IntraSight Plus is a multi-function system intended to be used as an adjunct diagnostic tool for qualitative and quantitative evaluation of vascular morphology in coronary arteries and peripheral vasculature in patients eligible for endovascular procedures. The system, when configured with optional features, can also be used for: - quantitative evaluation of coronary physiology using pressure-derived ratios, - quantitative evaluation of coronary artery dimensions, - enhanced visualization of stent or balloon placement in coronary arteries, and - co-registration of an x-ray image with coronary intravascular ultrasound and/or coronary pressure-derived ratios in patients eligible for percutaneous coronary intervention. The IntraSight Plus IVUS function is indicated for use in patients undergoing conventional angiographic procedure to assess vascular morphology and to assess the need for additional treatment post percutaneous intervention. IVUS is obtained with a compatible ultrasound catheter and used as an adjunct to examine vascular morphology in the coronary arteries and vessels of the peripheral vasculature. The IntraSight Plus FFR and iFR functions are indicated for use in patients with signs of coronary artery disease undergoing coronary catheterization procedures to assess the hemodynamic significance of coronary lesions that may contribute to myocardial ischemia. FFR and iFR are obtained with a compatible pressure guide wire.
Device Story
Multi-modality platform for imaging, physiology, and co-registration; integrates with interventional X-ray systems. Inputs: ultrasound signals from compatible IVUS catheters; pressure data from compatible guide wires; X-ray images. Transforms inputs via software applications (IVUS, FFR, iFR, co-registration, QCA, VE, DD) to provide qualitative/quantitative vascular morphology and physiological assessments. Used in cardiac catheterization labs by physicians, nurses, and technologists. Output: visual displays of vessel lumen/wall structures, pressure-derived ratios, and co-registered images. Assists clinicians in assessing vascular morphology, stent/balloon placement, and hemodynamic significance of lesions; supports clinical decision-making regarding treatment post-intervention. Benefits: improved visualization and diagnostic accuracy for endovascular procedures.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical testing, including software, cybersecurity, interoperability, electromagnetic compatibility, electrical/mechanical/thermal safety, usability, and packaging validation.
Technological Characteristics
Multi-modality platform; integrates with interventional X-ray. Components: Bedside Utility Box (BUB), Patient Interface Modules (PIMs), touchscreen module. Supports IVUS catheters (Eagle Eye, Visions PV, Refinity, etc.) and pressure guide wires (OmniWire). Connectivity: Networked/integrated with X-ray systems. Software-based applications for IVUS, FFR, iFR, co-registration, QCA, VE, and DD. Complies with standards for software, cybersecurity, and electrical/mechanical safety.
Indications for Use
Indicated for adult patients eligible for endovascular procedures or percutaneous coronary intervention, including those with signs of coronary artery disease undergoing coronary catheterization to assess vascular morphology and hemodynamic significance of coronary lesions.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
February 24, 2026
Philips Image Guided Therapy Corporation
Jeff Rongero
Principal Regulatory Affairs Specialist
9965 Federal Dr.
Colorado Springs, Colorado 80921
Re: K253714
Trade/Device Name: IntraSight Plus
Regulation Number: 21 CFR 892.1560
Regulation Name: Ultrasonic pulsed echo imaging system
Regulatory Class: Class II
Product Code: IYO, DSK, OWB, DSA
Dated: November 21, 2025
Received: November 24, 2025
Dear Jeff Rongero:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253714 - Jeff Rongero
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K253714 - Jeff Rongero
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Aneesh S. Deoras -S
Aneesh Deoras
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253714 | ? |
| Please provide the device trade name(s). | | ? |
| IntraSight Plus | | |
| Please provide your Indications for Use below. | | ? |
| IntraSight Plus is a multi-function system intended to be used as an adjunct diagnostic tool for qualitative and quantitative evaluation of vascular morphology in coronary arteries and peripheral vasculature in patients eligible for endovascular procedures. The system, when configured with optional features, can also be used for: - quantitative evaluation of coronary physiology using pressure-derived ratios, - quantitative evaluation of coronary artery dimensions, - enhanced visualization of stent or balloon placement in coronary arteries, and - co-registration of an x-ray image with coronary intravascular ultrasound and/or coronary pressure-derived ratios in patients eligible for percutaneous coronary intervention. The IntraSight Plus IVUS function is indicated for use in patients undergoing conventional angiographic procedure to assess vascular morphology and to assess the need for additional treatment post percutaneous intervention. IVUS is obtained with a compatible ultrasound catheter and used as an adjunct to examine vascular morphology in the coronary arteries and vessels of the peripheral vasculature. The IntraSight Plus FFR and iFR functions are indicated for use in patients with signs of coronary artery disease undergoing coronary catheterization procedures to assess the hemodynamic significance of coronary lesions that may contribute to myocardial ischemia. FFR and iFR are obtained with a compatible pressure guide wire. Please select the types of uses (select one or both, as applicable). | | |
| | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
PHILIPS
IntraSight Plus
Traditional 510(k)
Administrative Documentation
014_05: 510(k) Summary
As Required by 21 CFR 807.92:
| Submitter: | Philips Image Guided Therapy Corporation |
| --- | --- |
| Address: | 3721 Valley Centre Drive
Suite 500
San Diego, CA 92130
United States |
| Telephone Number: | +1 (617) 245-5416 |
| Contact Person: | Jeff Rongero |
| Date Prepared: | February 12, 2026 |
| --- | --- |
| Device Name: | IntraSight Plus | | | | |
| --- | --- | --- | --- | --- | --- |
| Trade Name: | IntraSight Plus | | | | |
| Common Name: | System, Imaging, Pulsed Echo, Ultrasonic | | | | |
| Classification Name: | Panel | Cardiovascular | | Radiology | |
| | 21 CFR Regulation Number | 870.1110 | 870.2900 | 892.1560 | 892.1650 |
| | 21 CFR Regulation Description | Blood pressure computer | Patient transducer and electrode cable (including connector) | Ultrasonic pulsed echo imaging system | Interventional fluoroscopic x-ray system |
| | Class | II | II | II | II |
| | Product Code | DSK | DSA | IYO | OWB |
| Predicate Device(s): | 510(k) No. | Device Name | Product Code(s) |
| --- | --- | --- | --- |
| | K190078 | IntraSight System | IYO, DSA, DSK |
| | K190626 | SyncVision System | OWB |
## Device Description:
Philips IntraSight Plus is a multi-modality, application-based platform providing a range of imaging, physiology and co-registration tools. It can be integrated with a compatible interventional X-ray system.
Philips IntraSight Plus provides qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vessels of the peripheral vasculature in adult patients eligible for endovascular procedures. It is used to support conventional angiographic procedures, providing images of vessel lumen and wall structures.
Philips Image Guided Therapy Corporation
K253714 - Page 1 of 4
{5}
PHILIPS
IntraSight Plus
Traditional 510(k)
Administrative Documentation
# Intended Use and Indications for Use:
IntraSight Plus is a multi-function system intended to be used as an adjunct diagnostic tool for qualitative and quantitative evaluation of vascular morphology in coronary arteries and peripheral vasculature in patients eligible for endovascular procedures.
The system, when configured with optional features, can also be used for:
- quantitative evaluation of coronary physiology using pressure-derived ratios,
- quantitative evaluation of coronary artery dimensions,
- enhanced visualization of stent or balloon placement in coronary arteries, and
- co-registration of an x-ray image with coronary intravascular ultrasound and/or coronary pressure-derived ratios
in patients eligible for percutaneous coronary intervention.
The IntraSight Plus IVUS function is indicated for use in patients undergoing conventional angiographic procedure to assess vascular morphology and to assess the need for additional treatment post percutaneous intervention. IVUS is obtained with a compatible ultrasound catheter and used as an adjunct to examine vascular morphology in the coronary arteries and vessels of the peripheral vasculature.
The IntraSight Plus FFR and iFR functions are indicated for use in patients with signs of coronary artery disease undergoing coronary catheterization procedures to assess the hemodynamic significance of coronary lesions that may contribute to myocardial ischemia. FFR and iFR are obtained with a compatible pressure guide wire.
| Comparison of Technological Characteristics: | | | |
| --- | --- | --- | --- |
| Aspect | IntraSight Plus (New Device) | IntraSight (K190078) SyncVision (K190626) | Comparison |
| Intended Users | • Physician
• Nurse
• Technologist | • Physician
• Nurse
• Technologist | Same |
| Intended Environment of Use | • Indoor
• Healthcare Facilities
• Cardiac Catheterization Laboratory (Sterile Procedure Room and Non-Sterile Control Room) | • Indoor
• Healthcare Facilities
• Cardiac Catheterization Laboratory (Sterile Examination Room and Non-Sterile Control Room) | Same |
| Total Workstations (Control Room) | • 1 tower | • 2 towers (1 each for IntraSight and SyncVision) | Different |
| Total Displays (Control Room) | • 1 display | • 2 displays (1 each for IntraSight and SyncVision) | Different |
| Total Keyboards and Mice (Control Room) | • 1 keyboard
• 1 mouse | • 2 keyboards (1 each for IntraSight and SyncVision)
• 2 mice (1 each for | Different |
Philips Image Guided Therapy Corporation
K253714 - Page 2 of 4
{6}
PHILIPS
IntraSight Plus
Traditional 510(k)
Administrative Documentation
| Bedside Control Options (Examination / Procedure Room) | • 1 Touchscreen Module | • 1 Touchscreen Module
• 1 Joystick | Different |
| --- | --- | --- | --- |
| Bedside Utility Box (BUB) (Examination / Procedure Room) | • Connects devices in the examination (procedure) room to the workstation in the control room | • Connects devices in the examination (procedure) room to the workstation in the control room | Same |
| Patient Interface Modules (PIMs) (Examination / Procedure Room) | • Digital IVUS PIM (also referred to SA-PIM)
• Rotational IVUS PIM (also referred to as SpinVision, PIMr)
• Functional Measurement (FM) PIM (FM-PIM) | • Digital IVUS PIM (also referred to SA-PIM)
• Rotational IVUS PIM (also referred to as SpinVision, PIMr)
• Functional Measurement (FM) PIM (FM-PIM) | Same |
| IVUS Catheters (Examination / Procedure Room) | • Eagle Eye Platinum
• Eagle Eye Platinum ST (Short Tip)
• Visions PV .014P
• Visions PV .014P RX
• Visions PV .018
• Visions PV .035
• Visions PV .035 Non-Hospital
• Reconnaissance PV .018 OTW
• Pioneer Plus
• Refinity ST (Short Tip) | • Eagle Eye Platinum
• Eagle Eye Platinum ST (Short Tip)
• Visions PV .014P
• Visions PV .014P RX
• Visions PV .018
• Visions PV .035
• Visions PV .035 Non-Hospital
• Reconnaissance PV .018 OTW
• Pioneer Plus
• Refinity ST (Short Tip)
• Refinity
• Revolution | Different |
| Pressure Guide Wires (Examination / Procedure Room) | OmniWire | • Omniwire
• Verrata PLUS
• Verrata | Different |
| Software Applications and Functionality | • IVUS
• Functional Measurements (FFR, iFR)
• Co-Registration / Tri-Registration
• QCA
• VE
• DD | • IVUS
• Functional Measurements (FFR, iFR)
• Co-Registration / Tri-Registration
• QCA
• VE
• DD | Same |
## Non-Clinical Testing:
The results of non-clinical testing were submitted, referenced, and relied upon to demonstrate the safety and effectiveness of IntraSight Plus and its substantial equivalence to the predicate devices. IntraSight Plus complies with and accounts for FDA recognized consensus standards addressing the
Philips Image Guided Therapy Corporation
K253714 - Page 3 of 4
{7}
PHILIPS
IntraSight Plus
Traditional 510(k)
Administrative Documentation
following:
- Software
- Cybersecurity
- Interoperability
- Electromagnetic compatibility
- Electrical, mechanical, and thermal safety
- Usability
- Packaging validation
Clinical Testing:
Not Applicable – IntraSight Plus has the same intended use and indications for use as the predicate devices. Differences in technological characteristics compared to the predicate devices do not raise new or different questions of safety or effectiveness. Therefore, clinical testing was not deemed necessary for IntraSight Plus to demonstrate its safety and effectiveness.
Conclusion:
IntraSight Plus has the same intended uses, indications for use, and is intended for the same users within the same intended environments of use as the predicate devices. It meets design verification and validation requirements, including non-clinical testing for software, cybersecurity, interoperability, electromagnetic compatibility, electrical safety, mechanical safety, thermal safety, usability, and packaging.
Differences in technological characteristics primarily reflect the reduction in hardware IntraSight Plus represents due to its merging of predicate device hardware and rendering some of it redundant. The differences do not introduce new intended uses or new indications for use and do not raise new or different questions of safety or effectiveness. IntraSight Plus addresses the differences through non-clinical testing and compliance with appropriate standards, the results of which demonstrate its safety and effectiveness.
Therefore, IntraSight Plus is substantially equivalent to the predicate device IntraSight System (K190078) and predicate device SyncVision System (K190626) with regard to intended uses, indications for use, and technological characteristics.
Philips Image Guided Therapy Corporation
K253714 - Page 4 of 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.