GM60A-32S & GM60A-40S

K150097 · Samsung Electronics Co., Ltd. · IZL · Feb 19, 2015 · Radiology

Device Facts

Record IDK150097
Device NameGM60A-32S & GM60A-40S
ApplicantSamsung Electronics Co., Ltd.
Product CodeIZL · Radiology
Decision DateFeb 19, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1720
Device ClassClass 2

Indications for Use

The GM60A Digital Mobile X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.

Device Story

Mobile X-ray system; captures radiographic images of human anatomy. Components: high voltage generator, X-ray tube, collimator, detector, DAP, barcode scanner. X-rays pass through patient; detector converts to electrical signals; signal processing device performs amplification and digital conversion; operation software saves as DICOM files. Images transmitted to PACS server for radiologist review. Used in clinical settings by physicians or technicians. Benefits: provides diagnostic imaging capabilities in a mobile format.

Clinical Evidence

Clinical images provided to support system performance. Images evaluated by a board-certified radiologist and found equivalent to the predicate device. Non-clinical data included MTF and DQE measurements per IEC 62220-1, showing no difference from the predicate.

Technological Characteristics

Mobile X-ray system; High Frequency generator (32kW/40kW); 40-150kVp range; 10-500mA range. Detectors: CsI, indirect, 140um pixel pitch. Connectivity: Wired/Wireless. Standards: ES 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-28, IEC 60601-2-54, ISO 14971, 21 CFR 1020.30/31.

Indications for Use

Indicated for generating radiographic images of human anatomy. Intended for use by qualified/trained doctors or technicians. Not for mammographic applications.

Regulatory Classification

Identification

A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three heads facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 19, 2015 Samsung Electronics Co., Ltd. % Chulsin Kim Regulatory Affairs Manager 129, Samsung-ro, Yeongtong-gu Suwon-si, Gyeonggi-do 443-742 REPUBLIC OF KOREA Re: K150097 Trade/Device Name: GM60A-32S & GM60A-40S Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Code: IZL Dated: January 14, 2015 Received: January 20, 2015 Dear Chulsin Kim: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Robert A Ochs Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use 510(k) Number (if known) K150097 Device Name GM60A-32S & GM60A-40S Indications for Use (Describe) The GM60A Digital Mobile X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications. | Type of Use (Select one or both, as applicable) | | | | |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|--| | X Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) | | | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | | | This section applies only to requirements of the Paperwork Reduction Act of 1995.<br>*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the<br>time to review instructions, search existing data sources, gather and maintain the data needed and complete<br>and review the collection of information. Send comments regarding this burden estimate or any other aspect<br>of this information collection, including suggestions for reducing this burden, to: | | | | | | Food and Druq Administration<br>Office of Chief Information Officer<br>Paperwork Reduction Act (PRA) Staff<br>PRAStaff(@fda.hhs.gov | Department of Health and Human Services | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Form Approved: OMB No. 0910-0120 {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The word is placed on a blue, oval background that is tilted slightly. 510(k) Premarket Notification - Traditional ## Section 5: 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted accordance with requirements of 21 CFR 807.92 - 1. Date: January 14, 2015 - Submitter 2. - A. Company Name: SAMSUNG ELECTRONICS Co., Ltd. - B. Address: 129, Samsung-ro, Yeongtong-gu, Suwon-si, Gyeonggi-do, 443-742, Republic of Korea #### Primary Contact Person 3. - A. Name: KIM, CHULSIN - B. Title: Regulatory Affairs Manager - C. Phone Number: +82-31-200-7661 - D. FAX Number: +82-31-200-1199 E-Mail: chulsin.kim@samsung.com #### Secondary Contact Person 4. - A. Name: Ninad Gujar - B. Title: Regulatory Affairs Manager - C. Phone Number: 978-564-8503 - D. FAX Number: 978-750-6677 E-Mail: ngujar@samsungneurologica.com #### 5. Identification Device - A. Trade Name: GM60A - B. Device Name: GM60A-32S, GM60A-40S - C. Common Name: Digital Diagnostic Mobile X-ray System - D. Classification Name: Mobile X-ray System - ட். Product Code: IZL - F. Regulation: 21 CFR 892.1720 #### Predicate Device 6. - A. Manufacturer: SAMSUNG ELECTRONICS Co., Ltd. - B. Trade Name: GM60A - C. Device Name: GM60A-32S, GM60A-40S - D. Classification Name: Mobile X-ray system - E. Product Code: IZL - F. 510(k) Number: K142492 - G. 510(k) Decision Date: December 17, 2014 #### Device Description 7. {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters against a blue background. The logo is simple and recognizable. ### 510(k) Premarket Notification - Traditional The GM60A Digital Mobile X-ray imaging system consists of High voltage generator (HVG), Xray tube, Collimator, Detector, DAP and Barcode scanner. This system is used to capture images by transmitting X-ray to a patient's body. The X-ray passing through a patient's body is sent to the detector and then converted into electrical signals. These signals go through the process of amplification and digital data conversion in the signal process device before being sent to the operation software and saved in DICOM file, a standard for medical imaging. The captured images are sent to the Picture Archiving & Communication System (PACS) server, and can be used for reading images. - 8. Intended Use The GM60A Digital Mobile X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications. ### 9. Summary of Substantial Equivalence Discussion The proposed GM60A adds a new detector (S3025-W) to the predicate device GM60A (K142492), and it has same materials, energy source or technological characteristics compare to the predicate device. The S3025-W detector has same pixel pitch and resolution from those detectors (S4343-W & S4335-W) used for the predicate device, and has just in differences of size. Comparisons of the following technological characteristics were assessed and the results demonstrate the substantial equivalence to the predicates. | Specification | Proposed Device | Predicate Device | Discussion | |---------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------| | Device Name | GM60A | GM60A | | | Manufacturer | SAMSUNG ELECTRONICS | SAMSUNG ELECTRONICS | | | 510(k) Number | N/A | K142492 | | | Appearances | Image: Proposed Device Appearance | Image: Predicate Device Appearance | Same | | Intended Use | The GM60A Digital Mobile X-ray Imaging<br>System is intended for use in generating<br>radiographic images of human anatomy by<br>a qualified/trained doctor or technician.<br>This device is not intended for<br>mammographic applications. | The GM60A Digital Mobile X-ray Imaging<br>System is intended for use in generating<br>radiographic images of human anatomy by<br>a qualified/trained doctor or technician.<br>This device is not intended for<br>mammographic applications. | Same | | Manufacturer Contents | GM60A | GM60A(K142492) | Discussion | | |-----------------------|-------------------|----------------|--------------|------| | (1)High | Voltage Generator | | | | | Type | High Frequency | High Frequency | Same | | | Max. Power | 32kW / 40kW | 32kW / 40kW | Same | | | Output<br>RANGE | kVp Range | 40 to 150kVp | 40 to 150kVp | Same | | | mA Range | 10 to 500mA | 10 to 500mA | Same | {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The background is a blue, rounded rectangle. 510(k) Premarket Notification - Traditional | Manufacturer | Contents | GM60A | GM60A(K142492) | Discussion | |--------------|---------------|-------------|----------------|------------| | | Exposure Time | 1msec-10sec | 1msec-10sec | Same | | Manufacturer Contents | GM60A | GM60A(K142492) | Discussion | | |-----------------------|-----------------|-----------------------------|-----------------------------|------| | (2)Tube assembly | | | | | | Moving<br>Range | Horizontal | 700~1250 | 700~1250 | Same | | | Vertical | 550~2020 / 550~1890(Option) | 550~2020 / 550~1890(Option) | Same | | Rotation<br>Range | Column | $\pm$ 315° | $\pm$ 315° | Same | | | Tube(Arm axis) | $\pm$ 180° | $\pm$ 180° | Same | | | Tube(Tube axis) | -30°~90° | -30°~90° | Same | | Collimator | | RALCO R221 DHHS | RALCO R221 DHHS | Same | | Manufacturer<br>Contents | GM60A | GM60A(K142492) | | Discussion | |-------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|--------------------------|--------------------------|--------------| | (3) Detectors<br>*NOTE: S3025-W detector as an option is added in the list of detectors that are S4335-W, S4343-W (cleared with K142492). | | | | | | Name | S4335-W | S4343-W | S3025-W | Different(1) | | Detector Type | Csl | Csl | Csl | Same | | | Indirect | Indirect | Indirect | Same | | Detector Area | 14"X17"<br>(345mmX425mm) | 17"X17"<br>(425mmX425mm) | 10"X12"<br>(245mmX295mm) | Different(2) | | Number of pixels | 2466X3040 | 3036X3040 | 1750X2108 | | | Pixel Pitch(um) | 140 | 140 | 140 | Same | | High Contrast<br>Limiting Resolution<br>(LP/mm) | 3.57 | 3.57 | 3.57 | Same | | Communication | Wired /<br>Wireless | Wired /<br>Wireless | Wired /<br>Wireless | Same | | (4) Grid | | | | | | Lines/cm | 84.6 | | | Same | | Grid mechanism | Stationary | | | Same | | Removability | Removable | | | Same | | Note | Differences | Explanation | |------|----------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | (1) | New detector,S3025-W, is<br>included | Small size of new detector, S3025-W is included as an option. | | (2) | Detector area is different from<br>those of predicate device also<br>number of pixels. | Detector area and number of pixels are different from those of<br>predicate device but new detector has same technical<br>characteristics, pixel pitch and resolution of detectors of<br>predicate device. The new detector has just differences in size<br>from predicate device and these differences do not contribute<br>any adverse impacts to the predicate device's safety and<br>performance. | In non-clinical data, the proposed detector shows curves and measurements of MTF and DQE {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The background is a solid, dark blue color. #### 510(k) Premarket Notification - Traditional that do not differ from the predicate device. In clinical data, the proposed GM60A has been shown a substantially equivalent to the predicate device. ### 10. Safety, EMC and Performance Data Electrical, mechanical, environmental safety and performance testing according to standard ES 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-28, IEC 60601-2-54, ISO14971, 21CFR1020.30 and 21CFR1020.31 were performed, and EMC testing was conducted in accordance with standard IEC 60601-1-2:2007. Wireless function was tested and verified followed by guidance, Radio frequency Wireless Technology in Medical Devices. All test results were satisfied as the standard. ### 11. Non-clinical data Non-clinical testing data was provided in conformance to the FDA "Guidance for the Submission of 510(k)'s for Solid-State X-ray Imaging Devices", which includes MTF and DQE measurements as tested by IEC 62220-1. The proposed device shows no difference in nonclinical testing data such as MTF and DQE measurements from the predicate device. ### 12. Clinical data Clinical images were provided; these images were not necessary to establish substantial equivalence based on the modifications to the device but they provide further evidence in addition to the laboratory performance data to show that the complete system works as intended. These images were evaluated by a radiologist with equivalent U.S. board certification and found to be equivalent to the predicate device. ### 13. Conclusions The non-clinical and clinical data demonstrate that the device is as safe, as effective, and performs as well as the legally marketed device identified in paragraph 6. - 14. Samsung Electronics Co., Ltd. will update and include in this summary any other information deemed seasonably necessary by the FDA
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...