The FONA Pan/Ceph device, comprising models FONA XPan DG Plus, FONA ART Plus, and FONA ART Plus C, is an extraoral source dental X-ray system intended to perform panoramic or cephalometric exams with production of diagnostic images in the dento-maxillo-facial region and in subregions, for general and pediatric dentistry, as well as carpal images.
Device Story
FONA Pan/Ceph is an extraoral dental X-ray system for panoramic and cephalometric imaging. It uses a rotating arm with an X-ray tube and a digital image receptor (CCD with CsI or CMOS with Cd(Zn)Te) to scan the patient's head. The device is operated by dental professionals in a clinical setting. Patient positioning is supported by laser lights, bite blocks, chin rests, and temple supports. The operator controls exposure via a manual hand-switch with dead-man functionality. An Ethernet connection transmits data to a side computer where software handles image acquisition, processing, display, and archiving. The system provides diagnostic images of the dento-maxillo-facial region and carpal structures, aiding clinicians in dental diagnosis and treatment planning.
Clinical Evidence
Bench testing only. Performance was evaluated using anthropomorphic phantoms for both panoramic and cephalometric modes. Image quality was assessed qualitatively against predicate devices. Technical performance metrics, including DQE (Detective Quantum Efficiency) and MTF (Modulation Transfer Function), were provided for the CCD and CMOS image receptors to demonstrate equivalence. Electrical safety and electromagnetic compatibility testing were conducted per IEC 60601-1 and IEC 60601-1-2.
Technological Characteristics
Extraoral X-ray system; anode voltage 61-85 kV; anode current 4-10 mA. Models use CCD (CsI scintillator) or CMOS (Cd(Zn)Te scintillator) digital receptors. Features motorized vertical carriage, motorized collimator, and cephalometric arm. Connectivity via Ethernet to external PC. Biocompatibility per ISO 10993-1:2010. Complies with IEC 60601-1 and IEC 60601-1-2.
Indications for Use
Indicated for patients requiring panoramic or cephalometric dental X-ray exams, including general and pediatric dentistry patients, and for carpal imaging.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, which is a traditional symbol of medicine.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 3, 2016
FONA Srl % Ms. Elisa Gusberti Quality and Regulatory Affairs Manager Via Galileo Galilei 11 Assago, Milan 20090 ITALY
Re: K161131 Trade/Device Name: FONA Pan/Ceph Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: MUH Dated: September 14, 2016 Received: September 16, 2016
Dear Ms. Gusberti:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D.'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K161131
# Device Name
FONA Pan/Ceph
Indications for Use (Describe)
The FONA Pan/Ceph device, comprising models FONA XPan DG Plus, FONA ART Plus, and FONA ART Plus C, is an extraoral source dental X-ray system intended to perform panoramic or cephalometric exams with production of diagnostic images in the dento-maxillo-facial region and in subregions, for general and pediatric dentistry, as well as carpal images.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------|------------------------------------------------------------|
| <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
PRAStaff@fda.hhs.gov
{4}------------------------------------------------
## 1 510(k) Summary
#### 1.1 Company and correspondent making the submission
| Name: | FONA Srl |
|------------------|-----------------------------------------------------------------------------------------------------------------|
| Address: | Via Galileo Galilei 11, 20090 Assago, Milan, Italy |
| Contact person: | Ms. Elisa Gusberti<br>Quality and Regulatory Affairs Manager |
| Contact details: | Phone +39 02 4571 2171<br>Mobile +39 344 2201285<br>Fax +39 02 4570 3385<br>E-mail elisa.gusberti@fonadental.it |
| Date prepared: | April 15, 2016 |
# 1.2 Device
| Device trade name: | FONA Pan/Ceph |
|--------------------|-----------------------------------------------------------------|
| Device models: | FONA XPan DG, FONA XPan DG Plus, FONA ART Plus, FONA ART Plus C |
| Regulation number: | 21CFR 872.1800 |
| Regulation name: | Extraoral source x-ray system |
| Classification: | II |
| Product code: | MUH |
#### 1.3 Predicate devices
The new FONA Pan/Ceph device is substantially equivalent to the following legally marketed predicate devices:
- PantOs (K031801) manufactured by Blue X Imaging Srl, the former company name of FONA Srl, primary . predicate device:
- Regulation name: Extraoral source X-ray system 트
- Product codes: MUH, EHD, MQB.
- ORTHOPHOS SL (K150217) manufactured by Sirona Dental Systems GmbH, which incorporates, for the ● panoramic and cephalometric modalities, the same types of image receptors used in the new FONA Pan/Ceph device:
- Requlation name: Computed Tomography X-ray system
- . Product code: OAS.
#### Indications for use 1.4
The FONA Pan/Ceph device, comprising models FONA XPan DG, FONA XPan DG Plus, FONA ART Plus, and FONA ART Plus C, is an extraoral source dental X-ray system intended to perform panoramic or cephalometric exams with production of diagnostic images in the dento-maxillo-facial region and in subregions, for general and pediatric dentistry, as well as carpal images.
#### 1.5 Device description
The FONA Pan/Ceph device is an X-ray equipment designed for:
- Dental panoramic radiography
- Cephalometry, when completed with lateral Ceph arm. .
The FONA Pan/Ceph device comprises four models:
- FONA XPan DG and FONA XPan DG Plus, with CCD type of image receptor with CsI scintillator
- FONA ART Plus, FONA ART Plus C, with CMOS type of image receptor with Cd(Zn)Te scintillator. ●
All the models allow for panoramic projections and two of them are extended with cephalometric functionality:
- . The models FONA XPan DG and FONA ART Plus feature panoramic projections only (type Pan Solo) and are thus equipped with a fixed image receptor mounted on the rotating arm.
{5}------------------------------------------------
- The models FONA XPan DG Plus and FONA ART Plus C feature panoramic and cephalometric projections ● (type Pan Ceph) and are equipped with a detachable image receptor to be placed on the rotating arm, for the panoramic modality, or on the cephalometric arm, for the cephalometric modality.
The control panel on the unit contains keys and signaling lights for a complete control of the operation and for the setting of the desired technique factors.
Class I LASER aiming lights support positioning of patient's head. which is stabilized through the use of bite blocks, chin rest, and, in case, temple supports.
The user controls the exposure using a manual hand-switch, implementing the dead man functionality.
An Ethernet connection cable allows the FONA Pan/Ceph device to interface with a computer for image acquisition, processing and storage.
The user may thus view and store the radiographic images using a software installed on a side computer with peripheral components such as a mouse, keyboard, and monitor. The software takes care of acquisition, processing, displaying, and archiving the image data.
A TWAIN interface is also available for interfacing to other existing programs with processing and storage capabilities, according to the user's needs.
#### 1.5.1 Scientific concept
The panoramic and cephalometric radiographs are based on traditional technologies which have been used for a long time.
#### 1.5.2 Physical and performance characteristics
The FONA Pan/Ceph device has a vertical carriage with a rotating arm plus a lateral arm for the cephalometric models, to which the power supply cord, the Ethernet cable and the exposure hand-switch are connected.
The equipment can be raised or lowered during patient positioning with the motorized movement along a column.
The rotating arm is equipped with the tube housing assembly, the housing for the image receptor and the control panel.
The arm carrying the Cephalostat and the housing for the digital sensor is fixed to the vertical carriage.
The panoramic radiographic image of the dental arch is made with a thin vertical X-ray beam out of the rotating arm which turns around and scans the head of the patient.
The cephalometric radiographic image, on the applicable models, is produced with a thin vertical X-ray beam that horizontally scans the head of the patient.
The FONA Pan/Ceph device features an anode voltage from 61 to 85 KV (at constant potential), with anode current from 4 to 10 mA (direct current, DC).
In panoramic modality the source to image receptor distance is of 51.3 cm (20.2″), and the exposure time ranges from 4.4 s up to 14.2 s, depending on the selected panoramic program.
In cephalometric modality, for the applicable models, the source to image receptor distance is of 165 cm (65"), and he exposure time is of 8 or of 10 s, depending on the cephalometric program.
Specific accessories like bite block, chin rest, nasal support, are provided to permit correct patient positioninq. A head rest made of two temporal resting bars is optionally provided.
The cephalometric models come with cephalostat with nasion reference and side bars with ear plugs to stabilize the head during exposure plus a hand-rest panel for the radiograph of the hand.
These patient positioning items are likely to enter in contact with the patient during the normal use of the FONA Pan/Ceph device. The use of standard hygienic protective sleeves (not included in the scope of supply) to avoid cross-contamination among patients, operators and other persons, is recommended.
The FONA Pan/Ceph device with its associated accessories fulfils the biocompatibility requirements of ISO 10993-1:2010.
{6}------------------------------------------------
### 1.6 Summary of technological characteristics
The new FONA Pan/Ceph device is directly derived from the PantOs (K031801), model PantOs DG xp, making use of:
- Same X-ray generator and tube housing assembly
- Same electronic boards for power control
- Same mechanical structure (carriage and column) and type of stepping motors for carriage movements ●
- . Same core of embedded software (FW) for control of mechanical movements, X-ray generation, service functionalities
- . Same LASER sources for aiming lights to support patient positioning.
The new FONA Pan/Ceph device incorporates changes to the PantOs device, model PantOs DG xp, comprising:
- Addition of a Ceph arm with scanning system driven by the same control unit
- . Addition of a motorized collimator driven by the same control unit
- Addition of motorization of the vertical travel of the carriage on the column for an easier pasitioning
- Adoption of up-to-grade image receptors based on CsI or Cd(Zn)Te high yield scintillators instead of the traditional Gd2O2S:Tb (Gadolinium oxide, Terbium doped, of LANEX™ screens by Kodak) less efficient (thus giving a less favorable signal to noise ratio).
The new FONA Pan/Ceph device incorporates the same basic features of the predicate device ORTHOPHOS SL (K150217), models ORTHOPHOS SL 2D and ORTHOPHOS SL 2D Ceph:
- Panoramic radiology with selected projections .
- . Cephalometric radiology with scanning modality
- . Use of CCD image receptor with CsI scintillator
- Use of CMOS image receptor with Cd(Zn)Te scintillator ●
- . Use of motorized collimator
- Use of motorization of the vertical travel of the carriage. ●
#### 1.7 Non-clinical performance testing
The new device FONA Pan/Ceph has been tested for Electrical Safety and Electromagnetic Compatibility and complies with the standards:
- . IEC 60601-1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2 Medical electrical equipment Part 1-2: General requirements for basic safety and essen-. tial performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.
A qualitative assessment of the image quality of the new FONA Pan/Ceph device has been carried out in comparison to the predicate devices. As a result, it has been confirmed that the image receptors used in the new FONA Pan/Ceph device and in the predicate devices, all allow for good diagnosis, both in panoramic and cephalometric images. However, the new types of image receptors, based on CSI or Cd(Zn)Te scirtillators, adopted in the new FONA Pan/Ceph device, provide clearer pictures due to a better signal to noise ratio, compared to those based on traditional Gd2O2S:Tb (Gadolinium of LANEX™ screens by Kodak) scintillators. Additional tests have been performed taking into account FDA Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices, Document issued on: August 6, 1999". In particular, DQE and MTF data have been provided for the CCD image receptors based on CsI(TI) scintillator, optically coupled, and for the direct conversion CMOS ones with Cd(Zn)Te absorber material, to demonstrate substantial equivalence.
These additional tests have been integrated with bench testing performed with an anthropomorphic phantom for the panoramic and cephalometric modes on the new device and on the predicate ones.
Software verification and validation testing of the new device has been performed according to the FDA Guidance "Guidance for the Content of Premarket Submissions for Software in Medical Devices".
### 1.8 Conclusion
Based on the comparison of intended use, technological characteristics and performance specifications. FONA Srl believes that the new FONA Pan/Ceph device is safe and effective and substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.