K111912 · Siemens Medical Solutions USA, Inc. · IZL · Sep 19, 2011 · Radiology
Device Facts
Record ID
K111912
Device Name
MOBILETT MIRA
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
IZL · Radiology
Decision Date
Sep 19, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1720
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Mobilett Mira is a radiographic system designed for use in wards, intensive care and premature-birth wards, pediatric and emergency departments, operating theatres as well as the central X-ray department.
Device Story
Mobilett Mira is a mobile X-ray system for healthcare professionals. Input: X-ray radiation directed at anatomical areas. Operation: System positioned at bedside; X-ray beam collimated to area of interest; solid-state detector captures image data; data transmitted wirelessly to mobile unit. Features: 35KW generator; rotatable suspension arm; motor-driven movement; rechargeable batteries; remote exposure control. Usage: Used in wards, ICU, pediatric/emergency departments, ORs, and radiology departments. Output: Digital radiographic images displayed for clinician review. Clinical impact: Enables bedside imaging for patients unable to be transported to central radiology; facilitates rapid diagnostic decision-making.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Mobile X-ray system; 35KW generator; solid-state image detector; wireless data transmission; motor-driven movement; rechargeable battery power; rotatable suspension arm; collimator with light field. System includes visual/audible safety warnings and automated error monitoring/blocking.
Indications for Use
Indicated for radiographic imaging of patients in clinical settings including wards, ICU, premature-birth wards, pediatric departments, emergency departments, operating theatres, and central X-ray departments.
Regulatory Classification
Identification
A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
K173187 — MobileDiagnost M50 · Philips India Limited · Nov 16, 2017
K192936 — Soltus 500 · Sedecal., Sa. · Nov 15, 2019
K103476 — BRIVOXR285AMX, OPTIMA XR200AMX, OPTIMAXR220AMX · Ge Medical Systems, LLC · Jun 16, 2011
K212134 — Diagnostic X-ray System/PLX5200A · Nanjing Perlove Medical Equipment Co., Ltd. · Aug 25, 2021
K222339 — uDR 380i Pro · Shanghai United Imaging Healthcare Co., Ltd. · Sep 30, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
#### Special 510(k) Submission: Mobilett Mira
## 510(k) Summary: Mobilett Mira
SEP 1 9 2011
Company: Siemens Medical Systems. Inc. 1 Valley Stream Parkway Malvern, PA 19355
111912
Date Prepared: June 30, 2011
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
#### General Information: 1.
### lmporter / Distributor: Siemens Medical Solutions, Inc.
51 Valley Stream Parkway, E-50 Malvern, PA 19355 Establishment Registration Number: 2240869
#### Manufacturing Site:
Siemens S.A. Parque Empresarial La Carpetana, Avenida Leonardo da Vinci, 15 E-28906 Getafe (Madrid), Spain Headquarters: Siemens AG Wittelsbacherplatz 2 D-80333 Munich 2, Germany Establishment Registration Number: 1000342169
#### 2. Contact Person:
Ms. Patricia D Jones Technical Specialist, Regulatory Submissions Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway G-01 Malvern, PA 19355 Phone: (610) 448 -3536 Fax: (610) 448-1787 Email: patricia.d.jones@siemens.com
{1}------------------------------------------------
K111912
Page 2 of 3
# SIEMEN
#### Special 510(k) Submission: Mobilett Mira
- 3. Device Name and Classification: Trade Name: Classification Name: Classification Panel: Classification Regulation: Device Class: Product Code:
Mobilett Mira Mobile X-ray System Radiology 21 CFR §892.1720 Class II IZL
## 4. Device Description:
The Mobilett Mira is a mobile X-ray system with a solid state detector. The system is designed to provide X-ray imaging for healthcare professionals. Cable less operation and motor-driven movements are supported with build in rechargeable batteries. The mobile generator is positioned at bedside and the X-ray system is directed to the anatomical area to be imaged. The image detector is placed perpendicular to the central beam behind the anatomical area of interest. The system features a collimator with a light field that mimics the x-ray field to limit the field of exposure to the area to be imaged. Exposure may be released via remote control. The image data acquired by the detector are sent wireless to the mobile unit.
The Mobilett Mira is a modification of the Mobilett XP originally cleared under Premarket Notification K033238 on 11/14/2003. The modification consists of a rotatable suspension arm, a solid state image detector, digital image acquisition system and a 35KW generator. This modified Mobilett XP will be marketed under the trade name Mobilett Mira. The modification does not affect the intended use of the device nor does it alter its fundamental scientific technology.
### 5. Intended Use:
The Mobilett Mira is a radiographic system designed for use in wards, intensive care and premature-birth wards, pediatric and emergency departments, operating theatres as well as the central X-ray department.
### 6. Substantial Equivalence:
The Mobilett Mira is substantially equivalent to the following devices:
| Predicate Device<br>Name and<br>Manufacturer | 510(k)<br>Number | Clearance<br>Date | Comparable Properties |
|----------------------------------------------|------------------|-------------------|------------------------------------------------------------|
| Mobilett XP<br>Mobilett XP hybrid | K033238 | 11/14/2003 | • Indications for use<br>• X-ray Generator<br>• X-ray tube |
| Ysio | K081722 | 08/22/2008 | • Solid State detector<br>• Imaging system |
Siemens Medical Systems, Inc.
{2}------------------------------------------------
KII 912
SIEMENS
## 7. Summary of Technological Characteristics of the Subject Device as Compared with the Predicate Device:
The modified Mobilett XP will be marketed under the trade name Mobilett Mira. The modifications consist of a rotatable suspension arm, a solid state image detector, digital image acquisition system and a 35KW generator. These modifications are 510(k)cleared in the mentioned predicate devices. Several of the components used in the Mobilett Mira are either commercially available with current Siemens systems or include minor modifications to existing components.
#### 8. General Safety and Effectiveness Concerns:
Instructions for use are included within the device labeling, and the information provided will enable the user to operate the device in a safe and effective manner. Several safety features including visual and audible warnings are incorporated into the system design. In addition the Mobilett Mira is continually monitored, and if an error occurs, the system functions will be blocked and an error message will be displayed.
Furthermore the operators are health care professionals familiar with and responsible for the X-ray examinations to be performed. To minimize electrical, mechanical and radiation hazards, Siemens adheres to recognized and established industry practice, and all equipment is subject to final performance testing.
#### 9. Conclusion as to Substantial Equivalence:
The Mobilett Mira is intended for the same indications for use as the predicate Mobilett XP / XP hybrid. The solid state imaging detector replaces the conventional cassette with film/screen combination. The portfolio of accessories is the same as with the predicate Mobilett XP. It is Siemens opinion, that the Mobilett Mira is substantially equivalent to the Mobilett XP.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a seal for the Department of Health & Human Services USA. The seal is circular and contains the department's name around the perimeter. In the center of the seal is an abstract image of an eagle.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Ms. Patricia D. Jones Technical Specialist, Regulatory Submissions Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway, E-50 MALVERN PA 19355
Re: K111912
SEP 1 9 2011
Trade/Device Name: Mobilett Mira Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Code: IZL Dated: August 29, 2011 Received: August 30, 2011
Dear Ms. Jones:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR, Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket noutication. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Mary Patel
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use Form
510(k) Number (if known): _ K | | 19 | 2 __
Device Name: Mobilett Mira
Indications for Use:
The Mobilett Mira is a radiographic system designed for use in wards, intensive care and premature-birth wards, pediatric and emergency departments, operating theatres as well as the central X-ray department.
Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Jordant D. O'Hern
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K111912
Page 1 of
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.