sam X1 Long Duration Ultrasound Device

K211513 · ZetrOZ Systems, LLC · PFW · Aug 18, 2021 · Physical Medicine

Device Facts

Record IDK211513
Device Namesam X1 Long Duration Ultrasound Device
ApplicantZetrOZ Systems, LLC
Product CodePFW · Physical Medicine
Decision DateAug 18, 2021
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 890.5300
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The sam X1 Long Duration Ultrasound Device is intended for home use to apply ultrasonic energy to generate deep heat within body tissues for the treatment of selected medical conditions such as the relief of pain, the relief of muscle spasms, the treatment of joint contractures, and the local increase in circulation.

Device Story

Device applies ultrasonic energy to generate deep heat in body tissues; used for pain relief, muscle spasms, joint contractures, and circulation increase. System comprises rechargeable power controller/timer and ultrasound-generating applicators (piezoelectric) applied via disposable coupling patches. Operates at 3 MHz; delivers 0.65 W (single applicator) or 1.3 W (dual applicator) in continuous wave mode. Intended for prescription home use by patients. Firmware manages timing and power button inputs. Output provides therapeutic deep heat; benefits include symptom management for specified medical conditions. Smaller battery pack design compared to predicate.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including power output evaluation, total treatment time, battery conformance, electrical safety (IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11), EMC, device operation validation, usability, software evaluation, and cybersecurity assessment.

Technological Characteristics

Ultrasonic diathermy device; piezoelectric generation using Lead Zirconate-Titanate crystals. 3 MHz frequency; 0.65 W/0.65 W/cm² (single) or 1.3 W/0.264 W/cm² (dual) output. Continuous wave (100% duty cycle). ABS plastic applicator with TPX lens. Powered by lithium battery with 5V DC input. Dimensions: 3.4 x 3.4 x 1.4 cm. Non-sterile. Software/firmware for battery level measurement and power control.

Indications for Use

Indicated for home use in patients requiring deep heat therapy for pain relief, muscle spasms, joint contractures, and local circulation increase.

Regulatory Classification

Identification

An ultrasonic diathermy for use in applying therapeutic deep heat for selected medical conditions is a device that applies to specific areas of the body ultrasonic energy at a frequency beyond 20 kilohertz and that is intended to generate deep heat within body tissues for the treatment of selected medical conditions such as relief of pain, muscle spasms, and joint contractures, but not for the treatment of malignancies.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below. August 18, 2021 ZetrOZ Systems, LLC Sabrina Lewis QAR Director 56 Quarry Road Trumbull, Connecticut 06611 ## Re: K211513 Trade/Device Name: sam X1 Long Duration Ultrasound Device Regulation Number: 21 CFR 890.5300 Regulation Name: Ultrasonic Diathermy Regulatory Class: Class II Product Code: PFW Dated: August 5, 2021 Received: August 6, 2021 ### Dear Sabrina Lewis: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) ### K211513 Device Name sam X1 Long Duration Ultrasound Device #### Indications for Use (Describe) The sam X I Long Duration Ultrasound Device is intended for home use to apply ultrasonic energy to generate deep heat within body tissues for the treatment of selected medical conditions such as the relief of muscle spasms, the treatment of joint contractures, and the local increase in circulation | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------|---------------------------------------------------------------------------------| | <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(K) SUMMARY | Device Trade Name: | sam X1 Long Duration Ultrasound Device | |--------------------|---------------------------------------------------------------------------------------| | Manufacturer: | ZetrOZ Systems, LLC<br>56 Quarry Road<br>Trumbull, CT 06611 | | Contact: | Ms. Sabrina Lewis<br>QAR Director<br>Phone: 888-202-9831<br>Email: sabrina@zetroz.com | | Date Prepared: | 5/13/2021 | | Regulation Number: | 21 CFR 890.5300 | | Regulation Name: | Ultrasonic diathermy | | Regulation Class: | Class II | | Product Code: | PFW | | Predicate Device: | sam 2.0 Long Duration Ultrasound System (K191568) | ## Indications for Use: The sam X1 Long Duration Ultrasound Device is intended for home use to apply ultrasonic energy to generate deep heat within body tissues for the treatment of selected medical conditions such as the relief of pain, the relief of muscle spasms, the treatment of joint contractures, and the local increase in circulation. ## Device Description: The sam X1 Long Duration Ultrasound Device consists of: - Rechargeable Power Controller and Timer - Ultrasound Generating Applicators ● Single use disposable accessories - . Ultrasonic Coupling Patch The Power Controller can be used to power the Applicator to generate ultrasonic energy at one frequency (3 MHz) and one power setting (0.65 W) per Applicator. In single Applicator mode, the {4}------------------------------------------------ patient can receive 0.65 W at 3 MHz. In dual Applicator mode, the patient can receive 1.3 W at 3 MHz. The Applicators are applied to the skin with onetime use Ultrasonic Coupling Patches. The sam X1 Long Duration Ultrasound Device contains firmware that controls the timing display and inputs from the power button to activate/deactivate the device. There is no control function of this firmware. The system is intended for prescription home use to apply ultrasonic energy for a long duration (1 hours) to generate deep heat within body tissues and treat selected medical conditions such as the relief of pain, the relief of muscle spasms, the treatment of joint contractures, and the local increase in circulation. ## Substantial Equivalence: The sam X1 Long Duration Ultrasound Device is substantially equivalent to the predicate devices cited on the previous page with respect to indications, design, function, and performance. The purpose of this special 510(k) is to allow a smaller battery back to be used. The output and strength of the signal are identical to the predicate. The only difference in the max duration of the subject device is 1 hour where the predicate device can be used for 1 to 4 hours. However, this change does not create a new intended use or different questions of safety or effectiveness as the timeframe of use is within the use of the predicate device. | Information | ZetrOZ sam X1 Long<br>Duration Ultrasound<br>Device<br>(Subject) | ZetrOZ<br>sam 2.0 Long Duration<br>Ultrasound Device<br>(K191568) | Comparison<br>(to sam 2.0) | |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------| | Classification Name | Ultrasonic Diathermy<br>Device | Ultrasonic Diathermy<br>Device | Identical | | Service Type | Physical Medicine | Physical Medicine | Identical | | Classification | 21 CFR 890.5300 | 21 CFR 890.5300 | Identical | | Class | II | II | Identical | | Indications | The sam X1 Long<br>Duration Ultrasound<br>Device is intended for<br>home use to apply<br>ultrasonic energy to<br>generate deep heat within<br>body tissues for the<br>treatment of selected<br>medical conditions such as<br>the relief of pain, the relief<br>of muscle spasms, the<br>treatment of joint<br>contractures, and the local<br>increase in circulation. | The sam 2.0 Long<br>Duration Ultrasound<br>Device is intended for<br>home use to apply<br>ultrasonic energy to<br>generate deep heat within<br>body tissues for the<br>treatment of selected<br>medical conditions such as<br>the relief of pain, the relief<br>of muscle spasms, the<br>treatment of joint<br>contractures, and the local<br>increase in circulation. | Identical | | Manufacturer | ZetrOZ | ZetrOZ | Identical | | Console/Generator<br>Dimensions (L x W x H<br>cm) | 3.4 cm x 3.4 cm x 1.4 cm | 6.10 cm L x 7.09 cm H x<br>1.88cm W | Similar<br>Console/generator<br>sized to accommodate<br>smaller<br>rechargeable<br>battery. | | Information | ZetrOZ sam X1 Long<br>Duration Ultrasound<br>Device<br>(Subject) | ZetrOZ<br>sam 2.0 Long Duration<br>Ultrasound Device<br>(K191568) | Comparison<br>(to sam 2.0) | | Treatment Head<br>Dimensions (L x W x H<br>cm) | 3.81 cm L x 3.30 cm W x<br>1.14 cm H | 3.81 cm L x 3.30 cm W x<br>1.14 cm H | Identical | | Console/Generator Weight<br>(kg) | 0.02 kg | 0.10 kg | Similar<br>Lesser weight<br>primarily driven by<br>use of smaller<br>rechargeable battery. | | Treatment Head Weight<br>(kg) | 0.01 kg | 0.01 kg | Identical | | Power Supply | 120/240 VAC with 5V DC<br>Input Power Jack and<br>Lithium Battery Powered | 120/240 VAC with 5V DC<br>Input Power Jack and<br>Lithium Battery Powered | Identical | | Leakage Current | 0.3 mA | 0.3 mA | Identical | | Crystal Material | Lead Zirconate-Titanate | Lead Zirconate-Titanate | Identical | | Technology of ultrasound<br>generation (e.g.,<br>piezoelectric.<br>magnetoconstructive) | Piezoelectric | Piezoelectric | Identical | | Treatment Mode(s) | Two discrete settings of<br>power at same Frequency | Two discrete settings of<br>power at same Frequency | Identical | | Beam Type (collimated or<br>divergent) | Divergent | Divergent | Identical | | Transducer Diameter (cm) | 5 cm | 5 cm | Identical | | Acoustic Working<br>Frequency and Accuracy<br>(MHz) | 3MHz ± 20% | 3MHz ± 20% | Identical | | Effective Radiating Area<br>and Accuracy (cm2) | One: 6 cm2<br>Two:12 cm2<br>± 20% | One: 6 cm2<br>Two:12 cm2<br>± 20% | Identical | | Beam Nonuniformity Ratio<br>and Accuracy | BNR: <5:1 ± 20% | BNR: <5:1 ± 20% | Identical | | Output Mode: (Continuous<br>Wave/Amplitude -<br>Modulated Wave) | Continuous Wave - 100%<br>duty cycle | Continuous Wave - 100%<br>duty cycle | Identical | | Maximum Timer Setting<br>and Accuracy | 1 Hour +/- 1 minute | 4 Hours +/- 1 minute | Similar<br>Predicate device<br>permits treatment<br>durations of 1, 2, 3, or<br>4 hours. | | Beam Maximum Intensity<br>and Accuracy (W/cm2) | 0.132 W/cm² ± 20% | 0.132 W/cm² ± 20% | Identical | | Maximum Value of the<br>Output Power (Rated<br>Output Power) and<br>Accuracy (W) | Single Applicator:<br>0.65W ± 20%<br>Dual Applicator:<br>1.3W ± 20% | Single Applicator:<br>0.65W ± 20%<br>Dual Applicator:<br>1.3W ± 20% | Identical | | Maximum Value of the<br>Effective Intensity and<br>Accuracy (Not to exceed 3<br>W/cm2 *) | 0.264 W/cm² ± 20% | 0.264 W/cm² ± 20% | Identical | | Information | ZetrOZ sam X1 Long<br>Duration Ultrasound<br>Device<br>(Subject) | ZetrOZ<br>sam 2.0 Long Duration<br>Ultrasound Device<br>(K191568) | Comparison<br>(to sam 2.0) | | For Amplitude Modulated<br>Waves | Not Amplitude Modulated | Not Amplitude Modulated | Identical | | Peak Temperature Rise vs.<br>Time and Tissue Depth to<br>Maximum Treatment Time<br>(for fixed Treatment Head<br>Placement) (deg C) | 7°C at 1 cm<br>4°C at 3 cm<br>2°C at 5 cm<br>Treatment time: 1 hour | 7°C at 1 cm<br>4°C at 3 cm<br>2°C at 5 cm<br>Treatment time: 1 hour | Identical | | Maximum Patient Contact<br>Surface Temperature of<br>Treatment Head under<br>Simulated or Actual Use<br>Conditions for all<br>Operating Conditions<br>(Continually operated for<br>maximum treatment time)<br>(deg C) | 44 °C | 44 °C | Identical | | Therapeutically Applied | Ultrasound Coupling Patch | Ultrasound Coupling Patch | Identical | | Applicator Type | Up to two circular<br>Applicators with 3 MHz<br>output | Up to two circular<br>Applicators with 3 MHz<br>output | Identical | | Applicator Type<br>Applicator Emitting<br>Surface Areas (cm2) | Up to two circular<br>Applicators<br>One Applicator : 5 cm2<br>Two Applicators :10 cm2 | Up to two circular<br>Applicators<br>One Applicator : 5 cm2<br>Two Applicators :10 cm2 | Identical | | Coupling Bandage | ABS Plastic with<br>integrated coupling<br>medium | ABS Plastic with<br>integrated coupling<br>medium | Identical | | Applicator Lens Material | TPX | TPX | Identical | | Environmental - Operating<br>Temperature Range | 0°C to +50°C (32°F to<br>+122°F) | 0°C to +50°C (32°F to<br>+122°F) | Identical | | Performance Standards | 21 CFR 1050.10 | 21 CFR 1050.10 | Identical | | Sterility | Non Sterile | Non Sterile | Identical | | Designed to meet Electrical<br>Safety Standards | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11 | Identical | | Biocompatibility | Yes | Yes | Identical | | Mechanical safety | Yes | Yes | Identical | | Radiation safety (if not<br>radioactive state as such) | Not Radioactive | Not Radioactive | Identical | | Software/Firmware | Yes - For measurement of<br>battery level | Yes - For measurement of<br>battery level | Identical | | Output Channels | Two Independent Power<br>Channels | Two Independent Power<br>Channels | Identical | {5}------------------------------------------------ {6}------------------------------------------------ ## Preclinical Testing: The sam X1 Long Duration Ultrasound Device was evaluated to ensure the smaller power control can still deliver the appropriate treatment for the duration of treatment (1 hour). Testing included: - Power Output Evaluation . - Evaluation of Total Treatment Time . {7}------------------------------------------------ - Battery Conformance Evaluation . - Electrical Safety and EMC Evaluation - Device Operation Validation ● - Powering On/Off . - Evaluation of Usability ● - Software Evaluation - Cybersecurity Evaluation . ## Clinical Testing: Clinical data was not necessary to support substantial equivalence. ## Conclusion: The sam X1 Long Duration Ultrasound Device possesses the same intended use and similar technological characteristics as the predicate devices. The sam X1 Long Duration Ultrasound Device is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...