /2, /3, /4, /5, /6, /7, /8, v7.0
#### 7. COMMON NAME:
System, X-ray, Computed Tomography
#### 8. DEVICE CLASSIFICATION:
Class II (per 21 CFR 892.1750, Computed Tomography X-ray System)
## 9. PRODUCT CODE / DESCRIPTION:
90JAK / Computed Tomography X-ray System
#### 10. PERFORMANCE STANDARD:
This device conforms to applicable Performance Standards for Ionizing Radiation Emitting Products [21 CFR, Subchapter J, Part 1020]
{4}------------------------------------------------
## 11. PREDICATE DEVICES:
| Product | Marketed by | 510(k) Number | Clearance Date |
|------------------------------------------------------------------|---------------------------------|---------------|-------------------|
| Aquilion ONE Vision, TSX-<br>301C/1 and 301C/2, v6.00 | Toshiba America Medical Systems | K132222 | November 7, 2013 |
| Aquilion ONE Vision, TSX-<br>301C/3, 301C/4 and 301C/5,<br>v6.00 | Toshiba America Medical Systems | K133497 | February 10, 2014 |
## 12. REASON FOR SUBMISSION:
Modification of a cleared device
## 13. DEVICE DESCRIPTION:
The Aquilion ONE Vision, TSX-301C/1, /2, /3, /4, /5, /6, /7, /8, v7.0 are whole body CT scanners that capture cross sectional volume data sets used to perform specialized studies, using indicated software/hardware, by a trained and qualified physician. These systems are based upon the technology and materials of previously marketed Toshiba CT systems.
#### 14. INDICATIONS FOR USE:
This device is indicated to acquire and display cross sectional volumes of the whole the head, with the capability to image whole organs in a single rotation. Whole organs include but are not limited to brain, heart, pancreas, etc.
The Aquilion ONE has the capability to provide volume sets of the entire organ. These volume sets can be used to perform specialized studies, using indicated software/hardware, of the whole organ by a trained and qualified physician.
## 15. SUBSTANTIAL EQUIVALENCE:
The Aquilion ONE Vision, TSX-301C/1, /2, /3, /4, /5, /6, /7, /8, v7.0 is substantially equivalent to Aquilion ONE Vision, TSX-301C/1 and 301C/2, v6.00 and Aquilion ONE Vision, TSX-301C/3, 301C/4 and 301C/5, v6.00, which were cleared via Pre-Market Notification 510(k), K132222 and K133497, respectively. The Aquilion ONE Vision, TSX-301C/1, /2, /3, /4, /5, /6, /7, /8, v7.0, incorporates modifications to the cleared device which include a new AIDR 3D selection (AIDR 3D Enhanced), the introduction of two new clinical applications, the availability of Single Energy Metal Artifact Reduction in Helical scan mode, the addition of SurekV with SURE Exposure 3D and scan protocol management. The method of operation and manufacturing process of the CT system remain unchanged from the cleared device.
| Item | Aquilion ONE Vision,<br>TSX-301C/1, /2, /3,<br>/4, /5, /6, /7, /8, v7.0 | Aquilion ONE<br>Vision<br>TSX-301C/1, /2<br>v6.00 | Aquilion ONE Vision<br>TSX-301C/3, /4, /5,<br>v6.00 | Comments |
|-------------------------------|-------------------------------------------------------------------------|---------------------------------------------------|-----------------------------------------------------|-----------------------------------|
| 510(k) Number | This submission | K132222 | K133497 | |
| Noise reduction<br>processing | AIDR 3D<br>AIDR 3D Enhanced | AIDR 3D | AIDR 3D | New selection<br>added (Enhanced) |
| Metal artifact<br>reduction | Volume Scan<br>Helical Scan | Volume Scan | Volume Scan | Helical scan<br>added |
A complete comparison table is included in this submission. See below for a brief summary of changes:
{5}------------------------------------------------
| ltem | Aquilion ONE Vision,<br>TSX-301C/1, /2, /3,<br>/4, /5, /6, /7, /8, v7.0 | Aquilion ONE<br>Vision<br>TSX-301C/1, /2<br>v6.00 | Aquilion ONE Vision<br>TSX-301C/3, /4, /5,<br>v6.00 | Comments |
|-------------------------------------------------|-------------------------------------------------------------------------|---------------------------------------------------|-----------------------------------------------------|--------------------------------|
| Dose modulation | SURE Exposure 3D<br>Sure kV | SURE Exposure 3D | SURE Exposure 3D | Feature<br>enhancement |
| 4D Orthopedic<br>Analysis | Optional | N/A | N/A | New application |
| 4D Cerebral Artery<br>Morphological<br>Analysis | Optional | N/A | N/A | New application |
| Detector | Up to 10% noise<br>reduction * | N/A | Up to 10% noise<br>reduction * | * Applicable<br>configurations |
## 16. SAFETY:
The device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards. This device is in conformance with the applicable parts of the IEC60601-1, IEC60601-1-1, IEC60601-1-2, IEC60601-1-3, IEC60601-1-4, IEC60601-1-6, IEC60601-1-9, IEC60601-2-28, IEC60601-2-32, IEC60601-2-44, IEC60825-1, IEC62304, IEC62366, NEMA PS 3.1-3.18, NEMA XR-25 and NEMA XR-26. Additionally, this device complies with all applicable requirements of the radiation safety performance standards, as outlined in 21 CFR §1010 and §1020.
## 17. TESTING
Risk analysis and verification/validation testing conducted through bench testing are included in this submission which demonstrates that the requirements for the modifications made to the system have been met. The modified system was evaluated, utilizing phantoms, to demonstrate that the image quality metric of AIDR 3D Enhanced versus current AIDR 3D is substantially equivalent to or better than the predicate device with regard to spatial resolution, CT number and contrast-to-noise ratio, noise properties and standard deviation of noise. Bench studies were also conducted to confirm that the SurekV feature selects the appropriate kV, to confirm that SEMAR functions similarly, with regard to image quality, when used in Volume and in Helical scan modes and that applicable configurations utilizing the "UREVision detector demonstrate improved noise reduction.
Additionally, clinical evaluations were conducted to demonstrate that the new applications 4D Orthopedic Analysis and 4D Cerebral Artery Morphological Analysis, perform as intended. The results confirmed that the applications were comparable to manual measurements and/or segmentations.
Software Documentation for a Moderate Level of Concern, per the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005, is also included as part of this submission.
Additionally, testing of the modified system was conducted in accordance with the applicable standards published by the International Electrotechnical Commission (IEC) for Medical Devices and CT Systems.
{6}------------------------------------------------
## 18. CONCLUSION
The modifications incorporated into the Aquilion ONE Vision, TSX-301C/1, /2, /3, /4, /5, /6, /7, /8, v7.0 do not change the indications for use or the intended use of the device. Based upon this information, conformance to standards, successful completion of software validation, application of risk management and design controls and the performance data presented in this submission it is concluded that the subject device is substantially equivalent in safety and effectiveness to the predicate devices.