K203010 · Apelem-Dms Group · JAA · Jan 22, 2021 · Radiology
Device Facts
Record ID
K203010
Device Name
Platinum dRF Imaging System
Applicant
Apelem-Dms Group
Product Code
JAA · Radiology
Decision Date
Jan 22, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Platinum dRF Imaging System is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a digital flat detector, it can perform a range of applications including: general radiology and diagnostic fluoroscopy examinations, conventional linear tomography and pediatrics examinations. The Platinum dRF is a device intended to visualize anatomical structures by converting a pattern of X-ray into a visible image. The system has medical applications ranging from, but not limited to, gastrointestinal, cranial, skeletal, thoracic, lung and urogenital tract examinations. It may also be used in: lymphography, endoscopy, myelography, venography and arthrography. The Platinum dRF may also be used for outpatient or emergency services, as well as for mobile transport examinations (wheelchair and bed). The Platinum dRF is not indicated for use in interventional radiology.
Device Story
The Platinum dRF Imaging System is a universal diagnostic X-ray system for radiographic and fluoroscopic studies. It utilizes a digital flat panel detector (Vieworks Vivix-D 1717G) to capture X-ray patterns, which are converted into visible images via the ADAM Imaging System software. The system is operated by licensed clinicians (radiologists, orthopedic surgeons, radiological technicians) in hospital or clinical settings, including outpatient and emergency departments. The ADAM software acquires, processes, displays, and exports images via DICOM protocol. Clinicians use these images to visualize anatomical structures for diagnosis. The device is integrated into a remote-controlled radiological table. By providing high-quality digital images, the system assists in the diagnosis of various conditions across multiple medical specialties, benefiting patients through accurate, non-interventional diagnostic imaging.
Clinical Evidence
Bench testing only. No clinical data provided. Performance was validated through comparative bench testing of the new detector (Vieworks Vivix-D 1717G) against the predicate detector (Thales RF4343 FL), evaluating pixel pitch, resolution (line pairs/mm), MTF, and DQE. Results confirmed the new detector's performance is sufficient for general radiographic procedures and equivalent to the predicate.
Technological Characteristics
System includes a remote-controlled radiological table, X-ray generator, and digital flat panel detector (Vieworks Vivix-D 1717G, CsI scintillator). Connectivity via DICOM network. Software (ADAM Imaging System) performs image acquisition and processing. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-54, and IEC 62304 standards.
Indications for Use
Indicated for universal diagnostic radiographic and fluoroscopic imaging in patients requiring gastrointestinal, cranial, skeletal, thoracic, lung, urogenital, lymphographic, endoscopic, myelographic, venographic, or arthrographic examinations. Includes pediatric applications. Not indicated for interventional radiology.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Apelem-DMS Group Platinum dRF Imaging System (K160301)
Submission Summary (Full Text)
{0}------------------------------------------------
January 22, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Apelem-DMS Group % Mr. Scott Blood Senior Regulatory Affairs Consultant MEDIcept, Inc. 200 Homer Avenue ASHLAND MA 01721
Re: K203010
Trade/Device Name: Platinum dRF Imaging System Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: JAA Dated: December 21, 2020 Received: December 28, 2020
Dear Mr. Blood:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203010
Device Name Platinum dRF Imaging System
### Indications for Use (Describe)
The Platinum dRF Imaging System is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a digital flat detector, it can perform a range of applications including: general radiology and diagnostic fluoroscopy examinations, conventional linear tomography and pediatrics examinations.
The Platinum dRF is a device intended to visualize anatomical structures by converting a pattern of X-ray into a visible image. The system has medical applications ranging from, but not limited to, gastrointestinal, cranial, skeletal, thoracic, lung and urogenital tract examinations.
It may also be used in: lymphography, endoscopy, myelography, venography and arthrography.
The Platinum dRF may also be used for outpatient or emergency services, as well as for mobile transport examinations (wheelchair and bed).
The Platinum dRF is not indicated for use in interventional radiology.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for MEDICEPT. The word "MEDI" is in large, bold, blue letters, while "CEPT" is in a cursive, purple font. Below the company name is the phrase "Trusted Solutions, Rapid Response" in a smaller font.
### 1. 510(K) SUMMARY K203010
This 510(k) Summary is in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content of this 510(k) summary is provided in conformance with 21 CFR ട്ല 807.92.
#### 1.1. Submitter's Information
Name: Apelem-DMS Group Address: Parc Scientifique Georges Besse, 175 Allee Von Neumann, 30035 Nimes cedex 1 France
33 466 29 09 07 Phone:
## Official FDA Contact:
MEDIcept Inc. 200 Homer Ave Ashland, MA 01721 Scott Blood sblood@medicept.com 978-729-5978
Date Prepared: November 4th, 2020
### 1.2. Device Administrative Information
| Trade Name: | Platinum dRF Imaging System |
|-----------------------|----------------------------------------------------------------------|
| Device Name: | Solid State X-Ray Imager (Flat Panel/Digital Imager) |
| Common Name: | System, X-Ray Fluoroscopic, Image-Intensified |
| Regulatory Class: | Class II per 21 CFR 892.1650 |
| Product Code: | JAA- 21 CFR 892.1650, Image-intensified fluoroscopic<br>x-ray system |
| Classification Panel: | Radiology |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for MEDICEPT. The word "MEDI" is in large, bold, blue letters, while the word "CEPT" is in a cursive, purple font. Below the company name is the phrase "Trusted Solutions, Rapid Response" in a smaller, gray font.
### 2. LEGALLY MARKETED PREDICATE DEVICE
The following is the identified predicate device:
The predicate referenced below has not been subject to a design-related recall.
Apelem-DMS Group Group – K160301–Platinum dRF Imaging System
- Indications for Use: ●
## Indications for Use for DMS Platinum dRF Imaging System:
The Platinum dRF Imaging System is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a digital flat detector, it can perform a range of applications including general R/F, diagnostic fluoroscopy, conventional linear tomography, angiography and pediatric examinations.
The Platinum dRF is a device intended to visualize anatomical structures by converting a pattern of X-ray into a visible image. The system has medical applications ranging from but not limited to gastrointestinal examinations, cranial, skeletal, thoracic and lung exposures as well as examination of the urogenital tract. The units may also be used in lymphography, endoscopy, myelography, venography, pediatrics, arthrography, digital angiography and digital subtraction angiography (DSA).
The Platinum dRF may be used for outpatient and emergency treatment, as well as for mobile transport (wheelchair and bed) examinations.
The Platinum dRF is not indicated for use in interventional radiology.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for MEDICEPT. The word "MEDI" is in large, bold, blue letters. The word "CEPT" is in a stylized purple font. Below the words is the phrase "Trusted Solutions, Rapid Response" in a smaller font, with "Trusted Solutions" in gray and "Rapid Response" in purple.
### 3. DEVICE DESCRIPTION
The subject of this Special 510(k) application is the change to the imaging software system to the ADAM Imaging System. As the resources are now available, Apelem-DMS Group was able to develop their own Imaging System called the ADAM Imaging System. The ADAM Imaging System will now be developed by Apelem-DMS Group. The predicate device (K160301) had an Off the Shelf Software Imaging System called DUET DRF Imaging System which was developed by Thales.
The ADAM Imaging System is a digital image acquisition system to be used in conjunction with a detector during radiography or fluoroscopy X-ray examinations to acquire, display, process and export images according to DICOM protocol via a network connection.
The ADAM Imaging System software interfaces with an X-Ray detector to acquire raw pixel data and image processing algorithms which transform raw pixel data into images and image sequences to help medical professionals with viewing images to for patient diagnosis.
The Duet DRF requires the use of the THALES RF4343 FL detectors, whereas ADAM Imaging System requires the use of the VIEWORKS VIVIX-D 1717G detector. Both detectors employ the same state-of-the-art indirect conversion technology based on CsI scintillator and photo-diodes. The detector Vivix-D 1717G is connected to the x-ray system by cable. The integration with the PLATINUM dRF system has been established and appears adequate. The VIEWWORKS VIVIX-D 1717G detector is only compatible with the ADAM Imaging System.
The Duet DRF and ADAM Imaging System are designed to support general radiography (excluding mammography) and fluoroscopy imaging procedures.
The Apelem-DMS Group ADAM Imaging System is not a stand-alone device. It is integrated into the Platinum dRF Imaging System and functions as a platform for FDA cleared or registered components (i.e. generator, panel detector, detector collimator, X-ray tube and software imaging packages), that are installed with a Apelem-DMS Group manufactured radiological examination table, control panel with system controller software, and electrical panel.
The Platinum dRF remote controlled table is a radiologic table equipped with a flat panel electronic detector. This table is used to perform general digital radiological, fluoroscopy and peripheral angiography.
### 3.1. Indications for Use (Proposed Device)
## Indications for Use for DMS Platinum dRF Imaging System:
The Platinum dRF Imaging System is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a digital flat detector, it can perform a range of applications including: general radiology and diagnostic fluoroscopy examinations, conventional linear tomography and pediatrics examinations.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for MEDICEPT. The word "MEDI" is in large, bold, blue letters, while "CEPT" is in a stylized purple font. Below the logo is the text "Trusted Solutions, Rapid Response" in a smaller font, with "Trusted Solutions" in purple and "Rapid Response" in gray.
The Platinum dRF is a device intended to visualize anatomical structures by converting a pattern of X-ray into a visible image. The system has medical applications ranging from, but not limited to, gastrointestinal, cranial, skeletal, thoracic, lung and urogenital tract examinations.
It may also be used in: lymphography, endoscopy, myelography, venography and arthrography.
The Platinum dRF may also be used for outpatient or emergency services, as well as for mobile transport examinations (wheelchair and bed).
The Platinum dRF is not indicated for use in interventional radiology.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for MEDICEPT. The first part of the logo, "MEDI", is in blue, while the second part, "CEPT", is in purple. Below the logo is the text "Trusted Solutions, Rapid Response" in a smaller font size.
### SUBSTANTIAL EQUIVALENCE COMPARISON AND DISCUSSION 4.
| | Proposed Device<br>(This Submission):<br>ADAM Imaging System integrated<br>into the DMS Platinum dRF<br>Imaging System | Predicate Device:<br>(K160301)<br>Duet Imaging System integrated<br>into the DMS Platinum dRF<br>Imaging System |
|--------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| FDA Product<br>Code | JAA | JAA |
| Single<br>Use/Reusable | Reusable | Same |
| Indications for<br>Use | The Platinum dRF Imaging System is<br>intended to be used as a universal<br>diagnostic imaging system for<br>radiographic and fluoroscopic studies.<br>Using a digital flat detector, it can<br>perform a range of applications<br>including: general radiology and<br>diagnostic fluoroscopy examinations,<br>conventional linear tomography and<br>pediatrics examinations.<br><br>The Platinum dRF is a device<br>intended to visualize anatomical<br>structures by converting a pattern of<br>X-ray into a visible image. The<br>system has medical applications<br>ranging from, but not limited to,<br>gastrointestinal, cranial, skeletal,<br>thoracic, lung and urogenital tract<br>examinations.<br><br>It may also be used in: lymphography,<br>endoscopy, myelography, venography<br>and arthrography.<br><br>The Platinum dRF may also be used<br>for outpatient or emergency services,<br>as well as for mobile transport<br>examinations (wheelchair and bed).<br><br>The Platinum dRF is not indicated for<br>use in interventional radiology. | The Platinum dRF Imaging System is<br>intended to be used as a universal<br>diagnostic imaging system for<br>radiographic and fluoroscopic studies.<br>Using a digital flat detector, it can<br>perform a range of applications<br>including general R/F, diagnostic<br>fluoroscopy, conventional linear<br>tomography, angiography and<br>pediatric examinations.<br><br>The Platinum dRF is a device<br>intended to visualize anatomical<br>structures by converting a pattern of<br>X-ray into a visible image. The<br>system has medical applications<br>ranging from but not limited to<br>gastrointestinal examinations, cranial,<br>skeletal, thoracic and lung exposures<br>as well as examination of the<br>urogenital tract.<br><br>The units may also be used in<br>lymphography, endoscopy,<br>myelography, venography, pediatrics,<br>arthrography, digital angiography and<br>digital subtraction angiography<br>(DSA).<br><br>The Platinum dRF may be used for<br>outpatient and emergency treatment, |
| | Proposed Device<br>(This Submission):<br>ADAM Imaging System integrated<br>into the DMS Platinum dRF<br>Imaging System | Predicate Device:<br>(K160301)<br>Duet Imaging System integrated<br>into the DMS Platinum dRF<br>Imaging System |
| | | as well as for mobile transport<br>(wheelchair and bed) examinations.<br>The Platinum dRF is not indicated for<br>use in interventional radiology. |
| Use Environment | Hospital or clinical setting by a<br>licensed doctor, orthopedic<br>surgeon, radiologist, radiological<br>technician, or technician<br>authorized by the manufacturer | Same |
| Patient Contact | The patient does not come into<br>direct contact with the ADAM<br>Imaging System component or the<br>Detector. The patient comes in<br>contact with the examination table<br>only. | Same |
| Flat Detector | VIEWORKS VIVX-D 1717G | RF4343 FL |
| Frame Rate | 12f/sec (RAD)<br>18f/sec fluoroscopy (large field),<br>No change | Same |
| Image Specifications | No Change | Same |
| Lead Shield | Included | Same |
| Platinum<br>examination<br>remote control<br>table | No change | Same |
| Imaging System<br>Component | ADAM Imaging System | Duet DRF Imaging System |
| | Proposed Device<br>(This Submission):<br>ADAM Imaging System integrated<br>into the DMS Platinum dRF<br>Imaging System | Predicate Device:<br>(K160301)<br>Duet Imaging System integrated<br>into the DMS Platinum dRF<br>Imaging System |
| Control Panel<br>with System<br>Controller<br>Software | No change | Same |
| Electrical Panel | No change | Same |
| Platinum Software | No change | Same |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for MEDICEPT. The first part of the logo, "MEDI", is in blue, while the second part, "CEPT", is in purple. Below the logo is the text "Trusted Solutions, Rapid Response", with "Trusted Solutions" in purple and "Rapid Response" in gray.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for MEDICEPT. The word "MEDI" is in large, bold, blue letters, while "CEPT" is in a stylized, cursive, purple font. Below the logo, the words "Trusted Solutions, Rapid Response" are written in a smaller font, with "Trusted Solutions" in purple and "Rapid Response" in gray.
The technological characteristics of the Detector are summarized in the table below:
| Detector<br>Information | Proposed: ADAM Imaging<br>System Detector Information | Predicate: Duet Imaging<br>System Detector<br>Information (K160301) | Comment |
|-----------------------------|-------------------------------------------------------|---------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Manufacturer<br>of detector | VIEWORKS | THALES | Different<br>manufacturers<br>but similar<br>technology |
| Model Name | VIVIX-D 1717G | PIXIUM RF 4343 FL | Different<br>manufacturers<br>but similar<br>technology |
| Scintillator | CsI | CsI | Same |
| Pixel Pitch | 140µm | 148µm | ADAM<br>Imaging<br>System's<br>detector has a<br>smaller pixel<br>pitch which<br>means better<br>detail<br>detectability |
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for MEDICEPT. The word "MEDICEPT" is written in a combination of blue and purple colors. Below the logo is the text "Trusted Solutions, Rapid Response" in a smaller font size.
| Detector<br>Information | Proposed: ADAM Imaging<br>System Detector Information | Predicate: Duet Imaging<br>System Detector<br>Information (K160301) | Comment | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
|---------------------------------------------|--------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|-----|-------------------------------------------------------------------------------------------|--|--|-----------------------|--|--|--|--|--------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------|---------------------|--|--|--|--|--|--|--|--|--|--|--|-----------------------------------------------------------------------------------------------------|
| Pixel Matrix | 3072 x 3072 pixels | 2874 x 2840 pixels | Adam<br>Imaging<br>System's<br>detector has a<br>larger pixel<br>matrix which<br>means a<br>larger<br>detection<br>field. | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Image Size | 17" x 17" | 17" x 17" | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Grayscale | 16 bits | 16 bit | Same | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Resolution | 3.5 linepairs/mm | 3.4 linepairs/mm | The<br>performance<br>of both<br>detectors are<br>similar<br>enough<br>(>3%) for the<br>procedures<br>indicated | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Modulation<br>Transfer<br>Function<br>(MTF) | MTF IEC<br>0.5 lp/mm<br>MTF IEC<br>1 lp/mm<br>MTF IEC<br>2 lp/mm<br>MTF IEC<br>3 lp/mm<br>MTF IEC<br>Nyquist | Minimal Typical 78 % 83 % 50 % 55 % 20 % 25 % 5 % 10 % - - | | | | | | | | | | | | | Minimal Typical 79% | | | | | MTF IEC<br>0.5 lp/mm<br>MTF IEC<br>1 lp/mm<br>MTF IEC<br>2 lp/mm<br>MTF IEC<br>3 lp/mm<br>MTF IEC<br>Nyquist | Minimal Typical 55% 66% 25 % 35 % 10 % 19 % 7 % 15 % | | | | | | | | | | | | | Performance<br>of both<br>detectors are<br>sufficient for<br>general<br>radiographic<br>procedures. |
| Detective<br>Quantum<br>Efficiency<br>(DQE) | 75% | Minimal Typical DQE<br>0.5 lp/mm Minimal Typical 60 % 65 % | | | | | | | | | 78% | Minimal Typical DQE<br>0 lp/mm Minimal Typical 65 % 73 % | | | | | | | | | | See Notes<br>below. | | | | | | | | | | | | |
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the logo for MediCept. The logo has the word "MEDI" in blue and the word "CEPT" in purple. Below the logo is the text "Trusted Solutions, Rapid Response" in a smaller font size.
| Detector<br>Information | Proposed: ADAM Imaging<br>System Detector Information | Predicate: Duet Imaging<br>System Detector<br>Information (K160301) | Comment |
|-------------------------|-------------------------------------------------------|---------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Dimensions | 471 x 471 x 35 mm | 500 x 490 x 45.5 mm | VIEWORKS<br>detector is<br>smaller than<br>the predicate<br>which will<br>enable easier<br>installation. |
| Weight | 10 kg | 9.1kg | VIEWORKS<br>detector is<br>heavier than<br>the predicate<br>but as the<br>difference is<br>less than 1 kg<br>there is no<br>impact. |
Notes:
Differences in Detective Quantum Efficiency (DQE):
The predicate device has a DQE of 78% if the shot noise was the only cause to be taken into account, i.e. the proportion of photons that are detected. This is close to the value found for the subject device detector (75%). The difference between the two detector DQE values is low enough to be considered equivalent because:
- 1. The difference in the required dose to obtain a given image quality is small enough (ratio of 78% to 75%) not to have any clinical significance (<10% of variation)
- 2. The evaluation of DQE, which combines information about the signal and the noise, has error margins: no definitive conclusion can be given if the difference is <5%
- The DQE value can vary somewhat between different units of the same model, specially 3. from a small difference in the scintillator thickness. This can be seen from the predicate device detector specification, which indicates 5% of difference between the "minimal" and "typical" values of the DQE.
The image below shows the design of the proposed (Apelem) and predicate (Thales) Imaging Systems.
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the logo for MEDICEPT. The word "MEDI" is in bold, dark blue letters, while "CEPT" is in a lighter purple color with a swoosh effect. Below the logo, the words "Trusted Solutions, Rapid Response" are written in a smaller font, with "Trusted Solutions" in purple and "Rapid Response" in gray.
| Manufacturers | APELEM | THALES |
|---------------|------------------------------------|------------------------------------|
| Design | Image: APELEM monitor and computer | Image: THALES monitor and computer |
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the logo for MEDICEPT. The word "MEDI" is in large, bold, blue letters, and the word "CEPT" is in a cursive, purple font. Below the logo is the text "Trusted Solutions, Rapid Response" in a smaller, purple font.
### COMPLIANCE WITH DESIGN CONTROLS 5.
The results of assessment under Design Controls supports that the Apelem-DMS Group Platinum dRF Imaging System integrated with the ADAM Imaging System component is substantially equivalent to the predicate device. There are no new causes of risk or no new cause of existing risk, therefore the change does not raise different issues of safety or effectiveness.
### Compliance with Standards 5.1.
The Platinum dRF Imaging System with the ADAM Imaging System component and VIEWORKS detector comply with the following standards:
- 1. IEC 60601-1:2005/A1:2012 Medical Electrical Equipment: General Requirements for basic safety and essential performance
- 2. IEC 60601-1-2:2014 Medical Electrical Equipment Part 1-2: General Requirements for basic safety and essential performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests
- 3. IEC 60601-1-3:2008+A1:2013 Medical electrical equipment part 1-3: general requirements for basic safety and essential performance - collateral standard: radiation protection in diagnostic x-ray equipment.
- 4. IEC 60601-2-54:2009 Medical electrical equipment Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy
- 5. EN ISO 14971:2012 Application of Risk Management to Medical Devices
- 6. IEC 62304:2006 Medical Device Software Software life cycle processes
- 7. EN 1041:2008 Information supplied by the manufacturer of Medical devices
- EN ISO 15223-1:2016 Medical devices Symbols to be used with medical device labels, 8. labelling and information to be supplied - Part 1: General requirements.
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the logo for MEDICEPT. The word "MEDI" is in blue, and the word "CEPT" is in purple. Underneath the logo is the text "Trusted Solutions, Rapid Response" in a smaller font.
### CONCLUSION 6.
The Apelem-DMS Group Platinum dRF Imaging System integrated with the ADAM Imaging System has the same intended use, technology, materials, and uses the same components as the predicate device except for the ADAM Imaging System component and VIEWORKS Detector. The changes to the proposed device have little to no impact on the safety or performance device (i.e. image quality) and no additional questions regarding safety or effectiveness have been raised. The fundamental scientific technology of the proposed device included in this submission remains unchanged from the legally marketed predicate device. Therefore, the proposed device is substantially equivalent to the predicate device.