ULTRA 1040

K241996 · Ecoray Co., Ltd. · IZL · Apr 18, 2025 · Radiology

Device Facts

Record IDK241996
Device NameULTRA 1040
ApplicantEcoray Co., Ltd.
Product CodeIZL · Radiology
Decision DateApr 18, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1720
Device ClassClass 2

Indications for Use

The ULTRA 1040 Portable X-ray Unit is a portable X-ray device, intended for use by a qualified/trained physician or technician for the purpose of acquiring X-ray images of the patient's extremities. This device is not intended for mammography.

Device Story

Portable X-ray unit; emits X-rays for diagnostic imaging of extremities. Components: main unit, exposure hand switch, battery charger, collimator, high-voltage tank. Input: user-selected kV/mAs parameters via control panel. Operation: system control software manages circuit modules, loads APR parameters, controls high-voltage generation, and monitors battery/status. Output: X-ray emission for image acquisition; status indicators on display. Used in clinical settings by physicians or technicians. Benefits: provides portable diagnostic imaging capability for extremity exams.

Clinical Evidence

Bench testing only. Included radiation performance, collimator accuracy, battery performance, and imaging quality assessments. Clinical adequacy evaluated via task-based image quality study using radiographic images acquired in standing-mounted mode, reviewed by radiologists.

Technological Characteristics

Mobile X-ray system; 2.4kW high-frequency generator; Lithium Polymer battery (58.8 VDC). Dimensions: 384 x 227.5 x 199 mm; Weight: 12 kg. Tube: OX/100-L. Collimator: double slit, manual LED. Connectivity: supports external X-ray detector. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-1-6, IEC 60601-2-28, IEC 60601-2-54, IEC 62304, IEC 62366-1, IEC 62133-2, IEC 60825-1.

Indications for Use

Indicated for use by qualified physicians or technicians to acquire X-ray images of patient extremities. Not intended for mammography.

Regulatory Classification

Identification

A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION April 18, 2025 ECORAY Co., Ltd. % Jay Mansour Mansour Consulting LLC 845 Aronsom Lake Court ROSWELL, GA 30075 Re: K241996 Trade/Device Name: Ultra 1040 Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile X-Ray System Regulatory Class: Class II Product Code: IZL Dated: February 20, 2025 Received: February 20, 2025 Dear Jay Mansour: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K241996 – Jay Mansour Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K241996 – Jay Mansour Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, ![img-0.jpeg](img-0.jpeg) for Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K241996 | | | Device Name ULTRA 1040 | | | Indications for Use (Describe) | | | The ULTRA 1040 Portable X-ray Unit is a portable X-ray device, intended for use by a qualified/trained physician or technician for the purpose of acquiring X-ray images of the patient's extremities. This device is not intended for mammography. | | | Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} ECORAY Co., Ltd. 510(k) Premarket Notification- Traditional 510(k) Summary 1 /10 {5} ECORAY Co., Ltd. 510(k) Premarket Notification- Traditional K241996 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92 Date: April 15, 2025 1. INFORMATION 1.1. Submitter Information Submitter Name: ECORAY Co., Ltd. Address: (04778) #714~716, ForHu B/D, 58, Wangsimni-ro, Seongdong-gu, Seoul, Republic Of Korea Telephone Number: Tel: +82-70-7510-3407 Email: ra@ecoray.kr 1.2. Official Correspondent Name: Mr. Jay Mansour Company Name: Mansour Consulting LLC Address: 845 Aronson Lake Court. Roswell, GA 30075 Telephone Number: Tel: +1 (678) 908-8180 Email: jay@mansourconsulting.com 2. DEVICE INFORMATION Trade Name: ULTRA 1040 Common Name: System, X-ray, Mobile Classification Name: Mobile x-ray system Product Code: IZL Classification Regulation: 21 CFR Part 892.1720 Device Class: II 3. PREDICATE DEVICE Manufacturer: Shantou Institute of Ultrasonic Instruments Co., Ltd. (SIUI) Trade Name: SR-8100 Common Name: System, X-ray, Mobile Classification Name: Mobile x-ray system Product Code: IZL Classification Regulation: 21 CFR Part 892.1720 Device Class: II 510(k) Number: K200637 2 / 10 {6} ECORAY Co., Ltd. 510(k) Premarket Notification- Traditional ## 4. SUBJECT DEVICE DESCRIPTION This Portable X-ray Unit (Model: ULTRA 1040) consists of the following major components: an X-ray main unit, an X-ray exposure hand switch and a battery charger and other components. The X-ray main unit is mainly for emitting X-rays required for X-ray exams; the hand switch is for output control of X-ray emitting, and the battery charger is for charging the built-in battery in the X-ray main unit. The device can be used with an X-ray detector, a computer for receiving and detecting signal results and image processing software. The major components of the X-ray main unit include: handle, enclosure, control panel, system control board, high-voltage tank, collimator (beam limiter), lithium-ion battery and system control software running on the system control board. The system control software is for real-time interaction and control with various circuit modules inside the portable X-ray unit. The software responds to user operations on the control panel. The user can adjust and control the kV and mAs parameters, and the software will display the parameters or directly load the APR parameters. The software loads the control data from X-ray output into the high-voltage generation control circuit of the system control board and control the high-voltage tank to generate high-voltage to excite the ray tube inside to emit X-rays, control the switch of the collimator indicator, and monitor the working status of the device, the battery power status, and control the display of the status indicators. ## 5. INDICATIONS FOR USE The ULTRA 1040 Portable X-ray Unit is a portable X-ray device, intended for use by a qualified/trained physician or technician for the purpose of acquiring X-ray images of the patient's extremities. This device is not intended for mammography. ## 6. SUBSTANTIAL EQUIVALENCE ### 6.1. Comparison Chart of Features and Specifications | Item | Proposed Device | Predicate Device | | | --- | --- | --- | --- | | K Number | K241996 | K200637 | | | Manufacturer | ECORAY Co., Ltd. | SIUI | | | Trade Name | ULTRA 1040 | SR-8100 | | | Device Classification Name | Mobile x-ray system | Mobile x-ray system | Same | | Product Code | IZL | IZL | Same | | Regulation Number | 21CFR892.1720 | 21CFR892.1720 | Same | | 510(k) Review Panel | Radiology | Radiology | Same | 3 / 10 {7} ECORAY Co., Ltd. 510(k) Premarket Notification- Traditional | Item | Proposed Device | Predicate Device | | | --- | --- | --- | --- | | Indication for use | The ULTRA 1040 Portable X-ray Unit is a portable X-ray device, intended for use by a qualified/trained physician or technician for the purpose of acquiring X-ray images of the patient’s extremities. This device is not intended for mammography. | The SR-8100 Portable X-ray Unit is a portable X-ray device, intended for use by a qualified/trained physician or technician for the purpose of acquiring X-ray images of the desired parts of patient’s anatomy (including head, cervical spine, chest, abdomen, lumbar spine, pelvis and extremities). The device may be used for handheld diagnostic imaging of body extremities. The system is subject to the following limitations of use when stand-mounted: The device may be used for diagnostic imaging of head, cervical spine, abdomen, lumbar spine, pelvis or extremities. The device may be used for imaging of the chest when used without a grid. This device is not to be used on bariatric patients, unless imaging body extremities. This device is not intended for mammography. | Similar | | Principle of Operation | General Purpose Diagnostic X ray | General Purpose Diagnostic X ray | Same | | Size | 384 x 227.5 x 199 mm | 227 x 522 x 217 mm | Similar | | Weight | 12 kg (Including Collimator) | 12kg (Including Collimator) | Same | | Energy source | Lithium Polymer Battery (58.8 VDC) Charging voltage : 100V ~ 240~V, 50/60 Hz, 80VA 250 exposure per charge | Lithium-ion Rechargeable Battery (29.6VDC), 300 exposures per charge. | Similar | | Use Interface | Soft touch push buttons | Soft touch push buttons | Same | | Exposure times | 0.008 sec – 2.5 sec | 0.02 sec – 2.5 sec: R’10 sec Step | Similar | | Tube current [mA] | 40~50kV/ 40mA 51~60kV/ 35mA 61~70kV/ 30mA 71~80kV/ 30mA 81~90kV/ 25mA 91~100kV/ 20mA | 25 mA @ 40kVDC – 60kVDC (1kVP steps) 20 mA @ 61kVDC – 100kVDC (1kVP steps) | Similar | 4 / 10 {8} ECORAY Co., Ltd. 510(k) Premarket Notification- Traditional | Item | Proposed Device | Predicate Device | | | --- | --- | --- | --- | | | • All 1kV steps | | | | Tube Voltage Range [kVp] | 40 - 100kVp, 61 Step (1kV Step) | 40 - 100kVp, 61 Step (1kV Step) | Same | | Memory settings | 10 Memories | 5 memories | Slightly high | | HF Generator | High frequency | High frequency | Same | | Output power [kW] | 2.4kW | 2.0 kW | Slightly high | | X-ray Tube | OX/100-L (C.E.I.) | D-125SB | Different | | Collimator | Double slit type and manually operation with LED Light indicator | SIUI SR-8100-19L | Different | | Focal Spot Size | • 1.0mm | • 1.2mm | Similar | | Anode Heat Storage Capacity | • 19,600 HU (14000J) | • 50,000HU (35,000J) | Different | | FDA Performance Standard | • Complies | • Complies | Same | | X-ray switching frequency | • 100kHz | • 100kHz | Same | | Control | • 2 Point Control (kV, mAs) | • 2 Point Control (kV, mAs) | Same | | Anatomical Programs | • Preprogrammed 12or 7 APR data - User Programmable | • Preprogrammed 9 APR data-User Programmable | Similar | | Collimator | Model | ULTRA 1040BT Collimator | SR-8100-19L | | | Manufacturer | ECORAY Co.,Ltd. | SIUI | | | Control | 30 sec Lamp timer | Manual with 15, 30, 45, 60 sec. Lamp timer | | | Field Shape | Rectangular | Rectangular | | | Max. Field Size | 35x35 cm (at 66cm SID) | 44x44cm (at 100cm SID) | | | Inherent Filtration | 0.58 mmAl | 1.2mmAl | | | Standard | Rotating flange | Rotating flange | 5 / 10 {9} ECORAY Co., Ltd. 510(k) Premarket Notification- Traditional | Item | | Proposed Device | Predicate Device | | | --- | --- | --- | --- | --- | | | Option | Tape measure (Max.200cm) | Tape measure (Max.200cm) | Same | | | Electrical Rating | 3.25 VDC, 10 W | 3-12VDC,10W | Similar | | Performance standard | | • IEC 60601-2-28:2017 • IEC 60601-2-54:2022 | • IEC 60601-2-28:2010 • IEC 60601-2-54:2015 | Similar | | Electrical safety | | • IEC 60601-1:2005/AMD1:2012/AMD2:2020 • IEC 60601-1-2:2014/AMD1:2020 • IEC 60601-1-3:2008/AMD1:2013/AMD2:2021 • IEC 62304:2006/AMD1:2015 • IEC 60601-1-6:2010/AMD1:2013/AMD2:2020 • IEC 62366-1:2015, IEC 62366-1:2015/AMD1:2020 | • IEC 60601-1:2012 • IEC 60601-1-2:2014 • IEC 60601-1-3:2013 • IEC 60601-1-6:2015 • IEC 62304:2006+AMD1:2015 • IEC 62366-1:2015/COR1:2016 | Similar | | Compatible X-ray Detector | | • Device trade name : VIVIX-S VW • Model Name : FXRD-3643VAW • Manufacture : VIEWORKS Co., LTD • 510(k) Number : K200418 | - | | ## 6.2. The difference between the subject device and the predicate device The predicate device is intended for acquiring X-ray images of the patient head, cervical spine, abdomen, lumbar spine, pelvis and extremities, while the subject device is intended for diagnostic imaging of body extremities. The differences in specific limitations (e.g., stand-mounted restrictions, use on bariatric patients) do not significantly alter the primary intended use of the device, which is to perform diagnostic imaging on a range of anatomical regions for medical purposes. Thus, the ULTRA 1040 can be seen as having the same fundamental intended use as the SR-8100, without constituting a new intended use. Other differences between ULTRA 1040 and SR-8100 include differences in collimator and X-ray Tube. But the above differences are inherent characteristics of device. The tube current and tube voltage range are similar to predicate device. 6 / 10 {10} 7 / 10 # ECORAY Co., Ltd. 510(k) Premarket Notification- Traditional And Size, Weight, Energy source, Exposure times, Performance standard, Electrical safety are also similar to predicate device, but they are not items that significantly affect performance, and their intended use is equivalent. The proposed device, ULTRA 1040 has been tested for electrical safety, EMC, software and Li-ion polymer battery pack has been validated. {11} ECORAY Co., Ltd. 510(k) Premarket Notification- Traditional 7. SUMMARY OF NON-CLINICAL TESTING Testing was performed in accordance with the following international standards: | No. | Test Items | Standards | | --- | --- | --- | | 1 | Medical electrical equipment Part 1: General requirement for basic safety and essential performance | IEC 60601-1:2005/AMD2:2020 | | 2 | Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment | IEC 60601-1-3:2008+A2:2021 | | 3 | Medical electrical equipment Part 1-6: General requirements for safety – Collateral standard: Usability | IEC 60601-1-6:2010+AMD2:2020 | | 4 | Medical electrical equipment - Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis | IEC 60601-2-28: 2017 | | 5 | Medical electrical equipment - Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy | IEC 60601-2-54:2022 | | 6 | Medical electrical equipment Part1-2: General requirements Section1.2 Collateral standard: Electromagnetic compatibility | IEC 60601-1-2:2014/AMD1:2020 | | 7 | Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems | IEC 62133-2:2017 | | 8 | Safety of laser products - Part 1: Equipment classification and requirements | IEC 60825-1:2007 | | 9 | Diagnostic x-ray system and their major components | 21 CFR 1020.30 | | 10 | Radiographic Equipment | 21 CFR 1020.31 | | 11 | Medical device software - Software life cycle processes | IEC 62304:2006/A1:2015 | 8 / 10 {12} ECORAY Co., Ltd. 510(k) Premarket Notification- Traditional ## 7.1. Safety, EMC and Performance Data Electrical, mechanical, environmental safety and performance testing were conducted according to standard IEC 60601-1, IEC 60601-1-3, IEC 60601-1-6, IEC 60601-2-28, IEC 60601-2-54 and EMC testing was conducted according to IEC 60601-1-2. All test results were satisfying with the standards. ## 7.2. Bench Performance Testing The bench test for the Ultra 1040 Portable X-ray Unit assessed radiation performance, collimator accuracy, battery performance, imaging quality, and safety. The Ultra 1040 Portable X-ray Unit met bench testing acceptance criteria as defined in the test protocols. This testing supports substantial equivalence of the Ultra 1040 technological characteristics and performance compared to the predicate device. ## 7.3. Software The following tests were performed to assess effectiveness of software of the device. The test was performed in accordance with following standards. | No. | Test Items | Standards | | --- | --- | --- | | 1 | Medical device software - Software life cycle processes | IEC 62304:2006/A1:2015 | ## 8. SUMMARY OF CLINICAL PERFORMANCE TESTING A comprehensive, task-based image quality study was conducted to assess the clinical adequacy of the device's imaging performance. This study collected radiographic images for relevant anatomical indications stated in the Indications for Use. Images were acquired for standing mounted imaging. Radiologic technologists acquired images in all acquisition modes with appropriately selected techniques, and radiologist clinically evaluated the image quality. ## 9. SPECIFIC GUIDANCE DOCUMENT All applicable aspects of these guidance documents listed in this 510(k) summary have been addressed. The device also conforms to the following: - 21 CFR 1020 Subchapter J: Performance Standards for Ionizing Radiation Emitting Products, - 21 CFR 1020.30: Diagnostic x-ray systems and their major components, - 21 CFR 1020.31: Radiographic Equipment Furthermore, the following Specific Guidance Document was utilized in the device development to ensure the safety of this device for both the operators and patients: - Content of Premarket Submissions for Device Software Functions Guidance for Industry and Food and Drug Administration Staff June 2023 - Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices Guidance for Industry and Food and Drug Administration Staff September 2016 - Pediatric Information for X-ray Imaging Device Premarket Notifications Guidance for Industry and Food and Drug Administration Staff November 2017 9 / 10 {13} ECORAY Co., Ltd. 510(k) Premarket Notification- Traditional ## 10. CONCLUSIONS Under the comparing substantial equivalence between the subject devices and the predicate devices, there are the same points such as general information, some technical and material information. Although there are some differences, the safety and performance test reports support the safety and effectiveness of the subject device. In this regard, we conclude that the subject device is substantially equivalent to the predicate device. 10 / 10
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...