MyLabSix Ultrasound System

K161168 · Esaote Europe B.V. · IYN · Sep 2, 2016 · Radiology

Device Facts

Record IDK161168
Device NameMyLabSix Ultrasound System
ApplicantEsaote Europe B.V.
Product CodeIYN · Radiology
Decision DateSep 2, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1550
Device ClassClass 2
AttributesPediatric

Intended Use

Esaote's MyLabSix ultrasound system is a mainframe ultrasound system used to perform diagnostic general ultrasound studies including Cardiac. Transesophageal Cardiac, Peripheral Vascular, Neonatal Cephalic, Adult Cephalic, Small Organs, Musculoskeletal (Conventional and Superficial), Abdominal, Fetal, Transvaginal, Transrectal, Pediatric, Intraoperative Abdominal, Laparoscopic and Other: Urologic. The system provides imaging for guidance of biopsy and imaging to assist in the placement of needles in vascular or other anatomical structures as well as peripheral nerve blocks in Musculoskeletal applications.

Device Story

MyLabSix is a mainframe ultrasound system for diagnostic imaging and fluid flow analysis. It processes inputs from various ultrasound probes (phased, convex, linear, Doppler, volumetric) to produce real-time images. The system features B-Mode, Amplitude Doppler, Tissue Enhancement Imaging, Multi View, Tissue Velocity Mapping, Color Flow Mapping, Pulse Wave Doppler, and 3D/4D modes. Operated by clinicians in clinical settings, it includes a height-adjustable keyboard and dual LCD displays (one with touch screen). The upgrade adds motorized probe management for 3D/4D volumetric acquisition, laparoscopic application support, needle visibility enhancement, and an on-board 'MyLibrary' tutorial feature. Output is used by physicians to guide biopsies, assist needle placement, and perform diagnostic studies, potentially improving procedural accuracy and patient outcomes.

Clinical Evidence

No clinical tests were performed. Substantial equivalence is supported by bench testing, including acoustic output measurement, biocompatibility, and electrical/mechanical safety testing per IEC and NEMA standards.

Technological Characteristics

Mainframe ultrasound system with LCD displays and touch-screen control panel. Supports phased, convex, linear, Doppler, and volumetric probes. Connectivity via LAN/USB for digital storage. Conforms to IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-2-37 (ultrasound safety), and ISO 10993 (biocompatibility). Features include Needle Visibility, MyLibrary tutorial, and motorized probe management for 3D/4D acquisition.

Indications for Use

Indicated for diagnostic ultrasound imaging or fluid flow analysis of human body including cardiac, transesophageal cardiac, peripheral vascular, neonatal/adult cephalic, small organs, musculoskeletal (conventional/superficial), abdominal, fetal, transvaginal, transrectal, pediatric, intraoperative abdominal, laparoscopic, and urologic applications. Includes biopsy guidance and needle placement assistance for vascular/anatomical structures and peripheral nerve blocks.

Regulatory Classification

Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 2, 2016 Esaote Europe B.V. % Allison Scott, RAC Managing Consultant Navigant Consulting, Inc. 9001 Wesleyan Road, Suite 200 INDIANAPOLIS IN 46268 Re: K161168 Trade/Device Name: MyLabSix Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: August 1, 2016 Received: August 2, 2016 Dear Ms. Scott: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Michael D'Hara For Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K161168 Device Name MyLabSix Ultrasound System ### Indications for Use (Describe) Esaote's MyLabSix ultrasound system is a mainframe ultrasound system used to perform diagnostic general ultrasound studies including Cardiac, Transesophageal Cardiac, Peripheral Vascular, Neonatal Cephalic, Adult Cephalic, Small Organs, Musculoskeletal (Conventional and Superficial), Abdominal, Fetal, Transvaginal, Transrectal, Pediatric, Intraoperative Abdominal, Laparoscopic and Other: Urologic. The system provides imaging for guidance of biopsy and imaging to assist in the placement of needles in vascular or other anatomical structures as well as peripheral nerve blocks in Musculoskeletal applications. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Prescription Use Only Per 21 CFR 801 Part D Concurrence DRH, Office of In Vitro Diagnostics and Radiological Health | Clinical Application | B | M | PWD | CWD | Color Doppler | Amplitude Doppler (AD) | Combined [4] | Color Velocity Mapping (TVM) | Harmonic Imaging (TEI) | Other (specify) | |----------------------------------------------------------------------------|---|---|-----|-----|---------------|------------------------|--------------|------------------------------|------------------------|-----------------| | Ophthalmic | P | P | P | P | P | P | P | P | P | P: 5 N/A | | Fetal | P | P | P | P | P | P | P | P | P | P: 5 N/A | | Abdominal | P | P | P | P | P | P | P | P | P | P: 5 N/A | | Intraoperative (Abdominal) | P | P | P | P | P | P | P | P | P | P: 5 N/A | | Intraoperative Neurological | P | P | P | P | P | P | P | P | P | P: 5 N/A | | Pediatric | P | P | P | P | P | P | P | P | P | P: 5 N/A | | Small Organs [1] | P | P | P | P | P | P | P | P | P | P: 5 N/A | | Neonatal Cephalic | P | P | P | P | P | P | P | P | P | P: 5 N/A | | Adult Cephalic | P | P | P | P | P | P | P | P | P | P: 5 N/A | | Cardiac [2] | P | P | P | P | P | P | P | P | P | P: 5 N/A | | Transesophageal (Cardiac) | P | P | P | P | P | P | P | P | P | P: 5 N/A | | Transesophageal (Non Cardiac) | P | P | P | P | P | P | P | P | P | P: 5 N/A | | Transrectal | P | P | P | P | P | P | P | P | P | P: 5 N/A | | Transvaginal | P | P | P | P | P | P | P | P | P | P: 5 N/A | | Transurethral | P | P | P | P | P | P | P | P | P | N: 5 N/A | | Intravascular | P | P | P | P | P | P | P | P | P | P: 5 N/A | | Peripheral Vascular | P | P | P | P | P | P | P | P | P | P: 5 N/A | | Laparoscopic | N | N | N | N | N | N | N | N | N | N: 5 N/A | | Musculo-skeletal Conventional [3] | P | P | P | P | P | P | P | P | P | P: 5 N/A | | Musculo-skeletal Superficial [3] | P | P | P | P | P | P | P | P | P | P: 5 N/A | | Other (Urological) | P | P | P | P | P | P | P | P | P | P: 5 N/A | | 5420 is already cleared via K141486 and to be cleared with this submission | | | | | | | | | | | N: 7 N: 7 ഹ ഹ 5 5 ம் ம ம Mode of Operat ended use: Diagnostic ultrasound imaging or fluid flow analysis of human body as fol 6420 specify) 11:4 ம ഹ ഹ ഹ N: 7 21 New 11 deatlem in New Brain 2017 Bir Birder under Appendix B 21 Naturalian - New Black - Previous Pro 21 Natural - News Black - News Blomber - Pro 10 Domineral Prival Pro CTO, National Hemophilia Foundation Senior VP, Programs THOMAS BRENNAN <E-mail: TBrennan@hemophilia.org> and Restorative Therapies 112 Cardiac RIS Adult and Pediatric [3] - [4] Combined modoro R+M+RW CY ES-Compound Imaging (Moving Window) (b) includes contrast (Cm in Activity 16) and (c) includes contrast (Cm in Activity 17). Visualization of the left ventricular endocardial border using cardiac magnetic resonance imaging F1/3D/4D {4}------------------------------------------------ | SI2C41 | SL3116 | SE3133 | TLC 3-13 | IH 6-18 | IL 4-13 | LP 4-13 | SB3123 | SL3323 | SL3235 | SL2325 | SL3332 | SC3421 | AC2541 | SC3123 | SE3123 | AL2442 | SL1543 | SP2730 | S2MCW | S5MCW | SB2C41 | SHFCW | ST2612 | IOT342 | |--------|--------|--------|----------|---------|---------|---------|--------|--------|--------|--------|--------|--------|--------|--------|--------|--------|--------|--------|-------|-------|--------|-------|--------|--------| | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | 11 | 12 | 13 | 14 | 15 | 16 | 17 | 18 | 19 | 20 | 21 | 22 | 23 | 24 | 25 | {5}------------------------------------------------ # Prescription Use Only Per 21 CFR 801 Part D Concurrence of DRH, Office of In Vitro Diagnostics and Radiological Health (OlR | Clinical Application | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler<br>(AD) | Combined<br>[4] | Color Velocity<br>Mapping (TVM) | Harmonic<br>Imaging (TEI) | Other (specify) | |----------------------------------|---|---|-----|-----|------------------|------------------------------|-----------------|---------------------------------|---------------------------|-----------------| | Ophthalmic | E | E | E | E | E | E | E | E | E | E: 5 | | Fetal | E | E | E | E | E | E | E | E | E | E: 5 | | Abdominal | E | E | E | E | E | E | E | E | E | E: 5 | | Intraoperative (Abdominal) | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | Pediatric | E | E | E | E | E | E | E | E | E | E: 5 | | Small Organs [1] | E | E | E | E | E | E | E | E | E | E: 5 | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac [2] | | | | | | | | | | | | Transesophageal (Cardiac) | | | | | | | | | | | | Transesophageal (Non Cardiac) | | | | | | | | | | | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral Vascular | E | E | E | E | E | E | E | E | E | E: 5 | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal Conventional | E | E | E | E | E | E | E | E | E | E: 5 | | Musculo-skeletal Superficial [3] | E | E | E | E | E | E | E | E | E | E: 5 | | Other (Urological) | E | E | E | E | E | E | E | E | E | E: 5 | | The SI2C41 probe is added | | | | | | | | | | | | Clinical Application | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler<br>(AD) | Combined<br>[4] | Color Velocity<br>Mapping (TVM) | Harmonic<br>Imaging (TEI) | Other (specify) | | Ophthalmic | | | | | | | | | | | | Fetal | | | | | | | | | | | | Abdominal | | | | | | | | | | | | Intraoperative (Abdominal) | P | P | P | | P | P | P | | P | | | Intraoperative Neurological | | | | | | | | | | | | Pediatric | P | P | P | | P | P | P | | P | E: 5 | | Small Organs [1] | P | P | P | | P | P | P | | P | E: 5 | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac [2] | | | | | | | | | | | | Transesophageal (Cardiac) | | | | | | | | | | | | Transesophageal (Non Cardiac) | | | | | | | | | | | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral Vascular | P | P | P | | P | P | P | | P | E: 5 | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal Conventional | P | P | P | | P | P | P | | P | E: 5 | | Musculo-skeletal Superficial [3] | P | P | P | | P | P | P | | P | E: 5 | | Other (Urological) | | | | | | | | | | | | Mode of Operations | | | | | | | | | | | | Clinical Application | B | M | PWD | CWD | Color Doppler | Amplitude Doppler (AD) | Combined [4] | Color Velocity Mapping (TVM) | Harmonic Imaging (TEI) | Other (specify) | | Ophthalmic | | E | | | E | | E | | E | E: 5 | | Fetal | E | E | E | | E | E | E | | E | E: 5 | | Abdominal | | | | | | | | | | | | Intraoperative (Abdominal) | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | Pediatric | | | | | | | | | | | | Small Organs [1] | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac [2] | | | | | | | | | | | | Transesophageal (Cardiac) | | | | | | | | | | | | Transesophageal (Non Cardiac) | | | | | | | | | | | | Transrectal | E | E | E | | E | E | E | | E | E: 5 | | Transvaginal | E | E | E | | E | E | E | | E | E: 5 | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral Vascular | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal Conventional | | | | | | | | | | | | Musculo-skeletal Superficial [3] | | | | | | | | | | | | Other (Urological) | E | E | E | | E | E | E | | E | E: 5 | | *the SE3133 probe is added | | | | | | | | | | | | Clinical Application | B | M | PWD | CWD | Color Doppler | Amplitude Doppler (AD) | Combined [4] | Color Velocity Mapping (TVM) | Harmonic Imaging (TEI) | Other (specify) | | Ophthalmic | E | E | E | | E | E | E | | | | | Fetal | | | | | | | | | | E: 5 | | Abdominal | | | | | | | | | | | | Intraoperative (Abdominal) | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | Pediatric | | | | | | | | | | | | Small Organs [1] | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac [2] | | | | | | | | | | | | Transesophageal (Cardiac) | | | | | | | | | | | | Transesophageal (Non Cardiac) | | | | | | | | | | | | Transrectal | E | E | E | | E | E | E | | E | E: 5 | | Transvaginal | E | E | E | | E | E | E | | E | E: 5 | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral Vascular | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal Conventional | | | | | | | | | | | | Musculo-skeletal Superficial [3] | | | | | | | | | | | | Other (Urological) | E | E | E | | E | E | E | | E | E: 5 | | he TLC 3-13 probe is added und | | | | | | | | | | | SI2C41 Intended use: Diagnostic ultrasound imaging or fluid flow analysis of human body as foll The S12C41 probe is added under Appendix F and to be under Appendix E and to be N: New indication; P: Previously cleared by FDA; E:Added under Apper Testicles and Prostate Breast Studies Thyroid Studies Brain Scans SOME STRAINS include Brazil, Reunion and India. OIL contains both and Palmitic [3] Musculo Skeletal - Nerve Block [4] Cardiac is Auricular and Pendular [4] Combined modes are: B+M+PW+CW+CFM+ [5] Compound Imaging (M-view) alaba walinda anala mara baya MDA, Endada under Appendia E 21 Museum e Marce Black Proveders 13 Museum e Marre Black Proveders 13 Museum e Marria Provinco Merce Mericle operi [7] 3D/4D {6}------------------------------------------------ SL3116 Intended use: Diagnostic ultrasound imaging or fluid flow analysis of human body as follow The SESTTO probe is all cady cleared via K101605 and added under Appendix E under Appendix E under Appenda B (1) Small Organs includes Breast, Thyroid etc. (2) Cardiac is Adult and Pediatric [32] Musculo Skeletal - Nerve Block [4] Combined modes are: B-: B+: M-: M+: PW-: PW+: CV-: CV+ 1.5 Compound Imaging (M-View) [6] Includes contrast (CnTI) in Adult Cardiac for left ventricle o INSPECTION OF LINE FOR VERTICAL BORD AND HORIZONTAL BORD REHABILITATION OF MYOCARDIAL INFARCTION PATIENTS {7}------------------------------------------------ Prescription Use Only Per 21 CFR 801 Part D Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OlR) ntended use: Diagnostic ultrasound imaging or fluid flow analysis of human body as follo SE3133 under Appendix E and to be cleared via this submission N: New indication: P: Previously cleared by FDA: E:Added und Director: Production: Camera: Lighting: Sound: Art Director: Editor: Music: Cast: 2) Cardiac is Adult, Pediatric and Pediatric Cardiac [3] Musculo Skeletal - Nerve Block [4] Combined modes are: B+M+PW+CW+ [5] Compound Imaging (MView) ﺍﻟﻤﺴﺎﻋﺪ ﺍﻟﻤﺎﺿﻲ ﺍﻟﻤﺴﺎﻋﺪﺓ ﺍﻟﻤﺴﺎﻋﺪﺓ ﺍﻟﻤﺴﺎﻋﺪﺓ ﺍﻟﻤﺴﺎﻋﺪﺓ ﺍﻟﻤﺴﺎﻋﺪﺓ ﺍﻟﻤﺴﺎﻋﺪﺓ ﺍﻟﻤﺴﺎﻋﺪﺓ ﺍﻟﻤﺴﺎﻋﺪﺓ ﺍﻟﻤﺴﺎﻋﺪﺓ ﺍﻟﻤﺴﺎﻋﺪﺓ ﺍﻟﻤﺴﺎﻋﺪﺓ ﺍﻟﻤﺴﺘﻮﻳﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺎﺑﻌﺔ ﺍﻟﻤﺘﻌﺪﺩﺓ ﺍﻟﻤﺘﺎﺑﻌﺔ [7]3D/4D {8}------------------------------------------------ T L C 3-13 Intended use: Diagnostic ultrasound imaging or fluid flow analysis of human body as follow The TLC 3-13 probe is added under Appendix E and to be cleared via this submission SUBMISSIONS index Bauen Printed by EDA: Bridded under Abendix B 13 Marcula Skault Marce Brother Bridded under Abendix B 13 Marcula Skeule Broupe Cuvic Arter Bridget Price Archives and 17 11. Small Organs includes Breast, Thyroid [2] Cardiac is Adult and Pediatric [3] Musculo Skeletal - Nerve Block [4] Combined modes are: B+M, PW+M, C+V REC COMBINED INDEX (RCI) + TOTAL HIRS RAD imaging (mview) [6] Includes contrast (CTI) in Adult Cardiac fo INCREASED CONTRAST FORM IN FINAL SURFACE FOR ISOLATION of the left ventricular endocardial border [5.1A] DIVISION OF REAL ESTATE, PENNSYLVANIA DEPARTMENT OF STATE APPLICATION FOR REAL ESTATE SALESPERSON LICENSE {9}------------------------------------------------ IH 6-18 Intended use: Diagnostic ultrasound imaging or fluid flow analysis of human body as follow | Clinical Application | B | M | PWD | CWD | Color Doppler | Amplitude Doppler (AD) | Combined [4] | Color Velocity Mapping (TVM) | Harmonic Imaging (TEI) | Other (specify) | |-----------------------------------|---|---|-----|-----|------------------|------------------------------|-----------------|---------------------------------|---------------------------|-----------------| | Ophthalmic | E | E | E | | E | E | E | | E | E: 5 | | Fetal | E | E | E | | E | E | E | | E | E: 5 | | Abdominal | E | E | E | | E | E | E | | E | E: 5 | | Intraoperative (Abdominal) | E | E | E | | E | E | E | | E | E: 5 | | Intraoperative Neurological | | | | | | | | | | | | Pediatric | E | E | E | | E | E | E | | E | E: 5 | | Small Organs [1] | E | E | E | | E | E | E | | E | E: 5 | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac [2] | | | | | | | | | | | | Transesophageal (Cardiac) | | | | | | | | | | | | Transesophageal (Non Cardiac) | | | | | | | | | | | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral Vascular | E | E | E | | E | E | E | | E | E: 5 | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal Conventional | E | E | E | | E | E | E | | E | E: 5 | | Musculo-skeletal Superficial [3] | E | E | E | | E | E | E | | E | E: 5 | | Other (Urological) | E | E | E | | E | E | E | | E | E: 5 | | The IH 6-18 probe is added | | | | | | | | | | | | Clinical Application | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler<br>(AD) | Combined<br>[4] | Color Velocity<br>Mapping (TVM) | Harmonic<br>Imaging (TEI) | Other (specify) | | Ophthalmic | E | E | E | E | E | E | E | E | E | E: 5 | | Fetal | E | E | E | E | E | E | E | E | E | E: 5 | | Abdominal | E | E | E | E | E | E | E | E | E | E: 5 | | Intraoperative (Abdominal) | E | E | E | E | E | E | E | E | E | E: 5 | | Intraoperative Neurological | | | | | | | | | | | | Pediatric | E | E | E | E | E | E | E | E | E | E: 5 | | Small Organs [1] | E | E | E | E | E | E | E | E | E | E: 5 | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac [2] | | | | | | | | | | | | Transesophageal (Cardiac) | | | | | | | | | | | | Transesophageal (Non Cardiac) | | | | | | | | | | | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral Vascular | E | E | E | E | E | E | E | E | E | E: 5 | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal Conventional [3] | E | E | E | E | E | E | E | E | E | E: 5 | | Musculo-skeletal Superficial [3] | E | E | E | E | E | E | E | E | E | E: 5 | | Other (Urological) | | | | | | | | | | | the IH 6-16 probe is added under Appendix E and to be moner of 10 probe is and to be under Appendix E and to be cleared via this submission अध्यक्ष र काम बनाने बाली बात 2018 - 12:44 PM Previous Provinsia Provinsia P 11 News March March Proved by Provinsia Provinsia B 13 News Market Marian Prove Corporation a 11. Small Organs includes Breast, Thyroid and more! [2] Cardiac is Adult and Pediatric 3) Musculo Skeletal - Nerve Block [4] Combined modes are: B+M+PW+CW+C [S]-Compound Imaging Signal (MEAN) (Arbitrary) CALCULATES COMPRESSOR VENT FACTOR FOR VENT LINE Cardiovascular, Endocrine, Neurologic, Musculoskeletal, Pulmonary, Renal, Gastrointestinal, Hematologic, Immunologic, Dermatologic, Ocular, ENT, Dental, and Mental Disorders border radial and occular centric left of the adaptation realizes {10}------------------------------------------------ Prescription Use Only Per 21 CFR 801 Part D Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OlR) under Appendix E and to be cleared via this submission N: New indication; P: Previously cleared by FDA; E:Added under A - **Breast** - **Cancer** - **Articles** - **Including** - **Ocean** - **Small** 2) Cardiac is Adult and Pediatric [3] Musculo Skeletal - Nerve Block [4] Combined modes are: B+M+PW+CW+ [5] Compound Imaging (Nview) classed via antilation 11 Marcin Results Marceller 11 Marcula elemin Merceller 13 Marcula elemin Merce Marcel Marie Marie Maria Maria Marina and 13 Marcula Republican Particl F1/3D/4D {11}------------------------------------------------ _P 4-13 ntended use: Diagnostic ultrasound imaging or fluid flow analysis of human body as for | Mode of Operations | | | | | | | | | | | | |--------------------------------------------------------------------------------------------------------|--------------------|----------|-----|-----|------------------|------------------------------|-----------------|---------------------------------|---------------------------|-----------------|------| | Clinical Application | B | M | PWD | CWD | Color Doppler | Amplitude Doppler (AD) | Combined [4] | Color Velocity Mapping (TVM) | Harmonic Imaging (TEI) | Other (specify) | | | Ophthalmic | | | | | | | | | | | | | Fetal | | | | | | | | | | | | | Abdominal | N | N | N | | N | N | N | | N | N: 5 | | | Intraoperative (Abdominal) | | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | | Pediatric | | | | | | | | | | | | | Small Organs [1] | | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | Cardiac [2] | | | | | | | | | | | | | Transesophageal (Cardiac) | | | | | | | | | | | | | Transesophageal (Non Cardiac) | | | | | | | | | | | | | Transrectal | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | Transurethral | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | Peripheral Vascular | | | | | | | | | | | | | Laparoscopic | N | N | N | | N | N | N | | N | N: 5 | | | Musculo-skeletal Conventional | | | | | | | | | | | | | Musculo-skeletal Superficial [3] | | | | | | | | | | | | | Other (Urological) | | | | | | | | | | | | | | | | | | | | | | | | | | SB3123<br>Intended use: Diagnostic ultrasound imaging or fluid flow analysis of human body as follows: | | | | | | Mode of Operations | | | | | | | Clinical Application | B | M | PWD | CWD | Color Doppler | Amplitude Doppler (AD) | Combined [4] | Color Velocity Mapping (TVM) | Harmonic Imaging (TEI) | Other (specify) | | | Ophthalmic | N | N | N | | N | N | N | | N | N: 5, 7 | | | Fetal | | | | | | | | | | | | | Abdominal | | | | | | | | | | | | | Intraoperative (Abdominal) | | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | | Pediatric | | | | | | | | | | | | | Small Organs [1] | | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | Cardiac [2] | | | | | | | | | | | | | Transesophageal (Cardiac) | | | | | | | | | | | | | Transesophageal (Non Cardiac) | | | | | | | | | | | | | Transrectal | N | N | N | | N | N | N | | N | N: 5, 7 | | | Transvaginal | N | N | N | | N | N | N | | N | N: 5, 7 | | | Transurethral | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | Peripheral Vascular | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | Musculo-skeletal Conventional | | | | | | | | | | | | | Musculo-skeletal Superficial [3] | | | | | | | | | | | | | Other (Urological) | N | N | N | | N | N | N | | N | N: 5, 7 | | | Mode of Operations | | | | | | | | | | | | | Clinical Application | B | M | PWD | CWD | Color Doppler | Amplitude Doppler (AD) | Combined [4] | Color Velocity Mapping (TVM) | Harmonic Imaging (TEI) | Other (specify) | | | Ophthalmic | | | | | | | | | | | | | Fetal | | | | | | | | | | | | | Abdominal | P | P | P | P | P | P | P | | P | P: 5 | | | Intraoperative (Abdominal) | P | P | P | P | P | P | P | | P | P: 5 | | | Intraoperative Neurological | | | | | | | | | | | | | Pediatric | P | P | P | P | P | P | P | | P | P: 5 | | | Small Organs [1] | | P | P | | P | P | P | | P | P: 5 | | | Neonatal Cephalic | | P | P | | P | P | P | | P | P: 5 | | | Adult Cephalic | | | | | | | | | | | | | Cardiac [2] | P | P | P | P | P | P | P | | P | P: 5 | | | Transesophageal (Cardiac) | | | | | | | | | | | | | Transesophageal (Non Cardiac) | | | | | | | | | | | | | Transrectal | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | Transurethral | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | Peripheral Vascular | P | P | P | | P | P | P | | P | P: 5 | | | Laparoscopic | | | | | | | | | | | | | Musculo-skeletal Conventional | P | P | P | | P | P | P | | P | P: 5 | | | Musculo-skeletal Superficial [3] | P | P | P | | P | P | P | | P | P: 5 | | | Other (Urological) | | | | | | | | | | | | | the SL3323 probe is already | | | | | | | | | | | | | Mode of Operations | | | | | | | | | | | | | Clinical Application | B | M | PWD | CWD | Color Doppler | Amplitude Doppler (AD) | Combined [4] | Color Velocity Mapping (TVM) | Harmonic Imaging (TEI) | Other (specify) | | | Ophthalmic | | | | | | | | | | | | | Fetal | | | | | | | | | | | | | Abdominal | | | | | | | | | | | | | Intraoperative (Abdominal) | | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | | Pediatric | P | P | P | | P | P | P | | P | P: 5 | | | Small Organs [1] | P | P | P | | P | P | P | | P | P: 5 | | | Neonatal Cephalic | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | Cardiac [2] | | | | | | | | | | | | | Transesophageal (Cardiac) | | | | | | | | | | | | | Transesophageal (Non Cardiac) | | | | | | | | | | | | | Transrectal | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | Transurethral | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | Peripheral Vascular | P | P | P | | P | P | P | | P | P: 5 | | | Laparoscopic | | | | | | | | | | | | | Musculo-skeletal Conventional | P | P | P | | P | P | P | | P | P: 5 | | | Musculo-skeletal Superficial [3] | P | P | P | | P | P | P | | P | P: 5 | | | Other (Urological) | | | | | | | | | | | | | Intended use: Diagnostic ultrasound Imaging or fluid flow analysis of human body as follows: | | | | | | Mode of Operations | | | | | | | Clinical Application | B | M | PWD | CWD | Color Doppler | Amplitude Doppler (AD) | Combined [4] | Color Velocity Mapping (TVM) | Harmonic Imaging (TEI) | Other (specify) | | | Ophthalmic | | | | | | | | | | | | | Fetal | | | | | | | | | | | | | Abdominal | | | | | | | | | | | | | Intraoperative (Abdominal) | | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | | Pediatric | E | E | E | | E | E | E | | E | E: 5 | | | Small Organs [1] | E | E | E | | E | E | E | | E | E: 5 | | | Neonatal Cephalic | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | Cardiac [2] | | | | | | | | | | | | | Transesophageal (Cardiac) |…
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...