HS30 Diagnostic Ultrasound System

K202556 · Samsung Medison Co., Ltd. · IYN · Oct 30, 2020 · Radiology

Device Facts

Record IDK202556
Device NameHS30 Diagnostic Ultrasound System
ApplicantSamsung Medison Co., Ltd.
Product CodeIYN · Radiology
Decision DateOct 30, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1550
Device ClassClass 2
AttributesPediatric

Intended Use

The diagnostic ultrasound system and probes are designed to obtain ultrasound images and analyze body fluids. The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel. It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, clinics and similar care environment for clinical diagnosis of patients.

Device Story

HS30 is a mobile, software-controlled diagnostic ultrasound system. It acquires ultrasound data via various transducers (linear, curved, endocavity, phased array) and displays images in multiple modes: 2D, M, Color/Power/PW/CW Doppler, Tissue Doppler, and 3D/4D. Used in hospitals and clinics by trained healthcare professionals for clinical diagnosis. The system provides real-time acoustic output (MI/TI) and includes analysis packages for anatomical measurements. Healthcare providers use these images and measurements to inform clinical decision-making. Benefits include non-invasive diagnostic imaging and fluid analysis.

Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including acoustic output, biocompatibility, cleaning/disinfection effectiveness, and thermal, electromagnetic, and mechanical safety testing in accordance with recognized standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-1, NEMA UD 2).

Technological Characteristics

Mobile cart-based ultrasound system with LED monitor. Supports linear, curved, endocavity, and phased array transducers (1.0-20.0 MHz). Modes: 2D, M, Doppler (Color, Power, PW, CW, Tissue), 3D/4D. Connectivity: DICOM. Safety: IEC 60601-1, IEC 60601-2-37, IEC 60601-1-2. Software-controlled with automated analysis features (Auto IMT+, Needle Mate+, Strain+, etc.).

Indications for Use

Indicated for fetal/obstetrics, abdominal, gynecology, pediatric, small organ, neonatal/adult cephalic, trans-rectal, trans-vaginal, musculoskeletal (conventional/superficial), urology, cardiac (adult/pediatric), and peripheral vessel imaging in patients requiring diagnostic ultrasound.

Regulatory Classification

Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. October 30, 2020 Samsung Medison Co., LTD % Ji Yea Lee Regulatory Affairs Specialist 3366, Hanseo-ro, Nam-myeon, Hongcheon-gun, Gangwon-do 25108 REPUBLIC OF KOREA Re: K202556 Trade/Device Name: HS30 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: August 31, 2020 Received: September 3, 2020 Dear Ji Yea Lee: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K202556 Device Name HS30 Diagnostic Ultrasound System ### Indications for Use (Describe) The diagnostic ultrasound system and probes are designed to obtain ultrasound images and analyze body fluids. The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel. It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, clinics and similar care environment for clinical diagnosis of patients. Modes of Operation: 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, Combined modes, Multi-Image mode(Dual, Quad), 3D/4D mode. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ 510(k) Premarket Notification - Traditional ## 5. 510(K) Summarv: K202556 In accordance with 21 CFR 807.92 the following summary of information is provided: - Date Prepared August 31, 2020 1. - 2. Manufacturer SAMSUNG MEDISON CO., LTD. 3366, Hanseo-ro, Nam-myeon, Hongcheon-gun, Gangwon-do 25108, REPUBLIC OF KOREA - 3. Primary Contact Person Ji Yea Lee Requlatory Affairs Specialist Phone: +82.2.2.2194.1594 Fax: +82.31.8017.9573 Email: jiyea722.lee@samsungmedison.com - 4. Secondary Contact Person Ninad Gujar Vice President Phone: +1.978.564.8632 Fax: +1.978.564.8677 Email: ngujar@neurologica.com - Proposed Device 5. - Common/Usual Name: Diagnostic Ultrasound System and Accessories - Proprietary Name: HS30 Diagnostic Ultrasound System - - Common Name: Diagnostic Ultrasound System । - Classification Names: system, imaqing, pulsed doppler, ultrasonic । - । Product Code: IYN, IYO, ITX - -Regulation: 21 CFR 892.1550, 892.1560, 892.1570 ### 6. Predicates - HS30 Diagnostic Ultrasound System (K182632) Primary - - HS40 Diagnostic Ultrasound System (K200339) - - 7. Device Description The HS30 is a general purpose, mobile, software controlled, diagnostic ultrasound system, Its function is to acquire ultrasound data and to display the data as 2D mode, M mode, Color Doppler imaging, Power Doppler imaging (including Directional Power Doppler mode; S-Flow), PW Spectral Doppler mode, CW Spectral Doppler mode, Harmonic imaging(S-Harmonic), Tissue Doppler imaging, Tissue Doppler Wave, Panoramic Imaging, Freehand 3D, Elastoscan Mode or as a combination of these modes. The HS30 also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The HS30 has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed. - 8. Intended Use The diagnostic ultrasound system and probes are designed to obtain ultrasound images {4}------------------------------------------------ 510(k) Premarket Notification – Traditional and analyze body fluids. The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel. It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, private practices, clinics and similar care environment for clinical diagnosis of patients. Modes of Operation: 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, Combined modes, Multi-Image mode(Dual, Quad), 3D/4D mode. - 9. Technology The HS30 employs the same fundamental scientific technology as the predicates. - 10. Determination of Substantial Equivalence The HS30 is substantially equivalent to the predicate devices with regard to intended use, imaging capabilities, technological characteristics and safety and effectiveness. | Feature | HS30<br>(Under Review) | HS30<br>(K182632)<br>Primary | HS40<br>(K200339) | |---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indication for<br>Use | - Fetal/Obstetrics<br>- Abdominal<br>- Gynecology<br>- Pediatric<br>- Small Organ<br>- Neonatal Cephalic<br>- Adult Cephalic<br>- Trans-rectal<br>- Trans-vaginal<br>- Musculo-skel.<br>(Conventional)<br>- Musculo-skel.<br>(Superficial)<br>- Urology<br>- Cardiac Adult<br>- Cardiac Pediatric<br>- Peripheral vessel | - Fetal/Obstetrics<br>- Abdominal<br>- Gynecology<br>- Pediatric<br>- Small Organ<br>- Neonatal Cephalic<br>- Adult Cephalic<br>- Trans-rectal<br>- Trans-vaginal<br>- Musculo-skel.<br>(Conventional)<br>- Musculo-skel.<br>(Superficial)<br>- Urology<br>- Cardiac Adult<br>- Cardiac Pediatric<br>- Peripheral vessel | - Fetal/Obstetrics<br>- Abdominal<br>- Gynecology<br>- Pediatric<br>- Small Organ<br>- Neonatal Cephalic<br>- Adult Cephalic<br>- Trans-rectal<br>- Trans-vaginal<br>- Musculo-skel.<br>(Conventional)<br>- Musculo-skel.<br>(Superficial)<br>- Urology<br>- Cardiac Adult<br>- Cardiac Pediatric<br>- Peripheral vessel | | Scanhead<br>Types: | - Linear Array<br>- Curved Linear Array<br>- Endocavity<br>- Phased Array<br>- Static Probes | - Linear Array<br>- Curved Linear Array<br>- Endocavity<br>- Phased Array<br>- Static Probes | - Linear Array<br>- Curved Linear Array<br>- Endocavity<br>- Phased Array<br>- Static Probes | | Scanhead Frequency | 1.0 ~ 20.0 MHz | 1.0 ~ 20.0 MHz | 1.0 ~ 20.0 MHz | | Feature | HS30<br>(Under Review) | HS30<br>(K182632)<br>Primary | HS40<br>(K200339) | | Acoustic<br>Output Display<br>& FDA Limits: | - Display Feature for<br>Higher Output-Track3<br>- MI Output Display<br>- TI Output Display | - Display Feature for<br>Higher Output-Track3<br>- MI Output Display<br>- TI Output Display | - Display Feature for<br>Higher Output-Track3<br>- MI Output Display<br>- TI Output Display | | Modes of<br>Operation: | - 2D Mode<br>- Color Doppler Mode<br>- Power Doppler (PD)<br>mode<br>- M mode<br>- Pulsed Wave (PW)<br>Doppler mode<br>- Continuous Wave<br>(CW) Doppler mode<br>- Tissue Doppler<br>Imaging (TDI) mode<br>- Tissue Doppler Wave<br>(TDW) mode<br>- ElastoScan Mode<br>- Combined modes<br>- Multi-Image mode<br>(Dual, Quad)<br>- 3D/4D mode | - 2D Mode<br>- Color Doppler Mode<br>- Power Doppler (PD)<br>mode<br>- M mode<br>- Pulsed Wave (PW)<br>Doppler mode<br>- Continuous Wave<br>(CW) Doppler mode<br>- Tissue Doppler<br>Imaging (TDI) mode<br>- Tissue Doppler Wave<br>(TDW) mode<br>- ElastoScan Mode<br>- Combined modes<br>- Multi-Image mode<br>(Dual, Quad) | - 2D Mode<br>- Color Doppler Mode<br>- Power Doppler (PD)<br>mode<br>- M mode<br>- Pulsed Wave (PW)<br>Doppler mode<br>- Continuous Wave<br>(CW) Doppler mode<br>- Tissue Doppler<br>Imaging (TDI) mode<br>- Tissue Doppler Wave<br>(TDW) mode<br>- ElastoScan Mode<br>- Combined modes<br>- Multi-Image mode<br>(Dual, Quad)<br>- 3D/4D mode | | 510(k) Track | Track 3 | Track 3 | Track 3 | | System<br>Characteristics: | - Mobile cart<br>- LED Monitor<br>- 256 gray shades on<br>monitor<br>- 100-120V, 60 Hz;<br>- 200-240V, 50 Hz | - Mobile cart<br>- LED Monitor<br>- 256 gray shades on<br>monitor<br>- 100-120V, 60 Hz;<br>- 200-240V, 50 Hz | - Mobile cart<br>- LED Monitor<br>- 256 gray shades on<br>monitor<br>- 100-120V, 60 Hz;<br>- 200-240V, 50 Hz | | Product Safety<br>Certification | - IEC 60601-1<br>- IEC60601-2-37 | - IEC 60601-1<br>- IEC60601-2-37 | - IEC 60601-1<br>- IEC60601-2-37 | | EMC<br>Compliance | IEC60601-1-2 | IEC60601-1-2 | IEC60601-1-2 | | Functionality | - DICOM<br>- Quick Scan (QScan)<br>- ClearVision<br>- MultiVision<br>- Auto IMT+<br>- Elastoscan<br>- Panoramic<br>- Needle Mate+<br>- Strain +<br>- EzExam+<br>- 3D Imaging<br>(Freehand 3D)<br>- 3D Rendering (MPR) | - DICOM<br>- Quick Scan (QScan)<br>- ClearVision<br>- MultiVision<br>- Auto IMT+<br>- Elastoscan<br>- Panoramic<br>- Needle Mate+<br>- Strain +<br>- EzExam+<br>- 3D Imaging<br>(Freehand 3D)<br>- 3D Rendering (MPR) | - DICOM<br>- Quick Scan (QScan)<br>- ClearVision<br>- MultiVision<br>- Auto IMT+<br>- Elastoscan<br>- Panoramic<br>- Needle Mate+<br>- Strain +<br>- EzExam+<br>- 3D Imaging<br>(Freehand 3D)<br>- 3D Rendering (MPR) | | Feature | HS30<br>(Under Review) | HS30<br>(K182632)<br>Primary | HS40<br>(K200339) | | | 3D MagiCut<br>3D XI(MSV, Oblique<br>View, XI VOCAL)<br>XI STIC<br>2D NT<br>BiometryAssist<br>LaborAssist | 3D MagiCut | 3D MagiCut<br>3D XI (MSV, Oblique<br>View, XI VOCAL)<br>XI STIC<br>2D NT<br>BiometryAssist<br>LaborAssist | | Transducers | [Linear array] | [Linear array] | [Linear array] | | | -LN5-12<br>-L5-12/50 | -LN5-12<br>-L5-12/50 | -LA3-16AD<br>-LN5-12<br>-L5-12/50<br>-L4-7<br>-LS6-15 | | | [Curved array]<br>-CF4-9<br>-C2-8<br>-C2-5<br>-CA2-6BM | [Curved array]<br>-CF4-9<br>-C2-8<br>-C2-5 | [Curved array]<br>CA2-8AD<br>-CF4-9<br>-C2-8<br>-C2-5<br>-CA2-6BM | | | [Endo Cavity]<br>-EVN4-9<br>-ER4-9 | [Endo Cavity]<br>-EVN4-9<br>-ER4-9 | [Endo Cavity]<br>-EVN4-9<br>-ER4-9 | | | [Phased array]<br>-PN2-4<br>-SP3-8 | [Phased array]<br>PN2-4 | [Phased array]<br>-PN2-4<br>-SP3-8 | | | [3D]<br>-VN4-8<br>-EV2-10A | | [3D]<br>-VN4-8<br>-V5-9<br>-EV2-10A | | | [CW]<br>DP2B | [CW]<br>DP2B | [CW]<br>DP2B<br>-DP8B | | Biopsy Guides | BP-KIT-024 | BP-KIT-024 | - BP-KIT-024<br>- BP-KIT-029 | | | - BP-KIT-035 [BP-KIT-<br>035-NG]<br>- BP-KIT-040 [BP-KIT-<br>040-NG]<br>- BP-KIT-045 [BP-KIT-<br>045-NG]<br>- BP-KIT-047 [BP-KIT-<br>047-NG]<br>- BP-KIT-049 [BP-KIT-<br>049-NG]<br>- BP-KIT-061<br>- BP-KIT-085<br>- BP-KIT-086 | - BP-KIT-035 [BP-KIT-<br>035-NG]<br>- BP-KIT-040 [BP-KIT-<br>040-NG]<br>- BP-KIT-045 [BP-KIT-<br>045-NG]<br>- BP-KIT-047 [BP-KIT-<br>047-NG]<br>- BP-KIT-061 | - BP-KIT-054 [BP-KIT-<br>054-NG]<br>- BP-KIT-055 [BP-KIT-<br>055-NG]<br>- BP-KIT-049 [BP-KIT-<br>049-NG]<br>- BP-KIT-060<br>- BP-KIT-061<br>- BP-KIT-068 [BP-KIT-<br>068-NG]<br>- BP-KIT-043<br>- BP-KIT-085<br>- BP-KIT-086 | | Feature | HS30<br>(Under Review) | HS30<br>(K182632)<br>Primary | HS40<br>(K200339) | | optional<br>devices | - Digital B/W Video<br>Printer<br>- Digital Color Video<br>Printer<br>- USB Printer<br>DVD recorder (DVR) | - Digital B/W Video<br>Printer<br>- Digital Color Video<br>Printer<br>- USB Printer<br>DVD recorder (DVR) | - Digital B/W Video<br>Printer<br>- Digital Color Video<br>Printer<br>- USB Printer<br>DVD recorder (DVR) | | Accessories | - Foot Switch<br>- ECG<br>Gel Warmer | - Foot Switch<br>- ECG<br>Gel Warmer | - Foot Switch<br>- ECG<br>Gel Warmer | {5}------------------------------------------------ # SAMSUNG 510(k) Premarket Notification – Traditional {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the word "SAMSUNG" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, which makes the blue letters stand out. 510(k) Premarket Notification – Traditional {7}------------------------------------------------ Image /page/7/Picture/1 description: The image shows the word "SAMSUNG" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, which makes the blue letters stand out. The word is horizontally oriented and centered. 510(k) Premarket Notification - Traditional ## 11. Summary of Non-Clinical Test The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electromagnetic and mechanical safety, and has been found to conform with applicable medical device safety standards. The HS30 and its applications comply with voluntary standards. | Reference No. | Title | |--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | IEC 60601-1 | ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012,<br>C1:2009/(R)2012 and A2:2010 /(R)2012<br>Medical Electrical Equipment - Part 1: General Requirements for<br>basic safety and essential performance. | | IEC 60601-1-2 | IEC60601-1-2: 2014(4th Edition), Medical electrical equipment - Part<br>1-2: General requirements for basic safety and essential<br>performance - EMC | | IEC 60601-2-37 | IEC60601-2-37:2007 + A1:2015, Particular requirements for the<br>safety of ultrasonic medical diagnostic and monitoring equipment | | ISO10993-1 | ISO 10993-1, Biological evaluation of medical devices -- Part 1:<br>Evaluation and testing within a risk management process. | | ISO14971 | ISO 14971:2007, Medical devices - Application of risk management<br>to medical devices | | NEMA UD 2-<br>2004 | NEMA UD 2-2004 (R2009)<br>Acoustic Output Measurement Standard for Diagnostic Ultrasound<br>Equipment Revision 3 | - 12. Summary of Clinical Tests The subject of this premarket submission, HS30 is not required clinical studies to support substantial equivalence. - 13. Conclusion Intended uses and other key features are consistent with traditional clinical practices and FDA guidelines. The design, development and quality process of the manufacturer confirms with 21 CFR 820 and ISO 13485. The device is designed to conform to applicable medical device safety standards and compliance. Therefore, SAMSUNG MEDISON CO., LTD. considers the HS30 to be as safe, as effective, and performance is substantially equivalent to the predicate devices. ### END of 510(K) Summary i
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...