K092837 · Opko Instrumentation, Inc. · IYO · Dec 23, 2009 · Radiology
Device Facts
Record ID
K092837
Device Name
OTI-SCAN 3000
Applicant
Opko Instrumentation, Inc.
Product Code
IYO · Radiology
Decision Date
Dec 23, 2009
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1560
Device Class
Class 2
Intended Use
The OTI-Scan 3000 ultrasound system is a multi-purpose personal-computer based ultrasonic diagnosis system for ophthalmic applications, intended to both visualize the interior of the eye by means of ultrasound and to make measurements inside the eye, including measurement of the axial length for determination of IOL power.
Device Story
OTI-Scan 3000 is a PC-based ultrasonic diagnostic system for ophthalmic use; utilizes pulsed echo ultrasound (sonar principles) to visualize eye interior and measure axial length. System operates in A-mode and B-mode. Used in clinical settings by eye care professionals. Input consists of ultrasonic echoes from ocular structures; system processes signals to generate visual images and biometric measurements. Output displayed on PC monitor for clinician review; assists in IOL power calculation and ocular diagnosis. Benefits include non-invasive assessment of ocular anatomy and precise biometric data for surgical planning.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included accuracy verification, ultrasonic emissions testing, electrical safety testing, and software validation.
Technological Characteristics
PC-based ultrasonic pulsed echo imaging system. Operates in A-mode and B-mode. Includes 20 MHz HF B-Probe transducer. Connectivity via PC interface. Software-controlled signal processing. Complies with 21 CFR 892.1560.
Indications for Use
Indicated for ophthalmic patients requiring visualization of the interior of the eye and measurement of axial length for IOL power determination.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
K183173 — LIGHTMED Ultrasound System · Lightmed USA, Inc. · Jul 1, 2019
K021683 — CINESCAN S · Quantel Medical · Jun 21, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
DEC 2 3 2009
OPKO Instrumentation OTI-Scan 3000 510(k) Summarv
| 510(k) Summary | |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company: | OPKO Instrumentation, LLC.<br>621 W 20th Street<br>Hialeah, FL 33010-2432<br>Phone: 888-268-6756 or 305-575-4178 |
| Contact: | Mario Arbesu<br>Director, Compliance and Regulatory Affairs<br>OPKO Instrumentation, LLC.<br>4400 Biscayne Boulevard<br>Miami, FL 33137<br>Phone 305.575.4631<br>Fax 305.575.4140 |
| Trade Name: | OTI-Scan 3000 |
| Device Type: | Ultrasonic Pulsed Echo Imaging System |
| Classification Regulation: | 892.1560 |
| Class: | II |
| Panel: | Radiology |
| Product Code: | IYO |
| Predicate Devices: | OTI I-Scan (K960622)<br>OTI Scan with Transducers (K030770)<br>OTI-Scan HF Module System (K031391) |
| Device Description: | The OTI-Scan 3000 is an Ultrasonic ophthalmic A-Scan<br>and B-Scan system that uses the principles of sonar<br>(pulsed ultrasound) to measure the axial length of the<br>eyes and to visualize the interior of the eye. |
| Indications for use: | The OTI-Scan 3000 ultrasound system is a multi-purpose<br>personal-computer based ultrasonic diagnosis system for<br>ophthalmic applications, intended to both visualize the<br>interior of the eye by means of ultrasound and to make<br>measurements inside the eye, including measurement of<br>the axial length for determination of IOL power. |
: 上一
14 .
for .
·
.
..............................................................................................................................................................................
.
{1}------------------------------------------------
Performance Data:
·
..
:
1. September 2017
: 上一篇: 上一篇:
. .
Bench-top testing has been performed on the OTI-Scan 3000 including accuracy tests, ultrasonic emissions teats, electrical safety tests and software validation tests.
:
.
. . . . .
.
.
.
.
이 대한민국 대학교 대학교 대학교 대학교
and the comments of the comments of
:
··
:
:
:
.
:
이 대한민국 대학교 대학교 대학교 대학교 대학교 대학교 대학 1982년 1992년 1992년 1992년 19
:
:
:
ાંકે
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image is a seal for the Department of Health & Human Services USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
Mr. Mario Arbesu Direction, Compliance and Regulatory Affairs OPKO Instrumentation, Inc. 4400 Biscayne Blvd. MIAMI FL 33137
DEC 2 3 2009
Re: K092837
Trade/Device Name: OTI-Scan 3000 Regulation Number: 21 CFR §892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: II Product Codes: IYO and ITX Dated: August 28, 2009 Received: November 30, 2009
Dear Mr. Arbesu
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the OTI-Scan 3000, as described in your premarket notification:
Transducer Model Number
OTI - Scan 20 MHz HF B-Probe
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Mario Arbesu
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
If you have any questions regarding the content of this letter, please contact Mr. Paul Hardy at (301) 796-6542.
Sincerely yours,
Joshua C Myppi
for
Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
{4}------------------------------------------------
### System: OTI-Scan 3000
Intended Use: The OTI-Scan 3000 ultrasound system is a multi-purpose personal-computer based Intended Use: The OTF-Scall 3000 unrasound system a misualize the interior of the eye ulfrasonic dignosis system for opilliamite applications, measurement of the axial length for determination of IOL power.
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | P | | | | | | P<br>(A-Mode) |
| Fetal Imaging &<br>Other | Fetal | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (Specify) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (Specify) | | | | | | | |
N = new indication; P = previous cleared by FDA; E = added under this appendix
* Examples of other modes of operation may include: A-mode, Amplitude Doppler, 3-D Imaging, Tissue Motion Doppler, and Color Velocity ాల
| Prescription Use | X |
|-----------------------------|---|
| (Part 21 CFR 801 Subpart D) | |
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|---------------------------------------------------------------|---------|
| (Division Sign-Off) | |
| Division of Reproductive, Abdominal, and Radiological Devices | |
| 510(k) Number | K092837 |
{5}------------------------------------------------
### System: OTI-Scan 20 MHz HF B-Probe
Intended Use: Transducer for B-Scans with the OTI-Scan 3000 Ultrasound Imaging System.
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | N | | | | | | |
| Fetal Imaging &<br>Other | Fetal | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (Specify) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (Specify) | | | | | | | |
N = new indication; P = previous cleared by FDA; E = added under this appendix
* Examples of other modes of operation may include. A-mode, Amplitude Doppler, 3-D Imaging, Tissue Motion Doppler, and Color Velocity
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
hormi m why
(Division Sign-O Division in Reproductive, Abdominal, and Kadsoingical Devices 510/k) Mumber
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.