K211473 · Oxos Medical, Inc. · IZL · Jun 4, 2021 · Radiology
Device Facts
Record ID
K211473
Device Name
Micro C Medical Imaging System, M01
Applicant
Oxos Medical, Inc.
Product Code
IZL · Radiology
Decision Date
Jun 4, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1720
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The M01 System is a hand-held X-ray system designed to aid clinicians with point of care visualization through diagnostic X-rays of extremities.
Device Story
Handheld, portable X-ray system for point-of-care diagnostic imaging of distal extremities. Comprises three subsystems: Emitter (X-ray tube, computer vision camera, operator controls), Cassette (digital X-ray detector, positioning lights), and Control Unit (high voltage generator, computing, display interface). Operator positions emitter over patient anatomy using computer vision positioning system; captures static/serial radiographs. System interfaces with external monitor, keyboard, mouse; supports PACS transfer. Used in clinical/surgical environments by trained clinicians. Provides diagnostic images to assist clinical decision-making. Benefits include portable, point-of-care visualization of extremities.
Clinical Evidence
No clinical data. Bench testing only, including usability testing for surgical environments and cleaning/disinfection validation.
Technological Characteristics
Handheld mobile X-ray system. Components: Emitter, Cassette (6x6" digital detector, Cesium Iodide scintillator), Control Unit. Energy: 120 VAC/60 Hz. Fixed mA (1.0), selectable kVp (40, 50, 60). Exposure times: 33ms, 66ms, 99ms. Connectivity: WiFi, Ethernet, USB 2.0, DICOM 3.0 compliant. Software: Custom validated, includes user interface for image adjustment (zoom, rotate, brightness, contrast, sharpness).
Indications for Use
Indicated for use by qualified/trained clinicians on adult and pediatric patients for taking diagnostic static and serial radiographic exposures of extremities. Not intended to replace radiographic systems with variable tube current and voltages (kVp) required for full optimization of image quality and radiation exposure for different exam types.
Regulatory Classification
Identification
A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
K203658 — Micro C Medical Imaging System, M01 · Oxos Medical, Inc. · Jan 12, 2021
K193535 — EZER, Portable X-ray System · Livermoretech, Inc. · May 7, 2020
K212654 — Micro C Medical Imaging System, M01 · Oxos Medical, Inc. · Feb 4, 2022
K230581 — TOPA12 Portable X-ray Unit · Neuf, Inc. · Aug 16, 2023
K252068 — MC2 Portable X-ray System · Oxos Medical · Dec 22, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
June 4, 2021
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
OXOS Medical, Inc. % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive. Suite 510k SAINT PAUL MN 55114
Re: K211473
Trade/Device Name: Micro C Medical Imaging System, M01 Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: Class II Product Code: IZL Dated: May 11, 2021 Received: May 12, 2021
Dear Prithul Bom:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211473
Device Name Micro C Medical Imaging System, M01
#### Indications for Use (Describe)
The Micro C Medical Imaging System, M01 is a handheld and portable general purpose X-ray system that is indicated for use by qualified/trained clinicians on adult and pediatric patients for taking diagnostic static and serial radiographic exposures of extremities. The device is not intended to replace a radiographic system that has both variable tube current and voltages (kVp) in the range that may be required for full optimization of image quality and radiation exposure for different exam types.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size:14px">☒</span> Prescription Use (Part 21 GER 801 Subpart D) </div> | <div> <span style="font-size:14px">☐</span> Over-The-Counter Use (21 CER 801 Subpart C) </div> |
|----------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for OXOS Medical. The logo consists of four symbols: a circle with two gaps, an X, another circle with two gaps, and the letter S. Below the symbols, the word "MEDICAL" is written in a smaller font.
# Section 5 510(k) Summary K211473
In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the Micro C Medical Imaging System, M01 Traditional 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document, 'Format for Traditional and Abbreviated 510(k)s', issued on August 12, 2005.
| Submitter: | OXOS Medical, Inc<br>1230 Peachtree St NE<br>Suite #300<br>Atlanta, GA 30309<br>Tel: 1-855-SEE-XRAY |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submission Contact: | Mo Khosravanipour<br>Director of Program Management<br>OXOS Medical, Inc.<br>Email: Mo@oxos.com |
| | Grace Powers, MS, MBA, RAC<br>Founder/Principal Consultant<br>Powers Regulatory Consulting<br>Email: grace@powersregulatory.com |
| Submission Date: | April 16, 2021 |
| Subject Device: | Trade Name: Micro C Medical Imaging System, M01<br>Common Name: System, X-Ray, Mobile<br>Regulation: 21 CFR § 892.1720<br>Regulatory Classification: 2<br>Product Code: IZL<br>Classification Panel: Radiology |
| Predicate Device: | Legally marketed device to which substantial equivalence is claimed:<br>Micro C Medical Imaging System, M01 (K203658)<br>Common Name: System, X-Ray, Mobile<br>Regulation: 21 CFR § 892.1720<br>Regulatory Classification: 2<br>Product Code: IZL<br>Classification Panel: Radiology |
{4}------------------------------------------------
## Section 5 510(k) Summary
## Device Description:
The Micro C Medical Imaging System, M01 (subject device) is a handheld X-ray system designed to aid clinicians with point of care visualization through diagnostic X-rays of distal extremities. The device allows a clinician to select desired technique factors best suited for their patient anatomy. The Micro C Medical Imaging System, M01 consists of three major subsystems: The Emitter, Cassette, and Control Unit. The System is intended to interface an external Monitor (touchscreen or non-touchscreen display), keyboard and a mouse, and can provide a remote operator interface over the network to a laptop. The Micro C Medical Imaging System, M01 utilizes a computer vision positioning system to allow the emitter to be positioned above the patient anatomy and aligned to the cassette by the operator. The device is used in a clinical environment. A description of the three major sub-systems is listed below.
- Emitter: This component contains the operator control panel, X-ray tube, and computer vision camera. The control panel allows the operator to control the major functions of the device, including the technique factors. This component is controlled and held in the operator's hand.
- Cassette: This component contains the X-ray detector that collects the X-ray energy and provides a digital representation to the control unit for eventual display. This component also contains status lights and IR lights to assist in X-ray field positioning. The patient anatomy of interest is placed on top of this module.
- . Control Unit: This component contains the High Voltage generator, computing power, monitor and keyboard inputs, and other electronics required for the functioning of the device. This module is typically placed on a shelf, cart, counter, or other flat surface convenient to the operator and environment.
The system is intended to work in conjunction with a DICOM monitor, keyboard and a mouse and the mains power outlet. These are accessories are not included in the submission.
The Micro C Medical Imaging System, M01 has custom validated software that includes a user interface that allows the operator to view and adjust captured radiographs and transfer radiographs to a PACS server or flash drive. The device software is based on the predicate device and no changes to the software were performed for this submission.
Image /page/4/Figure/9 description: The image shows a white box connected to a white handheld device and a white square device via cables. Two blue cables connect the box to the handheld device and the square device. A gray cable connects the box to the square device.
## Figure 5-1: Subject Device - Micro C Medical Imaging System, M01
{5}------------------------------------------------
Image /page/5/Picture/2 description: The image shows a computer setup with a monitor, keyboard, mouse, and a white box-shaped device. The monitor is connected to the white box, which also has connections for the keyboard, mouse, and an Ethernet cable. The power cords for the monitor and the white box are plugged into wall outlets.
## Intended Use:
The M01 System is a hand-held X-ray system designed to aid clinicians with point of care visualization through diagnostic X-rays of extremities.
## Indications for Use:
The Micro C Medical Imaging System, M01 is a handheld and portable general purpose X-ray system that is indicated for use by qualified/trained clinicians on adult and pediatric patients for taking diagnostic static and serial radiographic exposures of extremities.
The device is not intended to replace a radiographic system that has both variable tube current and voltages (kVp) in the range that may be required for full optimization of image quality and radiation exposure for different exam types.
## Technological Characteristics
The Micro C Medical Imaging System, M01 is identical to the predicate device cleared via K203658. The purpose of this submission is to add the pediatric population to the indication and to allow use in surgery.
The table below compares the subject and predicate device.
| | | Comparison | |
|-----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|
| Subject Device:<br>Micro C Medical Imaging<br>System, M01 | Predicate Device: Micro<br>C Medical Imaging<br>System, M01 (K203658) | | |
| Product Code<br> | IZL (Mobile X-Ray<br>System) | IZL (Mobile X-Ray<br>System) | Identical |
| Regulation | 21 CFR 892.1720 | 21 CFR 892.1720 | Identical |
| Classification<br>Name | Mobile X-Ray System | Mobile X-Ray System | Identical |
| Classification | Class 2 | Class 2 | Identical |
| Indication for<br>Use | The Micro C Medical<br>Imaging System, M01 is a<br>handheld and portable<br>general purpose X-ray<br>system that is indicated | The Micro C Medical<br>Imaging System, M01 is<br>a handheld and<br>portable general<br>purpose X-ray system | Different- The subject device<br>can be used in pediatric<br>populations. |
| | for use by<br>qualified/trained<br>clinicians on adult<br>patients for taking<br>diagnostic static and<br>serial radiographic<br>exposures of extremities.<br>The device is not<br>intended to replace a<br>radiographic system that<br>has both variable tube<br>current and voltages<br>(kVp) in the range that<br>may be required for full<br>optimization of image<br>quality and radiation<br>exposure for different<br>exam types. | that is indicated for use<br>by qualified/trained<br>clinicians on adult and<br>pediatric patients for<br>taking diagnostic static<br>and serial radiographic<br>exposures of<br>extremities.<br>The device is not<br>intended to replace a<br>radiographic system<br>that has both variable<br>tube current and<br>voltages (kVp) in the<br>range that may be<br>required for full<br>optimization of image<br>quality and radiation<br>exposure for different<br>exam types. | |
| Age of Device<br>Use | Adults and Pediatric | Adults | Different- The subject device<br>can be used in pediatric<br>populations. |
| Principle of<br>Operation | General purpose<br>diagnostic X-ray | General purpose<br>diagnostic X-ray | Identical |
| Image type<br>produced | Static, serial radiographic<br>and photographic images<br>for convenience. | Static, serial<br>radiographic and<br>photographic images for<br>convenience. | Identical |
| Detector | 6 x 6" digital detector | 6 x 6" digital detector | Identical |
| Collimator | The removable fixed<br>collimators (referred to<br>as pucks) | The removable fixed<br>collimators (referred to<br>as pucks) | Identical |
| Weight | Emitter: 2.86kg (6.3lbs)<br>Cassette: 6.5kg (14.3lbs)<br>Control Unit: 8.6kg<br>(19.0lbs) | Emitter: 2.86kg (6.3lbs)<br>Cassette: 6.5kg (14.3lbs)<br>Control Unit: 8.6kg<br>(19.0lbs) | Identical |
| Dimension/<br>Size | Emitter: 9.3"H x 3.5"W x<br>8.3"L (excluding SSD<br>Cone)<br>Cassette: 15.5"H x<br>15.5"W x 2.8"L<br>Control Unit: 16.4"H x<br>12.9"W x 5.5"L | Emitter: 9.3"H x 3.5"W x<br>8.3"L (excluding SSD<br>Cone)<br>Cassette: 15.5"H x<br>15.5"W x 2.8"L<br>Control Unit: 16.4"H x<br>12.9"W x 5.5"L | Identical |
| Triggering<br>Mechanism | Two stage triggering | Two stage triggering | Identical |
| Minimum<br>Source to<br>skin distance<br>(SSD) | 20 cm SSD Cone ensures<br>minimum SSD of 20 cm | 20 cm SSD Cone ensures<br>minimum SSD of 20 cm | Identical |
| Source to<br>Detector<br>distance | 20 - 45 cm | 20 - 45 cm | Identical |
| Light Field | Virtual light field on<br>Monitor UI. No projected<br>light field. | Virtual light field on<br>Monitor UI. No<br>projected light field. | Identical |
| Energy<br>Source | 120 VAC / 60 Hz (no<br>rechargeable battery) | 120 VAC / 60 Hz (no<br>rechargeable battery) | Identical |
| Exposure<br>Time | 33ms, 66ms and 99ms | 33ms, 66ms and 99ms | Identical |
| mA | 1.0 mA fixed | 1.0 mA fixed | Identical |
| kVp | 40kVp, 50kVp, and 60kVp | 40kVp, 50kVp, and<br>60kVp | Identical |
| Scintillator | Cesium lodide (CsI) | Cesium lodide (CsI) | Identical |
| Resolution/<br>Pixel size | 99 μm | 99 μm | Identical |
| DQE @<br>0Lp/mm | 70% | 70% | Identical |
| MTF @ 1<br>Lp/mm,<br>RQA5 | 60% | 60% | Identical |
| Ingress<br>Protection<br>Rating | IPOO | IPOO | Identical |
| Image<br>Processing | User Interface can be<br>used to drag, zoom,<br>rotate and also adjust<br>brightness, contrast, and<br>sharpness. | User Interface can be<br>used to drag, zoom,<br>rotate and also adjust<br>brightness, contrast,<br>and sharpness. | Identical |
| Connectivity<br>Options | WiFi, Ethernet, Four USB<br>2.0 ports | WiFi, Ethernet, Four<br>USB 2.0 ports | Identical |
| DICOM | Yes- DICOM 3.0<br>Compliant | Yes- DICOM 3.0<br>Compliant | Identical |
| Device<br>Package<br>Contents | ● Cassette<br>● Control Unit<br>● Emitter<br>● Collimation Pucks<br>● SSD Cone<br>● Cassette Power<br>Cable<br>● Cassette Data Cable<br>● Control Unit Power<br>Cable | ● Cassette<br>● Control Unit<br>● Emitter<br>● Collimation Pucks<br>● SSD Cone<br>● Cassette Power<br>Cable<br>● Cassette Data Cable<br>● Control Unit Power<br>Cable | Identical |
| | Connector Covers Instructions for Use Case | Connector Covers Instructions for Use Case | |
## Table 5-2: Device Comparison
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
#### Non-Clinical Performance Data
Testing specific to the use in surgery was performed and includes:
- Usability testing for use in surgery
- Cleaning and Disinfection Studies ●
The following specific guidance document was utilized in the device development to ensure the safety of this device for both the operators and patients:
- Guidance for Medical X-ray Imaging Devices Conformance with IEC Standards ●
- . Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices
- Radiation Safety Considerations for X-ray Equipment Designed for Hand-held Use
- Pediatric Information for X-ray Imaging Device Premarket Notifications
#### Conclusion
The Micro C Medical Imaging System, M01 is identical in design to the legally marketed predicate device. The updated indication and contraindication are acceptable as demonstrated by the performance data and does not raise different questions of safety and effectiveness.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.