RS80A Diagnostic Ultrasound System

K151663 · Samsung Medison Co., Ltd. · IYN · Aug 12, 2015 · Radiology

Device Facts

Record IDK151663
Device NameRS80A Diagnostic Ultrasound System
ApplicantSamsung Medison Co., Ltd.
Product CodeIYN · Radiology
Decision DateAug 12, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1550
Device ClassClass 2
AttributesPediatric, 3rd-Party Reviewed

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
Thyroid lesion analysis
Breast lesion analysis
Tissue stiffness estimation
Lesion analysis

Indications for Use

The RS80A Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body. The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel.

Device Story

RS80A is a mobile, software-controlled diagnostic ultrasound system. It acquires ultrasound data via various transducers and displays it in modes including B-mode, Color/Power Doppler, PW/CW Spectral Doppler, Harmonic imaging, Tissue Doppler, 3D/4D imaging, and Elastoscan. The system allows operators to perform anatomical measurements and analysis. It is used by healthcare professionals in clinical settings to aid diagnosis. The system displays real-time mechanical and thermal indices. New software features include Natural Vue (3D rendering), S-Shearwave (tissue stiffness), S-Fusion (image registration), Virtual Track (needle tracking), S-Detect (lesion analysis), and E-Thyroid/E-Breast (elastography). These tools assist clinicians in visualizing structures, assessing tissue properties, and guiding procedures like biopsies. The device benefits patients by providing non-invasive diagnostic imaging and fluid flow analysis.

Clinical Evidence

Bench testing only. The device was evaluated for acoustic output, biocompatibility, thermal, electrical, electromagnetic, and mechanical safety, conforming to standards including IEC 60601-1, IEC 60601-2-37, NEMA UD-2, NEMA UD-3, and ISO 10993-1.

Technological Characteristics

Mobile ultrasound system; supports linear, curved, endocavity, phased array, and static probes (1.0–20.0 MHz). Materials tested to ISO 10993-1. Connectivity includes USB wireless LAN adapter. Software features include B-mode, M-mode, Doppler (PW/CW/Color/Power), 3D/4D, Elastoscan, S-Shearwave, S-Fusion, S-Detect, and needle guidance. Sterilization/disinfection follows standard transducer protocols.

Indications for Use

Indicated for diagnostic ultrasound imaging and fluid analysis in fetal/obstetrics, abdominal, gynecology, pediatric, small organ, neonatal/adult cephalic, trans-rectal, trans-vaginal, musculoskeletal, urology, cardiac, and peripheral vessel applications.

Regulatory Classification

Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of a stylized caduceus, which is a traditional symbol of medicine, with three figures facing to the right. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 12, 2015 SAMSUNG MEDISON CO., LTD. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313 Re: K151663 Trade/Device Name: RS80A Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO and ITX Dated: July 29, 2015 Received: July 30, 2015 Dear Mr. Job: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Michael O'Hara For Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K151663 Device Name RS80A Diagnostic Ultrasound System Indications for Use (Describe) The RS80A Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body. The clinical applications include: Fetal/Obstetics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ 510(k) No.: ## Device Name: RS80A Diagnostic Ultrasound System Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | | | Mode of Operation (*includes simultaneous B-mode) | | | | | | |---------------------------|-------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|-------------------------------------------| | General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) | | Ophthalmic | Ophthalmic | | | | | | | | | | Fetal/Obstetrics (See Note 3) | P | P | P | | P | Note 1 | Notes 2, 4, 7, 8, 9, 11, 14 | | | Abdominal (See Note 10) | P | P | P | | P | Note 1 | Notes 2, 4, 5, 6, 7, 8, 9, 11, 12, 14, 15 | | | Intra-operative (See Note 6) | | | | | | | | | | Intra-operative (Neuro.) | | | | | | | | | Fetal Imaging<br>& Other | Laparoscopic | | | | | | | | | | Pediatric | P | P | P | N | P | Note 1 | Notes 8, 9, 11 | | | Small Organ (See Note 5) | P | P | P | | P | Note 1 | Note 2, 5, 6, 7, 8, 9, 11, 12, 14 | | | Neonatal Cephalic | P | P | P | | P | Note 1 | Notes 8, 9, 11 | | | Adult Cephalic | P | P | P | P | P | Note 1 | Notes 7 | | | Trans-rectal | P | P | P | | P | Note 1 | Notes 2, 7, 8, 9, 11, 12 | | | Trans-vaginal | P | P | P | | P | Note 1 | Notes 2, 7, 8, 9, 11, 12 | | | Trans-urethral | | | | | | | | | | Trans-esoph. (non-Cardiac) | | | | | | | | | | Musculo-skel. (Convent.) | P | P | P | | P | Note 1 | Note 2, 5, 6, 7, 8, 9, 11, 14 | | | Musculo-skel. (Superfic.) | P | P | P | | P | Note 1 | Note 2, 5, 6, 7, 8, 9, 11, 14 | | | Intra-luminal | | | | | | | | | | Other (See Note 13) | P | P | P | | P | Note 1 | Notes 2, 7, 8, 9, 12 | | | Cardiac Adult | P | P | P | P | P | Note 1 | Notes 4, 7 | | Cardiac | Cardiac Pediatric | P | P | P | P | P | Note 1 | Notes 4, 7 | | | Trans-esophageal (Cardiac) | | | | | | | | | | Other (spec.) | | | | | | | | | Peripheral<br>Vessel | Peripheral vessel | P | P | P | P | P | Note 1 | Note 2, 5, 6, 7, 8, 9, 11, 14 | | | Other (spec.) | | | | | | | | N= new indication; P= previously cleared by FDA K141620; E= added under Appendix E #### Additional Comments: Color Doppler includes Power (Amplitude) Doppler Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+PPI, B+CW, B+C+PW, B+PD+PW, B+PPH+PW, B+PPI+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C - Note 2: Includes imaging for guidance of biopsy Note 3: Includes infertility monitoring of follicle development Note 4: Color M-mode Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients Note 6: Abdominal organs and peripheral vessel Note 7: Tissue Harmonic Imaging (THI) - Note 8: 3D imaging Note 9: Spatial Compound Imaging Note 10: Includes Renal, Gynecology/Pelvis Note 11: Panoramic imaging Note 12: ElastoScan Note 13: Includes Urology/Prostate Note 14: S-Fusion Note 15: S-Shearwave Concurrence of Center for Devices and Radiological Health (CDRH) Prescription Use (Per 21 CFR 801.109) {4}------------------------------------------------ 510(k) No.: ## Device Name: CA1-7A for use with RS80A Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | | Mode of Operation (*includes simultaneous B-mode) | | | | | | | | |---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|--------------------------|--| | General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) | | | Ophthalmic | Ophthalmic | | | | | | | | | | | Fetal/Obstetrics (See Note 3) | P | P | P | P | P | Note 1 | Notes 2,4,7,8,9,11,14 | | | | Abdominal (See Note 10) | P | P | P | P | P | Note 1 | Notes 2,6,7,8,9,11,14,15 | | | | Intra-operative (See Note 6) | | | | | | | | | | | Intra-operative (Neuro.) | | | | | | | | | | Fetal Imaging<br>& Other | Laparoscopic | | | | | | | | | | | Pediatric | | | | | | | | | | | Small Organ (See Note 5) | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | Trans-rectal | | | | | | | | | | | Trans-vaginal | | | | | | | | | | | Trans-urethral | | | | | | | | | | | Trans-esoph. (non-Cardiac) | | | | | | | | | | | Musculo-skel. (Convent.) | | | | | | | | | | | Musculo-skel. (Superfic.) | | | | | | | | | | | Intra-luminal | | | | | | | | | | | Other (See Note 13) | | | | | | | | | | | Cardiac Adult | | | | | | | | | | Cardiac | Cardiac Pediatric | | | | | | | | | | | Trans-esophageal (Cardiac) | | | | | | | | | | | Other (spec.) | | | | | | | | | | Peripheral<br>Vessel | Peripheral vessel | | | | | | | | | | | Other (spec.) | | | | | | | | | N= new indication; P= previously cleared by FDA K141620; E= added under Appendix E #### Additional Comments: Color Doppler includes Power (Amplitude) Doppler Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+TD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPI+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C - Note 2: Includes imaging for guidance of biopsy Note 3: Includes infertility monitoring of follicle development Note 4: Color M-mode Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients Note 6: Abdominal organs and peripheral vessel Note 7: Tissue Harmonic Imaging (THI) Note 8: 3D imaging Note 9: Spatial Compound Imaging Note 10: Includes Renal, Gynecology/Pelvis Note 11: Panoramic imaging Note 12: ElastoScan Note 13: Includes Urology/Prostate Note 14: S-Fusion {5}------------------------------------------------ 510(k) No.: ### Device Name: C2-6 for use with RS80A Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | | Mode of Operation (*includes simultaneous B-mode) | | | | | | | | |---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|-------------------------|--| | General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) | | | Ophthalmic | Ophthalmic | | | | | | | | | | Fetal Imaging<br>& Other | Fetal/Obstetrics (See Note 3) | P | P | P | | P | Note 1 | Notes 2, 4, 7, 8, 9, 11 | | | | Abdominal (See Note 10) | P | P | P | | P | Note 1 | Notes 2, 6, 7, 8, 9, 11 | | | | Intra-operative (See Note 6) | | | | | | | | | | | Intra-operative (Neuro.) | | | | | | | | | | | Laparoscopic | | | | | | | | | | | Pediatric | | | | | | | | | | | Small Organ (See Note 5) | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | Trans-rectal | | | | | | | | | | | Trans-vaginal | | | | | | | | | | | Trans-urethral | | | | | | | | | | | Trans-esoph. (non-Cardiac) | | | | | | | | | | | Musculo-skel. (Convent.) | | | | | | | | | | | Musculo-skel. (Superfic.) | | | | | | | | | | | Intra-luminal | | | | | | | | | | | Other (See Note 13) | | | | | | | | | | Cardiac | Cardiac Adult | | | | | | | | | | | Cardiac Pediatric | | | | | | | | | | | Trans-esophageal (Cardiac) | | | | | | | | | | | Other (spec.) | | | | | | | | | | Peripheral<br>Vessel | Peripheral vessel | | | | | | | | | | | Other (spec.) | | | | | | | | | N= new indication; P= previously cleared by FDA K141620; E= added under Appendix E #### Additional Comments: Color Doppler includes Power (Amplitude) Doppler Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+PPI, B+CW, B+C+PW, B+PD+PW, B+PPH+PW, B+PPI+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C - Note 2: Includes imaging for guidance of biopsy Note 3: Includes infertility monitoring of follicle development Note 4: Color M-mode Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients Note 6: Abdominal organs and peripheral vessel Note 7: Tissue Harmonic Imaging (THI) - Note 8: 3D imaging Note 9: Spatial Compound Imaging Note 10: Includes Renal, Gynecology/Pelvis Note 11: Panoramic imaging Note 12: ElastoScan Note 13: Includes Urology/Prostate Note 14: S-Fusion Note 15: S-Shearwave {6}------------------------------------------------ 510(k) No.: ### Device Name: CF4-9 for use with RS80A Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | | | Mode of Operation (*includes simultaneous B-mode) | | | | | | |---------------------------|-------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|------------------| | General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) | | Ophthalmic | Ophthalmic | | | | | | | | | Fetal Imaging<br>& Other | Fetal/Obstetrics (See Note 3) | | | | | | | | | | Abdominal (See Note 10) | P | P | P | | P | Note 1 | Notes 8, 9, 11 | | | Intra-operative (See Note 6) | | | | | | | | | | Intra-operative (Neuro.) | | | | | | | | | | Laparoscopic | | | | | | | | | & Other | Pediatric | P | P | P | | P | Note 1 | Notes 8, 9, 11 | | | Small Organ (See Note 5) | | | | | | | | | | Neonatal Cephalic | P | P | P | | P | Note 1 | Notes 8, 9, 11 | | | Adult Cephalic | | | | | | | | | | Trans-rectal | | | | | | | | | | Trans-vaginal | | | | | | | | | | Trans-urethral | | | | | | | | | | Trans-esoph. (non-Cardiac) | | | | | | | | | | Musculo-skel. (Convent.) | | | | | | | | | | Musculo-skel. (Superfic.) | | | | | | | | | | Intra-luminal | | | | | | | | | | Other (See Note 13) | | | | | | | | | Cardiac | Cardiac Adult | | | | | | | | | | Cardiac Pediatric | | | | | | | | | | Trans-esophageal (Cardiac) | | | | | | | | | | Other (spec.) | | | | | | | | | Peripheral<br>Vessel | Peripheral vessel | P | P | P | | P | Note 1 | Notes 8, 9, 11 | | | Other (spec.) | | | | | | | | N= new indication; P= previously cleared by FDA K141620; E= added under Appendix E #### Additional Comments: Color Doppler includes Power (Amplitude) Doppler Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+PPI, B+CW, B+C+PW, B+PD+PW, B+PPH+PW, B+PPI+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C - Note 2: Includes imaging for guidance of biopsy Note 3: Includes infertility monitoring of follicle development Note 4: Color M-mode Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients Note 6: Abdominal organs and peripheral vessel Note 7: Tissue Harmonic Imaging (THI) Note 8: 3D imaging Note 9: Spatial Compound Imaging Note 10: Includes Renal, Gynecology/Pelvis Note 11: Panoramic imaging Note 12: ElastoScan Note 13: Includes Urology/Prostate Note 14: S-Fusion Note 15: S-Shearwave Image /page/6/Figure/24 description: The image contains text that states "Concurrence of Center for Devices and Radiological Health (CDRH)". The image also states "Prescription Use (Per 21 CFR 801.109)". The text is centered on the page. {7}------------------------------------------------ 510(k) No.: ### Device Name: E3-12A for use with RS80A Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | | | Mode of Operation (*includes simultaneous B-mode) | | | | | | | | |---------------------------|-------------------------------|--|---------------------------------------------------|---|-----|-----|-------------------|----------------------|----------------------|--| | General<br>(Track I only) | Specific<br>(Tracks I & III) | | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) | | | Ophthalmic | Ophthalmic | | | | | | | | | | | | Fetal/Obstetrics (See Note 3) | | P | P | P | P | P | Note 1 | Notes 2, 7, 8, 9 | | | | Abdominal (See Note 10) | | P | P | P | P | P | Note 1 | Notes 2, 7, 8, 9, 12 | | | Fetal Imaging<br>& Other | Intra-operative (See Note 6) | | | | | | | | | | | | Intra-operative (Neuro.) | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Pediatric | | | | | | | | | | | | Small Organ (See Note 5) | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Trans-rectal | | P | P | P | P | P | Note 1 | Notes 2, 7, 8, 9, 12 | | | | Trans-vaginal | | P | P | P | P | P | Note 1 | Notes 2, 7, 8, 9, 12 | | | | Trans-urethral | | | | | | | | | | | | Trans-esoph. (non-Cardiac) | | | | | | | | | | | | Musculo-skel. (Convent.) | | | | | | | | | | | | Musculo-skel. (Superfic.) | | | | | | | | | | | | Intra-luminal | | | | | | | | | | | | Other (See Note 13) | | P | P | P | P | P | Note 1 | Notes 2, 7, 8, 9, 12 | | | Cardiac | Cardiac Adult | | | | | | | | | | | | Cardiac Pediatric | | | | | | | | | | | | Trans-esophageal (Cardiac) | | | | | | | | | | | | Other (spec.) | | | | | | | | | | | Peripheral<br>Vessel | Peripheral vessel | | | | | | | | | | | | Other (spec.) | | | | | | | | | | N= new indication; P= previously cleared by FDA K141620; E= added under Appendix E #### Additional Comments: Color Doppler includes Power (Amplitude) Doppler Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+PPI, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPI+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C - Note 2: Includes imaging for guidance of biopsy Note 3: Includes infertility monitoring of follicle development Note 4: Color M-mode Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients Note 6: Abdominal organs and peripheral vessel Note 7: Tissue Harmonic Imaging (THI) Note 8: 3D imaging Note 9: Spatial Compound Imaging Note 10: Includes Renal, Gynecology/Pelvis Note 11: Panoramic imaging Note 12: ElastoScan Note 13: Includes Urology/Prostate Note 14: S-Fusion Note 15: S-Shearwave > Concurrence of Center for Devices and Radiological Health (CDRH) Prescription Use (Per 21 CFR 801.109) Indication for use {8}------------------------------------------------ 510(k) No.: ## Device Name: L3-12A for use with RS80A Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | | Mode of Operation (*includes simultaneous B-mode) | | | | | | | |---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|-----------------------------------| | General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) | | Ophthalmic | Ophthalmic | | | | | | | | | Fetal Imaging<br>& Other | Fetal/Obstetrics (See Note 3) | | | | | | | | | | Abdominal (See Note 10) | | | | | | | | | | Intra-operative (See Note 6) | | | | | | | | | | Intra-operative (Neuro.) | | | | | | | | | | Laparoscopic | | | | | | | | | | Pediatric | | | | | | | | | | Small Organ (See Note 5) | P | P | P | | P | Note 1 | Note 2, 5, 6, 7, 8, 9, 11, 12, 14 | | | Neonatal Cephalic | | | | | | | | | | Adult Cephalic | | | | | | | | | | Trans-rectal | | | | | | | | | | Trans-vaginal | | | | | | | | | | Trans-urethral | | | | | | | | | | Trans-esoph. (non-Cardiac) | | | | | | | | | | Musculo-skel. (Convent.) | P | P | P | | P | Note 1 | Note 2, 5, 6, 7, 8, 9, 11, 14 | | | Musculo-skel. (Superfic.) | P | P | P | | P | Note 1 | Note 2, 5, 6, 7, 8, 9, 11, 14 | | | Intra-luminal | | | | | | | | | | Other (See Note 13) | | | | | | | | | Cardiac | Cardiac Adult | | | | | | | | | | Cardiac Pediatric | | | | | | | | | | Trans-esophageal (Cardiac) | | | | | | | | | | Other (spec.) | | | | | | | | | Peripheral<br>Vessel | Peripheral vessel | P | P | P | | P | Note 1 | Note 2, 5, 6, 7, 8, 9, 11, 14 | | | Other (spec.) | | | | | | | | N= new indication; P= previously cleared by FDA K141620; E= added under Appendix E #### Additional Comments: Color Doppler includes Power (Amplitude) Doppler Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+TD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPI+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C - Note 2: Includes imaging for guidance of biopsy Note 3: Includes infertility monitoring of follicle development Note 4: Color M-mode Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients Note 6: Abdominal organs and peripheral vessel Note 7: Tissue Harmonic Imaging (THI) Note 8: 3D imaging Note 9: Spatial Compound Imaging Note 10: Includes Renal, Gynecology/Pelvis Note 11: Panoramic imaging Note 12: ElastoScan Note 13: Includes Urology/Prostate Note 14: S-Fusion {9}------------------------------------------------ 510(k) No.: ### Device Name: L5-13 for use with RS80A Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | | Mode of Operation (*includes simultaneous B-mode) | | | | | | | | |---------------------------|----------------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|-------------------------------|--| | General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) | | | Ophthalmic | Ophthalmic | | | | | | | | | | | Fetal/Obstetrics ( <i>See Note 3</i> ) | | | | | | | | | | | Abdominal ( <i>See Note 10</i> ) | | | | | | | | | | | Intra-operative ( <i>See Note 6</i> ) | | | | | | | | | | | Intra-operative (Neuro.) | | | | | | | | | | Fetal Imaging<br>& Other | Laparoscopic | | | | | | | | | | | Pediatric | | | | | | | | | | | Small Organ ( <i>See Note 5</i> ) | P | P | P | | P | Note 1 | Note 2, 5, 6, 8, 7, 9, 11, 12 | | | | Neonatal Cephalic | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | Trans-rectal | | | | | | | | | | | Trans-vaginal | | | | | | | | | | | Trans-urethral | | | | | | | | | | | Trans-esoph. (non-Cardiac) | | | | | | | | | | | Musculo-skel. (Convent.) | P | P | P | | P | Note 1 | Note 2, 5, 6, 7, 8, 9, 11 | | | | Musculo-skel. (Superfic.) | P | P | P | | P | Note 1 | Note 2, 5, 6, 7, 8, 9, 11 | | | | Intra-luminal | | | | | | | | | | | Other ( <i>See Note 13</i> ) | | | | | | | | | | Cardiac | Cardiac Adult | | | | | | | | | | | Cardiac Pediatric | | | | | | | | | | | Trans-esophageal (Cardiac) | | | | | | | | | | | Other (spec.) | | | | | | | | | | Peripheral<br>Vessel | Peripheral vessel | P | P | P | | P | Note 1 | Note 2, 5, 6, 7, 8, 9, 11 | | | | Other (spec.) | | | | | | | | | N= new indication; P= previously cleared by FDA K141620; E= added under Appendix E #### Additional Comments: Color Doppler includes Power (Amplitude) Doppler Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+PPI, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C - Note 2: Includes imaging for guidance of biopsy Note 3: Includes infertility monitoring of follicle development Note 4: Color M-mode Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients Note 6: Abdominal organs and peripheral vessel Note 7: Tissue Harmonic Imaging (THI) - Note 8: 3D imaging - Note 9: Spatial Compound Imaging - Note 10: Includes Renal, Gynecology/Pelvis - Note 11: Panoramic imaging - Note 12: ElastoScan - Note 13: Includes Urology/Prostate Note 14: S-Fusion Note 15: S-Shearwave {10}------------------------------------------------ 510(k) No.: ### Device Name: L7-16 for use with RS80A Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | | | Mode of Operation (*includes simultaneous B-mode) | | | | | | | |--------------------------|-------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|---------------------------|--| | General | Specific | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) | | | (Track I only) | (Tracks I & III) | | | | | | | | | | Ophthalmic | Ophthalmic | | | | | | | | | | | Fetal/Obstetrics (See Note 3) | | | | | | | | | | | Abdominal (See Note 10) | | | | | | | | | | | Intra-operative (See Note 6) | | | | | | | | | | | Intra-operative (Neuro.) | | | | | | | | | | Fetal Imaging<br>& Other | Laparoscopic | | | | | | | | | | | Pediatric | | | | | | | | | | | Small Organ (See Note 5) | P | P | P | P | P | Note 1 | Note 2, 5, 6, 7, 8, 9, 11 | | | | Neonatal Cephalic | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | Trans-rectal | | | | | | | | | | | Trans-vaginal | | | | | | | | | | | Trans-urethral | | | | | | | | | | | Trans-esoph. (non-Cardiac) | | | | | | | | | | | Musculo-skel. (Convent.) | P | P | P | P | P | Note 1 | Note 2, 5, 6, 7, 8, 9, 11 | | | | Musculo-skel. (Superfic.) | P | P | P | P | P | Note 1 | Note 2, 5, 6, 7, 8, 9, 11 | | | | Intra-luminal | | | | | | | | | | | Other (See Note 13) | | | | | | | | | | | Cardiac Adult | | | | | | | | | | Cardiac | Cardiac Pediatric | | | | | | | | | | | Trans-esophageal (Cardiac) | | | | | | | | | | | Other (spec.) | | | | | | | | | | Peripheral | Peripheral vessel | P | P | P | P | P | Note 1 | Note 2, 5, 6, 7, 8, 9, 11 | | | Vessel | Other (spec.) | | | | | | | | | N= new indication; P= previously cleared by FDA K141620; E= added under Appendix E #### Additional Comments: Color Doppler includes Power (Amplitude) Doppler Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+TD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPI+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C - Note 2: Includes imaging for guidance of biopsy Note 3: Includes infertility monitoring of follicle development Note 4: Color M-mode Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients Note 6: Abdominal organs and peripheral vessel Note 7: Tissue Harmonic Imaging (THI) Note 8: 3D imaging Note 9: Spatial Compound Imaging Note 10: Includes Renal, Gynecology/Pelvis Note 11: Panoramic imaging Note 12: ElastoScan Note 13: Includes Urology/Prostate Note 14: S-Fusion {11}------------------------------------------------ 510(k) No.: ### Device Name: LA2-9A for use with RS80A Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | | Mode of Operation (*includes simultaneous B-mode) | | | | | | | |---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|---------------------------| | General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) | | Ophthalmic | Ophthalmic | | | | | | | | | | Fetal/Obstetrics (See Note 3) | | | | | | | | | | Abdominal (See Note 10) | P | P | P | | P | Note 1 | Note 2, 5, 6, 7, 8, 9, 11 | | | Intra-operative (See Note 6) | | | | | | | | | | Intra-operative (Neuro.) | | | | | | | | | Fetal Imaging<br>& Other | Laparoscopic | | | | | | | | | | Pediatric | | | | | | | | | | Small Organ (See Note 5) | P | P | P | | P | Note 1 | Note 2, 5, 6, 7, 8, 9, 11 | | | Neonatal Cephalic | | | | | | | | | | Adult Cephalic | | | | | | | | | | Trans-rectal | | | | | | | | | | Trans-vaginal | | | | | | | | | | Trans-urethral | | | | | | | | | | Trans-esoph. (non-Cardiac) | | | | | | | | | | Musculo-skel. (Convent.) | P | P | P | | P | Note 1 | Note 2, 5, 6, 7, 8, 9, 11 | | | Musculo-skel. (Superfic.) | P | P | P | | P | Note 1 | Note 2, 5, 6, 7, 8, 9, 11 | | | Intra-luminal | | | | | | | | | | Other (See Note 13) | | | | | | | | | | Cardiac Adult | | | | | | | | | Cardiac | Cardiac Pediatric | | | | | | | | | | Trans-esophageal (Cardiac) | | | | | | | | | | Other (spec.) | | | | | | | | | Peripheral<br>Vessel | Peripheral vessel | P | P | P | | P | Note 1 | Note 2, 5, 6, 7, 8, 9, 11 | | | Other (spec.) | | | | | | | | N= new indication; P= previously cleared by FDA K141620; E= added under Appendix E #### Additional Comments: Color Doppler includes Power (Amplitude) Doppler Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+PPI, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C - Note 2: Includes imaging for guidance of biopsy Note 3: Includes infertility monitoring of follicle development Note 4: Color M-mode Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients Note 6: Abdominal organs and peripheral vessel Note 7: Tissue Harmonic Imaging (THI) - Note 8: 3D imaging Note 9: Spatial Compound Imaging - Note 10: Includes Renal, Gynecology/Pelvis Note 11: Panoramic imaging - Note 12: ElastoScan - Note 13: Includes Urology/Prostate Note 14: S-Fusion {12}------------------------------------------------ 510(k) No.: ### Device Name: PE2-4 for use with RS80A Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | | | Mode of Operation (*includes simultaneous B-mode) | | | | | | |--------------------------|-------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|------------------| | General | Specific | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) | | (Track I only) | (Tracks I & III) | | | | | | | | | Ophthalmic | Ophthalmic | | | | | | | | | | Fetal/Obstetrics (See Note 3) | | | | | | | | | | Abdominal (See Note 10) | P | P | P | P | P | Note 1 | Note 7 | | | Intra-operative (See Note 6) | | | | | | | | | | Intra-operative (Neuro.) | | | | | | | | | Fetal Imaging<br>& Other | Laparoscopic | | | | | | | | | | Pediatric | | | | | | | | | | Small Organ (See Note 5) | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | Adult Cephalic | P | P | P | P | P | Note 1 | Note 7 | | | Trans-rectal | | | | | | | | | | Trans-vaginal | | | | | | | | | | Trans-urethral | | | | | | | | | | Trans-esoph. (non-Cardiac) | | | | | | | | | | Musculo-skel. (Convent.) | | | | | | | | | | Musculo-skel. (Superfic.) | | | | | | | | | | Intra-luminal | | | | | | | | | | Other (See Note 13) | | | | | | | | | | Cardiac Adult | P | P | P | P | P | Note 1 | Note 4, 7 | | Cardiac | Cardiac Pediatric | P | P | P | P | P | Note 1 | Note 4, 7 | | | Trans-esophageal (Cardiac) | | | | | | | | | | Other (spec.) | | | | | | | | | Peripheral<br>Vessel | Peripheral vessel | | | | | | | | | | Other (spec.) | | | | | | | | N= new indication; P= previously cleared by FDA K141620; E= added under Appendix E #### Additional Comments: Color Doppler includes Power (Amplitude) Doppler Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+PPI, B+CW, B+C+PW, B+PD+PW, B+PPH+PW, B+PPI+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C - Note 2: Includes imaging for guidance of biopsy Note 3: Includes infertility monitoring of follicle development Note 4: Color M-mode Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients Note 6: Abdominal organs and peripheral vessel - Note 8: 3D imaging - Note 9: Spatial Compound Imaging - Note 10: Includes Renal, Gynecology/Pelvis - Note 11: Panoramic imaging - Note 12: ElastoScan - Note 13: Includes Urology/Prostate Note 14: S-Fusion - Note 15: S-Shearwave {13}------------------------------------------------ 510(k) No.: ### Device Name: PM1-6A for use with RS80A Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | | Mode of Operation (*includes simultaneous B-mode) | | | | | | | |---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------| | General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) | | Ophthalmic | Ophthalmic | | | | | | | | | | Fetal/Obstetrics (See Note 3) | | | | | | | | | | Abdominal (See Note 10) | P | P | P | P | P | Note 1 | Note 7 | | | Intra-operative (See Note 6) | | | | | | | | | | Intra-operative (Neuro.) | | | | | | | | | Fetal Imaging<br>& Other | Laparoscopic | | | | | | | | | | Pediatric | | | | | | | | | | Small Organ (See Note 5) | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | Adult Cephalic | P | P | P | P | P | Note 1 | Note 7 | | | Trans-rectal | | | | | | | | | | Trans-vaginal | | | | | | | | | | Trans-urethral | | | | | | | | | | Trans-esoph. (non-Cardiac) | | | | | | | | | | Musculo-skel. (Convent.) | | | | | | | | | | Musculo-skel. (Superfic.) | | | | | | | | | | Intra-luminal | | | | | | | | | | Other (See Note 13) | | | | | | | | | Cardiac | Cardiac Adult | P | P | P | P | P | Note 1 | Note 4, 7 | | | Cardiac Pediatric | P | P | P | P | P | Note 1 | Note 4, 7 | | | Trans-esophageal (Cardiac) | | | | | | | | | | Other (spec.) | | | | | | | | | Peripheral<br>Vessel | Peripheral vessel | | | | | | | | | | Other (spec.) | | | | | | | | N= new indication; P= previously cleared by FDA K141620; E= added under Appendix E #### Additional Comments: Color Doppler includes Power (Amplitude) Doppler Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C - Note 2: Includes imaging for guidance of biopsy Note 3: Includes infertility monitoring of follicle development Note 4: Color M-mode Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients Note 6: Abdominal organs and peripheral vessel Note 7: Tissue Harmonic Imaging (THI) Note 8: 3D imaging Note 9: Spatial Compound Imaging Note 10: Includes Renal, Gynecology/Pelvis Note 11: Panoramic imaging Note 12: ElastoScan Note 13: Includes Urology/Prostate Note 14: S-Fusion Note 15: S-Shearwave > Concurrence of Center for Devices and Radiological Health (CDRH) Prescription Use (Per 21 CFR 801.109) {14}------------------------------------------------ 510(k) No.: ### Device Name: V5-9 for use with RS80A Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | | Mode of Operation (*includes simultaneous B-mode) | | | | | | | |---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------| | General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) | | Ophthalmic | Ophthalmic | | | | | | | | | | Fetal/Obstetrics (See Note 3) | P | P | P | P | P…
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...