K240788 · Jkh Health Co., Ltd. · IMI · Jun 4, 2025 · Physical Medicine
Device Facts
Record ID
K240788
Device Name
Ultrasound Stimulator
Applicant
Jkh Health Co., Ltd.
Product Code
IMI · Physical Medicine
Decision Date
Jun 4, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Apply stationary use of ultrasound to: Generate deep heat within body tissues for the treatment of selected medical conditions such as the relief of pain, the relief of muscle spasms, the treatment of joint contractures, and the local increase in circulation. Apply continuous movement of ultrasound for: 1. Pain. 2. Pain relief, muscle spasms, and joint contractures. 3. Relief of pain, muscle spasms, and joint contractures that may be associated with: • Adhesive capsulitis, • Bursitis with slight calcification, • Myositis, • Soft tissue injuries, and • Shortened tendons due to past injuries and scar tissues. 4. Relief of pain, muscle spasms, and joint contractures resulting from: • Capsular tightness, and • Capsular scarring. 5. Localized increase in blood flow. 6. Increased range of motion of contracted joint using heat and stretch techniques.
Device Story
Portable, rechargeable ultrasound stimulator; used in home or clinical settings by patients under medical professional direction. Device applies ultrasonic energy to treatment sites via one or two applicators; utilizes piezoelectric PZT crystals to generate continuous wave ultrasound (1.5 MHz). System includes main console, applicator, charger, ultrasound gel, and adhesive pads. Microprocessor-controlled; features automatic alerts for improper application, automatic shut-off, and user override. Overheating protection limits treatment head temperature to <43°C. Provides deep heat to tissues; aids pain relief, muscle spasm reduction, joint contracture treatment, and increased blood flow. Healthcare providers use output to manage musculoskeletal conditions; patients benefit from non-invasive, home-based therapeutic heat and stretch therapy.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing, including power output evaluation, electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), thermal safety (IEC 60601-2-5), biocompatibility (ISO 10993-5/10), and software verification.
Technological Characteristics
Piezoelectric ultrasound generation using PZT crystals; 1.5 MHz frequency; continuous wave (100% duty cycle). ABS housing. Rechargeable 3.7Vdc battery; 100–240 Vac input. Microprocessor-controlled with automatic shut-off and overheating protection (<43°C). Non-sterile. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-5, and ISO 10993 standards.
Indications for Use
Indicated for patients requiring deep heat therapy for pain relief, muscle spasms, joint contractures, and localized circulation increase. Applicable for conditions including adhesive capsulitis, bursitis, myositis, soft tissue injuries, and scar tissue. Prescription use only.
Regulatory Classification
Identification
An ultrasonic diathermy for use in applying therapeutic deep heat for selected medical conditions is a device that applies to specific areas of the body ultrasonic energy at a frequency beyond 20 kilohertz and that is intended to generate deep heat within body tissues for the treatment of selected medical conditions such as relief of pain, muscle spasms, and joint contractures, but not for the treatment of malignancies.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 4, 2025
JKH Health Co., Ltd.
% Bill Dai
Managing Member
Jkh Usa LLC
14271 Jeffrey Rd.
#246
Irvine, California 92620
Re: K240788
Trade/Device Name: Ultrasound Stimulator
Regulation Number: 21 CFR 890.5300
Regulation Name: Ultrasonic Diathermy
Regulatory Class: Class II
Product Code: IMI, PFW
Dated: May 5, 2025
Received: May 5, 2025
Dear Dr. Bill Dai:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K240788 - Bill Dai
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K240788 - Bill Dai
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Amber T. Ballard -S
Amber Ballard, PhD
Assistant Director
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K240788
Device Name
Ultrasound Stimulator
Indications for Use (Describe)
Apply stationary use of ultrasound to:
Generate deep heat within body tissues for the treatment of selected medical conditions such as the relief of pain, the relief of muscle spasms, the treatment of joint contractures, and the local increase in circulation.
Apply continuous movement of ultrasound for:
1. Pain.
2. Pain relief, muscle spasms, and joint contractures.
3. Relief of pain, muscle spasms, and joint contractures that may be associated with:
• Adhesive capsulitis,
• Bursitis with slight calcification,
• Myositis,
• Soft tissue injuries, and
• Shortened tendons due to past injuries and scar tissues.
4. Relief of pain, muscle spasms, and joint contractures resulting from:
• Capsular tightness, and
• Capsular scarring.
5. Localized increase in blood flow.
6. Increased range of motion of contracted joint using heat and stretch techniques.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
> "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
Page 1 of 6
510(k) Summary
This 510(k) Summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
1. Submitter’s Information
Submitter: JKH Health Co., Ltd.
Address: 3-5F And 11F, Building 14, Tiangong Smart Valley Industrial Park, Fuhai Road, Xiagang Community, Chang'an Town, Dongguan, Guangdong, China
Contact Person: Pu Jiang
Tel: +86-755-27926589
Fax: +86-755-29970323
Email: sales@JKHhealth.com Date of Preparation: June 4, 2025
2. Subject Device
Trade/Device Name: Ultrasound Stimulator
Common Name: Ultrasonic Diathermy For Use In Applying Therapeutic Deep Heat Review
Panel: Physical Medicine
Product Code: IMI, PFW
Regulation Number: 21 CFR 890.5300
Device Class: II
Use: Prescription
3. Predicate device
Primary Predicate Device: ManaSport+
510(k) Number: K222098
Clearance Date: March 8, 2023
Submitter: ManaMed, Inc.
Predicate Device: sam 2.0 Long Duration Ultrasound System
510(k) Number: K191568
Clearance Date: March 6, 2020
Submitter: ZetrOZ Systems, LLC
4. Description of Subject Device
As a portable and rechargeable prescriptive device in the home or clinical/hospital setting, the subject device is intended to apply ultrasound to the patient’s treatment site. It is demonstrated ultrasound can be used to apply therapeutic deep heat for the treatment of medical conditions including relief of pain, muscle spasms, and joint contractures.
It is intended to be used for 30 minutes per cycle (up to 4 cycles) in the home or clinical/hospital setting by or under the direction of a medical professional to apply ultrasound to the patient’s treatment site. The device automatically alerts the patient in case of improper application or performance. All the control components and the rechargeable battery of the device are protectively housed, and the device is connected to one or two applicators of ultrasound.
The subject device consists of the main device, the applicator, the charger/adapter, the liquid and/or solid ultrasound gel, and the sticky pad. For the best result, use the ultrasound gel and the sticky pad supplied.
{5}
If you feel your skin is sensitive to the ultrasound gel and/or pad supplied, please contact your physician or use a different FDA-cleared ultrasound gel and/or pad.
## 5. Indications for Use
Apply stationary use of ultrasound to:
Generate deep heat within body tissues for the treatment of selected medical conditions such as the relief of pain, the relief of muscle spasms, the treatment of joint contractures, and the local increase in circulation.
Apply continuous movement of ultrasound for:
1. Pain.
2. Pain relief, muscle spasms, and joint contractures.
3. Relief of pain, muscle spasms, and joint contractures that may be associated with:
- Adhesive capsulitis,
- Bursitis with slight calcification,
- Myositis,
- Soft tissue injuries, and
- Shortened tendons due to past injuries and scar tissues.
4. Relief of pain, muscle spasms, and joint contractures resulting from:
- Capsular tightness, and
- Capsular scarring.
5. Localized increase in blood flow.
6. Increased range of motion of contracted joint using heat and stretch techniques.
## 6. Substantial Equivalence
The following Table 1 summarizes the comparison between the subject device and the predicate device, indicating the intended use and technical characteristics of the subject device are substantially equivalent to those of the predicate device.
Table 1. Comparison between the predicate device and the subject device
| Parameter & Predicate Device(s) | Subject Device | Primary Predicate Device | Predicate Device | Equivalence |
| --- | --- | --- | --- | --- |
| 510(k)/PMA Number | K240788 | K222098 | K191568 | N/A |
| Submitter /Manufacturer | JKH Health Co., Ltd. | ManaMed, Inc. | ZetrOZ Systems, LLC | N/A |
| Device Name/Model | Ultrasound Stimulator | ManaSport+ | sam 2.0 Long Duration Ultrasound System | N/A |
| Product Code of Ultrasound | IMI, PFW | IMI, PFW | PFW | Identical |
| Indications for Use | Apply stationary use of ultrasound to: Generate deep heat within body tissues for the treatment of selected medical conditions such as the relief of pain, the relief of muscle spasms, the treatment of joint contractures, and the local increase in circulation.
Apply continuous movement of ultrasound for:
1. Pain.
2. Pain relief, muscle spasms, and joint contractures.
3. Relief of pain, muscle | Apply stationary use of ultrasound to: Generate deep heat within body tissues for the treatment of selected medical conditions such as the relief of pain, the relief of muscle spasms, the treatment of joint contractures, and the local increase in circulation.
Apply continuous movement of ultrasound for:
1. Pain.
2. Pain relief, muscle spasms, and joint contractures. | The sam 2.0 Long Duration Ultrasound Device is intended for home use to apply ultrasonic energy to generate deep heat within body tissues for the treatment of selected medical conditions such as the relief of pain, the relief of muscle spasms, the treatment of joint contractures, and the local increase in circulation. | Identical |
{6}
| | | spasms, and joint contractures that may be associated with: • Adhesive capsulitis, • Bursitis with slight calcification, • Myositis, • Soft tissue injuries, and • Shortened tendons due to past injuries and scar tissues. 4. Relief of pain, muscle spasms, and joint contractures resulting from: • Capsular tightness, and • Capsular scarring. 5. Localized increase in blood flow. 6. Increased range of motion of contracted joint using heat and stretch techniques. | 3. Relief of pain, muscle spasms, and joint contractures that may be associated with: • Adhesive capsulitis, • Bursitis with slight calcification, • Myositis, • Soft tissue injuries, and • Shortened tendons due to past injuries and scar tissues. 4. Relief of pain, muscle spasms, and joint contractures resulting from: • Capsular tightness, and • Capsular scarring. 5. Localized increase in blood flow. 6. Increased range of motion of contracted joint using heat and stretch techniques. | | |
| --- | --- | --- | --- | --- | --- |
| Prescription or OTC | | Prescription | Prescription | Prescription | Identical |
| Power Source(s) | | 100–240 Vac with 5Vdc Input and rechargeable 3.7Vdc battery | 100–240 Vac with 5Vdc Input and rechargeable 3.7Vdc battery | 120/240 VAC with 5V DC Input Power Jack and Lithium Battery Powered | Identical |
| Number of Output | | 1 or 2 | 1 | 1-2 | Identical |
| Software /Firmware/ Microprocessor Control? | | Yes | Yes | Yes | Identical |
| Automatic No-Load Trip? | | Yes | Yes | Not Publicly Available | Identical |
| Automatic Shut Off? | | Yes | Yes | Not Publicly Available | Identical |
| User Override Control? | | Yes | Yes | Not Publicly Available | Identical |
| Indicator Display: | On/Off Status? | Yes | Yes | Not Publicly Available | Identical |
| | Low Battery? | Yes | Yes | Not Publicly Available | Identical |
| Timer Range (minutes) | | 30 | 20 | 240 | Longer than the primary predicate, and shorter than the predicate. There is an overheating temperature protection enforced to prevent the temperature of the treatment head going over 43°C during the normal use, so the possible range of temperature is lower than 43°C. The tests provided shows this does not affect the safety or effectiveness. |
| Compliance with Voluntary Standards? | | Yes | Yes | Yes | Identical |
| Biocompatibility? | | Yes | Yes | Yes | Identical |
| Sterility | | Non-Sterile | Non-Sterile | Non-Sterile | Identical |
| Housing construction material | | ABS | ABS | ABS | Identical |
| Console/Generator Dimensions (mm) [L x W x H] | | 138×62×22 | 140×56×24 | 61×70.9×18.8 | Similar, and do not affect the safety or effectiveness |
| Console/Generator Weight (kg) | | 0.18 | 0.13 | 0.01 | Similar, and do not affect the safety or effectiveness |
Page 3 of 6
{7}
| Treatment head dimensions (mm) [L x W x H] | 61 x 45 x 12 | 30 x 30 x 12 | 38.1 x 33 x 11.4 | Similar, and do not affect the safety or effectiveness |
| --- | --- | --- | --- | --- |
| Treatment Head Weight (kg) | 0.047 | 0.031 | 0.10 | Similar, and do not affect the safety or effectiveness |
| Functions and design | Ultrasound | Ultrasound | Ultrasound | Identical |
| Frequency | 1.5 MHz ± 10% | 1.5 | 3 MHz ± 20% | Identical |
| Leakage current | <0.3mA | <0.3mA | 0.3mA | Identical |
| Crystal material | PZT | PZT | Lead Zirconate-Titanate | Identical |
| Technology of Ultrasound generation | Piezoelectric | Piezoelectric | Piezoelectric | Identical |
| Output Mode | Continuous Wave - 100% duty cycle | Continuous Wave - 100% duty cycle | Continuous Wave - 100% duty cycle | Identical. |
| Ratio of temporal maximum output power to the output power | 1:1 | 1:1 | Not Publicly Available | Identical |
| Maximum Value of the Output Power (W ± 20%) | 0.60 | 0.60 | Single Applicator: 0.65W ± 20% Dual Applicator: 1.3W ± 20% | Similar |
| Temporal average power (W ± 20%) | 0.60 | 0.60 | Not Publicly Available | Identical |
| Maximum value of the effective intensity (W/cm2) | 0.15 ± 20% | 0.16 ± 20% | 0.264 ± 20% | Identical or similar. According to the IEC 61689, the deviation allowed is ± 30%. |
| Beam Maximum Intensity and Accuracy (W/cm2) | 0.15 ± 20% | 0.16 ± 20% | 0.132 ± 20% | Identical or similar. According to the IEC 61689, the deviation allowed is ± 30%. |
| Duty factor | 100% | 100% | Not Publicly Available | Identical |
| Beam Type | Collimated | Collimated | Divergent | Identical |
| Applicator size | Area: 3.9 cm2 | Area: 3.9 cm2 or 5cm2 | One Applicator: 5 cm2 Two Applicators: 10 cm2 | Identical |
| Maximum patient contact surface temperature of the treatment head under simulated or actual use conditions for all operating conditions (continually operated for maximum treatment time) (°C) | Meets IEC 60601-2-5, section 201.11 Protection against excessive temperature and other HAZARDS. | Meets IEC 60601-2-5, section 201.11 Protection against excessive temperature and other HAZARDS. | 44 °C | Identical |
| Peak Temperature Rise vs. Time and Tissue Depth to Maximum Treatment Time (for fixed Treatment Head Placement) (°C) | 8.6°C at 1 cm 6.1°C at 2 cm 2.1°C at 3 cm Max treatment time: 30 min per cycle and 4 cycles | 7.6°C at 1 cm 3.3°C at 2 cm 1.4°C at 3 cm Max treatment time: 20 min | 8°C at 1 cm 6°C at 3 cm 3°C at 5 cm Max treatment time: 4 hours | Identical or similar. There is an overheating temperature protection enforced to prevent the temperature of the treatment head going over 43°C during the normal use, so the possible range of temperature is lower than 43°C. The tests provided shows this does not affect the safety or effectiveness. |
| Beam Non-Uniformity Ratio | <5 | 4 ± 20% | <5:1 ± 20% | Identical |
| Effective Radiating Area (cm2) | 3.9 ± 20% | 3.9 or 5 ± 20% | One: 6 cm2 Two:12 cm2 ± 20% | Identical |
| | | | | |
| | | | | |
7. Summary of Substantial Equivalence
{8}
As shown in the above Table 1, the ultrasound from the subject device is identical or similar to the predicate device. There are no new safety or effectiveness issues concerning the subject device, which offers substantially equivalent technical specifications, features, intended use, safety, and effectiveness to the predicate device.
## 8. Non-Clinical Tests Performed
Non-clinical tests were performed on the subject device in order to validate the design and to assure conformance with the following voluntary design standards in connection with medical device electrical safety, and electromagnetic compatibility.
(a) ANSI AAMI ES60601-1 "Medical Electrical Equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1)".
(b) IEC 60601-1-2 "Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral standard: Electromagnetic Compatibility - Requirements and Tests".
(c) IEC 60601-2-5 "Medical electrical equipment - Part 2-5: Particular requirements for the basic safety and essential performance of ultrasonic physiotherapy equipment".
(d) ISO 10993-5 "Biological evaluation of medical devices -- Part 5: Tests for In Vitro Cytotoxicity".
(e) ISO 10993-10 "Biological evaluation of medical devices - Part 10: Tests for Irritation and Skin Sensitization".
(f) IEC 60601-1-6 "Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability".
(g) IEC 60601-1-11 "Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment"
In addition to the compliance of voluntary standards, the verification of software used in the subject device has been carried out according to the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, and the bench performance testing for the subject device and ultrasonic gel pad has also been conducted.
The subject device was evaluated to ensure all the non-clinical tests in this submission support a determination of substantial equivalence, including:
- Power output evaluation
- Evaluation of total treatment time
- Electrical safety and EMC evaluation
- Device Operation evaluation
- Powering On/Off
- Usability evaluation
- Biocompatibility evaluation
- Thermal safety evaluation
- Performance evaluation
Page 5 of 6
{9}
- Software evaluation
All the non-clinical tests performed above demonstrate the subject device is as safe and as effective as the legally marketed predicate devices. The detailed comparison between the subject device and the predicate devices in the above Table 1 demonstrates the intended use and technical characteristics of the subject device are substantially equivalent to those of the predicate devices.
## 9. Clinical Testing
Clinical testing was not provided in support of this 510(k) application.
## 10. Conclusion
The testing and comparison performed demonstrate the subject device is substantially equivalent to the predicate devices. Therefore, the subject device is as safe and effective as the predicate devices that have been legally marketed in the United States.
Page 6 of 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.