Browse hierarchy Radiology (RA) Subpart B — Diagnostic Devices 21 CFR 892.1550 Product Code IYN K060087 — ACCUVIX V7 DIAGNOSTIC ULTRASOUND SYSTEM
ACCUVIX V7 DIAGNOSTIC ULTRASOUND SYSTEM
K060087 · Medison Co., Ltd. · IYN · Jan 24, 2006 · Radiology
Device Facts
Record ID K060087
Device Name ACCUVIX V7 DIAGNOSTIC ULTRASOUND SYSTEM
Applicant Medison Co., Ltd.
Product Code IYN · Radiology
Decision Date Jan 24, 2006
Decision SESE
Submission Type Traditional
Regulation 21 CFR 892.1550
Device Class Class 2
Attributes Pediatric, 3rd-Party Reviewed
Intended Use
The ACCUVIX V7 system and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body. The clinical applications include: General, Obstetrics, Gynecology, Extremity, Urology, Breast. Renal, Vascular, Pediatric, Abdomen, Cardiac, Neonatal, Fetal Echo, Carotid, Varicose, Fetal Abdominal, Small organ, Intra-operative, Cephalic, Muscular-skeletal, Trans-rectal, Trans-vaginal, Contrast Agent, Transesophageal and Peripheral-vascular applications.
Device Story
ACCUVIX V7 is a mobile, software-controlled diagnostic ultrasound system. It acquires ultrasound data via various transducers and displays images in 2D, M-mode, Color/Power Doppler, Harmonic imaging, and PW Spectral Doppler. It also supports 3D imaging using mechanical scan probes. Used in clinical settings by healthcare professionals for anatomical measurement and diagnostic analysis. The system provides real-time acoustic output display (mechanical and thermal indices). Output is viewed on a CRT display to assist in clinical decision-making and patient diagnosis.
Clinical Evidence
Bench testing only. The device complies with UL 60601-1, CSA C22.2 No. 601.1, IEC 60601-2-37, NEMA UD 2-2004, NEMA UD 3-2004, IEC 61157, and ISO 10993 for biocompatibility. No clinical data was required for this 510(k) clearance.
Technological Characteristics
Mobile diagnostic ultrasound system. Sensing principle: ultrasonic pulsed echo/Doppler. Energy source: electrical. Connectivity: standard ultrasound system interfaces. Biocompatibility: ISO 10993. Safety standards: IEC 60601-2-37, NEMA UD 2/3. Software-controlled. Transducers: various models (e.g., C1-4EC, P2-4AC, 3D2-5EH).
Indications for Use
Indicated for diagnostic ultrasound imaging and fluid flow analysis in adult, pediatric, and neonatal patients across clinical applications including fetal, abdominal, intra-operative, pediatric, small organ, neonatal/adult cephalic, trans-rectal, trans-vaginal, trans-esophageal, musculoskeletal, cardiac, and peripheral vascular. Includes guidance for biopsy and infertility monitoring.
Regulatory Classification
Identification An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
ACCUVIX XO Diagnostic Ultrasound System (K052911 )
SA8000 Diagnostic Ultrasound System (K013627 )
Reference Devices
Related Devices
K070813 — ACCUVIX V10 DIAGNOSTIC ULTRASOUND SYSTEM · Medison Co., Ltd. · Apr 10, 2007
K052911 — ACCUVIX XQ DIAGNOSTIC ULTRASOUND SYSTEM · Medison Co., Ltd. · Oct 31, 2005
K092159 — ACCUVIX V20 DIAGNOSTIC ULTRASOUND SYSTEM · Medison Co., Ltd. · Jul 28, 2009
K063580 — SONOACE X8 DIAGNOSTIC ULTRASOUND SYSTEM · Medison Co., Ltd. · Dec 14, 2006
K080800 — ACCUVIX V20 DIAGNOSTIC ULTRASOUND SYSTEM · Medison Co., Ltd. · Apr 7, 2008
Submission Summary (Full Text)
{0}------------------------------------------------
K060087
jan 2 4 2006
510(k) Premarket Notification
ACCUVIX V7Diagnostic Ultrasound System
## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR, Part 807, Subpart E, Section 807.92.
### 1. Submitter's Information: 21 CFR 807.92(a)(1)
Medison Co. Ltd. 997-10, Daechi-dong, Gangnam-gu, Seoul 135-280, Korea
### Contact Person:
Mr. Kyung-Am, Shim Regulatory Affairs Manager
| Telephone: | 82.2.2194.1381 |
|------------|--------------------|
| Facsimile: | 82.2.2194.1399 |
| Email: | kashim@medison.com |
### Data Prepared:
December 15, 2005
### 2. Name of the device:
Common/Usual Name:
Diagnostic Ultrasound System and Accessories
Proprietary Name:
ACCUVIX V7 Diagnostic Ultrasound System
| Classification Names: | FR Number | Product Code |
|------------------------------------------|-----------|--------------|
| Ultrasonic Pulsed Doppler Imaging System | 892.1550 | 90-IYN |
| Ultrasound Pulsed Echo Imaging System | 892.1560 | 90-IYO |
| Diagnostic Ultrasound Transducer | 892.1570 | 90-ITX |
### 3. Identification of the predicate or legally marketed device:
K052911, 10/31/2005, ACCUVIX XO Diagnostic Ultrasound System K013627, 11/16/2001, SA8000 Diagnostic Ultrasound System
### 4. Device Description:
The ACCUVIX V7 is a general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as 2D mode, M mode, Color Doppler imaging, Power Doppler imaging, Harmonic imaging, and PW Spectral Doppler mode on the CRT display. It also provides the 3D imaging mode using the 3D probe in the Mechanical scan mode.
{1}------------------------------------------------
The ACCUVIX V7 has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed. The system also provides for the measurement of anatomical structures and for analysis packages that provide information used for clinical diagnostic purposes by competent health care professionals.
The ACCUVIX V7 has been designed to meet the following product safety standards:
- UL 60601-1. Safety requirements for Medical Equipment
- CSA C22.2 No. 601.1, Safety requirements for Medical Equipment
- IEC60601-2-37, Diagnostic Ultrasound Safety Standards
- EN/IEC60601-1, Safety requirements for Medical Equipment
- EN/IEC60601-1-2, EMC requirements for Medical Equipment y
- NEMA UD 2-2004 Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment
- NEMA UD 3-2004 Standard for Real Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment
- IEC 61157, Declaration of the acoustic output
- ISO10993, Biocompatibility
### 5. Intended Uses:
The ACCUVIX V7 system and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body.
The clinical applications include:
General, Obstetrics, Gynecology, Extremity, Urology, Breast. Renal, Vascular, Pediatric, Abdomen, Cardiac, Neonatal, Fetal Echo, Carotid, Varicose, Fetal Abdominal, Small organ, Intra-operative, Cephalic, Muscular-skeletal, Trans-rectal, Trans-vaginal, Contrast Agent, Transesophageal and Peripheral-vascular applications.
### 6. Technological Characteristics:
The ACCUVIX V7 is substantially equivalent to the ACCUVIX XQ Diagnostic Ultrasound System, cleared via K052911, and the SA8000 Diagnostic Ultrasound System, cleared via K013627. All systems transmit ultrasonic energy into patients, then perform post processing of received echoes to generate on-screen display of anatomic structures and fluid flow within the body. All system allow for specialized measurements of structures and flow, and calculations.
END of 510(K) Summary
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and features the department's name around the perimeter. In the center is a stylized image of an eagle or bird-like figure, composed of three curved lines that suggest wings or feathers. The logo is black and white.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 2 4 2006
Medison Co., Ltd. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25" Street NW BUFFALO MN 55313
Re: K060087
Trade Name: ACCUVIX V7 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Number: 21 CFR 892.1560 Regulatory Name: Ultrasonic pulsed echo imaging system Regulatory Number: 21 CFR 892.1570 Regulatory Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: January 10, 2006 Received: January 12, 2006
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the ACCUVIX V7 Diagnostic Ultrasound System, as described in your premarket notification:
Image /page/2/Picture/11 description: The image is a circular seal with the letters "FDA" in the center. Above the letters, the numbers "1XX-XXX" are printed. Below the letters, the word "Centennial" is printed. The seal is surrounded by a dotted line and some text that is difficult to read.
Jing Public .
{3}------------------------------------------------
### Transducer Model Number
| C1-4EC | MPT4-7AC |
|----------|----------|
| C2-6IC | P2-4AC |
| C3-7IM | P3-5AC |
| C5-2EL | P3-7AC |
| EC4-9ES | 3D2-5EH |
| EC4-9IS | 3D4-7EK |
| L4-7EL | 3D5-8EK |
| L6-12IS | VNW6-12 |
| L8-15IS | CW2.0 |
| LS5-13EC | CW4.0 |
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212.
Sincerely yours,
Nancyc Brogdon
Nancy C. Brogdon
Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{5}------------------------------------------------
### Section 4.3 INDICATIONS FOR USE
## DIAGNOSTIC ULTRASOUND INDICATIONS STATEMENT
| 510(k) No.: | | | | | | | | | |
|--------------------------------------------------------------------------------------------------|------------------------------|---|---|-----|-----|-------------------|---------------------------------------------------|--------------------|--|
| Device Name:<br>ACCUVIX V7 Diagnostic Ultrasound System | | | | | | | | | |
| Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | | | |
| | Clinical Application | | | | | | Mode of Operation (*includes simultaneous B-mode) | | |
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) | |
| Ophthalmic | Ophthalmic | | | | | | | | |
| | Fetal (See Note 3) | N | N | N | N | N | Note 1 | Notes 2, 7, 8 | |
| | Abdominal | N | N | N | N | N | Note 1 | Notes 2, 7, 8 | |
| | Intra-operative (See Note 6) | N | N | N | N | N | Note 1 | Note 8, 9 | |
| | Intra-operative (Neuro.) | N | N | N | N | N | Note 1 | Note 8, 9 | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | | |
| | Pediatric | N | N | N | N | N | Note 1 | Note 2,4,5,6,7,8,9 | |
| | Small Organ (See Note 5) | N | N | N | N | N | Note 1 | Note 2,5,6,8,9 | |
| | Neonatal Cephalic | N | N | N | N | N | Note 1 | | |
| | Adult Cephalic | N | N | N | N | N | Note 1 | Note 4, 7 | |
| | Trans-rectal | N | N | N | N | N | Note 1 | Note 2, 3, 8 | |
| | Trans-vaginal | N | N | N | N | N | Note 1 | Note 2, 3, 8 | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Cardiac) | N | N | N | N | N | | Note 7, 8 | |
| | Musculo-skel. (Convent.) | N | N | N | N | N | Note 1 | Note 2,5,6,8,9 | |
| | Musculo-skel. (Superfic.) | N | N | N | N | N | Note 1 | Note 2,5,6,8,9 | |
| | Intra-luminal | | | | | | | | |
| | Other (spec.) | | | | | | | | |
| | Cardiac Adult | N | N | N | N | N | Note 1 | Note 4, 7 | |
| Cardiac | Cardiac Pediatric | N | N | N | N | N | Note 1 | Note 4, 7 | |
| | Trans-esophageal (Cardiac) | N | N | N | N | N | Note 1 | Note 7, 8 | |
| | Other (spec.) | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | N | N | N | N | N | Note 1 | Note 5, 6, 9 | |
| | Other (spec.) | | | | N | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
#### Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B/M, B/PW Doppler, B/Color Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler/ B/Color Doppler/Color M
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Note 9: Panoramic imaging
Nancy c Brogdon
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K060082
{6}------------------------------------------------
510(k) No .:
| Device Name:<br>C1-4EC for use with ACCUVIX V7 | | | | | | | | |
|-----------------------------------------------------------------------------------------------------|------------------------------------------------------|---|---|-----|-----|-------------------|----------------------|------------------|
| Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:<br>Intended Use: | | | | | | | | |
| | Clinical Application<br>Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| General<br>(Track I only) | Ophthalmic | | | | | | | |
| Ophthalmic | Fetal (See Note 3) | N | N | N | | N | Note 1 | Notes 2, 7, 8 |
| | Abdominal | N | N | N | | N | Note 1 | Notes 2, 7, 8 |
| | Intra-operative (See Note 6) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | |
| | Pediatric | N | N | N | | N | Note 1 | Notes 2, 7, 8 |
| | Small Organ (See Note 5) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
#### Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B/M, B/PW Doppler, B/Color Doppler/PW Doppler/PW Doppler, B/Power Doppler/PW Doppler,
B/Color Doppler/Color M
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Note 9: Panoramic imaging
Nancy C Beardon
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number R060087
{7}------------------------------------------------
510(k) No.:
| Clinical Application | | | Mode of Operation (*includes simultaneous B-mode) | | | | | |
|---------------------------|------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal (See Note 3) | P | P | P | | P | Note 1 | Notes 2, 7, 8 |
| | Abdominal | P | P | P | | P | Note 1 | Notes 2, 7, 8 |
| | Intra-operative (See Note 6) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | | P | Note 1 | Notes 2, 7, 8 |
| | Small Organ (See Note 5) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by K052911 and K032329; E= added under Appendix E
### Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B/M, B/PW Doppler, B/Color Doppler, B/Color Doppler, B/Power Doppler/PW Doppler,
- B/Color Doppler/Color M
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Note 9: Panoramic imaging
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|---------------------------------------------------------------|---------|
| Prescription Use (Per 21 CFR 801.109) | |
| Nancy Brogdon | |
| (Division Sign-Off) | |
| Division of Reproductive, Abdominal, and Radiological Devices | |
| 510(k) Number | K060087 |
{8}------------------------------------------------
510(k) No.:
| C3-7IM for use with ACCUVIX V7 | | | | | | | | |
|------------------------------------------------------------------------------------|------------------------------|---|---|-----|-----|----------------|-------------------|---------------|
| Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | | |
| Clinical Application | | B | M | PWD | CWD | Color Doppler* | Combined* (Spec.) | Other (Spec.) |
| General<br>(Track I only) | Specific<br>(Tracks I & III) | | | | | | | |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal (See Note 3) | P | P | P | P | P | Note 1 | Notes 2, 7, 8 |
| | Abdominal | P | P | P | | P | Note 1 | Notes 2, 7, 8 |
| | Intra-operative (See Note 6) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | | P | Note 1 | Notes 2, 7, 8 |
| | Small Organ (See Note 5) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by K052911 and K032329; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B/M, B/PW Doppler, B/Color Doppler, B/Color Doppler, B/Power Doppler/PW Doppler, B/Color Doppler/Color M
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
ち:り火, Number
Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Note 9: Panoramic imaging
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
{9}------------------------------------------------
510(k) No.:
| C5-2EL for use with ACCUVIX V7<br>Device Name: | | | | | | | | |
|-----------------------------------------------------------------------------------------------------|------------------------------|---|---|-----|-----|-------------------|----------------------|------------------|
| Intended Use:<br>Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | | |
| Clinical Application | | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| General<br>(Track I only) | Specific<br>(Tracks I & III) | | | | | | | |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal (See Note 3) | P | P | P | | P | Note 1 | Notes 2, 7, 8 |
| | Abdominal | P | P | P | | P | Note 1 | Notes 2, 7, 8 |
| | Intra-operative (See Note 6) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | | P | Note 1 | Notes 2, 7, 8 |
| | Small Organ (See Note 5) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by K052911; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B/M, B/PW Doppler, B/Color Doppler, B/Color Doppler, B/Power Doppler/PW Doppler/PW Doppler,
B/Color Doppler/Color M
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Note 9: Panoramic imaging
Nancy Brogdon
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________
{10}------------------------------------------------
510(k) No .:
| Clinical Application | | | Mode of Operation (*includes simultaneous B-mode) | | | | | |
|-----------------------------------------------------------------------------------------------------|------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|------------------|
| Device Name:<br>EC4-9ES for use with ACCUVIX V7 | | | | | | | | |
| Intended Use:<br>Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | | |
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging<br>& Other | Fetal (See Note 3) | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (See Note 6) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (See Note 5) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | P | P | P | | P | Note 1 | Note 2, 8 |
| | Trans-vaginal | P | P | P | | P | Note 1 | Note 2, 8 |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | |
| | Other (spec.) | | | | | | | |
| | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by K013627 and K002185; E= added under Appendix E
#### Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B/M, B/PW Doppler, B/Color Doppler, B/Color Doppler, B/Power Doppler/PW Doppler,
- B/Color Doppler/Color M
Note 2: Includes imaging for guidance of biopsy
- Note 3: Includes infertility monitoring of follicle development
Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Note 9: Panoramic imaging
Nancy C brigdon
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices K060087
510(k) Number
{11}------------------------------------------------
510(k) No.:
| Device Name: | EC4-9IS for use with ACCUVIX V7 | | | | | | | |
|--------------------------|------------------------------------------------------------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------|
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | |
| Clinical Application | | Mode of Operation (*includes simultaneous B-mode) | | | | | | |
| General | Specific | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| (Track I only) | (Tracks I & III) | | | | | | | |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal (See Note 3) | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (See Note 6) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (See Note 5) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | P | P | P | | P | Note 1 | Note 2, 8 |
| | Trans-vaginal | P | P | P | | P | Note 1 | Note 2, 8 |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by K052911 and K032329; E= added under Appendix E
#### Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B/M, B/PW Doppler, B/Color Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler,
- B/Color Doppler/Color M
- Note 2: Includes imaging for guidance of biopsy
- Note 3: Includes infertility monitoring of follicle development
Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
- Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Note 9: Panoramic imaging
Nancy C. Burgdon
(Division Sign-Off)
Division of Reproductive and Radiological Devices 5 10(k) Number _______________________________________________________________________________________________________________________________________________________________
{12}------------------------------------------------
510(k) No.:
| Device Name: L4-7EL for use with ACCUVIX V7 | | | | | | | | |
|--------------------------------------------------------------------------------------------------|------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|------------------|
| Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | | |
| Clinical Application | | | Mode of Operation (*includes simultaneous B-mode) | | | | | |
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal (See Note 3) | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (See Note 6) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | |
| | Pediatric | N | N | N | | N | Note 1 | Note 2, 5, 6, 9 |
| | Small Organ (See Note 5) | N | N | N | | N | Note 1 | Note 2, 5, 6, 9 |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | |
| | Musculo-skel. (Convent.) | N | N | N | | N | Note 1 | Note 2, 5, 6, 9 |
| | Musculo-skel. (Superfic.) | N | N | N | | N | Note 1 | Note 2, 5, 6, 9 |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | N | N | N | | N | Note 1 | Note 5, 6, 9 |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B/M, B/PW Doppler, B/Color Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler, B/Color Doppler/Color M
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
Note 4: Color M-mode
Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Note 9: Panoramic imaging
Nancy C. Bragdon
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K060087
{13}------------------------------------------------
| 510(k) No.:<br>L6-12IS for use with ACCUVIX V7<br>Device Name: | | | | | | | | |
|-----------------------------------------------------------------------------------------------------|------------------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------|
| Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:<br>Intended Use: | | | | | | | | |
| Clinical Application | | Mode of Operation (*includes simultaneous B-mode) | | | | | | |
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal (See Note 3)<br>Abdominal | | | | | | | |
| | Intra-operative (See Note 6) | | | | | | | |
| Fetal Imaging<br>& Other | Intra-operative (Neuro.)<br>Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | | P | Note 1 | Note 2, 5, 6, 9 |
| | Small Organ (See Note 5) | P | P | P | | P | Note 1 | Note 2, 5, 6, 9 |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | |
| | Musculo-skel. (Convent.) | P | P | P | | P | Note 1 | Note 2, 5, 6, 9 |
| | Musculo-skel. (Superfic.) | P | P | P | | P | Note 1 | Note 2, 5, 6, 9 |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | P | P | P | P | Note 1 | Note 5, 6, 9 |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by K052911; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B/M, B/PW Doppler, B/Color Doppler, B/Color Doppler, B/Power Doppler/PW Doppler, B/Color Doppler/Color M
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
S ( "x) Number _______________________________________________________________________________________________________________________________________________________________
Note 4: Color M-mode
Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Note 9: Panoramic imaging
(Division Sign. Division of Reproductive, Abdom…