K101605 · Esaote Europe B.V. · IYN · Aug 31, 2010 · Radiology
Device Facts
Record ID
K101605
Device Name
MYLABONE ULTRASOUND SYSTEM
Applicant
Esaote Europe B.V.
Product Code
IYN · Radiology
Decision Date
Aug 31, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
Esaote's MyLabOne is an arm-held ultrasound system used to perform diagnostic general ultrasound studies including Cardiac, Peripheral Vascular, Neonatal Cephalic, Adult Cephalic, Small Organ, Musculoskeletal (Conventional and Superficial), Abdominal, Fetal, Obstetrics/Gynaecology, Transvaginal, Transrectal, Pediatric, Intraoperative Abdominal, and Other: Urology. The system provides imaging for guidance of biopsy and imaging to assist in t the placement of needles and catheters in vascular or other anatomical structures as well as peripheral nerve blocks in Musculoskeletal applications.
Device Story
MyLabOne is a battery-operated, portable, arm-held ultrasound system. It processes acoustic signals from phased, convex, and linear array probes to generate diagnostic images. Imaging modes include B-Mode, M-Mode, Amplitude Doppler, Tissue Enhancement Imaging (TEI), Color Flow Mapping, and Pulse Wave Doppler. The device features an LCD touchscreen for user interaction, data entry, and annotation. It includes internal storage and supports external archiving via LAN/USB. Used in clinical settings by healthcare professionals for diagnostic studies and procedural guidance (biopsy, needle/catheter placement, nerve blocks). Output is displayed on the integrated screen to assist clinical decision-making. Benefits include portability and real-time visualization for minimally invasive procedures.
Clinical Evidence
Bench testing only. Verification and validation conducted per 21 CFR 820.30. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-1, and AIUM/NEMA UD-2/UD-3 standards for acoustic output and thermal/mechanical indices.
Technological Characteristics
Portable, battery-powered ultrasound system with LCD touchscreen. Supports phased, convex, and linear array probes. Imaging modes: B, M, PWD, Color Doppler, Amplitude Doppler, TEI, and XView. Connectivity: LAN/USB for data storage. Safety standards: IEC 60601-1, IEC 60601-2-37. Biocompatibility: ISO 10993-1.
Indications for Use
Indicated for diagnostic ultrasound imaging in patients requiring Cardiac, Peripheral Vascular, Neonatal/Adult Cephalic, Small Organ (breast, thyroid, testicles), Musculoskeletal, Abdominal, Fetal, OB/GYN, Transvaginal, Transrectal, Pediatric, Intraoperative Abdominal, and Urological exams. Includes needle/catheter placement guidance and peripheral nerve blocks.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{1}------------------------------------------------
- K101605
AUG 31 2010
510(k) Summary
The following 510(k) summary has been prepared pursuant to requirements specified in 21CFR 1807.92(a).
807.92(a)(1)
# Submitter Information
Esaote Europe B.V. Philipsweg 1 Maastricht 6227AJ The Netherlands
Contact Person:
### Allison Scott, RAC 317.569.9500 x106 ascott@ansongroup.com
Date:
March 22, 2010
807.92(a)(2)
Trade Name:
MyLabOne Ultrasound System
Common Name:
Ultrasound Imaging System
Classification Name
| e(s): | Ultrasonic pulse Doppler imaging system | 892.1550 |
|-------|-----------------------------------------|----------|
| | Ultrasonic pulsed echo imaging system | 892.1560 |
| | Diagnostic ultrasonic transducer | 892.1570 |
Classification Number:
90IYN, 90IYO, 90ITX
l
{2}------------------------------------------------
### 807.92(a)(3)
# Predicate Device(s)
| K083882 | MyLabFive | Esaote Europe B.V. |
|---------|-----------|--------------------|
| K092058 | NanoMaxx | SonoSite, Inc. |
Additional substantial equivalence information is provided in the following substantial equivalence comparison table.
807.92 (a)(4)
# Device Description
The MyLabOne is a battery operated, portable ultrasound system used to perform diagnostic general ultrasound studies. Its primary modes of operation are: B-Mode, M-Mode, Amplitude Doppler (AD), Tissue Enhancement Imaging (TEI), Color Flow Mapping (CFM) and Pulse Wave Doppler.
The MyLabOne is equipped with a LCD Color Display. The LCD Display includes touch screen technology for a simple and intuitive activation of functions and data entry for patient information and screen annotations.
The MyLabOne can drive phased (PA), convex (CA) and linear array (LA) probes. The MyLabOne is equipped with an internal Hard Disk and with an optional external DVD-RW disk drive that can be used for image storage. Data can also be stored directly to external archiving media (hard-disk, PC, server) via a LAN/USB port. Optional accessory devices available for the MyLabOne include a monochrome or color page printer, a desk stand (for use on a desk or cart) and a mobile trolley equipped with four swiveling wheels and peripheral holder.
{3}------------------------------------------------
#### 807.92(a)(5)
### Intended Use(s)
Esaote's MyLabOne is an arm-held ultrasound system used to perform diagnostic general ultrasound studies including Cardiac, Peripheral Vascular, Neonatal Cephalic, Adult Cephalic, Small Organ, Musculoskeletal (Conventional and Superficial), Abdominal, Fetal, Obstetrics/Gynaecology, Transvaginal, Transrectal, Pediatric, Intraoperative Abdominal, and Other: Urology. The system provides imaging for guidance of biopsy and imaging to assist in t the placement of needles and catheters in vascular or other anatomical structures as well as peripheral nerve blocks in Musculoskeletal applications.
807.92(a)(6)
### Technological Characteristics
Esaote believes that the MyLabOne is substantial equivalent to Esaote's MyLabFive product (K083882) and SonoSite's Nanomaxx product (K092058).
- · Clinical uses for which the MyLabOne is designed are equivalent to those cleared for the Esaote MyLabFive and the SonoSite NanoMaxx.
- · The MyLabOne, the Esaote MyLabFive and the SonoSite NanoMaxx are designed to meet the IEC60601-1 and the IEC60601-2-37 safety requirements.
- · The MyLabOne, the Esaote MyLabFive and the SonoSite NanoMaxx provide an Acoustic Output Display feature per AIUM / NEMA standards, with equivalent Ispta and MI maximal values.
- · The MyLabOne, the Esaote MyLabFive and the SonoSite NanoMaxx provide a similar measurements and analysis package, with equal accuracy and precision.
- · The MyLabOne, the Esaote MyLabFive and the SonoSite NanoMaxx have digital storage capabilities, including Network connectivity.
- · The MyLabOne, the Esaote MyLabFive and the SonoSite NanoMaxx are designed to be powered by battery when no main power is available.
- · The MyLabOne image modes are available on other FDA cleared ultrasound systems. for instance the Esaote MyLabFive and Sonosite NanoMaxx.
3
{4}------------------------------------------------
807.92(b)(1)
# Non-Clinical Testing
Verification and validation tests have been conducted in accordance with design controls per CFR 820.30.
The system has been designed to meet the following standards:
- . IEC 60601-I
- IEC 60601-1-2 .
- IEC 60601-2-37 .
- ISO 10993-1 .
- AIUM/NEMA UD-3 Standard for Real Time Display of Thermal and Mechanical . Acoustic Output Indices on Diagnostic Ultrasound Equipment
- AIUM/NEMA UD-2 Acoustic Output Measurement Standard for Diagnostic . Ultrasound .
. .. .. . .
{5}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract image of an eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Esaote Europe B.V. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services 1394 25th Street NW BUFFALO MN 55313
2010 3 1
Re: K101605
Trade/Device Name: MyLabOne Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: August 18, 2010 Received: August 19, 2010
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the MyLabOne, as described in your premarket notification:
### Transducer Model Number
{6}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
If you have any questions regarding the content of this letter, please contact Shahram Vaezy at (301) 796-6242.
Sincerely yours.
signature
Jonald St. Pierre Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure(s)
{7}------------------------------------------------
510(k) Number (if known): |<{ 0(lof(005
Device Name: MyLabOne
Indications for Use:
Esaote's MyLabOne is an arm-held ultrasound system used to perform diagnostic general ultrasound studies including Cardiac, Peripheral Vascular, Neonatal Cephalic, Adult Cephalic, Small Organ, Musculoskeletal (Conventional and Superficial), Abdominal, Fetal, Obstetrics/Gynaecology, Transvaginal, Transrectal, Pediatric, Intraoperative Abdominal and Other: Urology.
The system provides imaging for guidance of biopsy and imaging to assist in the placement of needles and catheters in vascular or other anatomical structures as well as peripheral nerve blocks in Musculoskeletal applications.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
D.m.d. JHL
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K101605
MyLabOne Response to FDA Deficiencies
{8}------------------------------------------------
### MyLabOne
| | | Mode of Operations | | | | | | | | |
|-------------------------------------------------------------|---|--------------------|-----|-----|------------------|----------------------|-----------------|-----|----------------------------------------|-----------------|
| Clinical Application | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Combined<br>[3] | TVM | Tissue<br>Enhancement<br>Imaging (TEI) | Other (specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal/Obstetrics/Oyn | N | N | N | N | N | N | N | | N | 4 |
| Abdominal | N | N | N | N | N | N | N | | N | 4 |
| Intraoperative (Abdominal) | N | N | N | N | N | N | N | | N | 4 |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | N | N | N | N | N | N | N | | N | 4 |
| Small Organ [1] | N | N | N | N | N | N | N | | N | 4 |
| Neonatal Cephalic | N | N | N | N | N | N | N | | N | 4 |
| Adult Cephalic | N | N | N | N | N | N | N | | N | 4 |
| Cardiac [2] | N | N | N | N | N | N | N | | N | 4 |
| Transesophageal (Cardiac) | | | | | | | | | | |
| Transesophageal (Non Cardiac) | | | | | | | | | | |
| Transrectal | N | N | N | N | N | N | N | | N | 4 |
| Transvaginal | N | N | N | N | N | N | N | | N | 4 |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular (including<br>vascular access) | N | N | N | N | N | N | N | | N | 4 |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal Conventional<br>(including Nerve Blocking) | N | N | N | N | N | N | N | | N | 4 |
| Musculo-skeletal Superficial<br>(including Nerve Blocking) | N | N | N | N | N | N | N | | N | 4 |
| Other (Urological) | N | N | N | N | N | N | N | | N | 4 |
N=new indication; P=previously cleared by FDA; E= added under Appendix E Additional Comments:
- (1) Small Organs includes Breast, Thyroid and Testicles (2) Cardiac is Adult and Pediatric
- [3] Combined modes are: B + M + PW + CFM + PD
- XVicw
િનો
Prescription Use Only (Per 21 CFR 801 Part D) Concurrence of CDRH, Office of In Vitro Diagnostics Devices (OIVD)
(Division Sign-Off)
Division of Radiological Devices
Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
510K. K101605
{9}------------------------------------------------
### SL3323
| Clinical Application | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Combined<br>[3] | TVM | Tissue<br>Enhancement<br>Imaging (TEI) | Other (specify) |
|-------------------------------------------------------------|---|---|-----|-----|------------------|----------------------|-----------------|-----|----------------------------------------|-----------------|
| Ophthalmic | | | | | | | | | | |
| Fetal/Obstetrics/Gyn | | | | | | | | | | |
| Abdominal | N | N | N | | N | N | N | | N | 4 |
| Intraoperative (Abdominal) | N | N | N | | N | N | N | | N | 4 |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | N | N | N | | N | N | N | | N | 4 |
| Small Organ [1] | N | N | N | | N | N | N | | N | 4 |
| Neonatal Cephalic | N | N | N | | N | N | N | | N | 4 |
| Adult Cephalic | | | | | | | | | | |
| Cardiac [2] | | | | | | | | | | |
| Transesophageal (Cardiac) | | | | | | | | | | |
| Transesophageal (Non Cardiac) | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular (including<br>vascular access) | N | N | N | | N | N | N | | N | 4 |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal Conventional<br>(including Nerve Blocking) | N | N | N | | N | N | N | | N | 4 |
| Musculo-skeletal Superficial<br>(including Nerve Blocking) | N | N | N | | N | N | N | | N | 4 |
| Other (Urological) | | | | | | | | | | |
The SL3323 probe is to be cleared via
this submssion
N=new indication; P=previously cleared by FDA; E= added under Appendix E Additional Comments:
- (1) Small Organs includes Breast, Thyroid and Testicles [기 Cardiac is Adult and Pediatric [3] Combined modes are: B + M + PW + CFM + PD (4) XView
Prescription Use Only (Per 21 CFR 801 Part D) Concurrence of CDRH, Office of In Vitro Diagnostics Devices (OIVD)
ﯿ
.
(Division Sign-Off)
(Division of Radiological Devices Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
Collice of In Vitro Diagnostic Device Evaluation and Safety
K101605
510K.
{10}------------------------------------------------
### SL3116
| | Mode of Operations | | | | | | | | | |
|-------------------------------------------------------------|--------------------|---|-----|-----|------------------|----------------------|-----------------|-----|----------------------------------------|-----------------|
| Clinical Application | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Combined<br>[3] | TVM | Tissue<br>Enhancement<br>Imaging (TEI) | Other (specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal / Obstetrics / Gyn | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (Abdominal) | N | N | N | | N | N | N | | N | 4 |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | N | N | N | | N | N | N | | N | 4 |
| Small Organ [1] | N | N | N | | N | N | N | | N | 4 |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac [2] | | | | | | | | | | |
| Transesophageal (Cardiac) | | | | | | | | | | |
| Transesophageal (Non Cardiac) | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular (including<br>vascular access) | N | N | N | | N | N | N | | N | 4 |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal Conventional<br>(including Nerve Blocking) | N | N | N | | N | N | N | | N | 4 |
| Musculo-skeletal Superficial<br>(including Nerve Blocking) | N | N | N | | N | N | N | | N | 4 |
| Other (Urological) | | | | | | | | | | |
The SL3116 probe is to be cleared via
this submssion
N=new indication; P=previously cleared by FDA; E= added under Appendix E Additional Comments:
- [1] [2]
(3)
[4]
Small Organs includes Breast, Thyroid and Testicles Cardiac is Adult and Pediatric Combined modes are: B + M + PW + CFM + PD XView
> Prescription Use Only (Per 21 CFR 801 Part D) Concurrence of CDRH, Office of In Vitro Diagnostics Devices (OIVD)
(Division Sign-Off) Division of Radiological De Office of In ce Evaluation and Safety
510K K101605
{11}------------------------------------------------
### SC3121
| | Mode of Operations | | | | | | | | | |
|-------------------------------------------------------------|--------------------|---|-----|-----|------------------|----------------------|--------------|-----|----------------------------------------|-----------------|
| Clinical Application | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Combined [3] | TVM | Tissue<br>Enhancement<br>Imaging (TEI) | Other (specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal/Obstetrics/Gyn | N | N | N | | N | N | N | | N | 4 |
| Abdominal | N | N | N | | N | N | N | | N | 4 |
| Intraoperative (Abdominal) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | N | N | N | | N | N | N | | N | 4 |
| Small Organ [1] | | | | | | | | | | |
| Neonatal Cephalic | N | N | N | | N | N | N | | N | 4 |
| Adult Cephalic | | | | | | | | | | |
| Cardiac [2] | N | N | N | | N | N | N | | N | 4 |
| Transesophageal (Cardiac) | | | | | | | | | | |
| Transesophageal (Non Cardiac) | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular (including<br>vascular access) | N | N | N | | N | N | N | | N | 4 |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal Conventional<br>(including Nerve Blocking) | N | N | N | | N | N | N | | N | 4 |
| Musculo-skeletal Superficial<br>(including Nerve Blocking) | N | N | N | | N | N | N | | N | 4 |
| Other (Urological) | N | N | N | | N | N | N | | N | 4 |
The SC3121 probe is to be cleared via
this submssion
Nenew indication; P=previously cleared by FDA; E= added under Appendix E Additional Comments:
- (1) Small Organs includes Breast, Thyroid and Testicles (기 Cardiac is Adult and Pediatric (ર) Combined modes are: B + M + PW + CFM + PD
XView
- |4)
Prescription Use Only (Per 21 CFR 801 Part D) Concurrence of CDRH, Office of In Vitro Diagnostics Devices (OIVD)
(Division Sign-Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
510K.K101665
{12}------------------------------------------------
### SC3123
| | Mode of Operations | | | | | | | | | |
|-------------------------------------------------------------|--------------------|---|-----|-----|------------------|----------------------|-----------------|-----|----------------------------------------|-----------------|
| Clinical Application | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Combined<br>[3] | TVM | Tissue<br>Enhancement<br>Imaging (TEI) | Other (specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal/Obstetrics/Gyn | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (Abdominal) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | N | N | N | N | N | N | N | | N | 4 |
| Small Organ [1] | N | N | N | N | N | N | N | | N | 4 |
| Neonatal Cephalic | N | N | N | N | N | N | N | | N | 4 |
| Adult Cephalic | | | | | | | | | | |
| Cardiac [2] | N | N | N | N | N | N | N | | N | 4 |
| Transesophageal (Cardiac) | | | | | | | | | | |
| Transesophageal (Non Cardiac) | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular (including<br>vascular access) | N | N | N | N | N | N | N | | N | 4 |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal Conventional<br>(including Nerve Blocking) | N | N | N | N | N | N | N | | N | 4 |
| Musculo-skeletal Superficial<br>(including Nerve Blocking) | N | N | N | N | N | N | N | | N | 4 |
| Other (Urological) | | | | | | | | | | |
The SC3123 probe is to be cleared via
this submssion
[1]
[기
(3)
(ব)
N=new indication; P=previously cleared by FDA; E= added under Appendix E Additional Comments:
- Small Organs includes Breast, Thyroid and Testicles Cardiac is Adult and Pediatric Combined modes are: B + M + PW + CFM + PD XView
Prescription Use Only (Per 21 CFR 801 Part D) Concurrence of CDRH. Office of In Vitro Diagnostics Devices (OIVD)
S.D.R.
(Division Sign-Off)
Division of Radiological Devices Office of in Vitro Diagnostic Devices
Office of in Vitro Diagnostic Device Evaluation and Safety
510K Ki6605
{13}------------------------------------------------
#### SC3421
| | | Mode of Operations | | | | | | | | |
|-------------------------------------------------------------|---|--------------------|-----|-----|------------------|----------------------|-----------------|-----|----------------------------------------|-----------------|
| Clinical Application | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Combined<br>[3] | TVM | Tissue<br>Enhancement<br>Imaging (TEI) | Other (specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal/Obstetrics/Gyn | N | N | N | | N | N | N | | N | 4 |
| Abdominal | N | N | N | | N | N | N | | N | 4 |
| Intraoperative (Abdominal) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | N | N | N | | N | N | N | | N | 4 |
| Small Organ [1] | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac [2] | | | | | | | | | | |
| Transesophageal (Cardiac) | | | | | | | | | | |
| Transesophageal (Non Cardiac) | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular (including<br>vascular access) | N | N | N | | N | N | N | | N | 4 |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal Conventional<br>(including Nerve Blocking) | N | N | N | | N | N | N | | N | 4 |
| Musculo-skeletal Superficial<br>(including Nerve Blocking) | N | N | N | | N | N | N | | N | 4 |
| Other (Urological) | N | N | N | | N | N | N | | N | 4 |
The SC3421 probe is to be cleared via
this submasion
N=new indication; P=previously cleared by FDA; E= added under Appendix E Additional Comments:
.
- (1) Small Organs includes Breast, Thyroid and Testicles Cardiac is Adult and Pediatric (2) (3) Combined modes are: B + M + PW + CFM + PD (4) XVicw
Prescription Use Only (Per 21 CFR 801 Part D) Concurrence of CDRH, Office of In Vitro Diagnostics Devices (OIVD)
(Drision Sign-Off) (Simsion of Radiological Devices Office of the Diagnostic Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K K16605
{14}------------------------------------------------
#### SE3123
| | Mode of Operations | | | | | | | | | |
|-------------------------------------------------------------|--------------------|---|-----|-----|------------------|----------------------|-----------------|-----|----------------------------------------|-----------------|
| Clinical Application | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Combined<br>[3] | TVM | Tissue<br>Enhancement<br>Imaging (TEI) | Other (specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal/Obstetrics/Gyn | N | N | N | N | N | N | N | | N | 4 |
| Abdominal | | | | | | | | | | |
| Intraoperative (Abdominal) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ [1] | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac [2] | | | | | | | | | | |
| Transesophageal (Cardiac) | | | | | | | | | | |
| Transesophageal (Non Cardiac) | | | | | | | | | | |
| Transrectal | N | N | N | N | N | N | N | | N | 4 |
| Transvaginal | N | N | N | N | N | N | N | | N | 4 |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular (including<br>vascular access) | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal Conventional<br>(including Nerve Blocking) | | | | | | | | | | |
| Musculo-skeletal Superficial<br>(including Nerve Blocking) | | | | | | | | | | |
| Other (Urological) | N | N | N | N | N | N | N | | N | 4 |
The SE3123 probe is to be cleared via
this submssion
Nenew indication; P=previously cleared by FDA; E= added under Appendix E Additional Comments:
- (1) Small Organs includes Breast, Thyroid and Testicles (2) Cardiac is Adult and Pediatric (3) Combined modes are: B + M + PW + CFM + PD ' [4] XVicw
Prescription Use Only (Per 21 CFR 801 Part D) Concurrence of CDRH, Office of In Vitro Diagnostics Devices (OIVD)
Signature
(Division Sign-Off) Division of Radiological Devices Office of In Vitro Diagnostic Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K K116605
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.