ADVANTAGE WINDOWS TISSUE VOLUME OPTION

K963345 · GE Medical Systems · LNH · Oct 18, 1996 · Radiology

Device Facts

Record IDK963345
Device NameADVANTAGE WINDOWS TISSUE VOLUME OPTION
ApplicantGE Medical Systems
Product CodeLNH · Radiology
Decision DateOct 18, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1000
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

Advantage Windows Tissue Volume Option is intended to calculate the volume of any three dimensional structure from a set of CT images. The images are displayed and measured on the CT/MR Advantage Windows Diagnostic Workstation (K913770). The Tissue Volume Option will automate manual measurements and estimates that are currently made regarding the measure of volume of anatomy and pathology imaged with computed tomography.

Device Story

Software package for Advantage Windows Diagnostic Workstation (K913770); operates on Sun Sparc 20, Mod 40 or higher. Inputs: CT images transferred via DICOM or Ethernet. Operation: operator selects area of CT image; software automates volumetric calculation previously performed via manual distance tools. Output: volumetric measurement of selected anatomy/pathology. Used in clinical diagnostic settings by radiologists/technicians. Benefits: automates manual estimates; facilitates comparison of volumetric data across patient images.

Clinical Evidence

No clinical data. Bench testing only.

Technological Characteristics

Software-only package; runs on Sun Sparc 20, Mod 40 or higher workstation. Connectivity via DICOM or Ethernet. Principle of operation based on Advantage Windows 3D software platform.

Indications for Use

Indicated for calculating volume of 3D structures from CT images for anatomy and pathology assessment.

Regulatory Classification

Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{0} K963345 # Summary of Safety & Effectiveness ## Product Identification: **Name:** Advantage Windows Tissue Volume Option **Manufacturer:** General Electric Medical Systems 16800 W. Ryerson Road New Berlin, WI 53151 **Distributor:** Same as Manufacturer OCT 18 1996 ## Indications for Use: Advantage Windows Tissue Volume Option is intended to calculate the volume of any three dimensional structure from a set of CT images. The images are displayed and measured on the CT/MR Advantage Windows Diagnostic Workstation (K913770). The Tissue Volume Option will automate manual measurements and estimates that are currently made regarding the measure of volume of anatomy and pathology imaged with computed tomography. ## Device Description: The device is a software package to be used on the Advantage Windows Workstation (K913770). **Design:** The design is based on the same software platform as used for Advantage Windows 3D (VoxTool) and will operate on the Advantage Windows Operating System. The workstation hardware required for operation will be the Sun Sparc 20, Mod 40 or higher. The images used to measure volume can be captured by any CT scanner and transferred to Advantage Windows workstation by DICOM or Ethernet. **Energy Source and Exposure Levels:** There is no energy source associated with this package in and of itself. However, the energy source used to make the image being analyzed is the same used to take standard CT diagnostic images. **Principals of Operation:** The same as Advantage Windows 3D. **Features:** To calculate the volume of areas of a CT image selected by the operator. **Accessories:** None ## Marketing History: The Advantage Windows Tissue Volume Option is substantially equivalent to the currently marketed Advantage Windows 3D Option (K913770). ## Adverse Effects on Health: The package itself will not have any adverse affects on health. This tool measures and displays the volume of an outlined image. The outlining is in the hands of the operator who can choose to accept or reject the outlined region and then select to use that data to compare against earlier images or images from other patients. ## Conclusions: The Advantage Windows Tissue volume Option enhances the current Advantage Windows 3D package by automating volumetric measurements that are currently being made manually using the measure distance tools of the 3D package. It is substantially equivalent to the Advantage Windows 3D package in design, construction, principle of operation, and features.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...