K031801 · Blue X Imaging Srl · MUH · Sep 11, 2003 · Dental
Device Facts
Record ID
K031801
Device Name
PANTOS, MODELS: PANTOS 16, AND PANTOS DG
Applicant
Blue X Imaging Srl
Product Code
MUH · Dental
Decision Date
Sep 11, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.1800
Device Class
Class 2
Indications for Use
The PantOs equipment, model PantOs 16, and PantOs DG are indicated for individuals requiring extra-oral dental radiographic examinations and diagnosis of diseases of the teeth, jaw, and oral structures. It exposes and acquires radiographic images at the dentomaxillofacial region. Depending on the model, the anatomical structures are visualized either on a radiographic film or on a computer display, through a dedicated image-intensified fluoroscopic x-ray system.
Device Story
PantOs 16 and PantOs DG are dental panoramic X-ray systems; optional cephalometric lateral arm available. Device captures X-ray images of dentomaxillofacial region; PantOs 16 uses film cassettes; PantOs DG uses digital panoramic sensor for computer display visualization. Operator positions patient using bite block/chin rest; manual carriage adjustment aligns Frankfurt plane and front teeth via lateral light beam. High-frequency generator provides constant potential (60-86 kV, 4-10 mA). Clinicians use images for diagnosis of dental, jaw, and oral structure diseases. System allows carriage position adjustment to optimize focal trough for specific patient anatomy. Benefits include improved image stability and technique control compared to pulsed systems.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Extraoral dental X-ray system. High-frequency generator (60-86 kV, 4-10 mA). PantOs 16: film-based (15x30 cm flat cassette). PantOs DG: digital panoramic sensor. Adjustable rotating carriage for focal trough optimization. Manual patient positioning with bite block/chin rest and lateral light beam. Constant potential DC output.
Indications for Use
Indicated for individuals requiring extra-oral dental radiographic examinations and diagnosis of diseases of the teeth, jaw, and oral structures in the dentomaxillofacial region.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{1}------------------------------------------------
Exhibit V
## 510(k) Summary 031801
#### (1) MANUFACTURER SUBMITTING 510(K) NOTIFICATION
| Applicant | |
|-----------|--|
| | |
| Blue X Imaging Srl |
|-------------------------------------------------|
| Via Idiomi 3/16 |
| 20090 Assago Milan Italy |
| Registration Number: |
| Contact Person |
| Mr. Giuseppe Giacomini, CEO and General Manager |
| phone: + 39 02 4571 2171, fax + 39 02 4570 3385 |
| e-mail: giuseppe.giacomini@bluex.it |
Designated Agent, Initial distributor in the USA
Chicago X-Ray Systems, Inc. 251 E. Dundee Road Suite #6 Wheeling, IL 60090 Registration Number: 1421874 Contact Person: Mr. Al Sosa, President phone: 847 459 3889, fax: 847 459 9214 e-mail: chicagox-ray@att.net
#### Manufacturing site
Blue X Imaging Srl Via Idiomi 3/16 20090 Assago Milan Italy
Prepared
May 1, 2003
#### (2) DEVICE NAME
The proprietary name of the device is "PantOs". Trade name:
- "PantOs 16" for the versions which make use of traditional films with chemical . processing or storage phosphor plates for digital read out, and
"PantOs DG", for the versions which make use of a digital sensor. .
The classification names of this devices are:
- "UNIT, X-RAY, EXTRAORAL WITH TIMER"
- "SYSTEM,X-RAY,EXTRAORAL SOURCE,DIGITAL" ●
- "SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER". .
{2}------------------------------------------------
Ku31801
### (3) LEGALY MARKETED DEVICE TO WHICH EQUIVALENCE IS CLAIMED
In this Pre-market notification; Rotograph Plus, manufactured by Villa Sistem Medicali. is taken as device substantially Equivalent (SE) to PantOs 16.
The version of Rotograph Plus equipped with digital sensor DXIS by Signet or CDR Pan by Shick is substantially equivalent to PantOs DG.
- Rotograph Plus, with 510(k) number K972968, is manufactured by ● Villa Sistemi Medicali SpA. Via delle Azalee, 3 - 20090 Buccinasco Milan Italy Phone +39 02 488 591, Fax +39 02 488 1844.
- DXIS, digital panoramic sensor, with 510(k) number K983283, ● is manufactured by SIGNET S.A.S. 115 bis blv du Général Giraud - 94100 Saint Maur des Fossés, FRANCE Phone +33 1 4883 7300, Fax +33 1 4883 7310,
- CDR Pan, digital panoramic sensor, with 510(k) number K982661, is manufactured by Schick Technologies, Inc. 30-00 47th Avenue - Long Island City, NY 11101 Phone 718-482-2159, Fax: 718-937-5962
### (4) DEVICE DESCRIPTION
PantOs 16 is an X-ray equipment for dental panoramic radiography and for Cephalometry when completed with lateral arm (Ceph version.
PantOs 16 has an X-ray generator with anodic voltage from 60 to 86 kV (constant potential), anodic current from 4 to 10 mA (direct current).
Positioning the patient is done using a bite block or a chin rest, a few simple steps are required: once the height of the carriage is adjusted manually to bring the Frankfurt plane horizontally, correct rotation of the head is checked with the mirror. Front teeth alignment (to align to the in-focus-layer) is done by moving manually the carriage (backward or forward) till correct position is determined with the aid of the lateral light beam (the unit is adjusted without the need to move the patient).
PantOs DG performs like PantOs 16 and is equipped with a digital panoramic receptor instead of film cassette, thus producing panoramic images on the display of a computer system.
#### (5) INTENDED USE
The PantOs equipment, model PantOs 16 and PantOs DG are indicated for individuals requiring extra-oral dental radiographic examinations and diagnosis of diseases of the teeth, jaw, and oral structures. By using the system the user can expose and acquire radiographic images of the dentomaxillofacial region.
Depending on the model, the anatomical structures are visualized either on a radiographic film or on a computer display, through a dedicated image-intensified fluoroscopic x-ray system.
The intended use of PantOs 16 and PantOs DG is not altered but remains the same of the Rotograph Plus.
{3}------------------------------------------------
# K 031801 (6) MAIN DIFFERENCES
As detailed in the specific section, the main differences of the PantOs 16 with respect to the SE device are here summarized:
- Cassette type: PantOs uses a flat cassette of 15x30 cm (about 5.9x11.8 inches), while � Rotograph plus uses a curved one of 5x12 inches (smaller picture).
- . X-ray generator: PantOs uses a high frequency generator which allows for direct current supply (constant potential, instead of pulsed, also featuring fine control of radiographic technique factors (kV, mA, and mAs), and stability, independently from possible fluctuations of the mains voltage.
- . Focal Trough: PantOs gives the user the possibility to adjust the position of the rotating carriage to better fit patient anatomical structures into the focal trough, thus allowing for optimized panoramic projection on different clinical cases.
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo is circular in shape, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the top half of the circle. Inside the circle is a stylized image of a human figure with outstretched arms, resembling a bird in flight. The figure is composed of three curved lines, creating a sense of movement and dynamism.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Blue X Imaging Srl % Mr. Al Sosa Chicago X-Ray Systems, Inc. 251 E. Dundee Road Suite #6 WHEELING IL 60090
AUG 23 2013
Re: K031801
Trade/Device Name: PantOs 16 and GD Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: MUH, EHD, and MOB Dated: May 14, 2003 Received: June 13, 2003
Dear Mr. Sosa:
This letter corrects our substantially equivalent letter of September 11, 2003.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{5}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Exhibit III
### Indication for Use Statement
Page 1 of 1
K031801 510(k) Number (if known):
Device Name: PantOs
Indication For Use :
The PantOs equipment, model PantOs 16, and PantOs DG are indicated for individuals requiring extra-oral dental radiographic examinations and diagnosis of diseases of the teeth, jaw, and oral structures. It exposes and acquires radiographic images at the dentomaxillofacial region. Depending on the model, the anatomical structures are visualized either on a radiographic film or on a computer display, through a dedicated image-intensified fluoroscopic x-ray system.
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Optional Format 3-10-98)
**Prescription Use**
✓
Nancy C. Boydon
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number KD31801
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.