I-SCAN
Device Facts
| Record ID | K960622 |
|---|---|
| Device Name | I-SCAN |
| Applicant | Ophthalmic Technologies, Inc. |
| Product Code | IYO · Radiology |
| Decision Date | May 13, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.1560 |
| Device Class | Class 2 |
Indications for Use
The OTI i+tech "i-scan" ultrasound system is intended to both visualize the interior of the eye by means of ultrasound and to make measurements inside the eye, including the measurement of axial length for determination of IOL power.
Device Story
PC-based ultrasonic diagnostic system; utilizes three transducers for A-scan, B-scan, and Biometry (X-mode) imaging/measurement. Operator (clinician) uses hand-held or tonometer-mounted probes to capture ocular ultrasound signals. System processes signals to visualize internal structures and perform axial length measurements. Features automatic or manual measurement modes; includes IOL power calculation algorithms. Used in clinical settings to assist ophthalmologists in diagnostic assessment and surgical planning. Output displayed on screen for clinical decision-making regarding ocular health and IOL selection.
Clinical Evidence
Clinical testing compared i-scan against a predicate device on ten eyes; results demonstrated equivalent accuracy. Bench testing included phantom studies for automatic and manual measurement modes, biocompatibility testing of patient-contact materials, and electrical/emission safety testing per Canadian Standards Association requirements.
Technological Characteristics
PC-based ultrasound system; three transducers (A-scan, B-scan, Biometry). Patient-contact materials tested for biocompatibility. Connectivity via PC interface. Safety testing includes electromagnetic radiation and transducer emissions.
Indications for Use
Indicated for ophthalmic patients requiring visualization of internal eye structures and biometric measurements, including axial length for IOL power determination.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
- Sonomed Model A2500
- Biophysic Ophthalmocan B
Related Devices
- K960765 — MENTOR ADVENT A/B SYSTEM · Mentor Ophthalmics, Inc. · Aug 8, 1996
- K092837 — OTI-SCAN 3000 · Opko Instrumentation, Inc. · Dec 23, 2009
- K021683 — CINESCAN S · Quantel Medical · Jun 21, 2002
- K030770 — OTI SCAN WITH TRANSDUCERS · Ophthalmic Technologies, Inc. · Mar 27, 2003
- K100252 — MASTER-VU A-SCAN, MV4500 · Sonomed, Inc. · Apr 9, 2010