POTABLE AIR MASSAGER PUTIFINO, MODEL AM-7

K071596 · Nihon Seimitsu Sokki Co., Ltd. · IRP · Feb 15, 2008 · Physical Medicine

Device Facts

Record IDK071596
Device NamePOTABLE AIR MASSAGER PUTIFINO, MODEL AM-7
ApplicantNihon Seimitsu Sokki Co., Ltd.
Product CodeIRP · Physical Medicine
Decision DateFeb 15, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.5650
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Portable Air Massager putifino AM-7 is indicated for the temporary relief of minor muscle aches and pains and for temporary increase in circulation to the treated areas in people who are in good health. The Portable Air Massager putifino AM-7 simulates kneading and stroking of tissues by using an inflatable garment (Cuff).

Device Story

Portable Air Massager putifino AM-7 is a battery-operated, powered inflatable tube massager for calf use. Device consists of a controller unit and a cotton/velour massaging cuff with an internal inflatable bladder. User wraps cuff around calf; controller uses a microprocessor to manage an air pump and electric control valve to inflate/deflate the bladder. Three massage modes and three intensity levels are available; Mode 3 provides a cycle of massage and rest to promote circulation. Device is intended for OTC use by healthy individuals. Output is mechanical kneading/stroking of tissues. Benefits include temporary pain relief and increased local circulation.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Powered by 2x1.5V AA alkaline batteries. Materials: ABS (controller), cotton/polyester/nylon/velour (cuff). Sensing/actuation: Microprocessor-controlled air pump and electronic exhaust valve. Form factor: Portable, cuff-based. Connectivity: None. Sterilization: Not applicable.

Indications for Use

Indicated for healthy individuals seeking temporary relief of minor muscle aches and pains and temporary increase in local circulation in the calf area.

Regulatory Classification

Identification

A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K071596 #### 1. Company Identification NIHON SEIMITSU SOKKI CO., LTD. (Nissei) 2508-13 Nakago, Shibukawa, Gunma 377-0293 Japan Tel: +81-279-20-2311 Fax: +81-279-20-2411 #### 2. Official Correspondent Hideki Tomaru (Mr.) Assistant Manager of International Sales Div. #### 3. Date of Submission June 11, 2007 ## 4. Trade/Proprietary Name Portable Air Massager putifino AM-7 #### 5. Classification Number Class II, IRP, 21 CFR 890.5650 - Powered inflatable tube massager #### 6. Predicate Device Manufacturer : Salton, Inc Trade Name : Relaxor Perfect Touch Air Massaging System : K050697, Cleared on April 10, 2003 510(k) No. #### 7. Description of Device Portable Air Massager putifino AM-7 is a powered inflatable tube massager, which stimulates kneading and stroking of tissues with the hands by pressurizing and depressurizing the massaging garment (cuff) wrapped around the calf. The cotton and velour massaging cuff with inflatable bladder inside is assembled to the controller to facilitate the device's compactness and portability. The device is applicable to calves whose circumferences are between 8.7 to 17.7 inches (22 to 45 cm). With the included attachable extension hook, the applicable range extends to 19.7 inches (50 cm). The device owns three modes of massage pattern and three levels of massage intensity, which are controlled by the microprocessor. Operation of the mode button on the controller during massaging will switch the massage patterns. In MODE 1, massage is given by repetition of pressurization and depressurization of the cuff, where pressurization is set to one pressure level. In MODE 2, massage pattern of MODE 1 is repeated at two pressure levels. In MODE 3, which is intended to temporary increase blood circulation, approximately 10 minutes of massage at low pressure level and approximately 20 minutes of rest are repeated. In MODE 1 and 2, operation of the pressure button on the controller will switch the massage intensities: weak, moderate and hard. The user of the device can confirm both massage mode and level with the illuminated indicator on the controller. #### 8. Indication for use The Portable Air Massager putifino AM-7 is indicated for the temporary relief of minor muscle aches and pains and for temporary increase in circulation to the treated areas in people who are in good health. The Portable Air Massager putifino AM-7 simulates kneading and stroking of tissues by using an inflatable garment (Cuff). FEB 15 ======================================================================================================================================================================= {1}------------------------------------------------ #### 9. Technological characteristics of the subject device The device, powered with two AA alkaline batteries, achieves its intended use by inflating and deflating the bladder inside the massaging garment with the air pump and the electric control valve. The microprocessor controls inflation and deflation of the bladder to conduct message in three different modes and to provide massage at different intensities. The comparison table of the technological characteristics of the predicate device and the subject device is attached as Appendix 1. The predicate and subject devices share the same operational principal, that is, to provide massage by inflating and deflating massage garment wrapped around a part of the body. The differences between the devices arise in product features due to mechanical structures that the subject device is designed to be portable. The product features of the predicate device to change massage garments as mentioned above and to specify massage areas, which is enabled with the garment with separated chambers, are not equipped with the subject drevice, The other difference due to the structure is that the devices are operated with different power sources; the predicate is powered by AC adapter and the subject device by batteries. Although the product features are not identical, these characteristic differences should not affect the substantial equivalence of the subject device to the predicate device. #### 10. Conclusion The Portable Air Massager putifino AM-7 is substantially equivalent to Salton, Inc., Relaxor Perfect Touch Air Massaging System K050697. {2}------------------------------------------------ # Appendix 1: Comparison Table with Predicate Device | SUBJECT DEVICE:AM-7 | PREDICATE DEVICE:K030437 | | | | | | | | | | | | | | | |---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--| | INDICATION FOR USE | INDICATION FOR USE | | | | | | | | | | | | | | | | The Portable Air Massager putifino AM-7 is<br>indicated for the temporary relief of minor<br>muscle aches and pains and for temporary<br>increase in circulation to the treated areas in<br>people who are in good health. The Portable Air<br>Massager putifino AM-7 simulates kneading and<br>stroking of tissues by using an inflatable<br>garment (Cuff). | The Perfect Touch Air Massaging System is<br>indicated for the temporary relief of minor<br>muscle aches and pains and for temporary<br>increase in circulation to the treated areas in<br>people who are in good health. The Perfect<br>Touch simulates kneading and stroking of<br>tissues by using an inflatable garment. | | | | | | | | | | | | | | | | INDICATIONS | INDICATIONS | | | | | | | | | | | | | | | | Power<br>Pressure<br>Mode<br>MODE OF COMPRESSION | Speed<br>Intensity<br>Press Zone<br>MODE OF COMPRESSION | | | | | | | | | | | | | | | | Intermittent | Intermittent | | | | | | | | | | | | | | | | MASSAGE OPERATION | MASSAGE OPERATION | | | | | | | | | | | | | | | | Massage level:3 (Weak, Moderate, Strong)<br>Massage pattern:3(Mode1, Mode2, Mode3)<br>Mode3 is Blood circulation improvement mode | Speed:10 Speed settings<br>Massage level:6 Levels<br>Zone:16Zone | | | | | | | | | | | | | | | | MASSAGE TIME | MASSAGE TIME | | | | | | | | | | | | | | | | Mode1: 10 minutes<br>Mode2: 10 minutes<br>Mode3: Approximately 10 minutes of<br>massage and approximately 20 minutes of<br>rest are alternately repeated. | 15 minutes | | | | | | | | | | | | | | | | MASSAGE PRESSURE | MASSAGE PRESSURE | | | | | | | | | | | | | | | | Massage Pattern is Mode1:<br>Weak:9.3kPa (70mmHg)<br>Moderate:13.3kPa (100mmHg)<br>Strong:17.3kPa (130mmHg) | Intensity Settings 1 2 3 4 5 6 Garment Air Pressure 80 (mmHg) 104 (mmHg) 128 (mmHg) 152 (mmHg) 176 (mmHg) 200 (mmHg) | | | | | | | | | | | | | | | | Massage Pattern is Mode2:<br>Weak:9.3kPa (70mmHg),10.7kPa(80mmHg)<br>Moderate:13.3kPa(100mmHg),14.7kPa(110mm Hg)<br>Strong:17.3kPa(130mmHg),18.7kPa(140mmHg) | | | | | | | | | | | | | | | | | Blood stream improvement with Mode3:<br>Pressure is keep at 6.0kPa (45mmHg)<br>MASSAGE PART | MASSAGE PART | | | | | | | | | | | | | | | | Calves | Leg & Foot, Feet, Arm,<br>Neck & Shoulders, Lower Back, Hands | | | | | | | | | | | | | | | | PRESSURE RANGE | PRESSURE RANGE | | | | | | | | | | | | | | | | 0-200mmHg | 0-250mmHg | | | | | | | | | | | | | | | . {3}------------------------------------------------ | SUBJECT DEVICE:AM-7 | PREDICATE DEVICE:K030437 | |------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------| | NUNBER OF CHAMBERS | NUNBER OF CHAMBERS | | 1 | 1 to 12, Varies with Garments | | MATERIAL | MATERIAL | | Operation Unit: ABS<br>Massage Garment: Surface: Cotton / Polyester /<br>Nylon<br>Back: Velour | Operation Unit: ABS<br>Massage Garment: TPU<br>Nylon with TPU Backing 190X70 with TPU<br>Backing | | Pressure control | Pressure control | | Microprocessor and Pressure Sensor | Microprocessor | | Inflation | Inflation | | Pressurization pump<br>Deflation | Pressurization pump<br>Deflation | | Electronic control exhaust valve | exhaust valve | | POWER SOURCE | POWER SOURCE | | 2×1.5-Volt"AA" Alkali battery<br>Consumption 2W | Main body:120V 60Hz<br>Consumption 26W<br>AC Adapter:120V 60Hz<br>Consumption 36W | . . {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission to protect the health of all Americans and provide essential human services. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" in a circular arrangement. Public Health Service FEB 1 5 2008 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Nihon Seimitsu Sokki Co., Ltd. % Cosmos Corporation Mr. Koji Kubo Manager 3F, 2-17-6 Akebono-cho Tachikawa-shi Tokyo 190-0012 Japan Re: K071596 > Trade Name: Portable Air Massager putifino AM-7 Regulation Number: 21 CFR 890.5650 Regulation Name: Powered inflatable tube massager. Regulatory Class: Class II Product Code: IRP Dated: January 12, 2008 Received: January 16, 2008 Dear Mr. Kubo: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {5}------------------------------------------------ #### Mr. Koii Kubo forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. Mark N. Millerson Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ### Indications for Use 510(k) Number (if known): K071596 Device Name : Portable Air Massager putifino AM-7 Indications for Use: The Portable Air Massager putifino AM-7 is indicated for the temporary relief of minor muscle aches and pains and for temporary increase in circulation to the treated areas in people who are in good health. The Portable Air Massager putifino AM-7 simulates kneading and stroking of tissues by using an inflatable garment (Cuff). Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use X (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) R (Division Sign-Off) Division of General, Restorative, and Neurological Devices **510(k) Number** k071596
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...