The Volcano Angio-IVUS Mapping System is intended to provide a common frame of reference for intravascular ultrasound and angiographic images during diagnostic and/or interventional procedures for the coronary vasculature. The system provides indication of the location of the IVUS cross-sectional imaging plane for any given IVUS image as it relates to a two-dimensional (2D) angiographic image and/or the associated threedimensional (3D) model formulated from multiple 2D image projections.
Device Story
System integrates with In-Vision Gold Imaging console to process X-ray angiographic images; inputs include 2D angiographic images via DICOM network and IVUS pullbacks from In-Vision Gold System. Core function involves 3D reconstruction of coronary subtrees from two angiographic projections; maps IVUS pullback data onto 3D models and 2D angiograms. Used in catheterization labs by clinicians to improve IVUS orientation and interpretation. Output provides visual reference of IVUS cross-sectional plane location relative to angiographic anatomy; optionally integrates VH IVUS data. Benefits include enhanced anatomical visualization and guidance during coronary procedures.
Clinical Evidence
Bench testing only; no clinical data provided. Equivalence established through comparison of intended use and fundamental scientific technology to legally marketed predicate devices.
Technological Characteristics
Image acquisition and processing system; utilizes Ethernet connectivity for DICOM image transfer. Software-based 3D reconstruction from 2D angiographic projections. Compatible with In-Vision Gold Imaging console. Operates as an add-on to existing IVUS imaging systems.
Indications for Use
Indicated for patients undergoing diagnostic or interventional coronary vasculature procedures requiring spatial correlation between intravascular ultrasound (IVUS) and angiographic imaging.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
K110256 — THE IC-PRO SYSTEM · Paieon, Inc. · May 27, 2011
K160147 — CT TrueView · Philips Medical Systems Nederland B.V. · Aug 10, 2016
K253714 — IntraSight Plus · Philips Image Guided Therapy Corporation · Feb 24, 2026
K212490 — AVVIGO Guidance System II · Boston Scientific Corporation · Oct 20, 2021
K083745 — THE IC-PRO SYSTEM, MODEL VERSION 3.2 · Paieon, Inc. · Feb 17, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
K 060483
JUN 2 3 2006
Submitter: Volcano Corp. Volcano Angio-IVUS Mapping (AIM) System Traditional 510(k) Premarket Notification
# 510(k) SUMMARY
The 510(k) Summary is submitted as required by Section 807.92(a)
| Submitter Name: | Volcano Corp. |
|-----------------------------------------------|-------------------------------------------------------|
| Contact Person: | Michelle J. Badal, RAC<br>Manager, Regulatory Affairs |
| Address: | 2870 Kilgore Road<br>Rancho Cordova, CA 95670 |
| Phone Number: | 916-231-4497 |
| Fax Number: | 916-638-2647 |
| Date Prepared: | February 23, 2006 |
| Device Trade Name: | AIM - Angio-IVUS Mapping System |
| Device Common Name: | Intravascular Ultrasound Imaging System |
| Classification Name,<br>Number, Product Code: | Echocardiograph<br>21 CFR 870.2330, Product Code: DXK |
### Predicate Device:
The Volcano Angio-IVUS Mapping System is substantially equivalent to the Volcano In Vision System cleared under K031148 on May 28, 2003, and the Paieon CardiOp-B System cleared under K030139 on April 2, 2004.
### Device Description:
The Angio-IVUS Mapping System is an image acquisition and processing system on the In-Vision Gold Imaging console designed to process traditional X-ray angiographic images. A standard Ethernet cable links the In-Vision Gold Imaging System to the catheterization lab DICOM network, allowing the angiographic images to be transferred to the Angio-IVUS Mapping System. The system can be used to make a 3D reconstruction of a coronary subtree from two angiogram images taken at different angles. IVUS pullbacks acquired with the In-Vision Gold System can be mapped to the 3D reconstruction and the original 2D angiograms, resulting in a better tool for IVUS orientation and interpretation. In addition, Volcano's VH IVUS can (optionally) be performed, combining the detailed information of the arterial wall with the familiar angiographic roadmap.
{1}------------------------------------------------
# Intended Use:
The Volcano Angio-IVUS Mapping System is intended to provide a common frame of reference for intravascular ultrasound and angiographic images during diagnostic and/or interventional procedures for the coronary vasculature. The system provides indication of the location of the IVUS cross-sectional imaging plane for any given IVUS image as it relates to a two-dimensional (2D) angiographic image and/or the associated threedimensional (3D) model formulated from multiple 2D image projections.
# Performance Data:
The information provided in the premarket notification demonstrates that the Angio-IVUS Mapping System is substantially equivalent to the predicate devices, for which there is FDA clearance. This equivalence was demonstrated through comparison of intended use and fundamental scientific technology to a commercially available device. The information supplied in this premarket notification provides reasonable assurance that the Angio-IVUS Mapping System is safe and effective for the stated intended use.
### Conclusion:
The Volcano Angio-IVUS Mapping System has the same indication for use, and utilizes the same fundamental scientific technology as that of the predicate devices. The information provided in this premarket notification submission, along with the Declaration of Conformity to design controls supports a determination of substantial equivalence of the Volcano Angio-IVUS Mapping System to the predicate devices.
KOL 0483
page 2 of 2
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image contains the words "Public Health Service" in a bold, sans-serif font. The text is arranged vertically, with "Public Health" on the first line and "Service" on the second line. The words are black against a white background.
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" around the perimeter. In the center of the circle are three stylized human figures, depicted in a simple, abstract manner.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 2 3 2006
Volcano Corporation Ms. Michelle Badal Manager, Regulatory Affairs 2870 Kilgore Road Rancho Cordova CA 95670
Re: K060483
Trade/Device Name: Angio-IVUS Mapping System Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: Class II Product Code: IYO, IZI Dated: May 19, 2006 Received: May 22, 2006
Dear Ms. Badal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce. prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Dan); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the cuality
{3}------------------------------------------------
Page 2 - Ms. Michelle Badal
systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
BBlommansta
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K060483
Device Name:
Volcano Angio-IVUS Mapping System
Indications for Use:
The Volcano Angio-IVUS Mapping System is intended to provide a common frame of reference for intravascular ultrasound and angiographic images during diagnostic and/or interventional procedures for the coronary vasculature. The system provides indication of the location of the IVUS cross-sectional imaging plane for any given IVUS image as it relates to a two-dimensional (2D) angiographic image and/or the associated three-dimensional (3D) model formulated from multiple 2D image projections.
Prescription Use メ Over-The-Counter Use AND/OR (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
B.L. Sumner
(Posted November 13, 2003)
(División Sign-Off)
Division of Cardiovascular Devices
510(k) Number K060483
Page! 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.