HS30 Diagnostic Ultrasound System
K182632 · Samsung Medison Co., Ltd. · IYN · Jan 8, 2019 · Radiology
Device Facts
| Record ID | K182632 |
| Device Name | HS30 Diagnostic Ultrasound System |
| Applicant | Samsung Medison Co., Ltd. |
| Product Code | IYN · Radiology |
| Decision Date | Jan 8, 2019 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.1550 |
| Device Class | Class 2 |
| Attributes | Pediatric |
Intended Use
The HS30 Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body. The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel.
Device Story
HS30 is a mobile, software-controlled diagnostic ultrasound system; acquires ultrasound data via various linear, convex, phased, static, and volume probes (1.0–20.0 MHz). Transforms inputs into 2D, M-mode, Color/Power Doppler, PW/CW Spectral Doppler, Harmonic, Tissue Doppler, Panoramic, Freehand 3D, and Elastoscan images. Used in clinical settings by healthcare professionals for anatomical measurement and diagnostic analysis. Provides real-time acoustic output display (mechanical/thermal indices). Output aids clinicians in diagnosis and clinical decision-making; benefits patients through non-invasive diagnostic imaging.
Clinical Evidence
No clinical data; substantial equivalence demonstrated through bench testing, including acoustic output, thermal, electrical, electromagnetic, and mechanical safety evaluations, and software validation.
Technological Characteristics
Mobile cart-based ultrasound system; 128-channel beamformer; 64 transmit/receive channels. Supports linear, convex, phased, static, and volume probes (1.0–20.0 MHz). Features include B-mode, M-mode, Color/Power Doppler, PW/CW Doppler, Tissue Harmonic Imaging, 3D imaging, and Elastoscan. Connectivity via DICOM. Electrical safety per ANSI AAMI ES60601-1; EMC per IEC60601-1-2; Ultrasound safety per IEC60601-2-37 and NEMA UD 2/UD 3.
Indications for Use
Indicated for diagnostic ultrasound imaging and fluid analysis in fetal/obstetrics, abdominal, gynecology, pediatric, small organ, neonatal/adult cephalic, trans-rectal, trans-vaginal, musculoskeletal, urology, cardiac, and peripheral vessel applications.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
- HS40 Diagnostic Ultrasound System (K180409)
Related Devices
- K202556 — HS30 Diagnostic Ultrasound System · Samsung Medison Co., Ltd. · Oct 30, 2020
- K170493 — HS50 / HS60 Diagnostic Ultrasound System · Samsung Medison Co., Ltd. · Mar 28, 2017
- K210426 — HS40 Diagnostic Ultrasound System · Samsung Medison Co., Ltd. · May 13, 2021
- K152962 — HS70A Diagnostic Ultrasound System · Samsung Medison Co., Ltd. · Nov 2, 2015
- K112646 — SONOACE R7 DIAGNOSTIC ULTRASOUND SYSTEM · Samsung Medison Co., Ltd. · Oct 12, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.
Samsung Medison Co., Ltd. Jee Young Ju Regulatory Affairs Specialist 3366. Hanseo-ro. Nam-myeon. Hongcheon-gun, Gangwon-do 25108, REPUBLIC OF KOREA
Re: K182632
Trade/Device Name: HS30 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic Pulsed Doppler Imaging System Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: December 7, 2018 Received: December 10, 2018
Dear Jee Young Ju:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
January 8, 2019
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara For
Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K182632
Device Name HS30 Diagnostic Ultrasound System
#### Indications for Use (Describe)
The HS30 Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body.
The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
{3}------------------------------------------------
510(k) Premarket Notification - Traditional
Image /page/3/Picture/2 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white, set against a blue, oval-shaped background. The font is sans-serif and bold, giving the logo a clean and modern appearance. The oval shape is slightly tilted, adding a dynamic element to the design.
## DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
#### Device Name: HS30 Diagnostic Ultrasound System
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | | Mode of Operation (*includes simultaneous B-mode) | | | | | | |
|---------------------------|-------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|-------------------------|--|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) | |
| Ophthalmic | Ophthalmic | N | N | N | N | N | N | N | |
| Fetal Imaging<br>& Other | Fetal/Obstetrics (See Note 3) | N | N | N | N | N | Note 1 | Note 2, 7, 8, 9, 11 | |
| | Abdominal (See Note 10) | N | N | N | N | N | Note 1 | Note 2, 7, 8, 9, 10, 11 | |
| | Intra-operative (See Note 6) | N | N | N | N | N | N | | |
| | Intra-operative (Neuro.) | N | N | N | N | N | N | | |
| | Laparoscopic | N | N | N | N | N | N | | |
| | Pediatric | N | N | N | N | N | Note 1 | Note 2, 7, 8, 9, 11 | |
| | Small Organ (See Note 5) | N | N | N | N | N | Note 1 | Note 2, 7, 9, 11, 12 | |
| | Neonatal Cephalic | N | N | N | N | N | Note 1 | Note 8, 9, 11 | |
| | Adult Cephalic | N | N | N | N | N | Note 1 | Note 7 | |
| | Trans-rectal | N | N | N | N | N | Note 1 | Note 2, 7, 9, 12, 13 | |
| | Trans-vaginal | N | N | N | N | N | Note 1 | Note 2, 7, 9, 12 | |
| | Trans-urethral | N | N | N | N | N | N | N | |
| | Trans-esoph. (non-Cardiac) | N | N | N | N | N | N | N | |
| | Musculo-skel. (Convent.) | N | N | N | N | N | Note 1 | Note 2, 7, 8, 9, 11, 12 | |
| | Musculo-skel. (Superfic.) | N | N | N | N | N | Note 1 | Note 2, 7, 9, 11 | |
| | Intra-luminal | N | N | N | N | N | N | N | |
| | Other (See Note 13) | N | N | N | N | N | Note 1 | Note 2, 7, 9, 12 | |
| Cardiac | Cardiac Adult | N | N | N | N | N | Note 1 | Note 4, 7, 14 | |
| | Cardiac Pediatric | N | N | N | N | N | Note 1 | Note 4, 7, 14 | |
| | Trans-esophageal (Cardiac) | N | N | N | N | N | N | N | |
| | Other (spec.) | N | N | N | N | N | N | N | |
| Peripheral<br>Vessel | Peripheral vessel | N | N | N | N | N | Note 1 | Note 2, 7, 8, 9, 11 | |
| | Other (spec.) | N | N | N | N | N | N | N | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
#### Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+TD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
- Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
- Note 13: Includes Urology/Prostate
- Note 14: Tissue Doppler Imaging (TDI)
{4}------------------------------------------------
510(k) Premarket Notification - Traditional
Image /page/4/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The background is a blue oval shape.
## DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: CF4-9 for use with HS30
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | | Mode of Operation (*includes simultaneous B-mode) | | | | | | | |
|---------------------------|-------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|-------------------|--|--|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) | | |
| Ophthalmic | Ophthalmic | | | | | | | | | |
| Fetal Imaging<br>& Other | Fetal/Obstetrics (See Note 3) | P | P | P | | P | Note 1 | Notes 7, 8, 9, 11 | | |
| | Abdominal (See Note 10) | P | P | P | | P | Note 1 | Notes 7, 8, 9, 11 | | |
| | Intra-operative (See Note 6) | | | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | | | |
| | Laparoscopic | | | | | | | | | |
| | Pediatric | P | P | P | | P | Note 1 | Notes 7, 8, 9, 11 | | |
| | Small Organ (See Note 5) | | | | | | | | | |
| | Neonatal Cephalic | P | P | P | | P | Note 1 | Notes 7, 8, 9, 11 | | |
| | Adult Cephalic | | | | | | | | | |
| | Trans-rectal | | | | | | | | | |
| | Trans-vaginal | | | | | | | | | |
| | Trans-urethral | | | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | | | |
| | Musculo-skel. (Convent.) | P | P | P | | P | Note 1 | Notes 7, 8, 9, 11 | | |
| | Musculo-skel. (Superfic.) | | | | | | | | | |
| | Intra-luminal | | | | | | | | | |
| | Other (See Note 13) | | | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | | | |
| | Cardiac Pediatric | | | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | | | |
| | Other (spec.) | | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | P | P | | P | Note 1 | Notes 7, 8, 9, 11 | | |
| | Other (spec.) | | | | | | | | | |
N= new indication; P= previously cleared by FDA K180409; E= added under Appendix E
#### Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+TD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
Note 11: Panoramic imaging
Note 12: ElastoScan
Note 13: Includes Urology/Prostate
Note 14: Tissue Doppler Imaging (TDI)
{5}------------------------------------------------
510(k) Premarket Notification - Traditional
Image /page/5/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white letters on a blue oval background. The logo is simple and recognizable, and it is associated with the Samsung brand.
## DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: PN2-4 for use with HS30
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | | Mode of Operation (*includes simultaneous B-mode) | | | | | | | |
|---------------------------|-------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|------------------|--|--|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) | | |
| Ophthalmic | Ophthalmic | | | | | | | | | |
| | Fetal/Obstetrics (See Note 3) | | | | | | | | | |
| | Abdominal (See Note 10) | P | P | P | | P | Note 1 | Note 7 | | |
| | Intra-operative (See Note 6) | | | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | | | |
| | Pediatric | | | | | | | | | |
| | Small Organ (See Note 5) | | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | | |
| | Adult Cephalic | P | P | P | | P | Note 1 | Note 7 | | |
| | Trans-rectal (See Note 13) | | | | | | | | | |
| | Trans-vaginal (See Note 13) | | | | | | | | | |
| | Trans-urethral | | | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | | | |
| | Intra-luminal | | | | | | | | | |
| | Other (See Note 13) | | | | | | | | | |
| | Cardiac Adult | P | P | P | | P | Note 1 | Note 4, 7, 14 | | |
| Cardiac | Cardiac Pediatric | P | P | P | | P | Note 1 | Note 4, 7, 14 | | |
| | Trans-esophageal (Cardiac) | | | | | | | | | |
| | Other (spec.) | | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | | | |
| | Other (spec.) | | | | | | | | | |
N= new indication; P= previously cleared by FDA K180409; E= added under Appendix E
#### Additional Comments:
Note 1: B+M, B+PW, B+CW, B+C, B+PD, B+Elastoscan, B+C+PW, B+PD+PW, B+DPD+PW, B+TD+PW, B+C+M,
B+C+CW, B+PPI, B+TD, B+PPI+PW, Dual/Quad (B, B+C, B+PD, B+TD, B+DPD)
Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging
Note 9: MultiVision (Spatial Compound Imaging)
Note 10: Includes Renal, Gynecology/Pelvis
Note 11: Panoramic imaging
- Note 12: ElastoScan
Note 13: Includes Urology/Prostate
Note 14: Tissue Doppler Imaging (TDI)
Note 2: Includes imaging for guidance of biopsy
{6}------------------------------------------------
510(k) Premarket Notification - Traditional
Image /page/6/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The background is a blue, rounded oval shape.
## DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: EVN4-9 for use with HS30
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | | Mode of Operation (*includes simultaneous B-mode) | | | | | | | |
|---------------------------|-------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|----------------------|--|--|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) | | |
| Ophthalmic | Ophthalmic | | | | | | | | | |
| | Fetal/Obstetrics (See Note 3) | P | P | P | | P | Note 1 | Note 2, 7, 9 | | |
| | Abdominal (See Note 10) | | | | | | | | | |
| | Intra-operative (See Note 6) | | | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | | | |
| | Pediatric | | | | | | | | | |
| | Small Organ (See Note 5) | | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | | |
| | Adult Cephalic | | | | | | | | | |
| | Trans-rectal | P | P | P | | P | Note 1 | Note 2, 7, 9, 12, 13 | | |
| | Trans-vaginal | P | P | P | | P | Note 1 | Note 2, 7, 9, 12 | | |
| | Trans-urethral | | | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | | | |
| | Intra-luminal | | | | | | | | | |
| | Other (spec.) (See Note 13) | P | P | P | | P | Note 1 | Note 2, 7, 9, 12 | | |
| | Cardiac Adult | | | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | | | |
| | Other (spec.) | | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | | | |
| | Other (spec.) | | | | | | | | | |
N= new indication; P= previously cleared by FDA K180409; E= added under Appendix E
#### Additional Comments:
Note 1: B+M, B+PW, B+CW, B+C, B+PD, B+Elastoscan, B+C+PW, B+PD+PW, B+DPD+PW, B+TD+PW, B+C+M,
B+C+CW, B+PPI, B+TD, B+PPI+PW, Dual/Quad (B, B+C, B+PD, B+TD, B+DPD)
- Note 2: Includes imaging for guidance of biopsy
- Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
- Note 6: Abdominal organs and peripheral vessel
- Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging
Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
Note 13: Includes Urology/Prostate Note 14: Tissue Doppler Imaging (TDI)
Image /page/6/Figure/24 description: The image contains text that states "Concurrence of Center for Devices and Radiological Health (CDRH)". The image also states "Prescription Use (Per 21 CFR 801.109)". The text is centered on the page.
{7}------------------------------------------------
510(k) Premarket Notification - Traditional
Image /page/7/Picture/2 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white, set against a blue oval background. The font is sans-serif and appears to be bolded.
## DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
#### Device Name: DP2B for use with HS30
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | | Mode of Operation (*includes simultaneous B-mode) | | | | | |
|---------------------------|-------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging<br>& Other | Fetal/Obstetrics (See Note 3) | | | | | | | |
| | Abdominal (See Note 12) | | | | | | | |
| | Intra-operative (See Note 6) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (See Note 5) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | P | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (See Note 13) | | | | | | | |
| Cardiac | Cardiac Adult | | | | P | | | |
| | Cardiac Pediatric | | | | P | | | |
| | Trans-esophageal (Cardiac) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | P | | | |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by FDA K180409; E= added under Appendix E
#### Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+TD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
- Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
Note 13: Includes Urology/Prostate
Note 14: Tissue Doppler Imaging (TDI)
{8}------------------------------------------------
510(k) Premarket Notification - Traditional
Image /page/8/Picture/2 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white, set against a blue oval background. The oval is tilted slightly, giving the logo a dynamic appearance.
## DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: C2-8 for use with HS30
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation (*includes simultaneous B-mode) | | | | | | |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|----------------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging<br>& Other | Fetal/Obstetrics (See Note 3) | P | P | P | | P | Note 1 | Notes 2, 4, 7,8, 9, 11 |
| | Abdominal (See Note 10) | P | P | P | | P | Note 1 | Notes 2, 6, 7, 8, 9,10, 11 |
| | Pediatric | P | P | P | | P | Note 1 | Notes 2, 7, 8, 9, 11 |
| | Intra-operative (See Note 6) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Small Organ (See Note 5) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (See Note 13) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by FDA K180409; E= added under Appendix E
#### Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+TD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
- Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
- Note 13: Includes Urology/Prostate
- Note 14: Tissue Doppler Imaging (TDI)
{9}------------------------------------------------
510(k) Premarket Notification - Traditional
Image /page/9/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The background is a blue oval shape.
## DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: C2-5 for use with HS30
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation (*includes simultaneous B-mode) | | | | | | |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|-----------------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal/Obstetrics (See Note 3) | P | P | P | P | P | Note 1 | Notes 2, 4, 7, 8, 9, 11 |
| | Abdominal (See Note 10) | P | P | P | P | P | Note 1 | Notes 2, 6, 7, 8, 9, 10, 11 |
| | Intra-operative (See Note 6) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | P | P | Note 1 | Notes 2, 7, 8, 9, 11 |
| | Small Organ (See Note 5) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (See Note 13) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by FDA K180409; E= added under Appendix E
#### Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
- Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+TD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
- Note 2: Includes imaging for guidance of biopsy
- Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
- Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
- Note 6: Abdominal organs and peripheral vessel
- Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging
- Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
- Note 13: Includes Urology/Prostate
- Note 14: Tissue Doppler Imaging (TDI)
{10}------------------------------------------------
510(k) Premarket Notification - Traditional
Image /page/10/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white letters on a blue oval background. The logo is simple and recognizable, and it is often used on Samsung products and in advertising.
## DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
#### Device Name: LN5-12 for use with HS30
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation (*includes simultaneous B-mode) | | | | | | |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|-------------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging<br>& Other | Fetal/Obstetrics (See Note 3) | | | | | | | |
| | Abdominal (See Note 10) | | | | | | | |
| | Intra-operative (See Note 6) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | | P | Note 1 | Notes 2, 7, 8, 9, 11 |
| | Small Organ (See Note 5) | P | P | P | | P | Note 1 | Notes 2, 5, 7, 8, 9, 11 |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | |
| | Musculo-skel. (Convent.) | P | P | P | | P | Note 1 | Notes 2, 7, 8, 9, 11 |
| | Musculo-skel. (Superfic.) | P | P | P | | P | Note 1 | Notes 2, 7, 8, 9, 11 |
| | Intra-luminal | | | | | | | |
| | Other (See Note 13) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | P | P | | P | Note 1 | Notes 2, 7, 8, 9, 11 |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by FDA K180409; E= added under Appendix E
#### Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+TD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
Note 11: Panoramic imaging
Note 12: ElastoScan
Note 13: Includes Urology/Prostate
Note 14: Tissue Doppler Imaging (TDI)
{11}------------------------------------------------
510(k) Premarket Notification - Traditional
Image /page/11/Picture/2 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white, set against a blue, oval-shaped background. The oval is tilted slightly, giving the logo a dynamic appearance.
## DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
### Device Name: L5-12/50 for use with HS30
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation (*includes simultaneous B-mode) | | | | | | |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|-------------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging<br>& Other | Fetal/Obstetrics (See Note 3) | | | | | | | |
| | Abdominal (See Note 10) | | | | | | | |
| | Intra-operative (See Note 6) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | | P | Note 1 | Notes 2, 7, 8, 9, 11 |
| | Small Organ (See Note 5) | P | P | P | | P | Note 1 | Notes 2, 5, 7, 8, 9, 11 |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | |
| | Musculo-skel. (Convent.) | P | P | P | | P | Note 1 | Notes 2, 7, 8, 9, 11 |
| | Musculo-skel. (Superfic.) | P | P | P | | P | Note 1 | Notes 2, 7, 8, 9, 11 |
| | Intra-luminal | | | | | | | |
| | Other (See Note 13) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | P | P | | P | Note 1 | Notes 2, 7, 8, 9, 11 |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by FDA K180409; E= added under Appendix E
#### Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+TD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
- Note 13: Includes Urology/Prostate
- Note 14: Tissue Doppler Imaging (TDI)
{12}------------------------------------------------
510(k) Premarket Notification - Traditional
Image /page/12/Picture/2 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white, set against a blue, oval-shaped background. The font is sans-serif and bold, giving the logo a clean and modern appearance. The oval shape is slightly tilted, adding a dynamic element to the design.
## DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: ER4-9 for use with HS30
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation (*includes simultaneous B-mode) | | | | | | |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|-----------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal/Obstetrics (See Note 3) | P | P | P | | P | Note 1 | Notes 2, 7, 8 |
| | Abdominal (See Note 10) | | | | | | | |
| | Intra-operative (See Note 6) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (See Note 5) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | P | P | P | | P | Note 1 | Notes 2, 7, 9, 12, 13 |
| | Trans-vaginal | P | P | P | | P | Note 1 | Notes 2, 7, 9, 12 |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (See Note 13) | P | P | P | | P | Note 1 | Notes 2, 7, 9, 12 |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by FDA K180409; E= added under Appendix E
#### Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+TD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Note 9: MultiVision (Spatial Compound Imaging)
Note 10: Includes Renal, Gynecology/Pelvis
Note 11: Panoramic imaging
Note 12: ElastoScan
Note 13: Includes Urology/Prostate
Note 14: Tissue Doppler Imaging (TDI)
{13}------------------------------------------------
### 5. 510(K) Summary: K182632
In accordance with 21 CFR 807.92 the following summary of information is provided:
- Date Prepared December 07, 2018 1.
- 2. Manufacturer SAMSUNG MEDISON CO., LTD. 3366, Hanseo-ro, Nam-myeon, Hongcheon-gun, Gangwon-do 25108, REPUBLIC OF KOREA
- Primary Contact Person 3. Jee Young Ju Requlatory Affairs Specialist Phone: +82.2.2194.0861 Fax: +82.31.8017.9573 Email: jee.ju@samsungmedison.com
- 4. Secondary Contact Person Roberto Cunha Director of Regulatory & Quality Phone: +1.978.564.8503 Email: rcunha@samsungneurologica.com
- 5. Proposed Device
--
- Proprietary Name: HS30 Diagnostic Ultrasound System -
- Common Name: System, Imaging, Pulsed Doppler, Ultrasonic -
- System, Imaging, Pulsed Echo, Ultrasonic
- Transducer, Ultrasonic, Diagnostic
- Classification : 21 CFR 892.1550 Ultrasonic pulsed doppler imaging system - 21 CFR 892.1560 Ultrasonic pulsed echo imaging system 21 CFR 892.1570 Diagnostic ultrasonic transducer
- Product Code(s): IYN, IYO, ITX -
- Predicate Device 6.
- HS40 Diagnostic Ultrasound System (K180409) -
The predicate has not been the subject of a design-related recall.
- 7. Device Description
The HS30 is a general purpose, mobile, software controlled, diagnostic ultrasound system, Its function is to acquire ultrasound data and to display the data as 2D mode. M mode. Color Doppler imaging, Power Doppler imaging (including Directional Power Doppler mode; S-Flow), PW Spectral Doppler mode, CW Spectral Doppler mode, Harmonic imaging(S-Harmonic), Tissue Doppler imaging, Tissue Doppler Wave, Panoramic Imaging, Freehand 3D, Elastoscan Mode or as a combination of these modes. The HS30 also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The HS30 has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.
The HS30 uses digital multi-beam forming technology and supports a variety of Linear, Convex, Phased, Static and Volume probes for a wide variety of applications. It is an ultrasound scanner, which provides high resolution, high penetration performance, and various measurement functions. Probes are supported in frequencies from 1.0 MHz to 20.0 MHz. These probes can be applied to a variety of clinical applications such as Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult
{14}------------------------------------------------
Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel.
- Indications for Use 8.
The HS30 Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body.
The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel.
The indications for use of the proposed device are the same as the predicate HS40 (K180409) except Intra-Operative application.
Since the proposed device HS30 does not support the Intra-Operative application for any of the transducers, it does not raise new questions of safety and effectiveness.
#### 9. Technological Comparison to Predicate Devices
The proposed HS30 Diagnostic Ultrasound System and the currently marketed predicate device HS40 (K180409), employ the same fundamental scientific technology as all of the features are migrated from the predicate.
A comparison of the proposed HS30 Diagnostic Ultrasound System to the currently marketed and predicates HS40 (K180409) is provided in the Table below:
| Feature | HS30<br>(Under Review) | HS40<br>(K180409) |
|---------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use…