ADVANCED LUNG ANALYSIS-1

K013381 · General Electric Co. · JAK · Oct 26, 2001 · Radiology

Device Facts

Record IDK013381
Device NameADVANCED LUNG ANALYSIS-1
ApplicantGeneral Electric Co.
Product CodeJAK · Radiology
Decision DateOct 26, 2001
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1750
Device ClassClass 2
AttributesSoftware as a Medical Device, 3rd-Party Reviewed

Indications for Use

Advanced Lung Analysis (ALA-1) is intended to provide an optimized non-invasive application to measure abnormalities in the lung (for example, nodules, lessons etc.) form a set of computed Tomography (CT) images. The analysis is performed on the Advantage Windows Workstation (K913770). The software allows measurement of volume over time using a consistent standardized measurement protocol, thus providing an estimation of the volume doubling time. This may aid the physician in characterization of suspicious nodules and thus, patient management care decision process.

Device Story

Advanced Lung Analysis-1 (ALA-1) is an image analysis software package operating on the Advantage Windows Workstation. It processes CT image sets to identify and measure lung abnormalities like nodules or lesions. The device provides volumetric estimation of lesions over time, enabling calculation of volume doubling time. Key features include a bookmarking tool for tracking findings, a comparator tool for synchronized longitudinal review, and a database for detected abnormalities. Used by clinicians to enhance review speed and efficiency; output assists physicians in characterizing suspicious nodules and informing patient management decisions.

Clinical Evidence

Bench testing only. The device underwent software development, verification, and validation processes to ensure performance to specifications and adherence to industry standards. No clinical data provided.

Technological Characteristics

Image analysis software package; operates on Advantage Windows Workstation; provides volumetric measurement and longitudinal comparison of lung abnormalities; software-based; no specific hardware materials or energy sources defined as it is a software accessory to a CT system.

Indications for Use

Indicated for patients requiring measurement of lung abnormalities (e.g., nodules, lesions) from CT images to estimate volume doubling time and assist in characterization and management decisions.

Regulatory Classification

Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows a black and white logo of General Electric (GE). The logo consists of the letters 'GE' intertwined within a circular shape. The letters and the circle are in black, with some white highlights to give a sense of depth and dimension to the design. The overall design is simple, yet recognizable as the brand identity of General Electric. GE Medical Sys General Electric Com P O Box 414 Milwaukee, WI 53201 K013381 ### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.87(h). - Submitter Larry A. Kroger, Ph.D. Senior Regulatory Program Manager Telephone: (262) 544-3894, FAX: (262) 544-3863 Date Prepared: August 22, 2001 #### PRODUCT IDENTIFICATION | Name: | Advanced Lung Analysis-1 | |------------------------|---------------------------------------------------------------------------------------| | Classification Name: | Accessory to Computed Tomography System | | Classification Number: | 892.1750 | | Manufacturer : | General Electric Medical Systems<br>283, rue de la Miniere<br>78533 Buc Cedex, FRANCE | | Distributor: | General Electric Medical Systems, Milwaukee, WI | Marketed Devices Advanced Lung Analysis is substantially equivalent to the device listed below: | Model: | Tissue Volume | |---------------|-------------------------------------------------| | Manufacturer: | General Electric Medical Systems, Milwaukee, WI | | 510(k) #: | K963345 | #### Device Description: Advanced Lung Analysis-1 (ALA-1) is an image analysis software package that provides support to the medical professional in assessing abnormalities such as lesions and nodules and changes in their growth over time. This software allows volumetric estimation of a lesion or nodule size over time. It contains productivity tools such as the book marking tool to keep a record of previously found nodules/lesions, the comparator tool to allow 'synchromized' comparison of nodule/lesion over time, database of all the nodules/lesions detected and enhanced speed and efficiency of the review process. {1}------------------------------------------------ #### Indications for Use: Advanced Lung Analysis (ALA-1) is intended to provide an optimized non-invasive application to measure abnormalities in the lung (for example, nodules, lessons etc.) form a set of computed Tomography (CT) images. The analysis is performed on the Advantage Windows Workstation (K913770). The software allows measurement of volume over time using a consistent standardized measurement protocol, thus providing an estimation of the volume doubling time. This may aid the physician in characterization of suspicious nodules and thus, patient management care decision process. #### Comparison with Predicate: ALA-1 is an image analysis software package that allows extraction of a suspicious lesion/nodule area found in a series of CT images from the surrounding lung tissue. The tool is laid out to give volumetric estimation of a lesion or nodule size over time. The functional features of this package are substantially equivalent to that of the following device: | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>----------------<br><br>AND A FRIEND OF COLLECTION OF COLLECTION OF CONSULT OF CONSULT OF CONSULT OF CONSULTER OF CONSULTER OF CONSULTER OF CONSULTER OF CONSULTER OF CONSULTER OF CONSULTER OF CONSUL<br><br>CHANNELE FOR LED FOR LES LES LINE BRILDER<br>LAND MICH MILLE MINILLE E MILE LE MILE LE LE MILLER<br>. FOR COLLECT . LE . LE R LE SECURE SECTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTIO<br>----------------------------------- | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>A Comments of the of<br>1<br>on the control control concessful consideration of the consideration of the consideration of the consideration of the consideration of the consideration of the consideration<br>. | |---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>.<br>11 100 100 000 000 0 0 0 0 1 8 0 1 1 00 1 0 0 1 0 0 1 1 1 000 1 00 1 1 1 1 0000000 00 0 1 1 1 1 0000000 00 0 1 1 | ANTIQUE AND AND A<br>---------------------------------------------------------------------------------------<br><br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>. | #### Adverse Effects on Health : The potential hazards are identified in a risk management summary (hazard analysis) and are controlled by: - · Software Development, Validation and Verification Process to ensure performance to specifications, Federal Regulations and user requirements. - · Adherence to industry and international standards. #### Conclusions: ALA-1 does not result in any new potential safety risks and performs as well as devices currently on the market. GE considers features of the ALA-1 to be equivalent to those of Tissue Volume (K963345). {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with three stripes forming its body and wings. The eagle faces right and is positioned within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle. Food and Drug Administration 200 Corporate Boulevard Rockville MD 20850 General Electric Medical Systems % Mr. Wolfram Gmelin TUV Rheinland of North America 12 Commerce Road NEWTOWN CT 06470 Re: K013381 OCT 2 6 2001 Trade/Device Name: Advanced Lung Analysis-1 Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: 90 JAK Dated: October 8, 2001 Received: October 12, 2001 Dear Mr. Gmelin: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 8xx.1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Nancy C. Brogdon Nancy C. Brogdor Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ 510(k) Number (if known): Device Name: Advanced Lung Analysis (ALA-I) ## Indications for Use Advanced Lung Analysis (ALA-I) is intended to provide an optimized non-invasive application to Advanced Lung Analysis (ALA-i) is ifiended to promotes, lesions etc.) from a set of computed measure abhormalities in the lung (lor examply, the Advantage Windows Workstation The software allows measurement of volume over time using a consistent (K913770). (K913770). The Somware allows theasurement of estimation of the volume doubling time. standardized measurement protocol, thus providing and they and thus, natient standardized measurement protocol, thus providing an ostinator of the road. So and thus, patient management care decision process. # (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801-109) OR Over-The-Counter Use_ Nancy C brogdon Division Sign (Off) Division: Reproductive Abdominal, and Radiological Devices 510(k) Number K013381
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...