ARTHORALIX 8500, MODELS 110-0199G1, 110-0199G2, 110-0199G3, 110-0199G4

K050255 · Gendex Dental Systems · EHD · Mar 17, 2005 · Dental

Device Facts

Record IDK050255
Device NameARTHORALIX 8500, MODELS 110-0199G1, 110-0199G2, 110-0199G3, 110-0199G4
ApplicantGendex Dental Systems
Product CodeEHD · Dental
Decision DateMar 17, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.1800
Device ClassClass 2
AttributesPediatric

Intended Use

The Orthoralix 8500 Panoramic Dental X-ray System is to be used as an extraoral source of x-rays for imaging of the dento-maxillo-facial area.

Device Story

The Orthoralix 8500 is a rotational panoramic dental X-ray system. It consists of a wall-mounted column, a motor-driven overhead assembly, an X-ray tubehead with DC power supply, and a cassette drive system for 15 x 30 cm film. During operation, the tubehead and cassette holder rotate around the patient's head; the X-ray beam is collimated by a slit diaphragm. Patient positioning is facilitated by three lasers, a bite guide, and head stabilizers. The system features a standard panoramic mode and a child-specific mode to reduce radiation dosage. Movements are microprocessor-controlled. The device is intended for use by dental professionals to produce radiographic images for clinical diagnosis of the dento-maxillo-facial area.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on technological characteristics and intended use.

Technological Characteristics

Rotational panoramic X-ray system; DC power supply; microprocessor-controlled motor-driven assembly; slit diaphragm collimation; 15 x 30 cm film cassette drive; wall-mounted column; laser-guided patient positioning system.

Indications for Use

Indicated for imaging of the dento-maxillo-facial area in patients requiring panoramic dental radiography. Includes a child-specific mode to reduce radiation dosage.

Regulatory Classification

Identification

An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K050255 Image /page/0/Picture/1 description: The image shows the word "GENDEX" in a bold, sans-serif font. The letters are all capitalized and black. The "E" and "X" are connected at the bottom, and there is a registered trademark symbol next to the "X". The background is white. Gendex Dental Systems es Plaines, IL 60018-1884 847 640 4800 Tel 847 640 4970 Fax www.gendex.com MAR 1 7 2035 510(k) Summary Statement for the Gendex Orthoralix 8500 Panoramic Dental X-ray System #### General Information I. | Submitter: | Gendex Dental Systems<br>901 West Oakton Street<br>Des Plaines, IL 60018 | |---------------------------|--------------------------------------------------------------------------| | Telephone: | (847) 640-4800 – Company number<br>(847) 640-4924 - Contact person | | Fax: | (847) 640-4970 | | Contact Person: | John R. Miller<br>Director, Quality Assurance and Regulatory Affairs | | Summary Preparation Date: | January 31, 2005 | #### II. Names | Device Name: | Orthoralix 8500 Panoramic Dental X-ray System | |------------------------------|-----------------------------------------------| | Primary Classification Name: | 90EHD - Extraoral Source X-ray System | ## III. Predicate Devices - Gendex Orthoralix 9200 . - . Planmeca DIMAX2 - . Gendex Orthoralix 9200 DDE - Gendex GX-9000 . ### IV. Product Description The Orthoralix 8500 is a system for rotational panoramic radiography of the dentomaxillo-facial area. The system consists of: - Column to be mounted to the wall . - Motor-driven overhead assembly with controls for patient postioning, setting of . technique factors and radiographic projection geometry. - Cassette drvie system with flat cassette for 15 x 30 cm film . - X-ray tubehead, with DC power supply ● - Remote control box and handswitch ● {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the word "GENDEX" in a bold, sans-serif font. The letters are all capitalized and black. The "E" is slightly larger than the other letters. There is a registered trademark symbol to the right of the "X". dex Dental Systems During a panoramic exposure, the x-ray tubehead and cassette holder move around the During a panoramic exposure, the x-ray tube is collimated by a slit diaphragm. The flat patient's head. The beam from the x-fay tabe is commans locks the radiation scattered from the patient. Patient positioning is simple, accurate and consistent by means of three postioning lasers, a bite guide and head stabilizers. During the exposure, the patient postioning lasers, a bitc guite and nead backless. All movements for the panoramic remains still with e tholorized components retater traiser traotors, which are microprocessor controlled. Along wit the Standard Panoramic mode, the Orthoralix 8500 provides a special child Along with the Bandard Fanolania me adiation dosage to the child, and confining the image to the dentition. # V. Indications for Use/Rationale for Substantial Equivalence The Orthoralix 8500 Panoramic Dental X-ray System is to be used as an extraoral source of x-rays for imaging of the dento-maxillo-facial area. It share the same indications for use, similar materials, design, operational, and functional It share the same includio for act, equivalent to the predicate devices listed in Section II of this summary. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge. Inside the circle is a stylized image of an eagle with three wavy lines extending from its head. The seal appears to be a government emblem, likely representing the U.S. Department of Health and Human Services. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. John R. Miller IMI. John IC: Nince Director, Quality Assurance & Regulatory Affairs Gendex Dental Systems 901 W. Oakton Street DES PLAINES IL 60018 Re: K050255 KU30233 Trade/Device Name: Orthoralix 8500 Panoramic Dental X-ray System Regulation Number: 21 CFR §872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: EHD Dated: February 2, 2005 Received: February 3, 2005 Dear Mr. Miller: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced we nave revewed your becases by the promotially equivalent (for the indications for use stated in above and have decemined the do rest cate devices marketed in interstate commerce prior to the enclosure) to regarly manelocal Device Amendments, or to devices that have been May 20, 1970, dic Clacinent dato of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a prematic approval application (PMA). You may, therefore, market the do not require approval of a promation approvisions of the Act. The general controls provisions of the Act device, subject to the general oculess prior, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. MAR 1 7 2005 If your device is classified (see above) into either class II (Special Controls) or class III (Premarket IT your device is classified (300 a00 ro) into controls. Existing major regulations affecting your Apploval), it thay be subject to sach acerate Regulations, Title 21, Parts 800 to 898. In addition, FDA device can be round minouncements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that Please of advisor mar PDT is issuation of levice complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Federal statues and regulations administrated to registration and listing (21 CFR Part 807); labeling ACL STEQuirements, moduling, but not a more a vertirements as set forth in the quality systems (QS) (21 CFR Fart 801); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) This and the submit and the finding of unction of unction of your device to a le This letter will allow you to begin marketing your vice as usences an your device to a legally premarket notification. The FDA innding of substantial equivales your device y premarket notification. The FDA inding of substantial equivalence of your device to marketed predicate device results in a classification for your device and thus, permits y proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), pleased If you desire specific advice for your device on our labeling regaration (1) virus and on the regulation number at the top of this letter: | 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |-----------------|----------------------------------|--------------| | 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 892.xxxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Also, please note the regulation entitled, "Misoranion on your responsibilities under the Act from the 807.97). You may obtain other general information on your responsibility. Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800 DIVIsion of Sman Manakaland of at its Internet address 058-2041 01 (51 gov/cdrh/industry/support/index.html. Sincerely yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ 510(k) Number (if known): Not Assigned Device Name: Orthoralix 8500 Panoramic Dental X-Ray System Indications for Use: The Orthoralix 8500 Panoramic Dental X-Ray System is to be used as an I he Orthoranx 8500 Fanoraline Delital Ar Bay 99 extraoral source of x-rays for imaging of the dento-maxilofacial area. (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21CFR 801.109) : OR Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ Nancy C. Brogdon (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________ (Optional Format 1-2-96) 0052
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...