K143537 · Hitachi Medical Systems America, Inc. · IZL · Mar 6, 2015 · Radiology
Device Facts
Record ID
K143537
Device Name
Sirius Starmobile tiara
Applicant
Hitachi Medical Systems America, Inc.
Product Code
IZL · Radiology
Decision Date
Mar 6, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1720
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Mobile X-ray Unit Sirius Starmobile tiara is a general radiography system and is composed of the X-ray high voltage generator, X-ray tube, support unit, and digital radiograph device (DR-ID 800) made by Fujifilm Corporation. This device is designed for pediatric and adult patients. It is intended for use in general radiography of the head, body, or extremities including pediatric exams. The device output can provide an aid to diagnosis when used by a qualified physician.
Device Story
Battery-powered, mobile X-ray system; includes high-voltage generator, X-ray tube, support unit, and Fujifilm DR-ID 800 flat panel detector. Operated by technologists in clinical settings; provides rapid image availability to verify exam adequacy. Features motorized travel via dual rear motors and pantograph arm for positioning. Wireless connectivity enables RIS/HIS worklist updates. Output consists of digital radiographic images for physician diagnostic review. Benefits include reduced exam completion time and minimized return trips for repeat imaging.
Clinical Evidence
No clinical trials performed. Evidence consists of bench testing for safety and performance (IEC/AAMI standards) and sample clinical images to demonstrate image quality and diagnostic utility for general radiography applications.
Indicated for general radiography of the head, body, or extremities in pediatric and adult patients. Used as an aid to diagnosis by qualified physicians.
Regulatory Classification
Identification
A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Carestream DRX-Revolution Mobile X-ray System (K120062)
Reference Devices
FDR D-EVO Flat Panel Detector System (DR-ID600) (K132509)
Related Devices
K101517 — SEDECAL MOBILE DIGITAL DIAGNOSTIC X-RAY · Sedecal., Sa. · Aug 5, 2011
K090322 — EASY MOVING PLUS, SM-XXHF-YY · Sedecal., Sa. · Mar 17, 2009
K173018 — MDR17 Mobile Direct Radiographic System · Del Medical, Inc. · Nov 20, 2017
K213568 — DRX-Rise Mobile X-ray System · Carestream Health · Mar 23, 2022
K142383 — Optima XR200amx, Optima XR220amx · Ge Medical Systems, LLC · Sep 25, 2014
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the HHS logo, which consists of a stylized caduceus with three intertwined snakes and a staff. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the logo.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 6, 2015
Hitachi Medical Systems America, Inc. % Mr. Doug Thistlewaite Manager of Regulatory Affairs 1959 Summit Commerce Park TWINSBURG OH 44087
Re: K143537
Trade/Device Name: Sirius Starmobile tiara Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Code: IZL Dated: December 9, 2014 Received: December 15, 2014
Dear Mr. Thistlewaite:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K143537
Device Name Sirius Starmobile tiara
Indications for Use (Describe)
The Mobile X-ray Unit Sirius Starmobile tiara is a general radiography system and is composed of the X-ray high voltage generator, X-ray tube, support unit, and digital radiograph device (DR-ID 800) made by Fujifilm Corporation.
This device is designed for pediatric and adult patients.
It is intended for use in general radiography of the head, body, or extremities including pediatric exams. The device output can provide an aid to diagnosis when used by a qualified physician.
Type of Use (Select one or both, as applicable)
| <div> <svg height="16" width="16"> <path d="M2 2h12v12H2zM3 3v10h10V3H3zM4 4h8v8H4V4z" fill="black"></path> <path d="M4 4h8v8H4z" fill="white"></path> <path d="M4.7 11.3l-1.4-1.4 5.3-5.3-5.3 5.3zM11.3 4.7l-5.3 5.3 1.4 1.4 5.3-5.3z" fill="black"></path> </svg> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <svg height="16" width="16"> <path d="M2 2h12v12H2zM3 3v10h10V3H3z" fill="black"></path> </svg> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### Submitter Information
| Submitter: | Hitachi Medical Systems America, Inc.<br>1959 Summit Commerce Park<br>Twinsburg, Ohio 44087-2371<br>ph: (330) 425-1313<br>fax: (330) 963-0749 |
|------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Douglas J. Thistlethwaite |
| Date: | December 9, 2014 |
### Device Name
| Regulation Number: | 21 CFR 892.1720 |
|-------------------------|---------------------------------------------------------|
| Regulation Name: | Mobile X-ray System |
| Product Code | IZL-System, X-ray, Mobile |
| Trade/Proprietary Name: | Sirius Starmobile tiara System |
| Predicate Device(s): | Carestream DRX-Revolution Mobile X-ray System (K120062) |
## Device Intended Use
The Mobile X-ray Unit Sirius Starmobile tiara is a general radiography system and is composed of the X-ray high voltage generator, X-ray tube, support unit, and digital radiograph device (DR-ID 800) made by Fujifilm Corporation.
This device is designed for pediatric and adult patients.
It is intended for use in general radiography of the head, body, or extremities including pediatric exams. The device output can provide an aid to diagnosis when used by a qualified physician.
## Device Description
The Sirius Starmobile tiara battery powered, mobile x-ray system features a built-in Flat Panel Detector System: DR-ID800 which is a modification of the FDR D-EVO Flat Panel Detector System (DR-ID600), K132509. Because the Flat Panel Detector System is incorporated in the mobile equipment, the images are available to the technologist in a very short time, allowing the technologist to assure the exam has been performed adequately, minimizing return trips. Wireless communication is available, as an option, for updates to the patient worklist from the RIS/HIS.
The Sirius Starmobile tiara provides smooth and quiet motorized travel capability via rear wheels independently driven by dual motors, a versatile radiography range through the pantograph arm, and easy-to-operate positioning of flat detector providing sharp image quality with a short exam completion time.
# Device Technological Characteristics
The Sirius Starmobile tiara is substantially equivalent to the predicate device, the Carestream DRX-Revolution Mobile X-ray System. The Sirius Starmobile tiara and the Carestream DRX-Revolution Mobile X-ray System both consist of an x-ray generator, x-ray tube, collimator, and graphical user interface on an operator console with touch screen monitor. These components are mounted on a motorized cart that is battery powered to enable the device to be moved from location to location. Both systems utilize a digital flat panel detector for image capture and can also be used to expose CR storage phosphor or film cassettes. Both systems have operator console software with image processing capability.
{4}------------------------------------------------
# Summary of Clinical/Non-Clinical Testing
## Non-Clinical Testing
The Sirius Starmobile tiara was subjected to the following laboratory testing:
- AAMI ANSI ES60601-1:2005/(R) 2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 ● Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD)
- o IEC 60601-1-2 Edition 3: 2007 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
- IEC 60601-1-3 Edition 2.1 2013 Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment
- . IEC 60601-1-6 Edition 3.1 2013 Medical electrical equipment Part 1-6: General requirements for basic safety and essential performance Collateral standard: Usability
- . IEC 60601-2-28 Edition 2.0 2010 Medical electrical equipment - Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis
- . IEC 60601-2-54 Edition 1.0 2009 Medical electrical equipment - Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy
- IEC 62304 Edition 1.0, Medical device software Software life cycle processes .
#### Clinical Testing
The Sirius Starmobile tiara System submission includes sample clinical images in Section 10 Performance.
# Substantial Equivalence
Both the Sirius Starmobile tiara System (subject device) and the predicate device have the identical intended use for examinations of adult and pediatric patients over all parts of the body.
The key technological characteristics between the subject device and the predicate are substantially equivalent.
In regards to the x-ray tube, the subject device has a larger target angle, but the focal spot size is the same as the predicate, so beam coverage is equivalent. Anode heat capacity is less, but in mobile use the number of exams are less than conventional x-ray, so the lower heat capacity will not impact the ability to perform exams.
X-ray tube head movement is comparable between the subject device and the predicate. The maximum SID to floor is similar, but the subject device has a wider range of movement from the floor. While tube arm reach is slightly less than the predicate, all other motions of the subject device have a wider range.
All the detectors available for the subiect device have comparable sizes to the predicate. Detector pixel size of the subject device (0.150mm) is within the tvpical range for general radiography (0.139 to 0.200mm). Section 10, Performance, shows bench test results and sample clinical data indicating that the subject device covers the range of body and extremity imaging, with acceptable image quality, as compared to general mobile radiography applications available on the predicate device and others in its class.
{5}------------------------------------------------
# Conclusions
It is the opinion of Hitachi Medical Systems America, Inc. the Sirius Starmobile tiara is substantially equivalent with respect to hardware, base elements of the software, safety, effectiveness, and functionality to the Carestream DRX-Revolution Mobile X-ray System (K120062).
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.