K222991 · Xiamen Emoka Health Science & Technology Co., Ltd. · IRP · Mar 15, 2023 · Physical Medicine
Device Facts
Record ID
K222991
Device Name
Air Compression Leg Massager (model: EMK-701)
Applicant
Xiamen Emoka Health Science & Technology Co., Ltd.
Product Code
IRP · Physical Medicine
Decision Date
Mar 15, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5650
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Air Compression Leg Massager is indicated for the temporary relief of minor muscle aches and pains and for temporary increase in circulation to the treated areas in people who are in good health. Air Compression Leg Massager simulates kneading and stroking of tissues by using an inflatable garment.
Device Story
Portable, rechargeable air compression leg massager; intended for OTC use. Device consists of handheld controller with internal pump, battery, and microprocessor; connects via air hoses to leg wraps containing 2-chamber air bladders. User operates via power button and interface; controller inflates/deflates bladders in preset sequences (kneading/stroking simulation). Internal pressure sensor monitors inflation levels; pump stops at target pressure for rest periods. Used in home environment to increase circulation and relieve minor muscle aches. Output is mechanical compression of feet and calves; provides patient comfort via medical fabric wraps. Benefits include temporary pain relief and improved circulation.
Clinical Evidence
No clinical data. Bench testing only, including ANSI/AAMI ES60601-1, IEC 62133-2, IEC 60601-1-2, IEC 60601-1-11, ISO 10993-5, and ISO 10993-10.
Technological Characteristics
Powered inflatable tube massager; 2-chamber air bladders; Dacron medical fabric wraps. Microprocessor-controlled air pump; 11.1V 1000mAh internal battery; AC 100-240V input. Pressure range 120-210 mmHg. Biocompatibility per ISO 10993-5/10. Electrical safety per ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-11. Handheld controller with LED indicators.
Indications for Use
Indicated for temporary relief of minor muscle aches and pains and temporary increase in circulation to treated areas in healthy individuals.
Regulatory Classification
Identification
A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text on the right. The symbol on the left is a stylized representation of a human figure. The text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters. The words "U.S. FOOD & DRUG" are on the top line, and the word "ADMINISTRATION" is on the bottom line.
March 15, 2023
Xiamen Emoka Health Science & Technology Co., Ltd. Iris Fung Regulation Manager 6F, No.1 TianTai road, Science City, LuoGang District Guangzhou, Guangdong 361000 China
Re: K222991
Trade/Device Name: Air Compression Leg Massager (model: EMK-701) Regulation Number: 21 CFR 890.5650 Regulation Name: Powered inflatable tube massager Regulatory Class: Class II Product Code: IRP Dated: September 28, 2022 Received: September 28, 2022
Dear Iris Fung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Tushar Bansal -S
for Heather Dean, PhD Assistant Director THT5B3: Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K22991
Device Name
Air Compression Leg Massager (model: EMK-701)
Indications for Use (Describe)
Air Compression Leg Massager is indicated for the temporary relief of minor muscle aches and for temporary increase in circulation to the treated areas in good health. Air Compression Leg Massager simulates kneading and stroking of tissues by using an inflatable garment.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Date of the summary prepared: Mar 10, 2023
# 510(k) Summary
# 510K: K222991
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92. This is a Traditional 510(K) submission, and there were no prior submissions for the subject device.
# 1. Submitter'sInformation
## Sponsor
- � Company Name: Xiamen Emoka Health Science & Technology Co., Ltd.
- Address: D Building, No.100, Jinfu Road, Chengnan Industrial Zone, Tong'an District, � Xiamen, China
- � Phone: +86-592-7363861
- � Email: emk@emoka.cn
- Contact Person (including title): Zhang Feng (Quality Manager) �
# Application Correspondent:
- Xiamen Emoka Health Science & Technology Co., Ltd.. �
- � Address: D Building, No.100, Jinfu Road, Chengnan Industrial Zone, Tong'an District, Xiamen, China
- � Contact Person: Ms. Iris Fung
- Title: Regulation Manager �
- � Tel: +86-18588874857
- � Email: mdc-fs@foxmail.com; jianda-lee@foxmail.com
#### Subject Device Information 2.
- Type of 510(k) submission: Traditional �
- Classification: Powered Inflatable Tube Massager �
- � Trade Name: Air Compression Leg Massager
- � Model: EMK-701
- � Review Panel: Physical Medicine
- Product Code: IRP �
- Requlation Number: 21 CFR 890.5650 �
- Regulation Class: 2 �
# 3. Predicate Device Information
## Predicate Device I
- 510(k) number: K193354 �
{5}------------------------------------------------
- � Sponsor: Shenzhen Donqiilian Electronics Co., Ltd.
- Classification: Powered Inflatable Tube Massager �
- Trade Name: Air Compression Therapy Device � Model: S9019
- � Review Panel: Physical Medicine
- Product Code: IRP �
- � Regulation Number: 21 CFR 890.5650
- Regulation Class: 2 �
## Predicate Device II
- 510(k) number: K212935 �
- Sponsor: Xiamen Simo Electronic Co., Ltd �
- � Classification: Powered Inflatable Tube Massager
- Trade Name: Air Pressure Foot Massager �
- � Model: SM-512F
- Review Panel: Physical Medicine �
- � Product Code: IRP
- Regulation Number: 21 CFR 890.5650 �
- Regulation Class: 2 �
## 4. Device Description
Air Compression Leg Massager (Model: EMK-701) is a portable and rechargeable device. lt is intended to be an over-the-counter portable inflatable tube massage system which simulates kneading and stroking of tissue with the hands by use of inflatable pressure cuffs. It can be used to temporarily increase blood circulation and temporarily relieve minor muscle aches and pains.
Air Compression Leg Massager supplied clean and non-sterile, utilizes the pneumatically controlled leg wraps actuated by an electronically controlled air pump unit. A pump, battery and control components are protectively housed in a plastic case of Handheld Controller. Function buttons and light emitting diode (LED) indicators on the Handheld Controller make up the user interface. There are 3 ports at the bottom of Handheld Controller for connecting the alternating current (AC) adapter plug and two air hoses.
Each leg wrap has an air hose for connection to Handheld Controller, and both encase a 2chamber air bladder inside. Feet and calves can be wrapped and massaged separately by the two chambers. The soft medical fabric of wraps provides patient comfort and biocompatibility compliance.
In operation, the user turns the power on via the Power button. Then the Handheld Controller controls the inflating and deflating of the air bladders according to preset program
{6}------------------------------------------------
parameters. The air pressure is monitored by an internal pressure sensor and microprocessor. Once the pressure of the air bladder reaches the proper level, the pump is turned off for a rest period. The cycle of inflation and deflation repeats until the unit is turned off.
## 5. Intended Use / Indicationsfor Use
Air Compression Leg Massager is indicated for the temporary relief of minor muscle aches and pains and for temporary increase in circulation to the treated areas in people who are in good health. Air Compression Leg Massager simulates kneading and stroking of tissues by using an inflatable garment.
# 6. Test Summary
Air Compression Leg Massager has been evaluated for its safety and performance by lab bench testing as following:
- � ANSI/AAMI ES60601-1:2012, Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance
- � IEC 62133-2:2017, Secondary cells and batteries containing alkaline or other nonacid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems
- � IEC 60601-1-2:2014, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility
- ◆ IEC 60601-1-11:2015 Medical electrical equipment - part 1-11: general requirements for basic safety and essential performance - collateral standard: requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- � ISO 10993-5:2009, Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity
- � ISO 10993-10:2010, Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization
## 7. Comparison to predicate device and conclusion
The technological characteristics, features, specifications, materials, and intended use of Air Compression Leg Massager is substantially equivalent to the predicate devices quoted above. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.
| Elements of<br>comparison | Subject Device | Predicate Device I | Predicate Device II | Comparison |
|------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|
| Manufacturer | Xiamen Emoka Health<br>Science & Technology Co.,<br>Ltd. | Shenzhen Dongjilian<br>Electronics Co., Ltd. | Xiamen Simo Electronic Co.,<br>Ltd | -- |
| 510K number | K222991 | K193354 | K212935 | -- |
| Elements of | | | | |
| comparison | Subject Device | Predicate Device I | Predicate Device II | Comparison |
| Product Name | Air Compression Leg<br>Massager<br>BMK-701 | Air Compression Therapy<br>Device<br>S9019 | Air Pressure Foot Massager<br>SM-512F | -- |
| Classification<br>Name | Pow ered inflatable tube<br>massager | Pow ered inflatable tube<br>massager | Pow ered inflatable tube<br>massager | -- |
| Regulation<br>Class | 2 | 2 | 2 | -- |
| Product code | IRP | IRP | IRP | -- |
| Regulation<br>Number | 21 CFR 890.5650 | 21 CFR 890.5650 | 21 CFR 890.5650 | -- |
| OTC & Rx | OTC | OTC | OTC | Same |
| Indications for<br>Use | Air Compression Leg<br>Massager is indicated for the<br>temporary relief of minor<br>muscle aches and pains and<br>for temporary increase in<br>circulation to the treated areas<br>in people who are in good<br>health. Air Compression Leg<br>Massager simulates kneading<br>and stroking of tissues by<br>using an inflatable garment. | The Air Compression Therapy<br>Device is indicated for the<br>temporary relief of minor<br>muscle aches and pains and<br>for temporary increase in<br>circulation to the treated<br>areas in people w ho are in<br>good health. The Air<br>Compression Therapy Device<br>simulates kneading and<br>stroking of tissues by using<br>an inflatable garment. | Air Pressure Foot Massager is<br>intended to temporarily relieve<br>minor muscle aches and/or<br>pains, and to temporarily<br>increase circulation to the<br>treated areas. | Minor<br>difference<br>Note 1 |
| Pow er Source | Input: AC100-240V, 50/60Hz<br>Output: DC 13.5V, 1A<br>Internal battery: 11.1V,<br>1000mAh | 100~240V 50/60Hz | Voltage: DC5V 1A Lithium<br>Battery: 3.7V, 2400mah | Minor<br>difference<br>Note 2 |
| Pow er<br>consumption | 13.5 W | 12W | 5W | Minor<br>difference<br>Note 2 |
| Dimensions<br>(W*H*D) | 215*50*70 mm | 10.2*5.9 *25.6 inch | 211*55*50mm | Minor<br>difference<br>Note 2 |
| Photo | Image: Air Compression Leg Massager | Image: Air Compression Therapy Device | Not publicly available | Minor<br>difference<br>Note 2 |
| Weight | 2.0 Kg (4.4pounds) | 4.6 pounds | 1.15±0.1kg | Minor<br>difference<br>Note 2 |
| Housing<br>Materials | Molded ABS enclosure | Molded ABS enclosure | Not publicly available | Same |
| Sleeves<br>Dimensions | Leg Wrap: 730*468 mm | Leg Wrap: 730*260 mm | 340*365*360mm | Minor<br>difference |
| Elements of<br>comparison | Subject Device | Predicate Device I | Predicate Device II | Comparison |
| | Image: Subject Device | Image: Predicate Device I | | Note 3 |
| Application area | Leg (feet, calves) | Leg (feet, calve, thigh) | Feet | Minor<br>difference<br>Note 3 |
| Number of<br>Chambers | 2-chamber | 3-chamber | 2-chamber | Minor<br>difference<br>Note 3 |
| Sleeve<br>Materials | Dacron | Nylon with a Polyurethane<br>laminate | Terylene | Minor<br>difference<br>Note 3 |
| Mode of<br>Compression | Sequential | Sequential / Peristaltic | Sequential | Same |
| Air Pressure<br>Level | Low level: 120mmHg<br>Medium level: 170mmHg<br>High level: 210mmHg<br>Error range: ±25mmHg | Low level: 150mmHg;<br>Mid level: 185mmHg;<br>High Level: 215mmHg<br>Error range: ±25mmHg | 30-110mmHg | Minor<br>difference<br>Note 4 |
| Treatment Time | 15 minutes | 20 minutes | 0-30min, default as 15min | Minor<br>difference<br>Note 5 |
| Inflation time | 5 – 18 s | 3 – 30 s | Not publicly available | Minor<br>difference<br>Note 6 |
| Keep time | 2 - 5 s | 1 - 5s | Not publicly available | |
| Deflation time | 2 – 5 s | 1 - 5s…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.